[Code of Federal Regulations]
[Title 21, Volume 8]
[Revised as of April 1, 2001]
From the U.S. Government Printing Office via GPO Access
[CITE: 21CFR1020.32]

[Page 595-598]
 
                        TITLE 21--FOOD AND DRUGS
 
CHAPTER I--FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN 
                          SERVICES--(Continued)
 
PART 1020--PERFORMANCE STANDARDS FOR IONIZING RADIATION EMITTING PRODUCTS--Table of Contents
 
Sec. 1020.32  Fluoroscopic equipment.

    The provisions of this section apply to equipment for fluoroscopy 
and for the recording of images through an image intensifier except 
computed tomography x-ray systems manufactured on or after November 29, 
1984.
    (a) Primary protective barrier--(1) Limitation of useful beam. The 
fluoroscopic imaging assembly shall be provided with a primary 
protective barrier which intercepts the entire cross section of the 
useful beam at any SID. The x-ray tube used for fluoroscopy shall not 
produce x-rays unless the barrier is in position to intercept the entire 
useful beam. The exposure rate due to transmission through the barrier 
with the attenuation block in the useful beam combined with radiation 
from the image intensifier if provided, shall not exceed 
3.34 x 10-3 percent of the entrance exposure rate, at a 
distance of 10 centimeters from any accessible surface of the 
fluoroscopic imaging assembly beyond the plane of the image receptor. 
Radiation therapy simulation systems shall be exempt from this 
requirement provided the systems are intended only for remote control 
operation and the manufacturer sets forth instructions for assemblers 
with respect to control location as part of the information required in 
Sec. 1020.30(g). Additionally, the manufacturer shall provide to users, 
pursuant to Sec. 1020.30(h)(1)(i), precautions concerning the importance 
of remote control operation.
    (2) Measuring compliance. The entrance exposure rate shall be 
measured in accordance with paragraph (d) of this section. The exposure 
rate due to transmission through the primary barrier combined with 
radiation from the image intensifier shall be determined by measurements 
averaged over an area of 100 square centimeters with no linear dimension 
greater than 20 centimeters. If the source is below the tabletop, the 
measurement shall be made with the input surface of the fluoroscopic 
imaging assembly positioned 30 centimeters above the tabletop. If the 
source is above the tabletop and the SID is variable, the measurement 
shall be made with the end of the beam-limiting device or spacer as 
close to the tabletop as it can be placed, provided that it shall not be 
closer than 30 centimeters. Movable grids and compression devices shall 
be removed from the useful beam during the measurement. For all 
measurements, the attenuation block shall be positioned in the useful 
beam 10 centimeters from the point of measurement of entrance exposure 
rate and between this point and the input surface of the fluoroscopic 
imaging assembly.
    (b) Field limitation--(1) Nonimage-intensified fluoroscopy. (i) The 
x-ray field produced by nonimage-intensified fluoroscopic equipment 
shall not extend beyond the entire visible area of the image receptor. 
Means shall be provided for stepless adjustment of the field size. The 
minimum field size, at the greatest SID, shall be containable in a 
square of 5 centimeters by 5 centimeters.
    (ii) For equipment manufactured after February 25, 1978, when the 
angle between the image receptor and the beam axis of the x-ray beam is 
variable, means shall be provided to indicate when the axis of the x-ray 
beam is perpendicular to the plane of the image receptor. Compliance 
with paragraph (b)(1)(i) of this section shall be determined with the 
beam axis indicated to be perpendicular to the plane of the image 
receptor.
    (2) Image-intensified fluoroscopy. (i) For image-intensified 
fluoroscopic equipment other than radiation therapy simulation systems, 
neither the length nor the width of the x-ray field in the plane of the 
image receptor shall exceed that of the visible area of the image 
receptor by more than 3 percent of the SID. The sum of the excess length 
and the excess width shall be no greater than 4 percent of the SID.
    (ii) For rectangular x-ray fields used with circular image 
receptors, the error in alignment shall be determined along the length 
and width dimensions of the x-ray field which pass through the center of 
the visible area of the image receptor.
    (iii) For equipment manufactured after February 25, 1978, when the 
angle

