[Federal Register: September 7, 2007 (Volume 72, Number 173)]
[Rules and Regulations]               
[Page 51493-51531]
From the Federal Register Online via GPO Access [wais.access.gpo.gov]
[DOCID:fr07se07-17]                         


[[Page 51493]]

-----------------------------------------------------------------------

Part II





Environmental Protection Agency





-----------------------------------------------------------------------



40 CFR Parts 60, 72 and 75



Two Optional Methods for Relative Accuracy Test Audits of Mercury 
Monitoring Systems Installed on Combustion Flue Gas Streams and Several 
Amendments to Related Mercury Monitoring Provisions; Final Rule


[[Page 51494]]


-----------------------------------------------------------------------

ENVIRONMENTAL PROTECTION AGENCY

40 CFR Parts 60, 72 and 75

[EPA-HQ-OAR-2007-0164, FRL-8459-8]
RIN 2060-AO01

 
Two Optional Methods for Relative Accuracy Test Audits of Mercury 
Monitoring Systems Installed on Combustion Flue Gas Streams and Several 
Amendments to Related Mercury Monitoring Provisions

AGENCY: Environmental Protection Agency (EPA).

ACTION: Direct final rule.

-----------------------------------------------------------------------

SUMMARY: EPA is taking direct final action on two optional methods for 
relative accuracy audits of mercury monitoring systems installed on 
combustion flue gas streams and several amendments to related mercury 
monitoring provisions. This action approves two optional mercury (Hg) 
emissions test methods for potential use in conjunction with an 
existing regulatory requirement for Hg emissions monitoring, as well as 
several revisions to the mercury monitoring provisions themselves. This 
action is in regard to the testing and monitoring requirements for 
mercury specified in the Federal Register on May 18, 2005. Since that 
publication, EPA has received numerous comments concerning the 
desirability of EPA evaluating and allowing use of the measurement 
techniques addressed in the two optional methods in lieu of the methods 
identified in the cited Federal Register publication, as they can 
produce equally acceptable measures of the relative accuracy achieved 
by Hg monitoring systems. This action allows use of these two optional 
methods entirely at the discretion of the owner or operator of an 
affected emission source in place of the two currently specified 
methods. This direct final rule also amends Performance Specification 
12A by adding Methods 30A and 30B to the list of reference methods 
acceptable for measuring Hg concentration and the Hg monitoring 
provisions of May 18, 2005, to reflect technical insights since gained 
by EPA which will help to facilitate implementation including 
clarification and increased regulatory flexibility for affected 
sources.

DATES: This rule is effective on November 6, 2007 without further 
notice, unless EPA receives adverse comment by October 9, 2007. If EPA 
receives adverse comment, EPA will publish a timely withdrawal in the 
Federal Register informing the public that some or all of the 
amendments in this rule will not take effect.

ADDRESSES: Submit your comments, identified by Docket ID No. EPA-HQ-
OAR-2007-0164, by one of the following methods:
     http://www.regulations.gov. Follow the on-line instructions for 

submitting comments.
     E-mail: a-and-r-docket@epa.gov.
     Fax: (202) 566-9744.
     Mail: Two Optional Methods for Relative Accuracy Test 
Audits of Mercury Monitoring Systems Installed on Combustion Flue Gas 
Streams and Several Amendments to the Related Mercury Monitoring 
Provisions, Environmental Protection Agency, Mailcode: 2822T, 1200 
Pennsylvania Avenue, NW., Washington, DC 20460. Please include a total 
of two copies.
     Hand Delivery: EPA Docket Center, 1301 Constitution 
Avenue, NW., EPA Headquarters Library, Room 3334, EPA West Building, 
Washington, DC 20460. Such deliveries are only accepted during the 
Docket's normal hours of operation, and special arrangements should be 
made for deliveries of boxed information.
    Instructions: Direct your comments to Docket ID No. EPA-HQ-OAR-
2007-0164. EPA's policy is that all comments received will be included 
in the public docket without change and may be made available online at 
http://www.regulations.gov, including any personal information provided, 

unless the comment includes information claimed to be Confidential 
Business Information (CBI) or other information whose disclosure is 
restricted by statute. Do not submit information that you consider to 
be CBI or otherwise protected through http://www.regulations.gov or e-mail. 

The http://www.regulations.gov Web site is an ``anonymous access'' system, 

which means EPA will not know your identity or contact information 
unless you provide it in the body of your comment. If you send an e-
mail comment directly to EPA without going through http://www.regulations.gov, 

your e-mail address will be automatically captured and included as part 
of the comment that is placed in the public docket and made available 
on the Internet. If you submit an electronic comment, EPA recommends 
that you include your name and other contact information in the body of 
your comment and with any disk or CD-ROM you submit. If EPA cannot read 
your comment due to technical difficulties and cannot contact you for 
clarification, EPA may not be able to consider your comment. Electronic 
files should avoid the use of special characters, any form of 
encryption, and be free of any defects or viruses. For additional 
information about EPA's public docket, visit the EPA Docket Center 
homepage at http://www.epa.gov/epahome/dockets.htm.

    Docket: All documents in the docket are listed in the 
http://www.regulations.gov index. Although listed in the index, some 

information is not publicly available, e.g., CBI or other information 
whose disclosure is restricted by statute. Certain other material, such 
as copyrighted material, will be publicly available only in hard copy. 
Publicly available docket materials are available either electronically 
in http://www.regulations.gov or in hard copy at the Two Optional Methods for 

Relative Accuracy Audits of Mercury Monitoring Systems Installed on 
Combustion Flue Gas Streams Air and Radiation Docket, EPA/DC, EPA West 
Building, EPA Headquaters Library, Room 3334, 1301 Constitution Avenue, 
NW., Washington, DC. The Public Reading Room is open from 8:30 a.m. to 
4:30 p.m., Monday through Friday, excluding legal holidays. The 
telephone number for the Public Reading Room is (202) 566-1744, and the 
telephone number for the Air and Radiation Docket is (202) 566-1742.

FOR FURTHER INFORMATION CONTACT: Either Mr. William Grimley, Office of 
Air Quality Planning and Standards, Air Quality Assessment Division, 
Measurement Technology Group (E143-02), EPA, Research Triangle Park, NC 
27711, telephone (919) 541-1065, facsimile number (919) 541-0516, e-
mail address: grimley.william@epa.gov or Ms. Robin Segall, Office of 
Air Quality Planning and Standards, Air Quality Assessment Division, 
Measurement Technology Group (E143-02), EPA, Research Triangle Park, NC 
27711, telephone (919) 541-0893, facsimile number (919) 541-0516, e-
mail address: segall.robin@epa.gov.

SUPPLEMENTARY INFORMATION: 

I. Why is EPA using a Direct Final Rule?

    EPA is publishing this rule without a prior proposed rule because 
we view this as a noncontroversial action and anticipate no adverse 
comment. The most important benefit of direct final rulemaking for this 
action is to provide: (1) Additional reference method options, and (2) 
judicious revisions to mercury monitoring provisions specified in the 
Federal Register on May 18, 2005 that, if successful, relieve affected 
facilities of uncertainty regarding final emission monitoring 
requirements and certification details as opposed to waiting through a 
potentially protracted proposal/final

[[Page 51495]]

rulemaking process. Insofar as the two methods are concerned, EPA 
believes that they contain the necessary elements to generate 
acceptable data quality without being unduly burdensome. Through 
experience gained from developing existing performance based methods 
and trading rules, EPA has learned to identify test method criteria 
significant to effective rule implementation. EPA believes each of the 
two methods adopted in this action contain adequate specific criteria 
and procedures essential to the accurate measurement of Hg emissions, 
without adversely compromising the goals of performance-based 
methodology. EPA will continue to support and advance the principles 
and practicality of these methods by adding detailed method application 
information to facilitate their use to the Web site http://www.epa.gov/airmarkets/
 as it becomes available. Since use of either of these 

methods is not mandatory, but optional, there should be no objection to 
their availability. Regarding the amendments to the Hg emission 
monitoring provisions of 40 CFR parts 72 and 75, these amendments 
reflect EPA's increased technical understanding since the May 18, 2005 
rulemaking. However, in the ``Proposed Rules'' section of today's 
Federal Register, we are publishing a separate document that will serve 
as the proposed rule to approve provisions, if any, of this direct 
final rule that receive relevant adverse comments on this direct final 
rule. We will not institute a second comment period on this action. Any 
parties interested in commenting must do so at this time. For further 
information about commenting on this rule, see the ADDRESSES section of 
this document.
    If EPA receives adverse comment on one or more distinct provisions 
of this rulemaking, we will publish a timely withdrawal in the Federal 
Register indicating which provisions we are withdrawing and informing 
the public that those provisions will not take effect. The provisions 
that are not withdrawn will become effective on the date set out above, 
notwithstanding adverse comment on any other provision. We would 
address all public comments in a subsequent final rule based on the 
proposed rule.

II. Does This Action Apply to Me?

    Regulated Entities. The regulated categories and entities affected 
by this direct final rule include:

------------------------------------------------------------------------
                                                  Examples of regulated
            Category                NAICS \a\            entities
------------------------------------------------------------------------
Industry.......................          221112  Fossil fuel-fired
                                                  electric utility steam
                                                  generating units.
Federal government.............      \b\ 221122  Fossil fuel-fired
                                                  electric utility steam
                                                  generating units owned
                                                  by the Federal
                                                  government.
State/local governments........      \b\ 221122  Fossil fuel-fired
                                                  electric utility steam
                                                  generating units owned
                                                  by municipalities.
Tribal governments.............          921150  Fossil fuel-fired
                                                  electric utility steam
                                                  generating units in
                                                  Indian country.
------------------------------------------------------------------------
\a\ North American Industry Classification System.
\b\ Federal, State, or local government-owned and operated
  establishments are classified according to the activity in which they
  are engaged.

    This table is not intended to be exhaustive, but rather provides a 
guide for readers regarding entities likely to be affected by this 
direct final rule. If you have any questions regarding the 
applicability of this direct final rule to a particular entity, consult 
either the air permit authority for the entity or your EPA regional 
representative as listed in 40 CFR 63.13.

III. Where Can I Obtain a Copy of This Action?

    In addition to being available in the docket, an electronic copy of 
this direct final rule is also available on the World Wide Web through 
the Technology Transfer Network (TTN). Following signature, a copy of 
this direct final rule will be posted on the TTN's policy and guidance 
page for newly proposed or promulgated rules at the following address: 
http://www.epa.gov/ttn/oarpg. The TTN provides information and 

technology exchange in various areas of air pollution control.

