[Code of Federal Regulations]
[Title 40, Volume 19]
[Revised as of July 1, 2002]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR141.23]

[Page 353-363]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 141--NATIONAL PRIMARY DRINKING WATER REGULATIONS--Table of Contents
 
            Subpart C--Monitoring and Analytical Requirements
 
Sec. 141.23  Inorganic chemical sampling and analytical requirements.

    Community water systems shall conduct monitoring to determine 
compliance with the maximum contaminant levels specified in Sec. 141.62 
in accordance with this section. Non-transient, non-community water 
systems shall conduct monitoring to determine compliance with the 
maximum contaminant levels specified in Sec. 141.62 in accordance with 
this section. Transient, non-community water systems shall conduct 
monitoring to determine compliance with the nitrate and nitrite maximum 
contaminant levels in Secs. 141.11 and 141.62 (as appropriate) in 
accordance with this section.
    (a) Monitoring shall be conducted as follows:
    (1) Groundwater systems shall take a minimum of one sample at every 
entry point to the distribution system which is representative of each 
well after treatment (hereafter called a sampling point) beginning in 
the initial compliance period. The system shall take each sample at the 
same sampling point unless conditions make another sampling point more 
representative of each source or treatment plant.
    (2) Surface water systems shall take a minimum of one sample at 
every entry point to the distribution system after any application of 
treatment or in the distribution system at a point which is 
representative of each source after treatment (hereafter called a 
sampling point) beginning in the initial compliance period. The system 
shall take each sample at the same sampling point unless conditions make 
another sampling point more representative of each source or treatment 
plant.
    Note: For purposes of this paragraph, surface water systems include 
systems with a combination of surface and ground sources.
    (3) If a system draws water from more than one source and the 
sources are combined before distribution, the system must sample at an 
entry point to the distribution system during periods of normal 
operating conditions (i.e., when water is representative of all sources 
being used).
    (4) The State may reduce the total number of samples which must be 
analyzed by allowing the use of compositing. Composite samples from a 
maximum of five samples are allowed, provided that the detection limit 
of the method used for analysis is less than one-fifth of the MCL. 
Compositing of samples must be done in the laboratory.
    (i) If the concentration in the composite sample is greater than or 
equal to one-fifth of the MCL of any inorganic chemical, then a follow-
up sample must be taken within 14 days at each sampling point included 
in the composite. These samples must be analyzed for the contaminants 
which exceeded one-fifth of the MCL in the composite sample. Detection 
limits for each analytical method and MCLs for each inorganic 
contaminant are the following:

[[Page 354]]