[[Page 596]]

between the image receptor and beam axis is variable, means shall be 
provided to indicate when the axis of the x-ray beam is perpendicular to 
the plane of the image receptor. Compliance with paragraph (b)(2)(i) of 
this section shall be determined with the beam axis indicated to be 
perpendicular to the plane of the image receptor.
    (iv) Means shall be provided to permit further limitation of the 
field. Beam-limiting devices manufactured after May 22, 1979, and 
incorporated in equipment with a variable SID and/or the capability of a 
visible area of greater than 300 square centimeters shall be provided 
with means for stepless adjustment of the x-ray field. Equipment with a 
fixed SID and the capability of a visible area of no greater than 300 
square centimeters shall be provided with either stepless adjustment of 
the x-ray field or with a means to further limit the x-ray field size at 
the plane of the image receptor to 125 square centimeters or less. 
Stepless adjustment shall, at the greatest SID, provide continuous field 
sizes from the maximum obtainable to a field size containable in a 
square of 5 centimeters by 5 centimeters.
    (3) If the fluoroscopic x-ray field size is adjusted automatically 
as the SID or image receptor size is changed, a capability may be 
provided for overriding the automatic adjustment in case of system 
failure. If it is so provided, a signal visible at the fluoroscopist's 
position shall indicate whenever the automatic field adjustment is 
overridden. Each such system failure override switch shall be clearly 
labeled as follows:

                For X-ray Field Limitation System Failure

    (c) Activation of tube. X-ray production in the fluoroscopic mode 
shall be controlled by a device which requires continuous pressure by 
the operator for the entire time of any exposure. When recording serial 
fluoroscopic images, the operator shall be able to terminate the x-ray 
exposure(s) at any time, but means may be provided to permit completion 
of any single exposure of the series in process.
    (d) Entrance exposure rates. For fluoroscopic equipment manufactured 
before May 19, 1995, the following requirements apply:
    (1) Equipment with automatic exposure rate control (AERC). 
Fluoroscopic equipment that is provided with AERC shall not be operable 
at any combination of tube potential and current that will result in an 
exposure rate in excess of 2.58x10-3 coulomb per kilogram (C/
kg) per minute (10 roentgens per minute (10 R/min)) at the point where 
the center of the useful beam enters the patient, except:
    (i) During recording of fluoroscopic images, or
    (ii) When an optional high-level control is provided. When so 
provided, the equipment shall not be operable at any combination of tube 
potential and current that will result in an exposure rate in excess of 
1.29x10-3 C/kg per minute (5 R/min) at the point where the 
center of the useful beam enters the patient, unless the high-level 
control is activated. Special means of activation of high-level controls 
shall be required. The high-level control shall be operable only when 
continuous manual activation is provided by the operator. A continuous 
signal audible to the fluoroscopist shall indicate that the high-level 
control is being employed.
    (2) Equipment without AERC (manual mode). Fluoroscopic equipment 
that is not provided with AERC shall not be operable at any combination 
of tube potential and current that will result in an exposure rate in 
excess of 1.29x10-3 C/kg per minute (5 R/min) at the point 
where the center of the useful beam enters the patient, except:
    (i) During recording of fluoroscopic images, or
    (ii) When an optional high-level control is activated. Special means 
of activation of high-level controls shall be required. The high-level 
control shall be operable only when continuous manual activation is 
provided by the operator. A continuous signal audible to the 
fluoroscopist shall indicate that the high-level control is being 
employed.
    (3) Equipment with both an AERC mode and a manual mode. Fluoroscopic 
equipment that is provided with both an

[[Page 597]]