IV. How Is This Document Organized?

    The information presented in this preamble is organized as 
follows:

I. Why Is EPA Using a Direct Final Rule?
II. Does This Action Apply to Me?
III. Where Can I Obtain a Copy of This Action?
IV. How Is This Document Organized?
V. Background
VI. This Action
VII. Statutory and Executive Order Reviews
    A. Executive Order 12866: Regulatory Planning and Review
    B. Paperwork Reduction Act
    C. Regulatory Flexibility Act
    D. Unfunded Mandates Reform Act
    E. Executive Order: 13132: Federalism
    F. Executive Order 13175: Consultation and Coordination With 
Indian Tribal Governments
    G. Executive Order 13045: Protection of Children From 
Environmental Health and Safety Risks
    H. Executive Order 13211: Actions That Significantly Affect 
Energy Supply, Distribution, or Use
    I. National Technology Transfer and Advancement Act
    J. Executive Order 12898: Federal Actions To Address 
Environmental Justice in Minority Populations and Low-Income 
Populations
    K. Congressional Review Act

V. Background

    On May 18, 2005, in the preamble of the Clean Air Mercury Rule 
(CAMR) (70 FR 28608), EPA stated its intention to propose and 
promulgate an instrumental reference method as an alternative to the 
use of ASTM Method D6784-02 (the Ontario Hydro Method) to perform 
Relative Accuracy Test Audits (RATAs) of Hg continuous emission 
monitoring systems (CEMS) and sorbent trap monitoring systems used to 
monitor Hg emissions from coal-fired power plants.
    In comments on the proposed CAMR, commenters had two primary 
objections to the use of the Ontario Hydro Method as the reference test 
method for RATAs. Some expressed concern that the complexity of this 
wet chemical method could lead to results that would cause a properly 
functioning Hg CEMS to fail a RATA. Other commenters noted that, unlike 
instrumental reference methods used to audit CEMS for SO2 
and NOX that provide real-time values, test results from the 
Ontario Hydro Method can take weeks to be received from the laboratory. 
Commenters stated that this time lag can lead to implementation 
problems with regard to both missing data and emissions reporting.
    Since the CAMR was promulgated, EPA has proposed changes to 40 CFR 
part 75, which would allow the use of EPA Method 29, with enhanced 
quality-assurance procedures, as an alternative Hg reference method (71 
FR 49257; August 22, 2006). Although Method 29 is somewhat simpler than 
the Ontario Hydro Method and is more familiar to stack testers and 
State regulatory agencies, it is also a wet chemistry method and is, 
therefore, subject to the same limitations that make the Ontario Hydro 
method less than optimal for RATA testing.
    In view of these considerations, EPA believes that for RATA 
testing, an instrumental Hg reference method

[[Page 51496]]

would be preferable to both the Ontario Hydro Method and to Method 29. 
An instrumental method would provide real-time data that would best 
facilitate implementation of a mercury cap and trade program. 
Therefore, this action approves a performance-based instrumental 
reference method for measuring Hg emission concentrations.
    Another commenter to the proposed CAMR recommended that the sorbent 
trap monitoring approach, now specified in 40 CFR part 75, appendix K, 
be considered for use as a reference method. Although EPA did not 
commit to establishing a sorbent trap reference method at the time of 
CAMR promulgation, stakeholder interest in this methodology has 
increased significantly. In an August 22, 2006 Federal Register notice, 
EPA solicited comment on the use of sorbent trap technology for Hg 
reference method testing, and numerous supportive comments were 
received. In view of this, we initiated a review of available 
historical test data where concurrent measurements of Hg concentration 
were made with sorbent trap systems and either the Ontario Hydro Method 
or Method 29. These data, taken together with additional supporting 
data from recent field tests that were performed after the CAMR was 
promulgated, suggest that using the sorbent trap methodology for Hg 
reference method testing is viable. The Hg sorbent trap approach is 
less onerous to use than either Ontario Hydro or Method 29, and 
although it does not measure real-time Hg concentrations, a thermal 
technique can be used to analyze the samples on the same day that they 
are collected, facilitating RATA testing in the context of a cap and 
trade program. Therefore, this action also approves a sorbent trap 
reference method for Hg, as an alternative to the Ontario Hydro Method 
and Method 29.
    This direct final rule also includes several carefully considered 
amendments to the Hg emission monitoring provisions of 40 CFR parts 72 
and 75. EPA believes these amendments will facilitate implementation of 
the CAMR by clarifying portions of that rule and by providing added 
regulatory flexibility to the affected sources.

VI. This Action

    This direct final rule allows for the earliest possible use of two 
optional reference test methods for measuring total vapor phase mercury 
emissions from stationary sources as well as several related amendments 
to the Hg monitoring provisions of the CAMR. Both an instrumental test 
method and a sorbent trap test method for measurement of total vapor 
phase mercury emissions are being added to Appendix A-8 of 40 CFR part 
60 as approved alternatives to the Ontario Hydro Method and EPA Method 
29 to perform RATAs of installed mercury monitoring systems. The two 
methods are discussed below, and the related amendments are explained 
in detail later in this section.
    The first method being added to appendix A-8 of 40 CFR part 60 
today is titled ``Method 30A--Determination of Total Vapor Phase 
Mercury Emissions from Stationary Sources (Instrumental Analyzer 
Procedure).'' In Method 30A, a sample of the effluent gas is 
continuously extracted and conveyed to an analyzer capable of measuring 
the total vapor phase Hg concentration. Elemental and oxidized mercury 
(i.e., Hg\0\ and Hg+\2\) may be measured separately or 
simultaneously but, for purposes of this method, total vapor phase Hg 
is the sum of Hg\0\ and Hg+\2\. Method 30A provides test 
program-specific verification of method performance using a dynamic 
spiking approach, coupled with other performance criteria, which 
include system calibration, interference testing, and system integrity/
drift checks. The dynamic spiking requirement, which is a gaseous 
``method of standard additions,'' is the only part of Method 30A not 
parallel to the routinely applied instrumental reference methods used 
to perform relative accuracy testing of CEMS for SO2 and 
NOX. The dynamic spiking procedure is included in Method 30A 
to characterize measurement bias for Hg, which can be highly reactive 
on a site-specific basis (i.e., for each emissions sample matrix), with 
recovery criteria set to ensure that the bias is held to a minimal 
level. All performance requirements of Method 30A must be met for the 
data to be considered valid. The availability of an instrumental 
reference method for Hg testing is consistent with the approach EPA has 
taken in the successful Acid Rain and NOX Budget emissions 
trading programs.
    Method 30A is performance based in keeping with the criteria 
established under our Notice of Intent to Implement Performance Based 
Measurement Systems for Environmental Monitoring (62 FR 52098, October 
6, 1997). Use of the performance-based measurement approach will allow 
for continued development and application of new, improved, and more 
cost-effective Hg measurement technologies while ensuring the 
collection of data of known quality.
    Based on EPA's experience in conducting test programs to evaluate 
the procedures and performance criteria included in Method 30A, EPA 
recognizes that although prototypes of all equipment needed to perform 
this method have been successfully demonstrated in the field, at 
present the equipment needed to follow all procedures required by the 
method is commercially available only on a limited basis, and is being 
further refined. One of the issues of greatest concern in the 
development of an instrumental reference method for Hg has been the 
design of the sampling probe. Most of the commercially-available probes 
suitable for Hg measurement are very heavy (over 100 lbs.) making it 
difficult to move the probe from point-to-point and port-to-port for Hg 
stratification testing and/or sample traverses. Much progress is being 
made in probe redesign. One manufacturer has recently developed a probe 
that weighs less than 40 lbs., samples at significantly lower flow 
rates, and is suitable for dynamic spiking. Additional field testing of 
this probe and others currently under development is underway, and EPA 
plans to continue to actively encourage equipment development and 
evaluation. To encourage the use of Method 30A, including further 
development of the supporting equipment, which we believe will 
eventually enable source testers to perform Hg monitoring system RATAs 
more efficiently and will become the reference method of choice for 
many testing companies and affected sources, we are deferring the 
requirement for implementation of the dynamic spiking and Hg 
stratification test procedures until January 1, 2009. EPA believes this 
deferral is reasonable because Hg monitoring data reported to EPA in 
2009 will not be used in the trading of Hg allowances, as allowance 
accounting under the CAMR does not begin until 2010. Source testers are 
encouraged to use this time to acquire the necessary equipment and 
familiarize themselves with these procedures. Also, for all emissions 
test programs and RATAs performed under CAMR prior to January 1, 2009, 
we are allowing either: (1) A 12-point traverse for sulfur dioxide 
(SO2) to be substituted for a 12-point Hg traverse, in cases 
where stratification testing is used to determine the appropriate 
number and location of the reference method sampling points, or (2) use 
of the alternate three-point traverse line (0.4, 1.2, and 2.0 meters 
from the stack wall) as specified in section 8.1.3.2 of Performance 
Specification 2 (40 CFR part 60, appendix B). We

[[Page 51497]]

believe that in the short-term, these temporary deferrals will 
encourage the application of Method 30A and will help affected CAMR 
sources meet the January 1, 2009 deadline for initial certification of 
the required Hg monitoring systems. Several additional Method 30A 
development considerations are worthy of note. A preliminary draft of 
Method 30A was first available for public consideration on an EPA Web 
site (http://www.epa.gov/ttn/emc/) on February 28, 2006. Since that time, EPA 

and several stakeholder groups have evaluated the various technical 
aspects of the method. Based on the combined laboratory and field 
observations, EPA has been able to simplify several procedural 
requirements that we believe are essential to the method. The dynamic 
spiking requirement (for test program-specific verification of 
measurement system data quality) has been reduced to only a pretest 
requirement. The interference test has been made optional. The three-
point system calibration error test using Hg+\2\ has been 
streamlined to a system integrity check using a zero gas and a single 
upscale Hg+\2\ gas. Another change has been to relax the 
Hg\0\ calibration error specification from 2 percent to 5 percent of 
span, in recognition of the fact that this procedure is a check of the 
entire measurement system, as well as the current knowledge regarding 
the uncertainty of NIST traceable standards. EPA does plan, however, to 
reconsider this specification relaxation as more field data become 
available. A final consideration in development of Method 30A has been 
the requirement for calibration with both Hg\0\ and Hg+\2\. 
Some stakeholders have recommended that we eliminate the Hg\0\ 
calibration and rely solely on the Hg+\2\ calibration. EPA, 
however, believes this approach would not be adequate, because if only 
Hg+\2\ were used, instrument calibration response adjustment 
could compensate for an unknown amount of converter inefficiency, which 
would then result in an inaccurate total mercury measurement in 
situations where Hg\0\ is an appreciable fraction of the total stack 
gas Hg.
    The second method being added to appendix A-8 of 40 CFR part 60 
today is titled ``Method 30B--Use of Sorbent Traps to Measure Total 
Vapor Phase Mercury Emissions from Coal-Fired Combustion Sources.'' In 
Method 30B, a sample of the effluent gas is continuously drawn through 
a series of tubes containing activated carbon or another sorbent 
material. After sampling, the tubes are sealed. The Hg captured by the 
sorbent is then either: (1) Thermally desorbed and analyzed; or (2) the 
tubes are transferred to a laboratory for extraction of Hg and 
analysis. Like Method 30A, Method 30B is a performance-based method and 
contains performance specifications and procedures for hardware 
selection and calibration, sorbent spiking, and analytical recovery/
analysis which allow for development and application of new, improved, 
and more cost-effective Hg measurement technologies while still 
ensuring the collection of data of known quality. In particular, Method 
30B contains five key measurement performance tests designed to ensure: 
(1) Selection of a sorbent and analytical technique combination capable 
of quantitative collection and analysis of gaseous Hg, (2) collection 
during field testing of enough Hg on each sorbent trap to be reliably 
quantified, and (3) adequate performance of the method for each test 
program.
    In considering development of a sorbent trap-based reference 
method, EPA has reviewed historical emissions data where sorbent trap 
measurement systems were operated concurrently with either the Ontario 
Hydro Method or Method 29 (40 CFR part 60, appendix A-8). EPA has also 
conducted several field test evaluations of sorbent trap systems versus 
the Ontario Hydro Method in collaboration with the Electric Power 
Research Institute (EPRI). Based on these efforts, we have concluded 
that a sorbent trap-based technique coupled with appropriate 
performance criteria and QA procedures can provide Hg emissions data of 
quality comparable to that produced by the Ontario Hydro Method. Data 
supporting this conclusion are presented in the docket, EPA-HQ-OAR-
2007-0164.
    As we have done for Method 30A, for Method 30B emission tests and 
RATAs performed prior to January 1, 2009, we are allowing either: (1) A 
12-point traverse for sulfur dioxide (SO2) to be substituted 
for a 12-point Hg traverse for the stratification testing used to 
determine the number and location of the reference method sampling 
points, or (2) use of the alternate three-point traverse line (0.4, 
1.2, and 2.0 meters from the stack wall) as specified in section 
8.1.3.2 of Performance Specification 2 (40 CFR part 60, appendix B). We 
also intend to extend this temporary deferral of mercury stratification 
testing to application of the Ontario Hydro Method and Method 29. EPA 
believes this deferral is reasonable because Hg monitoring data 
reported to EPA in 2009 will not be used in the trading of Hg 
allowances, as allowance accounting under the CAMR does not begin until 
2010.
    This direct final rule also amends Performance Specification 12A of 
appendix B to part 60 by adding Methods 30A and 30B to the list of 
reference methods acceptable for relative accuracy testing of Hg 
emissions monitoring systems. Once this direct final rule becomes 
effective, the reference methods acceptable for Hg measurement in 
Performance Specification 12A will include Methods 29, 30A, 30B, and 
ASTM D6784-02.
    With today's action, EPA is taking the opportunity to include 
several considered revisions to the Hg emission monitoring provisions 
of 40 CFR parts 72 and 75 as described in detail below. EPA is 
including these revisions in this direct final rule because we believe 
that they will facilitate implementation of the Hg monitoring under 
CAMR.
    First, Sec.  75.81(a) is being revised to confirm that the Hg CEMS 
and sorbent trap monitoring systems required under subpart I of part 75 
are to measure the total vapor phase mass concentration of Hg in the 
flue gas, including both the elemental and oxidized forms of Hg, 
expressed in units of micrograms per standard cubic meter ([mu]g/scm). 
Although it is generally understood that total vapor phase Hg is the 
regulated pollutant under CAMR, it recently was brought to EPA's 
attention that subpart I of part 75 does not explicitly state that Hg 
monitoring systems only need to measure total vapor phase Hg. The 
amended language in Sec.  75.81(a) clarifies this.
    Second, paragraph (i) in Sec.  75.15 is being revised and a new 
paragraph (d)(2)(ix) is being added to Sec.  75.20, to codify the rules 
for using optional non-redundant (``cold'') backup Hg monitoring 
systems and like-kind replacement Hg analyzers, when the primary Hg 
monitoring system is unable to provide quality-assured data. For the 
other types of monitoring systems required by part 75, these monitoring 
options have been in place since May 1999 (see 64 FR 28597, May 26, 
1999). Today's action simply extends these provisions to Hg monitoring 
systems. Through the years, the regulated community has found these 
backup monitoring options to be beneficial, in that they minimize the 
use of missing data substitution procedures during outages of the 
primary monitoring system.
    In particular, Sec.  75.20(d)(2)(ix) specifies that a non-redundant 
backup Hg monitoring system can either be a Hg CEMS or a sorbent trap 
monitoring system. The non-redundant backup Hg