               Detection Limits for Inorganic Contaminants
------------------------------------------------------------------------
                                                             Detection
        Contaminant          MCL (mg/l)     Methodology     limit (mg/l)
------------------------------------------------------------------------
Antimony..................  0.006......  Atomic            0.003
                                          Absorption;
                                          Furnace.
                                         Atomic            0.0008 \5\
                                          Absorption;
                                          Platform.
                                         ICP-Mass          0.0004
                                          Spectrometry.
                                         Hydride-Atomic    0.001
                                          Absorption.
Arsenic...................  \6\ 0.01...  Atomic            0.001
                                          Absorption;
                                          Furnace.
                                         Atomic            \7\ 0.0005
                                          Absorption;
                                          Platform--
                                          Stabilized
                                          Temperature.
                                         Atomic            0.001
                                          Absorption;
                                          Gaseous Hydride.
                                         ICP-Mass          \8\ 0.0014
                                          Spectrometry.
Asbestos..................  7 MFL \1\..  Transmission      0.01 MFL
                                          Electron
                                          Microscopy.
Barium....................  2..........  Atomic            0.002
                                          Absorption;
                                          furnace
                                          technique.
                                         Atomic            0.1
                                          Absorption;
                                          direct
                                          aspiration.
                                         Inductively       0.002 (0.001)
                                          Coupled Plasma.
Beryllium.................  0.004......  Atomic            0.0002
                                          Absorption;
                                          Furnace.
                                         Atomic            0.00002 \5\
                                          Absorption;
                                          Platform.
                                         Inductively       0.0003
                                          Coupled Plasma
                                          \2\.
                                         ICP-Mass          0.0003
                                          Spectrometry.
Cadmium...................  0.005......  Atomic            0.0001
                                          Absorption;
                                          furnace
                                          technique.
                                         Inductively       0.001
                                          Coupled Plasma.
Chromium..................  0.1........  Atomic            0.001
                                          Absorption;
                                          furnace
                                          technique.
                                         Inductively       0.007 (0.001)
                                          Coupled Plasma.
Cyanide...................  0.2........  Distillation,     0.02
                                          Spectrophotomet
                                          ric \3\.
                                         Distillation,     0.005
                                          Automated,
                                          Spectrophotomet
                                          ric \3\.
                                         Distillation,     0.05
                                          Selective
                                          Electrode \3\.
                                         Distillation,     0.02
                                          Amenable,
                                          Spectrophotomet
                                          ric \4\.
Mercury...................  0.002......  Manual Cold       0.0002
                                          Vapor Technique.
                                         Automated Cold    0.0002
                                          Vapor Technique.
Nickel....................  xl.........  Atomic            0.001
                                          Absorption;
                                          Furnace.
                                         Atomic            0.0006 \5\
                                          Absorption;
                                          Platform.
                                         Inductively       0.005
                                          Coupled Plasma
                                          \2\.
                                         ICP-Mass          0.0005
                                          Spectrometry.
Nitrate...................  10 (as N)..  Manual Cadmium    0.01
                                          Reduction.
                                         Automated         0.01
                                          Hydrazine
                                          Reduction.
                                         Automated         0.05
                                          Cadmium
                                          Reduction.
                                         Ion Selective     1
                                          Electrode.
                                         Ion               0.01
                                          Chromatography.
Nitrite...................  1 (as N)...  Spectrophotometr  0.01
                                          ic.
                                         Automated         0.05
                                          Cadmium
                                          Reduction.
                                         Manual Cadmium    0.01
                                          Reduction.
                                         Ion               0.004
                                          Chromatography.
Selenium..................  0.05.......  Atomic            0.002
                                          Absorption;
                                          furnace.
                                         Atomic            0.002
                                          Absorption;
                                          gaseous hydride.
Thallium..................  0.002......  Atomic            0.001
                                          Absorption;
                                          Furnace.
                                         Atomic            0.0007 \5\
                                          Absorption;
                                          Platform.
                                         ICP-Mass          0.0003
                                          Spectrometry.
------------------------------------------------------------------------
\1\ MFL = million fibers per liter 10 [mu]m.
\2\ Using a 2X preconcentration step as noted in Method 200.7. Lower
  MDLs may be achieved when using a 4X preconcentration.
\3\ Screening method for total cyanides.
\4\ Measures ``free'' cyanides.
\5\ Lower MDLs are reported using stabilized temperature graphite
  furnace atomic absorption.
\6\ The value for arsenic is effective January 23, 2006. Unit then, the
  MCL is 0.05 mg/L.
\7\ The MDL reported for EPA method 200.9 (Atomic Absorption; Platform--
  Stablized Temperature) was determined using a 2x concentration step
  during sample digestion. The MDL determined for samples analyzed using
  direct analyses (i.e., no sample digestion) will be higher. Using
  multiple depositions, EPA 200.9 is capable of obtaining MDL of 0.0001
  mg/L.
\8\ Using selective ion monitoring, EPA Method 200.8 (ICP-MS) is capable
  of obtaining a MDL of 0.0001 mg/L.

    (ii) If the population served by the system is 3,300 
persons, then compositing may only be permitted by the State at sampling 
points within a single system. In systems serving [le]3,300 persons, the 
State may permit compositing among different systems provided the 5-
sample limit is maintained.
    (iii) If duplicates of the original sample taken from each sampling 
point used in the composite sample are available, the system may use 
these instead of resampling. The duplicates must be analyzed and the 
results reported to the State within 14 days after completing analysis 
of the composite sample, provided the holding time of the sample is not 
exceeded.
    (5) The frequency of monitoring for asbestos shall be in accordance 
with

