                        TITLE 21--FOOD AND DRUGS
 
CHAPTER I--FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN 
                          SERVICES (CONTINUED)
 
PART 814--PREMARKET APPROVAL OF MEDICAL DEVICES--Table of Contents
 
             Subpart B--Premarket Approval Application (PMA)
 
Sec. 814.20  Application.


    (a) The applicant or an authorized representative shall sign the 
PMA. If the applicant does not reside or have a place of business within 
the United States, the PMA shall be countersigned by an authorized 
representative residing or maintaining a place of business in the United 
States and shall identify the representative's name and address.
    (b) Unless the applicant justifies an omission in accordance with 
paragraph (d) of this section, a PMA shall include:
    (1) The name and address of the applicant.
    (2) A table of contents that specifies the volume and page number 
for each item referred to in the table. A PMA shall include separate 
sections on nonclinical laboratory studies and on clinical 
investigations involving human subjects. A PMA shall be submitted in six 
copies each bound in one or more numbered volumes of reasonable size. 
The applicant shall include information that it believes to be trade 
secret or confidential commercial or financial information in all copies 
of the PMA

[[Page 120]]

and identify in at least one copy the information that it believes to be 
trade secret or confidential commercial or financial information.
    (3) A summary in sufficient detail that the reader may gain a 
general understanding of the data and information in the application. 
The summary shall contain the following information:
    (i) Indications for use. A general description of the disease or 
condition the device will diagnose, treat, prevent, cure, or mitigate, 
including a description of the patient population for which the device 
is intended.
    (ii) Device description. An explanation of how the device functions, 
the basic scientific concepts that form the basis for the device, and 
the significant physical and performance characteristics of the device. 
A brief description of the manufacturing process should be included if 
it will significantly enhance the reader's understanding of the device. 
The generic name of the device as well as any proprietary name or trade 
name should be included.
    (iii) Alternative practices and procedures. A description of 
existing alternative practices or procedures for diagnosing, treating, 
preventing, curing, or mitigating the disease or condition for which the 
device is intended.
    (iv) Marketing history. A brief description of the foreign and U.S. 
marketing history, if any, of the device, including a list of all 
countries in which the device has been marketed and a list of all 
countries in which the device has been withdrawn from marketing for any 
reason related to the safety or effectiveness of the device. The 
description shall include the history of the marketing of the device by 
the applicant and, if known, the history of the marketing of the device 
by any other person.
    (v) Summary of studies. An abstract of any information or report 
described in the PMA under paragraph (b)(8)(ii) of this section and a 
summary of the results of technical data submitted under paragraph 
(b)(6) of this section. Such summary shall include a description of the 
objective of the study, a description of the experimental design of the 
study, a brief description of how the data were collected and analyzed, 
and a brief description of the results, whether positive, negative, or 
inconclusive. This section shall include the following:
    (A) A summary of the nonclinical laboratory studies submitted in the 
application;
    (B) A summary of the clinical investigations involving human 
subjects submitted in the application including a discussion of subject 
selection and exclusion criteria, study population, study period, safety 
and effectiveness data, adverse reactions and complications, patient 
discontinuation, patient complaints, device failures and replacements, 
results of statistical analyses of the clinical investigations, 
contraindications and precautions for use of the device, and other 
information from the clinical investigations as appropriate (any 
investigation conducted under an IDE shall be identified as such).
    (vi) Conclusions drawn from the studies. A discussion demonstrating 
that the data and information in the application constitute valid 
scientific evidence within the meaning of Sec. 860.7 and provide 
reasonable assurance that the device is safe and effective for its 
intended use. A concluding discussion shall present benefit and risk 
considerations related to the device including a discussion of any 
adverse effects of the device on health and any proposed additional 
studies or surveillance the applicant intends to conduct following 
approval of the PMA.
    (4) A complete description of:
    (i) The device, including pictorial representations;
    (ii) Each of the functional components or ingredients of the device 
if the device consists of more than one physical component or 
ingredient;
    (iii) The properties of the device relevant to the diagnosis, 
treatment, prevention, cure, or mitigation of a disease or condition;
    (iv) The principles of operation of the device; and
    (v) The methods used in, and the facilities and controls used for, 
the manufacture, processing, packing, storage, and, where appropriate, 
installation of the device, in sufficient detail so that a person 
generally familiar with current good manufacturing practice can make