[[Page 51498]]

monitoring system must be initially certified at each unit or stack 
location where it will be used, in accordance with Sec.  
75.20(d)(2)(i). For a non-redundant backup Hg CEMS, all of the initial 
certification tests specified in Sec.  75.20(c)(1) are required, except 
for the 7-day calibration error test. However, for ongoing quality 
assurance (QA), a RATA is required only once every two years (8 
calendar quarters), as specified in Sec.  75.20(d)(2)(vi). For a non-
redundant backup sorbent trap monitoring system, a RATA is required for 
initial certification, and once every two years thereafter for ongoing 
QA.
    When a certified non-redundant backup Hg CEMS or a like-kind 
replacement Hg analyzer is brought into service, a three-point 
linearity check with elemental Hg standards and a single-point system 
integrity check will be required. Alternatively, a three-level system 
integrity check may be performed instead of these two tests. When a 
certified non-redundant backup sorbent trap monitoring system is 
brought into service, only the routine sampling and QA procedures of 
Sec.  75.15 and appendix K of part 75 will be required.
    Each non-redundant backup Hg monitoring system and each like-kind 
replacement Hg analyzer will be subject to the applicable ongoing QA 
requirements, restrictions and conditions specified in Sec.  
75.20(d)(2). For certified non-redundant backup Hg CEMS and like-kind 
replacement Hg analyzers, the weekly system integrity checks described 
in section 2.6 of appendix B of 40 CFR part 75 will also be required as 
long as the system or analyzer remains in service, unless the daily 
calibration error tests of the analyzer are done using NIST-traceable 
oxidized Hg standards.
    Third, a new paragraph (k) is being added to Sec.  75.15 that: (1) 
Clarifies that, when the RATA of an appendix K sorbent trap monitoring 
system is performed, the type of sorbent material used in the appendix 
K sorbent traps must be the same as that used for daily operation of 
the appendix K monitoring system, and (2) allows the appendix K traps 
used during RATA testing to be smaller than the traps used for daily 
operation of the appendix K monitoring system. This change will be 
particularly advantageous at very low Hg concentrations as it will 
facilitate shorter RATA test run times. Parallel changes are being made 
to section 6.5.7 of appendix A of part 75 to be consistent with the 
provisions of Sec.  75.15(k). Section 6.5.7 currently requires the 
appendix K sorbent traps used for the RATA to be the same size as the 
traps used for daily operation of the appendix K monitoring system.
    Fourth, today's action revises a number of sections of part 75, 
appendix K, pertaining to the use of sorbent trap monitoring systems. 
EPA is withdrawing the requirement to use the percentage recovery of 
the elemental Hg spike in section 3 of each sorbent trap to adjust or 
``normalize'' the Hg mass collected in sections 1 and 2 of the trap. 
The requirement to spike the third section of each trap is being 
retained and data from each pair of traps must still be invalidated if 
either or both spike recovery percentages fall outside the acceptable 
limits;\1\ however, the results of the spike recoveries will no longer 
be used to adjust the Hg mass collected in the first two sections of 
the traps. EPA is making this rule change based on an analysis of 
recent spike recovery data from long-term appendix K field 
demonstrations. Although the vast majority of the spike recoveries in 
these studies have been within the currently acceptable limits of 75 to 
125 percent, the requirement to normalize based on spike recovery could 
affect data precision. For a given pair of traps, if one spike recovery 
was high (e.g., 110 percent) and the other one low (e.g., 90 percent), 
normalization of the Hg mass collected in the first two trap sections 
using third section spike recoveries could make it difficult for a pair 
of sorbent traps to meet the relative deviation (RD) specifications in 
appendix K. In the example cited, normalization of the data would cause 
the Hg concentrations measured by the traps to be adjusted by 10 
percent in opposite directions, i.e., one upward and one downward. 
Thus, two Hg concentrations that may have been in close agreement 
without normalization now might not be able to meet the RD 
specifications. In view of this, EPA has concluded that evaluating the 
spike recovery data on a pass/fail basis instead of using the percent 
recovery values to adjust the emissions data is more technically sound 
and is also consistent with the way in which the results of daily and 
quarterly QA assessments of CEMS are interpreted.
---------------------------------------------------------------------------

    \1\ On August 22, 2006, EPA proposed to amend Appendix K to 
allow the data from a pair of sorbent traps to be validated in cases 
where the third section spike recovery from only one of the traps 
meets the percent recovery specifications (see 71 FR 49275). EPA 
proposed to allow the results from the trap that meets the 
specifications to be used for reporting, provided that a single trap 
adjustment factor (STAF) of 1.222 is applied. EPA is evaluating the 
comments received on this proposal and expects to publish the final 
rule in the summer of 2007.
---------------------------------------------------------------------------

    Regarding the range of acceptable third section spike recoveries, 
EPA is not changing the 75 to 125 percent acceptance criteria. As 
previously noted, early field experience with appendix K monitoring 
systems has demonstrated that spike recoveries within this range are 
achievable. However, recent appendix K data indicate that more 
stringent acceptance criteria may be justifiable. It appears that there 
has been a marked improvement in third section spike recovery 
percentages. Recoveries in the range from 85 to 115 percent are 
consistently being achieved. If this trend continues, EPA may propose 
to tighten the spike recovery acceptance criteria in a future 
rulemaking. Toward that end, EPA will continue to collect and evaluate 
third section spike recovery data from appendix K monitoring systems in 
the months ahead.
    To effect these changes to appendix K, section 11.5 is being 
removed and reserved; section 10.4 is being revised; Equations K-6 and 
K-7 are being redesignated as Equations K-5 and K-6, respectively; and 
the definition of ``M*'' in redesignated Equation K-5 is being revised.
    EPA is also revising appendix K to allow the owner or operator to 
use other types of gas flow meters besides the conventional dry gas 
meter (DGM) to quantify sample gas volume. Since the publication of 
appendix K (see 70 FR 28695, May 18, 2005), numerous requests have been 
received from the regulated community to allow this flexibility. In 
response to these requests, EPA initiated an investigation of the 
feasibility of replacing the DGM in a sorbent trap monitoring system 
with a thermal mass flow meter. As a result of its investigation, EPA 
has concluded that a properly calibrated thermal mass flow meter can be 
at least as accurate as a DGM. The mass flow meter is also a more 
modern technology than the DGM; since it has no moving parts, it may be 
more reliable than a DGM for continuous duty.
    Having found one type of gas flow meter that can measure as 
accurately as a DGM, EPA is persuaded that there may be other 
commercially available gas flow meter technologies that are equally 
capable and may be suitable for appendix K applications. Accordingly, 
EPA has decided that a performance-based approach, rather than a 
prescriptive one, is more appropriate for appendix K gas flow meters. 
Today's action allows the use of any type of gas flow meter that is 
capable of accurately measuring gas volumes to within 2 percent.
    Section 9.2.2.1 of appendix K now requires the manufacturer of the 
gas flow meter to perform all necessary set-

[[Page 51499]]

up, testing, programming, etc. of the meter and to provide any 
necessary instructions so that for the particular field application, 
the meter will give an accurate readout of dry gas volume in units of 
standard cubic meters. Then, prior to its initial use, the flow meter 
must be calibrated at a minimum of three settings covering the expected 
range of sample flow rates for the appendix K system. The initial 
calibration may be performed either by the manufacturer or by the end 
user. The calibration of the gas flow meter must be checked quarterly 
thereafter, at an intermediate flow rate. For mass flow meters, the 
initial three-point calibration must be performed by using either a 
compressed gas mixture containing CO2, O2, and 
N2 in proportions representative of the stack gas 
composition or by using the actual stack gas. However, when the initial 
calibration is done with a compressed gas mixture, the mass flow meter 
may not be used until an additional on-site calibration check of the 
flow meter at an intermediate flow rate is performed and passed, using 
the actual stack gas.
    To calibrate the gas flow meter, the owner or operator may either 
follow the basic procedures in section 10.3 or section 16 of Method 5 
in appendix A-3 of part 60 for calibration of dry gas meters, or 
alternatively, may temporarily install a reference gas flow meter 
(RGFM) at the discharge of the appendix K monitoring system while the 
monitoring system is in operation and make concurrent measurements of 
dry stack gas volume with the RGFM and the appendix K gas flow meter. 
If the latter option is chosen, the RGFM may either be a gas flow 
metering device that has been calibrated according to section 10.3.1 or 
section 16 of Method 5 or a NIST-traceable volumetric calibration 
device with an accuracy of 1 percent. Note that this 
alternative calibration technique allows required QA checks to be 
performed with little or no disruption of the operation of the sorbent 
trap monitoring system.
    Regardless of which calibration approach is used, a calibration 
factor, Yi, must be obtained at each tested flow rate, where 
Yi is the ratio of the volume measured by the reference 
meter to the volume measured by the flow meter being calibrated. For 
the initial three-point calibration, the three Yi values 
must be averaged, and each individual Yi must be within 
 0.02 of the average value. The average value, Y, must then 
be used to correct the gas volumes measured by the gas flow meter. For 
single-level calibration checks (e.g., the quarterly checks performed 
for routine QA), the Yi value obtained at the tested flow 
rate must be compared with the current value of Y. If Yi 
differs from Y by more than 5 percent, a full three-point recalibration 
is then required to determine a new Y value.
    In this direct final action, the majority of the revised rule 
provisions pertaining to gas flow meters can be found in sections 5.1.5 
and 9.2 of appendix K. Minor revisions to sections 7.2.3 and 7.2.5, 
Figure K-1, and Table K-1 are being made to be consistent with the 
changes to sections 5.1.5 and 9.2. In several other places throughout 
part 75 and in the definition of ``Sorbent trap monitoring system'' in 
part 72, the term ``dry gas meter,'' when used in reference to a 
sorbent trap monitoring system, is being replaced with the more general 
term ``gas flow meter.'' Revisions to section 1.5.2 of appendix B of 
part 75 will require the gas flow meter calibration procedures and 
protocols for periodic recalibration of reference gas flow meters to be 
included in the QA plan for the affected unit.
    This direct final action, which approves the use of two optional 
methods (Methods 30A and 30B) for determining total vapor phase Hg 
emissions from stationary sources, is being taken in response to 
numerous public comments concerning the desirability of allowing the 
use of these types of methods to comply with the Hg emission monitoring 
requirements of the CAMR for electric utility steam generating units. 
In the May 18, 2005 final rule (70 FR 28636), we summarized the public 
comments that we received regarding the use of an instrumental method 
as an alternative to the Ontario Hydro Method specified in the proposed 
CAMR. As noted earlier in this preamble, the commenters primarily 
objected to the required use of the Ontario Hydro Method as the 
reference method for the RATAs of Hg monitoring systems and expressed 
concern about the complexities in the method and the amount of time 
that is required to perform the testing and to receive the results. 
Commenters pointed out that it could take days to complete the testing 
and weeks to receive the results from a laboratory. Commenters claimed 
that for the cap and trade program proposed under CAMR, these delays 
could lead to significant implementation problems with respect to the 
timely reporting of emissions data. Further, if a RATA should be failed 
or invalidated (e.g., if fewer than nine test runs meet the relative 
deviation criterion for the paired Ontario Hydro trains), data from the 
Hg monitoring system would be invalidated from the hour of the failed 
or invalidated test until the hour of completion of a successful RATA. 
Conservatively high substitute data values would have to be reported 
during that entire time period. In our response to those comments in 
the final CAMR rule, we stated that the alternative use of an 
instrumental method for the required RATAs of Hg monitoring systems and 
sorbent trap monitoring systems is allowed by the final rule but is 
subject to approval by the Administrator. We also stated our commitment 
to propose and promulgate a Hg instrumental reference method once 
sufficient supporting field test data become available. We further 
stated that ``A Hg instrumental reference method for RATA testing is 
vastly preferable to the Ontario Hydro Method and will greatly 
facilitate the implementation of a Hg cap-and-trade program.''
    Since promulgation of CAMR, we have continued to communicate with 
stakeholders interested in the Hg monitoring requirements of the rule, 
and we have come to more clearly understand that it is of great 
interest to the affected entities to have additional reference method 
options available for relative accuracy testing of installed Hg 
monitoring systems as soon as possible. Accordingly, at the end of 
2005, we began developing an instrumental test method for Hg and 
solicited feedback from the stakeholders on a working draft of the 
method (referred to as PRE-009 at http://www.epa.gov/ttn/emc/prelim.html
). More recently, we have been developing a viable sorbent 