[[Page 355]]

paragraph (b) of this section: the frequency of monitoring for antimony, 
arsenic, barium, beryllium, cadmium, chromium, cyanide, fluoride, 
mercury, nickel, selenium and thallium shall be in accordance with 
paragraph (c) of this section; the frequency of monitoring for nitrate 
shall be in accordance with paragraph (d) of this section; and the 
frequency of monitoring for nitrite shall be in accordance with 
paragraph (e) of this section.
    (b) The frequency of monitoring conducted to determine compliance 
with the maximum contaminant level for asbestos specified in 
Sec. 141.62(b) shall be conducted as follows:
    (1) Each community and non-transient, non-community water system is 
required to monitor for asbestos during the first three-year compliance 
period of each nine-year compliance cycle beginning in the compliance 
period starting January 1, 1993.
    (2) If the system believes it is not vulnerable to either asbestos 
contamination in its source water or due to corrosion of asbestos-cement 
pipe, or both, it may apply to the State for a waiver of the monitoring 
requirement in paragraph (b)(1) of this section. If the State grants the 
waiver, the system is not required to monitor.
    (3) The State may grant a waiver based on a consideration of the 
following factors:
    (i) Potential asbestos contamination of the water source, and
    (ii) The use of asbestos-cement pipe for finished water distribution 
and the corrosive nature of the water.
    (4) A waiver remains in effect until the completion of the three-
year compliance period. Systems not receiving a waiver must monitor in 
accordance with the provisions of paragraph (b)(1) of this section.
    (5) A system vulnerable to asbestos contamination due solely to 
corrosion of asbestos-cement pipe shall take one sample at a tap served 
by asbestos-cement pipe and under conditions where asbestos 
contamination is most likely to occur.
    (6) A system vulnerable to asbestos contamination due solely to 
source water shall monitor in accordance with the provision of paragraph 
(a) of this section.
    (7) A system vulnerable to asbestos contamination due both to its 
source water supply and corrosion of asbestos-cement pipe shall take one 
sample at a tap served by asbestos-cement pipe and under conditions 
where asbestos contamination is most likely to occur.
    (8) A system which exceeds the maximum contaminant levels as 
determined in Sec. 141.23(i) of this section shall monitor quarterly 
beginning in the next quarter after the violation occurred.
    (9) The State may decrease the quarterly monitoring requirement to 
the frequency specified in paragraph (b)(1) of this section provided the 
State has determined that the system is reliably and consistently below 
the maximum contaminant level. In no case can a State make this 
determination unless a groundwater system takes a minimum of two 
quarterly samples and a surface (or combined surface/ground) water 
system takes a minimum of four quarterly samples.
    (10) If monitoring data collected after January 1, 1990 are 
generally consistent with the requirements of Sec. 141.23(b), then the 
State may allow systems to use that data to satisfy the monitoring 
requirement for the initial compliance period beginning January 1, 1993.
    (c) The frequency of monitoring conducted to determine compliance 
with the maximum contaminant levels in Sec. 141.62 for antimony, 
arsenic, barium, beryllium, cadmium, chromium, cyanide, fluoride, 
mercury, nickel, selenium and thallium shall be as follows:
    (1) Groundwater systems shall take one sample at each sampling point 
during each compliance period. Surface water systems (or combined 
surface/ground) shall take one sample annually at each sampling point.
    (2) The system may apply to the State for a waiver from the 
monitoring frequencies specified in paragraph (c)(1) of this section. 
States may grant a public water system a waiver for monitoring of 
cyanide, provided that the State determines that the system is not 
vulnerable due to lack of any industrial source of cyanide.
    (3) A condition of the waiver shall require that a system shall take 
a minimum of one sample while the waiver is