[[Page 121]]

a knowledgeable judgment about the quality control used in the 
manufacture of the device.
    (5) Reference to any performance standard under section 514 of the 
act or the Radiation Control for Health and Safety Act of 1968 (42 
U.S.C. 263b et seq.) in effect or proposed at the time of the submission 
and to any voluntary standard that is relevant to any aspect of the 
safety or effectiveness of the device and that is known to or that 
should reasonably be known to the applicant. The applicant shall--
    (i) Provide adequate information to demonstrate how the device 
meets, or justify any deviation from, any performance standard 
established under section 514 of the act or under the Radiation Control 
for Health and Safety Act, and
    (ii) Explain any deviation from a voluntary standard.
    (6) The following technical sections which shall contain data and 
information in sufficient detail to permit FDA to determine whether to 
approve or deny approval of the application:
    (i) A section containing results of the nonclinical laboratory 
studies with the device including microbiological, toxicological, 
immunological, biocompatibility, stress, wear, shelf life, and other 
laboratory or animal tests as appropriate. Information on nonclinical 
laboratory studies shall include a statement that each such study was 
conducted in compliance with part 58, or, if the study was not conducted 
in compliance with such regulations, a brief statement of the reason for 
the noncompliance.
    (ii) A section containing results of the clinical investigations 
involving human subjects with the device including clinical protocols, 
number of investigators and subjects per investigator, subject selection 
and exclusion criteria, study population, study period, safety and 
effectiveness data, adverse reactions and complications, patient 
discontinuation, patient complaints, device failures and replacements, 
tabulations of data from all individual subject report forms and copies 
of such forms for each subject who died during a clinical investigation 
or who did not complete the investigation, results of statistical 
analyses of the clinical investigations, device failures and 
replacements, contraindications and precautions for use of the device, 
and any other appropriate information from the clinical investigations. 
Any investigation conducted under an IDE shall be identified as such. 
Information on clinical investigations involving human subjects shall 
include the following:
    (A) A statement with respect to each study that it either was 
conducted in compliance with the institutional review board regulations 
in part 56, or was not subject to the regulations under Sec. 56.104 or 
Sec. 56.105, and that it was conducted in compliance with the informed 
consent regulations in part 50; or if the study was not conducted in 
compliance with those regulations, a brief statement of the reason for 
the noncompliance.
    (B) A statement that each study was conducted in compliance with 
part 812 or part 813 concerning sponsors of clinical investigations and 
clinical investigators, or if the study was not conducted in compliance 
with those regulations, a brief statement of the reason for the 
noncompliance.
    (7) For a PMA supported solely by data from one investigation, a 
justification showing that data and other information from a single 
investigator are sufficient to demonstrate the safety and effectiveness 
of the device and to ensure reproducibility of test results.
    (8)(i) A bibliography of all published reports not submitted under 
paragraph (b)(6) of this section, whether adverse or supportive, known 
to or that should reasonably be known to the applicant and that concern 
the safety or effectiveness of the device.
    (ii) An identification, discussion, and analysis of any other data, 
information, or report relevant to an evaluation of the safety and 
effectiveness of the device known to or that should reasonably be known 
to the applicant from any source, foreign or domestic, including 
information derived from investigations other than those proposed in the 
application and from commercial marketing experience.
    (iii) Copies of such published reports or unpublished information in 
the possession of or reasonably obtainable by

