                        TITLE 21--FOOD AND DRUGS
 
CHAPTER I--FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN 
                          SERVICES (CONTINUED)
 
PART 806--MEDICAL DEVICES; REPORTS OF CORRECTIONS AND REMOVALS--Table of Contents
 
                     Subpart B--Reports and Records
 
Sec. 806.10  Reports of corrections and removals.


    (a) Each device manufacturer or importer shall submit a written 
report to FDA of any correction or removal of a device initiated by such 
manufacturer or importer if the correction or removal was initiated:
    (1) To reduce a risk to health posed by the device; or
    (2) To remedy a violation of the act caused by the device which may 
present a risk to health unless the information has already been 
provided as set forth in paragraph (f) of this section or the corrective 
or removal action is exempt from the reporting requirements under 
Sec. 806.1(b).
    (b) The manufacturer or importer shall submit any report required by 
paragraph (a) of this section within 10-working days of initiating such 
correction or removal.
    (c) The manufacturer or importer shall include the following 
information in the report:
    (1) The seven digit registration number of the entity responsible 
for submission of the report of corrective or removal action (if 
applicable), the month, day, and year that the report is made, and a 
sequence number (i.e., 001 for the first report, 002 for the second 
report, 003 etc.), and the report type designation ``C'' or ``R''. For 
example, the complete number for the first correction report submitted 
on June 1, 1997, will appear as follows for a firm with the registration 
number 1234567: 1234567-6/1/97-001-C. The second correction report 
number submitted by the same firm on July 1, 1997, would be 1234567-7/1/
97-002-C etc. For removals, the number will appear as follows: 1234567-
6/1/97-001-R and 1234567-7/1/97-002-R, etc. Firms that do not have a 
seven digit registration number may use seven zeros followed by the 
month, date, year, and sequence number (i.e. 0000000-6/1/97-001-C for 
corrections and 0000000-7/1/97-001-R for removals). Reports received 
without a seven digit

[[Page 57]]

registration number will be assigned a seven digit central file number 
by the district office reviewing the reports.
    (2) The name, address, and telephone number of the manufacturer or 
importer, and the name, title, address, and telephone number of the 
manufacturer or importer representative responsible for conducting the 
device correction or removal.
    (3) The brand name and the common name, classification name, or 
usual name of the device and the intended use of the device.
    (4) Marketing status of the device, i.e., any applicable premarket 
notification number, premarket approval number, or indication that the 
device is a preamendments device, and the device listing number. A 
manufacturer or importer that does not have an FDA establishment 
registration number shall indicate in the report whether it has ever 
registered with FDA.
    (5) The model, catalog, or code number of the device and the 
manufacturing lot or serial number of the device or other identification 
number.
    (6) The manufacturer's name, address, telephone number, and contact 
person if different from that of the person submitting the report.
    (7) A description of the event(s) giving rise to the information 
reported and the corrective or removal actions that have been, and are 
expected to be taken.
    (8) Any illness or injuries that have occurred with use of the 
device. If applicable, include the medical device report numbers.
    (9) The total number of devices manufactured or distributed subject 
to the correction or removal and the number in the same batch, lot, or 
equivalent unit of production subject to the correction or removal.
    (10) The date of manufacture or distribution and the device's 
expiration date or expected life.
    (11) The names, addresses, and telephone numbers of all domestic and 
foreign consignees of the device and the dates and number of devices 
distributed to each such consignee.
    (12) A copy of all communications regarding the correction or 
removal and the names and addresses of all recipients of the 
communications not provided in accordance with paragraph (c)(11) of this 
section.
    (13) If any required information is not immediately available, a 
statement as to why it is not available and when it will be submitted.
    (d) If, after submitting a report under this part, a manufacturer or 
importer determines that the same correction or removal should be 
extended to additional lots or batches of the same device, the 
manufacturer or importer shall within 10-working days of initiating the 
extension of the correction or removal, amend the report by submitting 
an amendment citing the original report number assigned according to 
paragraph (c)(1) of this section, all of the information required by 
paragraph (c)(2), and any information required by paragraphs (c)(3) 
through (c)(12) of this section that is different from the information 
submitted in the original report. The manufacturer or importer shall 
also provide a statement in accordance with paragraph (c)(13) of this 
section for any required information that is not readily available.
    (e) A report submitted by a manufacturer or importer under this 
section (and any release by FDA of that report or information) does not 
necessarily reflect a conclusion by the manufacturer, importer, or FDA 
that the report or information constitutes an admission that the device 
caused or contributed to a death or serious injury. A manufacturer or 
importer need not admit, and may deny, that the report or information 
submitted under this section constitutes an admission that the device 
caused or contributed to a death or serious injury.
    (f) No report of a correction or removal is required under this 
part, if a report of the correction or removal is required and has been 
submitted under parts 803, 804, or 1004 of this chapter.

[62 FR 27191, May 19, 1997, as amended at 63 FR 42232, Aug. 7, 1998]