[[Page 356]]

effective. The term during which the waiver is effective shall not 
exceed one compliance cycle (i.e., nine years).
    (4) The State may grant a waiver provided surface water systems have 
monitored annually for at least three years and groundwater systems have 
conducted a minimum of three rounds of monitoring. (At least one sample 
shall have been taken since January 1, 1990). Both surface and 
groundwater systems shall demonstrate that all previous analytical 
results were less than the maximum contaminant level. Systems that use a 
new water source are not eligible for a waiver until three rounds of 
monitoring from the new source have been completed.
    (5) In determining the appropriate reduced monitoring frequency, the 
State shall consider:
    (i) Reported concentrations from all previous monitoring;
    (ii) The degree of variation in reported concentrations; and
    (iii) Other factors which may affect contaminant concentrations such 
as changes in groundwater pumping rates, changes in the system's 
configuration, changes in the system's operating procedures, or changes 
in stream flows or characteristics.
    (6) A decision by the State to grant a waiver shall be made in 
writing and shall set forth the basis for the determination. The 
determination may be initiated by the State or upon an application by 
the public water system. The public water system shall specify the basis 
for its request. The State shall review and, where appropriate, revise 
its determination of the appropriate monitoring frequency when the 
system submits new monitoring data or when other data relevant to the 
system's appropriate monitoring frequency become available.
    (7) Systems which exceed the maximum contaminant levels as 
calculated in Sec. 141.23(i) of this section shall monitor quarterly 
beginning in the next quarter after the violation occurred.
    (8) The State may decrease the quarterly monitoring requirement to 
the frequencies specified in paragraphs (c)(1) and (c)(2) of this 
section provided it has determined that the system is reliably and 
consistently below the maximum contaminant level. In no case can a State 
make this determination unless a groundwater system takes a minimum of 
two quarterly samples and a surface water system takes a minimum of four 
quarterly samples.
    (d) All public water systems (community; non-transient, non-
community; and transient, non-community systems) shall monitor to 
determine compliance with the maximum contaminant level for nitrate in 
Sec. 141.62.
    (1) Community and non-transient, non-community water systems served 
by groundwater systems shall monitor annually beginning January 1, 1993; 
systems served by surface water shall monitor quarterly beginning 
January 1, 1993.
    (2) For community and non-transient, non-community water systems, 
the repeat monitoring frequency for groundwater systems shall be 
quarterly for at least one year following any one sample in which the 
concentration is <50 percent of the MCL. The State may allow a 
groundwater system to reduce the sampling frequency to annually after 
four consecutive quarterly samples are reliably and consistently less 
than the MCL.
    (3) For community and non-transient, non-community water systems, 
the State may allow a surface water system to reduce the sampling 
frequency to annually if all analytical results from four consecutive 
quarters are <50 percent of the MCL. A surface water system shall return 
to quarterly monitoring if any one sample is [gE]50 percent of the MCL.
    (4) Each transient non-community water system shall monitor annually 
beginning January 1, 1993.
    (5) After the initial round of quarterly sampling is completed, each 
community and non-transient non-community system which is monitoring 
annually shall take subsequent samples during the quarter(s) which 
previously resulted in the highest analytical result.
    (e) All public water systems (community; non-transient, non-
community; and transient, non-community systems) shall monitor to 
determine compliance with the maximum contaminant level for nitrite in 
Sec. 141.62(b).
    (1) All public water systems shall take one sample at each sampling

[[Page 357]]