[[Page 122]]

the applicant if an FDA advisory committee or FDA requests.
    (9) One or more samples of the device and its components, if 
requested by FDA. If it is impractical to submit a requested sample of 
the device, the applicant shall name the location at which FDA may 
examine and test one or more devices.
    (10) Copies of all proposed labeling for the device. Such labeling 
may include, e.g., instructions for installation and any information, 
literature, or advertising that constitutes labeling under section 
201(m) of the act.
    (11) An environmental assessment under Sec. 25.20(n) prepared in the 
applicable format in Sec. 25.40, unless the action qualifies for 
exclusion under Sec. 25.30 or Sec. 25.34. If the applicant believes that 
the action qualifies for exclusion, the PMA shall under Sec. 25.15(a) 
and (d) provide information that establishes to FDA's satisfaction that 
the action requested is included within the excluded category and meets 
the criteria for the applicable exclusion.
    (12) A financial certification or disclosure statement or both as 
required by part 54 of this chapter.
    (13) Such other information as FDA may request. If necessary, FDA 
will obtain the concurrence of the appropriate FDA advisory committee 
before requesting additional information.
    (c) Pertinent information in FDA files specifically referred to by 
an applicant may be incorporated into a PMA by reference. Information in 
a master file or other information submitted to FDA by a person other 
than the applicant will not be considered part of a PMA unless such 
reference is authorized in writing by the person who submitted the 
information or the master file. If a master file is not referenced 
within 5 years after the date that it is submitted to FDA, FDA will 
return the master file to the person who submitted it.
    (d) If the applicant believes that certain information required 
under paragraph (b) of this section to be in a PMA is not applicable to 
the device that is the subject of the PMA, and omits any such 
information from its PMA, the applicant shall submit a statement that 
identifies the omitted information and justifies the omission. The 
statement shall be submitted as a separate section in the PMA and 
identified in the table of contents. If the justification for the 
omission is not accepted by the agency, FDA will so notify the 
applicant.
    (e) The applicant shall periodically update its pending application 
with new safety and effectiveness information learned about the device 
from ongoing or completed studies that may reasonably affect an 
evaluation of the safety or effectiveness of the device or that may 
reasonably affect the statement of contraindications, warnings, 
precautions, and adverse reactions in the draft labeling. The update 
report shall be consistent with the data reporting provisions of the 
protocol. The applicant shall submit three copies of any update report 
and shall include in the report the number assigned by FDA to the PMA. 
These updates are considered to be amendments to the PMA. The time frame 
for review of a PMA will not be extended due to the submission of an 
update report unless the update is a major amendment under 
Sec. 814.37(c)(1). The applicant shall submit these reports--
    (1) 3 months after the filing date,
    (2) Following receipt of an approvable letter, and
    (3) At any other time as requested by FDA.
    (f) If a color additive subject to section 706 of the act is used in 
or on the device and has not previously been listed for such use, then, 
in lieu of submitting a color additive petition under part 71, at the 
option of the applicant, the information required to be submitted under 
part 71 may be submitted as part of the PMA. When submitted as part of 
the PMA, the information shall be submitted in three copies each bound 
in one or more numbered volumes of reasonable size. A PMA for a device 
that contains a color additive that is subject to section 706 of the act 
will not be approved until the color additive is listed for use in or on 
the device.
    (g) FDA has issued a PMA guidance document to assist the applicant 
in the arrangement and content of a PMA. This guidance document is 
available on the Internet at http://www.fda.gov/cdrh/

[[Page 123]]

dsma/pmaman/front.html. This guidance document is also available upon 
request from the Center for Devices and Radiological Health, Division of 
Small Manufacturers Assistance (HFZ-220), 1350 Piccard Dr., Rockville, 
MD 20850, FAX 301-443-8818.
    (h) If you are sending a PMA, PMA amendment, PMA supplement, or 
correspondence with respect to a PMA, you must send it to the Document 
Mail Center (HFZ-401), Center for Devices and Radiological Health, Food 