trap reference method. These efforts have resulted in Methods 30A and 
30B.
    The general beneficial impacts of this direct final rule to approve 
the two optional Hg test methods and amend targeted portions of 40 CFR 
parts 72 and 75 include: Allowing affected sources to choose the use of 
an alternative to the Ontario Hydro Method without the administrative 
burden of applying for Administrator approval on a case-by-case basis; 
providing the availability of real-time RATA results (Method 30A); 
reducing the overall RATA testing times; reducing costs relative to the 
Ontario Hydro Method; and providing additional flexibility in appendix 
K sorbent trap monitoring and backup monitoring approaches. The two 
optional methods being approved by this direct final rule are 
considered to be comparable to the Ontario Hydro Method in terms of the 
quality of the results produced. Over the last year, EPA has 
collaborated with EPRI and some of its members in a number of field 
test programs that have confirmed that the instrumental reference 
method approved/established in this notice will provide data comparable 
to or better

[[Page 51500]]

than that of the ``Ontario Hydro Method.''
    Assuming we do not receive adverse comment on this direct final 
rulemaking and Methods 30A and 30B become final, we plan to post 
information relevant to Method 30A and 30B applications and equipment 
advances on EPA's Web site at http://www.epa.gov/airmarkets.


VII. Statutory and Executive Order Reviews

A. Executive Order 12866: Regulatory Planning and Review

    This action is not a ``significant regulatory action'' under the 
terms of Executive Order (EO) 12866 (58 FR 51735, October 4, 1993) and 
is therefore not subject to review under the EO.

B. Paperwork Reduction Act

    This action does not impose an information collection burden under 
the provisions of the Paperwork Reduction Act, 44 U.S.C. 3501 et seq. 
Burden means the total time, effort, or financial resources expended by 
persons to generate, maintain, retain, or disclose or provide 
information to or for a Federal agency. This includes the time needed 
to review instructions; develop, acquire, install, and utilize 
technology and systems for the purposes of collecting, validating, and 
verifying information, processing and maintaining information, and 
disclosing and providing information; adjust the existing ways to 
comply with any previously applicable instructions and requirements; 
train personnel to be able to respond to a collection of information; 
search data sources; complete and review the collection of information; 
and transmit or otherwise disclose the information.
    An agency may not conduct or sponsor, and a person is not required 
to respond to a collection of information, unless it displays a 
currently valid OMB control number. The OMB control numbers for EPA's 
regulations in 40 CFR are listed in 40 CFR part 9.

C. Regulatory Flexibility Act

    The Regulatory Flexibility Act (RFA) generally requires an agency 
to prepare a regulatory flexibility analysis of any rule subject to 
notice and comment rulemaking requirements under the Administrative 
Procedure Act or any other statute unless the agency certifies that the 
rule will not have a significant economic impact on a substantial 
number of small entities. Small entities include small businesses, 
small organizations, and small governmental jurisdictions.
    For purposes of assessing the impacts of today's rule on small 
entities, small entity is defined as: (1) A small business whose parent 
company has fewer than 100 or 1,000 employees, or fewer than 4 billion 
kilowatt-hr per year of electricity usage, depending on the size 
definition for the affected North American Industry Classification 
System code; (2) a small governmental jurisdiction that is a government 
of a city, county, town, school district or special district with a 
population of less than 50,000; and (3) a small organization that is 
any not-for-profit enterprise which is independently owned and operated 
and is not dominant in its field.
    After considering the economic impacts of today's direct final rule 
on small entities, I certify that this action will not have a 
significant economic impact on a substantial number of small entities. 
This direct final rule will not impose any requirements on small 
entities because it does not impose any additional regulatory 
requirements, but rather provides clarification and additional 
regulatory flexibilty.

D. Unfunded Mandates Reform Act

    Title II of the Unfunded Mandates Reform Act of 1995 (UMRA), Pub. 
L. 104-4, establishes requirements for Federal agencies to assess the 
effects of their regulatory actions on State, local, and tribal 
governments and the private sector. Under section 202 of the UMRA, EPA 
generally must prepare a written statement, including a cost-benefit 
analysis, for proposed and final rules with ``Federal mandates'' that 
may result in expenditures to State, local, and tribal governments, in 
the aggregate, or to the private sector, of $100 million or more in any 
one year. Before promulgating an EPA rule for which a written statement 
is needed, section 205 of the UMRA generally requires EPA to identify 
and consider a reasonable number of regulatory alternatives and adopt 
the least costly, most cost-effective or least burdensome alternative 
that achieves the objectives of the rule. The provisions of section 205 
do not apply when they are inconsistent with applicable law. Moreover, 
section 205 allows EPA to adopt an alternative other than the least 
costly, most cost-effective, or least burdensome alternative if the 
Administrator publishes with the final rule an explanation why that 
alternative was not adopted. Before EPA establishes any regulatory 
requirements that may significantly or uniquely affect small 
governments, including tribal governments, it must have developed under 
section 203 of the UMRA a small government agency plan. The plan must 
provide for notifying potentially affected small governments, enabling 
officials of affected small governments to have meaningful and timely 
input in the development of EPA regulatory proposals with significant 
Federal intergovernmental mandates, and informing, educating, and 
advising small governments on compliance with the regulatory 
requirements.
    EPA has determined that this direct final rule does not contain a 
Federal mandate that may result in expenditures of $100 million or more 
for State, local, and tribal governments in the aggregate, or to the 
private sector in any 1 year, nor does this rule significantly or 
uniquely impact small governments, because it contains no requirements 
that impose new obligations upon them. Thus, this direct final rule is 
not subject to the requirements of sections 202 and 205 of the UMRA.

E. Executive Order 13132: Federalism

    Executive Order 13132, entitled ``Federalism'' (64 FR 43255, August 
10, 1999), requires EPA to develop an accountable process to ensure 
``meaningful and timely input by State and local officials in the 
development of regulatory policies that have federalism implications.'' 
``Policies that have federalism implications'' is defined in the 
Executive Order to include regulations that have ``substantial direct 
effects on the States, on the relationship between the national 
government and the States, or on the distribution of power and 
responsibilities among the various levels of government.''
    This direct final rule does not have federalism implications. It 
will not have substantial direct effects on the States, on the 
relationship between the national government and the States, or on the 
distribution of power and responsibilities among the various levels of 
government, as specified in Executive Order 13132. The use of these 
methods is optional on the part of the regulated entities listed. Thus, 
Executive Order 13132 does not apply to this direct final rule.

F. Executive Order 13175: Consultation and Coordination With Indian 
Tribal Governments

    Executive Order 13175, entitled ``Consultation and Coordination 
with Indian Tribal Governments'' (65 FR 67249, November 9, 2000), 
requires EPA to develop an accountable process to ensure ``meaningful 
and timely input by tribal officials in the development of regulatory 
policies that have tribal implications.'' This direct final rule does 
not have tribal implications, as specified in Executive Order 13175. It 
will not have substantial direct effects on tribal governments, on the

[[Page 51501]]

relationship between the Federal government and Indian tribes, or on 
the distribution of power and responsibilities between the Federal 
government and Indian tribes. Thus, Executive Order 13175 does not 
apply to this final rule.

G. Executive Order 13045: Protection of Children From Environmental 
Health and Safety Risks

    Executive Order 13045: ``Protection of Children from Environmental 
Health Risks and Safety Risks'' (62 FR 19885, April 23, 1997) applies 
to any rule that: (1) Is determined to be ``economically significant'' 
as defined under Executive Order 12866, and (2) concerns an 
environmental health or safety risk that EPA has reason to believe may 
have a disproportionate effect on children. If the regulatory action 
meets both criteria, the Agency must evaluate the environmental health 
or safety effects of the planned rule on children, and explain why the 
planned regulation is preferable to other potentially effective and 
reasonably feasible alternatives considered by the Agency. EPA 
interprets Executive Order 13045 as applying only to those regulatory 
actions that are based on health or safety risks, such that the 
analysis required under section 5-501 of the Order has the potential to 
influence the regulation. This rule is not subject to Executive Order 
13045 because it does not establish an environmental standard intended 
to mitigate health or safety risks.

H. Executive Order 13211: Actions That Significantly Affect Energy 
Supply, Distribution, or Use

    This rule is not subject to Executive Order 13211, ``Actions 
Concerning Regulations That Significantly Affect Energy Supply, 
Distribution, or Use'' (66 FR 28355, May 22, 2001) because it is not a 
significant regulatory action under Executive Order 12866.

I. National Technology Transfer Advancement Act

    Section 12(d) of the National Technology Transfer and Advancement 
Act of 1995 (NTTAA), Public Law No. 104-113, section 12(d) (15 U.S.C. 
272 note) directs EPA to use voluntary consensus standards in its 
regulatory activities unless to do so would be inconsistent with 
applicable law or otherwise impractical. Voluntary consensus standards 
are technical standards (e.g., materials specifications, test methods, 
sampling procedures, and business practices) that are developed or 
adopted by voluntary consensus standards bodies. The NTTAA directs EPA 
to provide Congress, through OMB, explanations when the Agency decides 
not to use available and applicable voluntary consensus standards. This 
rulemaking involves technical standards. Consistent with the NTTAA, EPA 
in a previous related rulemaking (70 FR 28606, May 18, 2005) identified 
an acceptable VCS for measuring Hg emissions. The standard ASTM D6784-
02, Standard Test Method for Elemental, Oxidized, Particle-Bound and 
Total Mercury Gas Generated from Coal-Fired Stationary sources (Ontario 
Hydro Method) was cited in that final rule for measuring Hg emissions. 
After today's action becomes effective, the Ontario Hydro Method will 
remain an acceptable method for measuring Hg emissions.

J. Executive Order 12898: Federal Actions To Address Environmental 
Justice in Minority Populations and Low-Income Populations

    Executive Order 12898 (59 FR 7629 (Feb. 16, 1994)) establishes 
federal executive policy on environmental justice. Its main provision 
directs federal agencies, to the greatest extent practicable and 
permitted by law, to make environmental justice part of their mission 
by identifying and addressing, as appropriate, disproportionately high 
and adverse human health or environmental effects of their programs, 
policies, and activities on minority populations and low-income 
populations in the United States.
    EPA has determined that this direct final rule will not have 
disproportionately high and adverse human health or environmental 
effects on minority or low-income populations because it does not 
affect the level of protection provided to human health or the 
environment. This direct final rule does not affect or relax the 
control measures on sources impacted by this rule and therefore will 
not cause emissions increases from these sources.

K. Congressional Review Act

    The Congressional Review Act, 5 U.S.C. 801 et seq., as added by the 
Small Business Regulatory Enforcement Fairness Act of 1996, generally 
provides that before a rule may take effect, the Agency promulgating 
the rule must submit a rule report, which includes a copy of the rule, 
to each House of the Congress and to the Comptroller General of the 
United States. EPA will submit a report containing this rule and other 
required information to the U.S. Senate, the U.S. House of 
Representatives, and the Comptroller General of the United States prior 
to publication of the rule in the Federal Register. A major rule cannot 
take effect until 60 days after it is published in the Federal 
Register. This action is not a ``major rule'' as defined by 5 U.S.C. 
804(2). This rule will be effective on November 6, 2007.