AERC mode and a manual mode shall not be operable at any combination of 
tube potential and current that will result in an exposure rate in 
excess of 2.58x10-3 C/kg per minute (10 R/min) in either mode 
at the point where the center of the useful beam enters the patient 
except:
    (i) During recording of fluoroscopic images, or
    (ii) When the mode or modes have an optional high-level control, in 
which case that mode or modes shall not be operable at any combination 
of tube potential and current that will result in an exposure rate in 
excess of 1.29x10-3 C/kg per minute (5 R/min) at the point 
where the center of the useful beam enters the patient, unless the high-
level control is activated. Special means of activation of high-level 
controls shall be required. The high-level control shall be operable 
only when continuous manual activation is provided by the operator. A 
continuous signal audible to the fluoroscopist shall indicate that the 
high-level is being employed.
    (4) Measuring compliance. Compliance with paragraph (d) of this 
section shall be determined as follows:
    (i) If the source is below the x-ray table, the exposure rate shall 
be measured at 1 centimeter above the tabletop or cradle.
    (ii) If the source is above the x-ray table, the exposure rate shall 
be measured at 30 centimeters above the tabletop with the end of the 
beam-limiting device or spacer positioned as closely as possible to the 
point of measurement.
    (iii) In a C-arm type of fluoroscope, the exposure rate shall be 
measured at 30 centimeters from the input surface of the fluoroscopic 
imaging assembly, with the source positioned at any available SID, 
provided that the end of the beam-limiting device or spacer is no closer 
than 30 centimeters from the input surface of the imaging assembly.
    (iv) In a lateral type of fluoroscope, the exposure rate shall be 
measured at a point 15 centimeters from the centerline of the x-ray 
table and in the direction of the x-ray source with the end of the beam-
limiting device or spacer positioned as closely as possible to the point 
of measurement. If the tabletop is movable, it shall be positioned as 
closely as possible to the lateral x-ray source, with the end of the 
beam-limiting device or spacer no closer than 15 centimeters to the 
centerline of the x-ray table.
    (5) Exemptions. Fluoroscopic radiation therapy simulation systems 
are exempt from the requirements set forth in paragraph (d) of this 
section.
    (e) Entrance exposure rate limits. For fluoroscopic equipment 
manufactured on and after May 19, 1995, the following requirements 
apply:
    (1) Fluoroscopic equipment operable at any combination of tube 
potential and current that results in an exposure rate greater than 
1.29x10-3 C/kg per minute (5 R/min) at the point where the 
center of the useful beam enters the patient shall be equipped with 
AERC. Provision for manual selection of technique factors may be 
provided.
    (2) Fluoroscopic equipment shall not be operable at any combination 
of tube potential and current that will result in an exposure rate in 
excess of 2.58x10-3 C/kg per minute (10 R/min) at the point 
where the center of the useful beam enters the patient except:
    (i) During the recording of images from an x-ray image-intensifier 
tube using photographic film or a video camera when the x-ray source is 
operated in a pulsed mode.
    (ii) When an optional high-level control is activated. When the 
high-level control is activated, the equipment shall not be operable at 
any combination of tube potential and current that will result in an 
exposure rate in excess of 5.16x10-3 C/kg per minute (20 R/
min) at the point where the center of the useful beam enters the 
patient. Special means of activation of high-level controls shall be 
required. The high-level control shall only be operable when continuous 
manual activation is provided by the operator. A continuous signal 
audible to the fluoroscopist shall indicate that the high-level control 
is being employed.
    (3) Measuring compliance. Compliance with paragraph (e) of this 
section shall be determined as follows:
    (i) If the source is below the x-ray table, the exposure rate shall 
be measured at 1 centimeter above the tabletop or cradle.

[[Page 598]]

    (ii) If the source is above the x-ray table, the exposure rate shall 
be measured at 30 centimeters above the tabletop with the end of the 
beam-limiting device or spacer positioned as closely as possible to the 
point of measurement.
    (iii) In a C-arm type of fluoroscope, the exposure rate shall be 
measured at 30 centimeters from the input surface of the fluoroscopic 
imaging assembly, with the source positioned at any available SID, 
provided that the end of the beam-limiting device or spacer is no closer 
than 30 centimeters from the input surface of the fluoroscopic imaging 
assembly.
    (iv) In a lateral type of fluoroscope, the exposure rate shall be 
measured at a point 15 centimeters from the centerline of the x-ray 
table and in the direction of the x-ray source with the end of the beam-
limiting device or spacer positioned as closely as possible to the point 
of measurement. If the tabletop is movable, it shall be positioned as 
closely as possible to the lateral x-ray source, with the end of the 
beam-limiting device or spacer no closer than 15 centimeters to the 
centerline of the x-ray table.
    (4) Exemptions. Fluoroscopic radiation therapy simulation systems 
are exempt from the requirements set forth in paragraph (e) of this 
section.
    (f) Indication of potential and current. During fluoroscopy and 
cinefluorography, x-ray tube potential and current shall be continuously 
indicated. Deviation of x-ray tube potential and current from the 
indicated values shall not exceed the maximum deviation as stated by the 
manufacturer in accordance with Sec. 1020.30(h)(3).
    (g) Source-skin distance. Means shall be provided to limit the 
source-skin distance to not less than 38 centimeters on stationary 
fluoroscopes and to not less than 30 centimeters on mobile and portable 
fluoroscopes. In addition, for image-intensified fluoroscopes intended 
for specific surgical application that would be prohibited at the 
source-skin distances specified in this paragraph, provisions may be 
made for operation at shorter source-skin distances but in no case less 
than 20 centimeters. When provided, the manufacturer must set forth 
precautions with respect to the optional means of spacing, in addition 
to other information as required in Sec. 1020.30(h).
    (h) Fluoroscopic timer. Means shall be provided to preset the 
cumulative on-time of the fluoroscopic tube. The maximum cumulative time 
of the timing device shall not exceed 5 minutes without resetting. A 
signal audible to the fluoroscopist shall indicate the completion of any 
preset cumulative on-time. Such signal shall continue to sound while x-
rays are produced until the timing device is reset. As an alternative to 
the requirements of this paragraph, radiation therapy simulation systems 
may be provided with a means to indicate the total cumulative exposure 
time during which x-rays were produced, and which is capable of being 
reset between x-ray examinations.
    (i) Mobile and portable fluoroscopes. In addition to the foregoing 
requirements of this section, mobile and portable fluoroscopes shall 
provide intensified imaging.

[58 FR 26404, May 3, 1993, as amended at 59 FR 26404, May 19, 1994]