List of Subjects

40 CFR Part 60

    Environmental protection, Administrative practice and procedures, 
Air pollution control, Continuous emission monitors, Electric 
utilities, Mercury, Test methods and procedures.

40 CFR Part 72

    Environmental protection, Administrative practice and procedures, 
Air pollution control, Continuous emission monitors, Electric 
utilities, Mercury, Test methods and procedures.

40 CFR Part 75

    Environmental protection, Administrative practice and procedures, 
Air pollution control, Continuous emission monitors, Electric 
utilities, Mercury, Test methods and procedures.

    Dated: August 17, 2007.
Stephen L. Johnson,
Administrator.

0
For the reasons set out in the preamble, title 40, chapter I, parts 60, 
72, and 75 of the Code of Federal Regulations are amended as follows:

PART 60--STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES

0
1. The authority citation for part 60 continues to read as follows:

    Authority: 42 U.S.C. 7401-7601.

Appendix A-8 [Amended]

0
2. Amend Appendix A-8 by revising the heading and adding in numerical 
order Methods 30A and 30B to read as follows:

APPENDIX A-8 TO PART 60--TEST METHODS 26 THROUGH 30B

* * * * *

Method 30A--Determination of Total Vapor Phase Mercury Emissions From 
Stationary Sources (Instrumental Analyzer Procedure)

1.0 Scope and Application

What Is Method 30A?

    Method 30A is a procedure for measuring total vapor phase 
mercury (Hg) emissions from stationary sources using an instrumental 
analyzer. This method is particularly appropriate for performing 
emissions testing and for conducting relative accuracy test audits 
(RATAs) of mercury continuous emissions monitoring systems (Hg CEMS) 
and sorbent trap monitoring systems at coal-fired combustion 
sources. Quality assurance and quality control

[[Page 51502]]

requirements are included to assure that you, the tester, collect 
data of known and acceptable quality for each testing site. This 
method does not completely describe all equipment, supplies, and 
sampling procedures and analytical procedures you will need but 
refers to other test methods for some of the details. Therefore, to 
obtain reliable results, you should also have a thorough knowledge 
of these additional methods which are also found in appendices A-1 
and A-3 to this part:
    (a) Method 1--Sample and Velocity Traverses for Stationary 
Sources.
    (b) Method 4--Determination of Moisture Content in Stack Gases.
    1.1 Analytes. What does this method determine? This method is 
designed to measure the mass concentration of total vapor phase Hg 
in flue gas, which represents the sum of elemental Hg (Hg\0\) and 
oxidized forms of Hg (Hg+\2\), in mass concentration 
units of micrograms per cubic meter ([mu]g/m\3\).

------------------------------------------------------------------------
              Analyte                   CAS No.          Sensitivity
------------------------------------------------------------------------
Elemental Hg (Hg\0\)..............       7439-97-6  Typically < 2% of
                                                     Calibration Span.
Oxidized Hg (Hg+\2\)..............  ..............  (Same).
------------------------------------------------------------------------

    1.2 Applicability. When is this method required? Method 30A is 
offered as a reference method for emission testing and for RATAs of 
Hg CEMS and sorbent trap monitoring systems at coal-fired boilers. 
Method 30A may also be specified for other source categories in the 
future, either by New Source Performance Standards (NSPS), National 
Emission Standards for Hazardous Air Pollutants (NESHAP), emissions 
trading programs, State Implementation Plans (SIP), or operating 
permits that require measurement of Hg concentrations in stationary 
source emissions to determine compliance with an applicable emission 
standard or limit, or to conduct RATAs of Hg CEMS and sorbent trap 
monitoring systems.
    1.3 Data Quality Objectives (DQO). How good must my collected 
data be? Method 30A has been designed to provide data of high and 
known quality for Hg emission testing and for relative accuracy 
testing of Hg monitoring systems including Hg CEMS and sorbent trap 
monitoring systems. In these and other applications, the principle 
objective is to ensure the accuracy of the data at the actual 
emission levels encountered. To meet this objective, calibration 
standards prepared according to an EPA traceability protocol must be 
used and measurement system performance tests are required.

2.0 Summary of Method

    In this method, a sample of the effluent gas is continuously 
extracted and conveyed to an analyzer capable of measuring the total 
vapor phase Hg concentration. Elemental and oxidized mercury (i.e., 
Hg\0\ and Hg+\2\) may be measured separately or 
simultaneously but, for purposes of this method, total vapor phase 
Hg is the sum of Hg\0\ and Hg+\2\. You must meet the 
performance requirements of this method (i.e., system calibration, 
interference testing, dynamic spiking, and system integrity/drift 
checks) to validate your data. The dynamic spiking requirement is 
deferred until January 1, 2009.

3.0 Definitions

    3.1 Calibration Curve means the relationship between an 
analyzer's response to the injection of a series of calibration 
gases and the actual concentrations of those gases.
    3.2 Calibration Gas means a gas standard containing Hg\0\ or 
HgCl2 at a known concentration that is produced and 
certified in accordance with an EPA traceability protocol for 
certification of Hg calibration standards.
    3.2.1 Zero Gas means a calibration gas with a concentration that 
is below the level detectable by the measurement system.
    3.2.2 Low-Level Gas means a calibration gas with a concentration 
that is 10 to 30 percent of the calibration span.
    3.2.3 Mid-Level Gas means a calibration gas with a concentration 
that is 40 to 60 percent of the calibration span.
    3.2.4 High-Level Gas means a calibration gas whose concentration 
is equal to the calibration span.
    3.3 Converter means a device that reduces oxidized mercury 
(Hg+\2\) to elemental mercury (Hg\0\).
    3.4 Calibration Span means the upper limit of valid instrument 
response during sampling. To the extent practicable the measured 
emissions are to be between 10 and 100 percent of the selected 
calibration span (i.e., the measured emissions should be within the 
calibrated range determined by the Low- and High-Level gas 
standards). It is recommended that the calibration span be at least 
twice the native concentration to accommodate the dynamic spiking 
procedure.
    3.5 Centroidal Area means the central area that has the same 
shape as the stack or duct cross section and is no greater than one 
percent of the stack or duct total cross-sectional area.
    3.6 Data Recorder means the equipment that permanently records 
the concentrations reported by the analyzer.
    3.7 Drift Check means the test to determine the difference 
between the measurement system readings obtained in a post-run 
system integrity check and the prior pre-run system integrity check 
at a specific calibration gas concentration level (i.e., zero, mid-
level, or high-level).
    3.8 Dynamic Spiking means a procedure in which a known mass or 
concentration of vapor phase HgCl2 is injected into the 
probe sample gas stream at a known flow rate, in order to assess the 
effects of the flue gas matrix on the accuracy of the measurement 
system.
    3.9 Gas Analyzer means the equipment that detects the total 
vapor phase Hg being measured and generates an output proportional 
to its concentration.
    3.10 Interference Test means the test to detect analyzer 
responses to compounds other than Hg, usually gases present in the 
measured gas stream, that are not adequately accounted for in the 
calibration procedure and may cause measurement bias.
    3.11 Measurement System means all of the equipment used to 
determine the Hg concentration. The measurement system may generally 
include the following major subsystems: sample acquisition, 
Hg+2 to Hg0 converter, sample transport, 
sample conditioning, flow control/gas manifold, gas analyzer, and 
data recorder.
    3.12 Native Concentration means the total vapor phase Hg 
concentration in the effluent gas stream.
    3.13 NIST means the National Institute of Standards and 
Technology, located in Gaithersburg, Maryland.
    3.14 Response Time means the time it takes for the measurement 
system, while operating normally at its target sample flow rate or 
dilution ratio, to respond to a known step change in gas 
concentration (from a low-level to a high-level gas) and to read 
within 5 percent of the stable high-level gas response.
    3.15 Run means a series of gas samples taken successively from 
the stack or duct. A test normally consists of a specific number of 
runs.
    3.16 System Calibration Error means the difference between the 
measured concentration of a low-, mid-, or high-level Hg\0\ 
calibration gas and the certified concentration of the gas when it 
is introduced in system calibration mode.
    3.17 System Calibration Mode means introducing the calibration 
gases into the measurement system at the probe, upstream of all 
sample conditioning components.
    3.18 Test refers to the series of runs required by the 
applicable regulation.

4.0 Interferences

    Interferences will vary among instruments and potential 
instrument-specific spectral and matrix interferences must be 
evaluated through the interference test and the dynamic spiking 
tests.

5.0 Safety

What safety measures should I consider when using this method?

    This method may require you to work with hazardous materials and 
in hazardous conditions. You are encouraged to establish safety 
procedures before using the method. Among other precautions, you 
should become familiar with the safety recommendations in the gas 
analyzer user's manual. Occupational Safety and Health 
Administration (OSHA) regulations concerning use of compressed gas 
cylinders and noxious gases may apply.

[[Page 51503]]

6.0 Equipment and Supplies

    6.1 What do I need for the measurement system? This method is 
intended to be applicable to multiple instrumental technologies. You 
may use any equipment and supplies that meet the following 
specifications.
    6.1.1 All wetted sampling system components, including probe 
components prior to the point at which the calibration gas is 
introduced, must be chemically inert to all Hg species. Materials 
such as perfluoroalkoxy (PFA) Teflon\TM\, quartz, treated stainless 
steel (SS) are examples of such materials. [Note: These materials of 
construction are required because components prior to the 
calibration gas injection point are not included in the system 
calibration error, system integrity, and interference tests.]
    6.1.2 The interference, system calibration error, system 
integrity, drift and dynamic spiking test criteria must all be met 
by the system used.
    6.1.3 The system must be capable of measuring and controlling 
sample flow rate.
    6.1.4 All system components prior to the Hg+\2\ to 
Hg\0\ converter must be maintained at a sample temperature above the 
acid gas dew point.
    6.2 Measurement System Components. Figure 30A-1 in Section 17.0 
is an example schematic of a Method 30A measurement system.
    6.2.1 Sample Probe. The probe must be made of the appropriate 
materials as noted in Section 6.1.1, heated when necessary (see 
Section 6.1.4), configured with ports for introduction of 
calibration and spiking gases, and of sufficient length to traverse 
all of the sample points.
    6.2.2 Filter or Other Particulate Removal Device. The filter or 
other particulate removal device is considered to be a part of the 
measurement system, must be made of appropriate materials as noted 
in Section 6.1.1, and must be included in all system tests.
    6.2.3 Sample Line. The sample line that connects the probe to 
the converter, conditioning system and analyzer must be made of 
appropriate materials as noted in Section 6.1.1.
    6.2.4 Conditioning Equipment. For dry basis measurements, a 
condenser, dryer or other suitable device is required to remove 
moisture continuously from the sample gas. Any equipment needed to 
heat the probe, or sample line to avoid condensation prior to the 
moisture removal component is also required. For wet basis systems, 
you must keep the sample above its dew point either by: (1) Heating 
the sample line and all sample transport components up to the inlet 
of the analyzer (and, for hot-wet extractive systems, also heating 
the analyzer) or (2) by diluting the sample prior to analysis using 
a dilution probe system. The components required to do either of the 
above are considered to be conditioning equipment.
    6.2.5 Sampling Pump. A pump is needed to push or pull the sample 
gas through the system at a flow rate sufficient to minimize the 
response time of the measurement system. If a mechanical sample pump 
is used and its surfaces are in contact with the sample gas prior to 
detection, the pump must be leak free and must be constructed of a 
material that is non-reactive to the gas being sampled (see Section 
6.1.1). For dilution-type measurement systems, an ejector pump 
(eductor) may be used to create a sufficient vacuum that sample gas 
will be drawn through a critical orifice at a constant rate. The 
ejector pump may be constructed of any material that is non-reactive 
to the gas being sampled.
    6.2.6 Calibration Gas System(s). One or more systems may be 
needed to introduce calibration gases into the measurement system. A 
system should be able to flood the sampling probe sufficiently to 
prevent entry of gas from the effluent stream.
    6.2.7 Dynamic Spiking Port. For the purposes of the dynamic 
spiking procedure described in Section 8.2.7, the measurement system 
must be equipped with a port to allow introduction of the dynamic 
spike gas stream with the sample gas stream, at a point as close as 
possible to the inlet of the probe so as to ensure adequate mixing. 
The same port used for system calibrations and calibration error 
checks may be used for dynamic spiking purposes.
    6.2.8 Sample Gas Delivery. The sample line may feed directly to 
a converter, to a by-pass valve (for speciating systems), or to a 
sample manifold. All valve and/or manifold components must be made 
of material that is non-reactive to the gas sampled and the 
calibration gas, and must be configured to safely discharge any 
excess gas.
    6.2.9 Hg Analyzer. An instrument is required that continuously 
measures the total vapor phase Hg in the gas stream and meets the 
applicable specifications in Section 13.0.
    6.2.10 Data Recorder. A recorder, such as a computerized data 
acquisition and handling system (DAHS), digital recorder, strip 
chart, or data logger, is required for recording measurement data.
    6.3 Moisture Measurement System. If correction of the measured 
Hg emissions for moisture is required (see Section 8.5), either 
Method 4 in appendix A-3 to this part or other moisture measurement 
methods approved by the Administrator will be needed to measure 
stack gas moisture content.

7.0 Reagents and Standards

    7.1 Calibration Gases. What calibration gases do I need? You 
will need calibration gases of known concentrations of Hg\0\ and 
HgCl2. Special reagents and equipment may be required to 
prepare the HgCl\2\ gas standards (e.g., a NIST-traceable solution 
of HgCl2 and a gas generator equipped with mass flow 
controllers).
    The following calibration gas concentrations are required:
    7.1.1 High-Level Gas. Equal to the selected calibration span.
    7.1.2 Mid-Level Gas. 40 to 60 percent of the calibration span.
    7.1.3 Low-Level Gas. 10 to 30 percent of the calibration span.
    7.1.4 Zero Gas. No detectable Hg.
    7.1.5 Dynamic Spike Gas. The exact concentration of the 
HgCl2 calibration gas used to perform the pre-test 
dynamic spiking procedure described in Section 8.2.7 depends on the 
native Hg concentration in the stack The spike gas must produce a 
spiked sample concentration above the native concentration, as 
specified in Section 8.2.7.2.2.
    7.2 Interference Test. What reagents do I need for the 
interference test? Use the appropriate test gases listed in Table 
30A-3 in Section 17.0 (i.e., the potential interferents for the 
source to be tested, as identified by the instrument manufacturer) 
to conduct the interference check. These gases need not be of 
protocol gas quality.

8.0 Sample Collection

Emission Test Procedure

    Figure 30A-2 in Section 17.0 presents an overview of the test 
procedures required by this method. Since you may choose different 
options to comply with certain performance criteria, you must 
identify the specific options and associated frequencies you select 
and document your results in regard to the performance criteria.
    8.1 Sample Point Selection. What sampling site and sampling 
points do I select?
    8.1.1 When this method is used solely for Hg emission testing 
(e.g., to determine compliance with an emission standard or limit), 
use twelve sampling points located according to Table 1-1 or Table 
1-2 of Method 1 in appendix A-1 to this part. Alternatively, you may 
conduct a stratification test as described in Section 8.1.3 to 
determine the number and location of the sampling points.
    8.1.2 When this method is used for relative accuracy testing of 
a Hg CEMS or sorbent trap monitoring system, follow the sampling 
site selection and sampling point layout procedures for gas monitor 
RATA testing described in the appropriate performance specification 
or applicable regulation (e.g., Performance Specification 2, section 
8.1.3 of appendix B to this part or section 6.5.6 of appendix A to 
part 75 of this chapter), with one exception. If you elect to 
perform stratification testing as part of the sampling point 
selection process, perform the testing in accordance with Section 
8.1.3 of this method (see also ``Summary Table of QA/QC 
Requirements'' in Section 9.0).
    8.1.3 Determination of Stratification. If you elect to perform 
stratification testing as part of the sampling point selection 
process and the test results show your effluent gas stream to be 
unstratified or minimally stratified, you may be allowed to sample 
at fewer points or at different points than would otherwise be 
required.
    8.1.3.1 Test Procedure. To test for stratification, use a probe 
of appropriate length to measure the total vapor phase Hg 
concentration at twelve traverse points located according to Table 
1-1 or Table 1-2 of Method 1 in appendix A-1 to this part. 
Alternatively, for a sampling location where stratification is 
expected (e.g., after a wet scrubber or at a point where dissimilar 
gas streams are combined together), if a 12-point Hg stratification 
test has been previously performed at that location and the results 
of the test showed the location to be minimally stratified or 
unstratified according to the criteria in section 8.1.3.2, you may 
perform an abbreviated 3-point or 6-point Hg stratification test at 
the points specified in

[[Page 51504]]

section 6.5.6.2(a) of appendix A to part 75 of this chapter in lieu 
of performing the 12-point test. Sample for a minimum of twice the 
system response time (see Section 8.2.6) at each traverse point. 
Calculate the individual point and mean Hg concentrations.
    8.1.3.2 Acceptance Criteria and Sampling Point Location.
    8.1.3.2.1 If the Hg concentration at each traverse point differs 
from the mean concentration for all traverse points by no more than: 
(a) 5 percent of the mean concentration; or (b) 0.2 [mu]g/m\3\ (whichever is less restrictive), the gas stream 
is considered to be unstratified and you may collect samples from a 
single point that most closely matches the mean.
    8.1.3.2.2 If the 5 percent or 0.2 [mu]g/m\3\ criterion in 
Section 8.1.3.2.1 is not met, but the Hg concentration at each 
traverse point differs from the mean concentration for all traverse 
points by no more than: (a)10 percent of the mean; or 
(b)0.5 [mu]g/m\3\ (whichever is less restrictive), the 
gas stream is considered to be minimally stratified, and you may 
take samples from three points, provided the points are located on 
the measurement line exhibiting the highest average Hg concentration 
during the stratification test. If the stack diameter (or equivalent 
diameter, for a rectangular stack or duct) is greater than 2.4 
meters (7.8 ft), locate the three sampling points at 0.4, 1.0, and 
2.0 meters from the stack or duct wall. Alternatively, if a RATA 
required by part 75 of this chapter is being conducted, you may 
locate the three points at 4.4, 14.6, and 29.6 percent of the duct 
diameter, in accordance with Method 1 in appendix A-1 to this part. 
For stack or duct diameters of 2.4 meters (7.8 ft) or less, locate 
the three sampling points at 16.7, 50.0, and 83.3 percent of the 
measurement line.
    8.1.3.2.3 If the gas stream is found to be stratified because 
the 10 percent or 0.5 [mu]g/m\3\ criterion in Section 8.1.3.2.2 is 
not met, then either locate three sampling points at 16.7, 50.0, and 
83.3 percent of the measurement line that exhibited the highest 
average Hg concentration during the stratification test, or locate 
twelve traverse points for the test in accordance with Table 1-1 or 
Table 1-2 of Method 1 in appendix A-1 to this part; or, if a RATA 
required by part 75 of this chapter is being conducted, locate six 
Method 1 points along the measurement line that exhibited the 
highest average Hg concentration.
    8.1.3.3 Temporal Variations. Temporal variations in the source 
Hg concentration during a stratification test may complicate the 
determination of stratification. If temporal variations are a 
concern, you may use the following procedure to normalize the 
stratification test data. A second Hg measurement system, i.e., 
either an installed Hg CEMS or another Method 30A system, is 
required to perform this procedure. Position the sampling probe of 
the second Hg measurement system at a fixed point in the stack or 
duct, at least one meter from the stack or duct wall. Then, each 
time that the Hg concentration is measured at one of the 
stratification test points, make a concurrent measurement of Hg 
concentration at the fixed point. Normalize the Hg concentration 
measured at each traverse point, by multiplying it by the ratio of 
CF,avg to CF, where CF is the 
corresponding fixed-point Hg concentration measurement, and 
CF,avg is the average of all of the fixed-point 
measurements over the duration of the stratification test. Evaluate 
the results of the stratification test according to section 8.1.3.2, 
using the normalized Hg concentrations.
    8.1.3.4 Stratification Testing Exemption. Stratification testing 
need not be performed at a test location where it would otherwise be 
required to justify using fewer sample points or different sample 
points, if the owner or operator documents that the Hg concentration 
in the stack gas is expected to be 3 [mu]g/m\3\ or less at the time 
of a Hg monitoring system RATA or an Hg emissions test. To 
demonstrate that a particular test location qualifies for the 
stratification testing exemption, representative Hg emissions data 
must be collected just prior to the RATA or emissions test. At least 
one hour of Hg concentration data is required for the demonstration. 
The data used for the demonstration shall be recorded at process 
operating conditions that closely approximate the operating 
conditions that will exist during the RATA or emissions test. It is 
recommended that collection of the demonstration data be integrated 
with the on-site pretest procedures required by the reference method 
being used for the RATA or emissions test (whether this method or 
another approved Hg reference method is used). Quality-assured data 
from an installed Hg monitoring system may also be used for the 
demonstration. If a particular test location qualifies for the 
stratification testing exemption, sampling shall be performed at 
three points, as described in section 8.1.3.2.2 of this method. The 
owner or operator shall fully document the method used to collect 
the demonstration data and shall keep this documentation on file 
with the data from the associated RATA or Hg emissions test.
    8.1.3.5 Interim Alternative Stratification Test Procedures. In 
the time period between the effective date of this method and 
January 1, 2009, you may follow one of the following two procedures. 
Substitute a stratification test for sulfur dioxide (SO2) 
for the Hg stratification test described in section 8.1.3.1. If this 
option is chosen, follow the test procedures in section 6.5.6.1 of 
appendix A to part 75 of this chapter. Evaluate the test results and 
determine the sampling point locations according to section 6.5.6.3 
of appendix A to part 75 of this chapter. If the sampling location 
is found to be minimally stratified or unstratified for 
SO2, it shall be considered minimally stratified or 
unstratified for Hg. Alternatively, you may forgo stratification 
testing, assume the gas stream is minimally stratified, and sample 
at three points as described in section 8.1.3.2.2 of this method.
    8.2 Initial Measurement System Performance Tests. What initial 
performance criteria must my system meet before I begin sampling? 
Before measuring emissions, perform the following procedures:
    (a) Interference Test;
    (b) Calibration Gas Verification;
    (c) Measurement System Preparation;
    (d) 3-Point System Calibration Error Test;
    (e) System Integrity Check;
    (f) Measurement System Response Time Test; and
    (g) Dynamic Spiking Test.
    8.2.1 Interference Test (Optional). Your measurement system 
should be free of known interferences. It is recommended that you 
conduct this interference test of your measurement system prior to 
its initial use in the field to verify that the candidate test 
instrument is free from inherent biases or interferences resulting 
from common combustion emission constituents. If you have multiple 
measurement systems with components of the same make and model 
numbers, you need only perform this interference check on one system 
and you may also rely on an interference test conducted by the 
manufacturer on a system having components of the same make and 
model(s) of the system that you use. The interference test procedure 
is found in Section 8.6 of this method.
    8.2.2 Calibration Gas Verification. How must I verify the 
concentrations of my calibration gases?
    8.2.2.1 Cylinder Gas Standards. When cylinder gas standards are 
used for Hg0, obtain a certificate from the gas 
manufacturer and confirm that the documentation includes all 
information required by an EPA traceability protocol (see Section 
16). Confirm that the manufacturer certification is complete and 
current. Ensure that the calibration gas certifications have not 
expired.
    8.2.2.2 Other Calibration Standards. All other calibration 
standards for HgCl2 and Hg0, such as gas 
generators, must meet the requirements of an EPA traceability 
protocol (see Section 16), and the certification procedures must be 
fully documented in the test report.
    8.2.2.3 Calibration Span. Select the calibration span (i.e., 
high-level gas concentration) so that the measured source emissions 
are 10 to 100 percent of the calibration span. This requirement is 
waived for applications in which the Hg concentrations are 
consistently below 1 [mu]g/m\3\; however, the calibration span for 
these low-concentration applications shall not exceed 5 [mu]g/m\3\.
    8.2.3 Measurement System Preparation. How do I prepare my 
measurement system for use? Assemble, prepare, and precondition the 
measurement system according to your standard operating procedure. 
Adjust the system to achieve the correct sampling rate or dilution 
ratio (as applicable). Then, conduct a 3-point system calibration 
error test using Hg0 as described in Section 8.2.4, an 
initial system integrity check using HgCl2 and a zero gas 
as described in Section 8.2.5, and a pre-test dynamic spiking test 
as described in Section 8.2.7.
    8.2.4 System Calibration Error Test. Conduct a 3-point system 
calibration error test before the first test run. Use Hg\0\ 
standards for this test. Introduce the low-, mid-, and high-level 
calibration gases in any order, in system calibration mode, unless 
you desire to determine the system response time during this test, 
in which case, inject the gases such that the high-level injection

[[Page 51505]]

directly follows the low-level injection. For non-dilution systems, 
you may adjust the system to maintain the correct flow rate at the 
analyzer during the test, but you may not make adjustments for any 
other purpose. For dilution systems, you must operate the 
measurement system at the appropriate dilution ratio during all 
system calibration error checks, and you may make only the 
adjustments necessary to maintain the proper ratio. After each gas 
injection, wait until a stable response has been obtained. Record 
the analyzer's final, stable response to each calibration gas on a 
form similar to Table 30A-1 in Section 17.0. For each calibration 
gas, calculate the system calibration error using Equation 30A-1 in 
Section 12.2. The calibration error specification in Section 13.1 
must be met for the low-, mid-, and high-level gases. If the 
calibration error specification is not met for all three gases, take 
corrective action and repeat the test until an acceptable 3-point 
calibration is achieved.
    8.2.5 System Integrity Check. Perform a two-point system 
integrity check before the first test run. Use the zero gas and 
either the mid- or high-level HgCl2 calibration gas for 
the check, whichever one best represents the total vapor phase Hg 
concentration levels in the stack. Record the data on a form similar 
to Table 30A-2 in Section 17.0. The system integrity check 
specification in Section 13.2 must be met for both the zero gas and 
the mid- or high-level gas. If the system integrity specification is 
not met for both gases, take corrective action and repeat the test 
until an acceptable system integrity check is achieved.
    8.2.6 Measurement System Response Time. The measurement system 
response time is used to determine the minimum sampling time for 
each sampling point and is equal to the time that is required for 
the measured Hg concentration to increase from the stable low-level 
calibration gas response to a value within 5 percent of the stable 
high-level calibration gas response during the system calibration 
error test in Section 8.2.4. Round off the measured system response 
time to the nearest minute.
    8.2.7 Dynamic Spiking Test. You must perform dynamic spiking 
prior to the first test run to validate your test data. The purpose 
of this procedure is to demonstrate that the site-specific flue gas 
matrix does not adversely affect the accuracy of the measurement 
system. The specifications in Section 13.5 must be met to validate 
your data. If these specifications are not met for the pre-test 
dynamic spiking, you may not proceed with the test until 
satisfactory results are obtained. For the time period between the 
effective date of this method and January 1, 2009, the dynamic 
spiking requirement is waived.
    8.2.7.1 How do I perform dynamic spiking? Dynamic spiking is a 
gas phase application of the method of standard additions, which 
involves injecting a known quantity of Hg into the measurement 
system upstream of all sample conditioning components, similar to 
system calibration mode, except the probe is not flooded and the 
resulting sample stream includes both effluent gas and the spike 
gas. You must follow a written procedure that details how the spike 
is added to the system, how the spike dilution factor (DF) is 
measured, and how the Hg concentration data are collected and 
processed.
    8.2.7.2 Spiking Procedure Requirements.
    8.2.7.2.1 Spiking Gas Requirements. The spike gas must also be a 
HgCl2 calibration gas certified by an EPA traceability 
protocol. You must choose concentrations that can produce the target 
levels while being injected at a volumetric flow rate that is < =20 
percent of the total volumetric flow rate through the measurement 
system (i.e., sample flow rate plus spike gas flow rate).
    8.2.7.2.2 Target Spiking Level. The target level for spiking 
must be 150 to 200 percent of the native Hg concentration; however, 
if the native Hg concentration is < 1 [mu]g/m\3\, set the target 
level to add between 1 and 4 [mu]g/m\3\ Hg+\2\ to the 
native concentration. Use Equation 30A-5 in Section 12.5 to 
calculate the acceptable range of spike gas concentrations at the 
target level. Then select a spike gas concentration in that range.
    8.2.7.2.3 Spike Injections. You must inject spikes in such a 
manner that the spiking does not alter the total volumetric sample 
system flow rate and dilution ratio (if applicable). You must 
collect at least 3 data points, and the relative standard deviation 
(RSD) specification in Section 13.5 must be met. Each data point 
represents a single spike injection, and pre- and post-injection 
measurements of the native Hg concentration (or diluted native 
concentration, as applicable) are required for each spike injection.
    8.2.7.2.4 Spike Dilution Factor (DF). For each spike injection, 
DF, the dilution factor must be determined. DF is the ratio of the 
total volumetric flow rate of gas through the measurement system to 
the spike gas flow rate. This factor must be >=5. The spiking mass 
balance calculation is directly dependent on the accuracy of the DF 
determination. As a result, high accuracy total volumetric flow rate 
and spike gas flowrate measurements are required. These flow rates 
may be determined by direct or indirect measurement. Calibrated flow 
meters, venturies, orifices or tracer gas measurements are examples 
of potential flow measurement techniques.
    8.2.7.2.5 Concentrations. The measurement system must record 
total vapor phase Hg concentrations continuously during the dynamic 
spiking procedure. It is possible that dynamic spiking at a level 
close to 200 percent of the native Hg concentration may cause the 
measured Hg concentration to exceed the calibration span value. 
Avoid this by choosing a lower spiking level or by recalibration at 
a higher span. The measurements shall not exceed 120 percent of the 
calibration span. The ``baseline'' measurements made between spikes 
may represent the native Hg concentration (if spike gas flow is 
stopped between injections) or the native Hg concentration diluted 
by blank or carrier gas flowing at the same rate as the spike gas 
(if gas flow cannot be stopped between injections). Each baseline 
measurement must include at least 4 readings or 1 minute (whichever 
is greater) of stable responses. Use Equation 30A-10 or 30A-11 in 
Section 12.10 (as applicable) to convert baseline measurements to 
native concentration.
    8.2.7.2.6 Recovery. Calculate spike recoveries using Equation 
30A-7 in Section 12.7. Mass recoveries may be calculated from stable 
responses based on injected mass flows or from integrated response 
peaks based on total mass injected. Calculate the mean and RSD for 
the three (or more) spike injections and compare to the 
specifications in Section 13.5. 
    8.2.7.2.7 Error Adjustment Option. You may adjust the 
measurement data collected during dynamic spiking for the system 
calibration error using Equation 30A-3 in Section 12. To do this, 
perform the initial system integrity check prior to the dynamic 
spiking test, and perform another system integrity check following 
the dynamic spiking test and before the first test run. If you 
choose this option, you must apply Equation 30A-3 to both the spiked 
sample concentration measurement (Css) and the baseline 
or native concentration measurement (Cnative), each 
substituted in place of Cavg in the equation.
    8.2.7.3 Example Spiking Procedure Using a Hot Vapor Calibration 
Source Generator.
    (a) Introduce the spike gas into the probe using a hot vapor 
calibration source generator and a solution of HgCl2 in 
dilute HC1 and HNO3. The calibrator uses a mass flow 
controller (accurate within 2 percent) to measure the gas flow, and 
the solution feed is measured using a top-loading balance accurate 
to 0.01g. The challenges of injecting oxidized Hg may make it 
impractical to stop the flow of gas between spike injections. In 
this case, operate the hot vapor calibration source generator 
continuously during the spiking procedure, swapping blank solutions 
for HgCl2 solutions when switching between spiking and 
baseline measurements.
    (b) If applicable, monitor the measurement system to make sure 
the total sampling system flow rate and the sample dilution ratio do 
not change during this procedure. Record all data on a data sheet 
similar to Table 30A-5 in Section 17.0. If the Hg measurement system 
design makes it impractical to measure the total volumetric flow 
rate through the system, use a spike gas that includes a tracer for 
measuring the dilution factor, DF (see Equation 30A-9 in Section 
12.9). Allow the measurements to stabilize between each spike 
injection, average the pre- and post-injection baseline 
measurements, and calculate the native concentration. If this 
measurement shifts by more than 5 percent during any injection, it 
may be necessary to discard that data point and repeat the injection 
to achieve the required RSD among the injections. If the spikes 
persistently show poor repeatability, or if the recoveries are not 
within the range specified in Section 13.5, take corrective action.
    8.2.8 Run Validation. How do I confirm that each run I conduct 
is valid?
    8.2.8.1 System Integrity Checks.
    (a) Before and after each test run, perform a two-point system 
integrity check using the same procedure as the initial system 
integrity check described in Section 8.2.5. You may use data from 
that initial system integrity

[[Page 51506]]

check as the pre-run data for the first test run, provided it is the 
most recent system integrity check done before the first run. You 
may also use the results of a successful post-run system integrity 
check as the pre-run data for the next test run. Do not make any 
adjustments to the measurement system during these checks, other 
than to maintain the target calibration gas flow rate and the proper 
dilution ratio.
    (b) As a time-saving alternative, you may, at the risk of 
invalidating multiple test runs, skip one or more integrity checks 
during a test day. Provided there have been no auto-calibrations or 
other instrument alterations, a single integrity check may suffice 
as a post-run check to validate (or invalidate) as many consecutive 
test runs as can be completed during a single test day. All 
subsequent test days must begin with a pre-run system integrity 
check subject to the same performance criteria and corrective action 
requirements as a post-run system integrity check.
    (c) Each system integrity check must meet the criteria for 
system integrity checks in Section 13.2. If a post-run system 
integrity check is failed, all test runs since the last passed 
system integrity check are invalid. If a post-run or a pre-run 
system integrity check is failed, you must take corrective action 
and pass another 3-point Hg\0\ system calibration error test 
(Section 8.2.4) followed by another system integrity check before 
conducting any additional test runs. Record the results of the pre- 
and post-run system integrity checks on a form similar to Table 30A-
2 in Section 17.0.
    8.2.8.2 Drift Check. Using the data from the successful pre- and 
post-run system integrity checks, calculate the zero and upscale 
drift, using Equation 30A-2 in Section 12.3. Exceeding the Section 
13.3 specification does not invalidate the run, but corrective 
action must be taken and a new 3-point Hg\0\ system calibration 
error test and a system integrity check must be passed before any 
more runs are made.
    8.3 Dilution-Type Systems--Special Considerations. When a 
dilution-type measurement system is used, there are three important 
considerations that must be taken into account to ensure the quality 
of the emissions data. First, the critical orifice size and dilution 
ratio must be selected properly so that the sample dew point will be 
below the sample line and analyzer temperatures. Second, a high-
quality, accurate dilution controller must be used to maintain the 
correct dilution ratio during sampling. The dilution controller 
should be capable of monitoring the dilution air pressure, orifice 
upstream pressure, eductor vacuum, and sample flow rates. Third, 
differences between the molecular weight of calibration gas 
mixtures, dilution air, and the stack gas molecular weight must be 
considered because these can affect the dilution ratio and introduce 
measurement bias.
    8.4 Sampling.
    (a) Position the probe at the first sampling point. Allow the 
system to flush and equilibrate for at least two times the 
measurement system response time before recording any data. Then, 
traverse and record measurements at all required sampling points. 
Sample at each traverse point for an equal length of time, 
maintaining the appropriate sample flow rate or dilution ratio (as 
applicable). For all Hg instrumental method systems, the minimum 
sampling time at each sampling point must be at least two times the 
system response time, but not less than 10 minutes. For 
concentrating systems, the minimum sampling time must also include 
at least 4 concentration measurement cycles.
    (b) After recording data for the appropriate period of time at 
the first traverse point, you may move the sample probe to the next 
point and continue recording, omitting the requirement to allow the 
system to equilibrate for two times the system response time before 
recording data at the subsequent traverse points. You must, however, 
sample at this and all subsequent traverse points for the required 
minimum amount of time specified in this section. If you must remove 
the probe from the stack for any reason, you must again allow the 
sampling system to equilibrate for at least two times the system 
response time prior to resuming data recording.
    (c) If at any point the measured Hg concentration exceeds the 
calibration span value, you must at a minimum identify and report 
this as a deviation from the method. Depending on the data quality 
objectives of the test, this event may require corrective action 
before proceeding. If the average Hg concentration for any run 
exceeds the calibration span value, the run is invalidated.
    8.5 Moisture Correction. If the moisture basis (wet or dry) of 
the measurements made with this method is different from the 
moisture basis of either: (1) The applicable emission limit; or (2) 
a Hg CEMS or sorbent trap monitoring system being evaluated for 
relative accuracy, you must determine the moisture content of the 
flue gas and correct the measured gas concentrations to a dry basis 
using Method 4 in appendix A-3 of this part or other appropriate 
methods, subject to the approval of the Administrator.
    8.6 Optional Interference Test Procedure.
    (a) Select an appropriate calibration span that reflects the 
source(s) to be tested and perform the interference check at 40 
percent of the lowest calibration span value anticipated, e.g., 10 
[mu]g/m\3\. Alternatively, successfully conducting the interference 
test at an absolute Hg concentration of 2 [mu]g/m\3\ will 
demonstrate performance for an equivalent calibration span of 5 
[mu]g/m\3\, the lowest calibration span allowed for Method 30A 
testing. Therefore, performing the interference test at the 2 [mu]/
m\3\ level will serve to demonstrate acceptable performance for all 
calibration spans greater than or equal to 5 [mu]g/m\3\.
    (b) Introduce the interference test gases listed in Table 30A-3 
in Section 17.0 into the measurement system separately or as a 
mixture. The interference test gases HCl and NO must be introduced 
as a mixture. The interference test gases must be introduced into 
the sampling system at the probe such that the interference gas 
mixtures pass through all filters, scrubbers, conditioners, and 
other components as would be configured for normal sampling.
    (c) The interference test must be performed using 
HgCl2, and each interference test gas (or gas mixture) 
must be evaluated in triplicate. This is accomplished by measuring 
the Hg response first with only the HgCl2 gas present and 
then when adding the interference test gas(es) while maintaining the 
HgCl2 concentration of the test stream constant. It is 
important that the equipment used to conduct the interference test 
be of sufficient quality so as to be capable of blending the 
HgCl2 and interference gases while maintaining the Hg 
concentration constant. Gas blending system or manifolds may be 
used.
    (d) The duration of each test should be for a sufficient period 
of time to ensure the Hg measurement system surfaces are conditioned 
and a stable output is obtained. Measure the Hg response of the 
analyzer to these gases in [mu]g/m3. Record the responses and 
determine the overall interference response using Table 30A-4 in 
Section 17.0 and the equations presented in Section 12.11. The 
specification in Section 13.4 must be met.
    (e) A copy of these data, including the date completed and a 
signed certification, must be included with each test report. The 
intent of this test is that the interference test results are 
intended to be valid for the life of the system. As a result, the Hg 
measurement system should be operated and tested in a configuration 
consistent with the configuration that will be used for field 
applications. However, if the system used for field testing is not 
consistent with the system that was interference-tested, the 
interference test must be repeated before it is used for any field 
applications. Examples of such conditions include, but are not 
limited to: major changes in dilution ratio (for dilution based 
systems), changes in catalyst materials, changes in filtering device 
design or materials, changes in probe design or configuration, and 
changes in gas conditioning materials or approaches.

9.0 Quality Control

What quality control measures must I take?

    The table which follows is a summary of the mandatory, 
suggested, and alternative quality assurance and quality control 
measures and the associated frequency and acceptance criteria. All 
of the QC data, along with the run data, must be documented and 
included in the test report.

[[Page 51507]]



                                       Summary Table of QA/QC Requirements
----------------------------------------------------------------------------------------------------------------
     Status \1\       Process or element    QA/QC specification     Acceptance criteria      Checking frequency
----------------------------------------------------------------------------------------------------------------
S..................  Identify Data User..  ....................  Regulatory Agency or       Before designing
                                                                  other primary end user     test.
                                                                  of data.
M..................  Analyzer Design.....  Analyzer range......  Sufficiently > high-level  ....................
                                                                  gas to allow
                                                                  determination of system
                                                                  calibration error.
S..................  ....................  Analyzer resolution   <  2.0 % of full-scale      Manufacturer design.
                                            or sensitivity.       range.
S..................  ....................  Interference          Overall response < = 3% of
                                            response.             calibration span.
                                                                 Alternatively, overall
                                                                  response < = 0.3 [mu]g/
                                                                  m\3\.
M..................  Calibration Gases...  Traceability          Validation of
                                            protocol.             concentration required.
M..................  ....................  High-level Hg\0\ gas  Equal to the calibration   Each calibration
                                                                  span.                      error test.
M..................  ....................  Mid-level Hg\0\ gas.  40 to 60% of calibration   Each calibration
                                                                  span.                      error test.
M..................  ....................  Low-level Hg\0\ gas.  10 to 30% of calibration   Each calibration
                                                                  span.                      error test.
M..................  ....................  High-level HgCl2 gas  Equal to the calibration   Each system
                                                                  span.                      integrity check (if
                                                                                             it better
                                                                                             represents Cnative
                                                                                             than the mid level
                                                                                             gas).
M..................  ....................  Mid-level HgCl2.....  40 to 60% of calibration   Each system gas
                                                                  span.                      integrity check (if
                                                                                             it better
                                                                                             represents Cnative
                                                                                             than the high level
                                                                                             gas).
M..................  ....................  Zero gas............  .........................  Each system
                                                                                             integrity check.
M..................  ....................  Dynamic spike gas     A high-concentration       Pre-test; dynamic
                                            (Cnative >= 1 [mu]g/  HgCl2 gas, used to         spiking not
                                            m\3\).                produce a spiked sample    required until 1/1/
                                                                  concentration that is      09.
                                                                  150 to 200% of the
                                                                  native concentration.
M..................  ....................  Dynamic spike gas     A high-concentration       Pre-test; dynamic
                                            (Cnative <  1 [mu]g/   HgCl2 gas, used to         spiking not
                                            m\3\).                produce a spiked sample    required until 1/1/
                                                                  concentration that is 1    09.
                                                                  to 2 [mu]g/m\3\ above
                                                                  the native concentration.
S..................  Data Recorder Design  Data resolution.....  < = 0.5% of full-scale....  Manufacturer design.
M..................  Sample Extraction...  Probe material......  Inert to sample            Each run.
                                                                  constituents (e.g., PFA
                                                                  Teflon, or quartz if
                                                                  stack > 500 [deg]F).
M..................  Sample Extraction...  Probe, filter and     For dry-basis analyzers,   Each run.
                                            sample line           keep sample above the
                                            temperature.          dew point, by heating
                                                                  prior to moisture
                                                                  removal.
                                                                 For wet-basis analyzers,
                                                                  keep sample above dew
                                                                  point at all times, by
                                                                  heating or dilution.
M..................  Sample Extraction...  Calibration valve     Inert to sample            Each test.
                                            material.             constituents (e.g., PFA
                                                                  Teflon or PFA Teflon
                                                                  coated).
S..................  Sample Extraction...  Sample pump material  Inert to sample            Each test.
                                                                  constituents.
M..................  Sample Extraction...  Manifold material...  Inert to sample            Each test.
                                                                  constituents.
M..................  Particulate Removal.  Filter inertness....  Pass calibration error     Each calibration
                                                                  check.                     error check.
M..................  System Calibration    System calibration    CE < = 5.0 % of the         Before initial run
                      Performance.          error (CE) test.      calibration span for the   and after a failed
                                                                  low-, mid-or high-level    system integrity
                                                                  Hg\0\ calibration gas.     check or drift
                                                                 Alternative                 test.
                                                                  specification: < = 0.5
                                                                  [mu]g/m\3\ absolute
                                                                  difference between
                                                                  system response and
                                                                  reference value.
M..................  System Calibration    System integrity      Error < = 5.0% of the       Before initial run,
                      Performance.          check.                calibration span for the   after each run, at
                                                                  zero and mid- or high-     the beginning of
                                                                  level HgCl2 calibration    subsequent test
                                                                  gas.                       days, and after a
                                                                 Alternative                 failed system
                                                                  specification: < = 0.5      integrity check or
                                                                  [mu]g/m\3\ absolute        drift test.
                                                                  difference between
                                                                  system response and
                                                                  reference value.
M..................  System Performance..  System response time  Used to determine minimum  During initial 3-
                                                                  sampling time per point.   point system
                                                                                             calibration error
                                                                                             test.
M..................  System Performance..  Drift...............  < = 3.0% of calibration     At least once per
                                                                  span for the zero and      test day.
                                                                  mid- or high-level gas.
                                                                 Alternative
                                                                  specification: < = 0.3
                                                                  [mu]g/m\3\ absolute
                                                                  difference between pre-
                                                                  and post-run system
                                                                  calibration error
                                                                  percentages..
M..................  System Performance..  Minimum sampling      The greater of two times   Each sampling point.
                                            time.                 the system response time
                                                                  or 10 minutes.
                                                                  Concentrating systems
                                                                  must also include at
                                                                  least 4 cycles.
M..................  System Performance..  Percentage spike      Percentage spike           Before initial
                                            recovery and          recovery, at the target    dynamic spiking not
                                            relative standard     level: 100     required until 1/1/
                                            deviation.            10%.                       09.
                                                                 Relative standard
                                                                  deviation: < = 5 percent.
                                                                 Alternative
                                                                  specification: absolute
                                                                  difference between
                                                                  calculated and measured
                                                                  spike values < = 0.5
                                                                  [mu]g/m\3\.

[[Page 51508]]


M..................  Sample Point          Number and Location   For emission testing       Prior to first run.
                      Selection.            of Sample Points.     applications, use 12
                                                                  points, located
                                                                  according to Method 1 in
                                                                  appendix A-1 to this
                                                                  part, unless the results
                                                                  of a stratification test
                                                                  allow fewer points to be
                                                                  used.
                     ....................  ....................  For Part 60 RATAs, follow
                                                                  the procedures in
                                                                  Performance
                                                                  Specification 2, section
                                                                  8.1.3, and for Part 75
                                                                  RATAs, follow the
                                                                  procedures in section
                                                                  6.5.6 of appendix A to
                                                                  Part 75. That is:
                     ....................  ....................   At any test
                                                                  location, you may use 3
                                                                  sample points located at
                                                                  16.7, 50.0, and 83.3% of
                                                                  a ``long'' measurement
                                                                  line passing through the
                                                                  centroidal area; or
                     ....................  ....................   At any test
                                                                  location, you may use 6
                                                                  sample points along a
                                                                  diameter, located
                                                                  according to Method 1
                                                                  (Part 75 RATAs, only);
                                                                  or
                     ....................  ....................   At a location
                                                                  where stratification is
                                                                  not expected and the
                                                                  measurement line is >
                                                                  2.4 m (7.8 ft), you may
                                                                  use 3 sample points
                                                                  located along a
                                                                  ``short'' measurement
                                                                  line at 0.4, 1.0, and
                                                                  2.0 m from the stack or
                                                                  duct wall or, for Part
                                                                  75 only, sample points
                                                                  may be located at 4.4,
                                                                  14.6, and 29.6% of the
                                                                  measurement line; or
                     ....................  ....................   After a wet
                                                                  scrubber or at a point
                                                                  where dissimilar gas
                                                                  streams are combined,
                                                                  either locate 3 sample
                                                                  points along the
                                                                  ``long'' measurement
                                                                  line or locate 6 Method
                                                                  1 points along a
                                                                  diameter (Part 75,
                                                                  only), unless the
                                                                  results of a
                                                                  stratification test
                                                                  allow you to use a
                                                                  ``short'' 3-point
                                                                  measurement line or to
                                                                  sample at a single point.
                     ....................  ....................   If it can be
                                                                  demonstrated that stack
                                                                  gas concentration is < =
                                                                  3 [mu]g/m\3\, then the
                                                                  test site is exempted
                                                                  from stratification
                                                                  testing. Use the 3-point
                                                                  ``short'' measurement
                                                                  line if the stack
                                                                  diameter is > 2.4 m (7.8
                                                                  ft) and the 3-point
                                                                  ``long'' line for stack
                                                                  diameters < = 2.4 m (7.8
                                                                  ft).
A..................  Sample Point          Stratification Test   If the Hg concentration    Prior to first run.
                      Selection.            (see Section 8.1.3).  \2\ at each traverse
                                                                  point during the