point in the compliance period beginning January 1, 1993 and ending 
December 31, 1995.
    (2) After the initial sample, systems where an analytical result for 
nitrite is <50 percent of the MCL shall monitor at the frequency 
specified by the State.
    (3) For community, non-transient, non-community, and transient non-
community water systems, the repeat monitoring frequency for any water 
system shall be quarterly for at least one year following any one sample 
in which the concentration is <50 percent of the MCL. The State may 
allow a system to reduce the sampling frequency to annually after 
determining the system is reliably and consistently less than the MCL.
    (4) Systems which are monitoring annually shall take each subsequent 
sample during the quarter(s) which previously resulted in the highest 
analytical result.
    (f) Confirmation samples:
    (1) Where the results of sampling for antimony, arsenic, asbestos, 
barium, beryllium, cadmium, chromium, cyanide, fluoride, mercury, 
nickel, selenium or thallium indicate an exceedance of the maximum 
contaminant level, the State may require that one additional sample be 
collected as soon as possible after the initial sample was taken (but 
not to exceed two weeks) at the same sampling point.
    (2) Where nitrate or nitrite sampling results indicate an exceedance 
of the maximum contaminant level, the system shall take a confirmation 
sample within 24 hours of the system's receipt of notification of the 
analytical results of the first sample. Systems unable to comply with 
the 24-hour sampling requirement must immediately notify persons served 
by the public water system in accordance with Sec. 141.202 and meet 
other Tier 1 public notification requirements under Subpart Q of this 
part. Systems exercising this option must take and analyze a 
confirmation sample within two weeks of notification of the analytical 
results of the first sample.
    (3) If a State-required confirmation sample is taken for any 
contaminant, then the results of the initial and confirmation sample 
shall be averaged. The resulting average shall be used to determine the 
system's compliance in accordance with paragraph (i) of this section. 
States have the discretion to delete results of obvious sampling errors.
    (g) The State may require more frequent monitoring than specified in 
paragraphs (b), (c), (d) and (e) of this section or may require 
confirmation samples for positive and negative results at its 
discretion.
    (h) Systems may apply to the State to conduct more frequent 
monitoring than the minimum monitoring frequencies specified in this 
section.
    (i) Compliance with Secs. 141.11 or 141.62(b) (as appropriate) shall 
be determined based on the analytical result(s) obtained at each 
sampling point.
    (1) For systems which are conducting monitoring at a frequency 
greater than annual, compliance with the maximum contaminant levels for 
antimony, asbestos, barium, beryllium, cadmium, chromium, cyanide, 
fluoride, mercury, nickel, selenium or thallium is determined by a 
running annual average at any sampling point. If the average at any 
sampling point is greater than the MCL, then the system is out of 
compliance. If any one sample would cause the annual average to be 
exceeded, then the system is out of compliance immediately. Any sample 
below the method detection limit shall be calculated at zero for the 
purpose of determining the annual average.
    (2) For systems which are monitoring annually, or less frequently, 
the system is out of compliance with the maximum contaminant levels for 
asbestos, antimony, barium, beryllium, cadmium, chromium, cyanide, 
fluoride, mercury, nickel, selenium or thallium if the level of a 
contaminant at any sampling point is greater than the MCL. If a 
confirmation sample is required by the State, the determination of 
compliance will be based on the average of the two samples.
    (3) Compliance with the maximum contaminant levels for nitrate and 
nitrate is determined based on one sample if the levels of these 
contaminants are below the MCLs. If the levels of nitrate and/or nitrite 
exceed the MCLs in

[[Page 358]]

the initial sample, a confirmation sample is required in accordance with 
paragraph (f)(2) of this section, and compliance shall be determined 
based on the average of the initial and confirmation samples.
    (4) Arsenic sampling results will be reported to the nearest 0.001 
mg/L.
    (j) Each public water system shall monitor at the time designated by 
the State during each compliance period.
    (k) Inorganic analysis:
    (1) Analysis for the following contaminants shall be conducted in 
accordance with the methods in the following table, or their equivalent 
as determined by EPA. Criteria for analyzing arsenic, barium, beryllium, 
cadmium, calcium, chromium, copper, lead, nickel, selenium, sodium, and 
thallium with digestion or directly without digestion, and other 
analytical test procedures are contained in Technical Notes on Drinking 
Water Methods, EPA-600/R-94-173, October 1994. This document also 
contains approved analytical test methods which remain available for 
compliance monitoring until July 1, 1996. These methods will not be 
available for use after July 1, 1996. This document is available from 
the National Technical Information Service, NTIS PB95-104766, U.S. 
Department of Commerce, 5285 Port Royal Road, Springfield, Virginia 
22161. The toll-free number is 800-553-6847.


----------------------------------------------------------------------------------------------------------------
   Contaminant and methodology \13\        EPA            ASTM \3\              SM \4\               Other
----------------------------------------------------------------------------------------------------------------
Alkalinity:
    Titrimetric......................  ...........  D1067-92B            2320 B               I-1030-85 \5\
    Electrometric titration..........  ...........  ...................  ...................  ..................
Antimony:
    ICP-Mass Spectrometry............    \2\ 200.8
    Hydride-Atomic Absorption........  ...........  D-3697-92            ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
    Atomic Absorption; Furnace.......  ...........  ...................  3113 B               ..................
Arsenic \14\:
    Inductively Coupled Plasma \15\..    \2\ 200.7  ...................  \15\ 3120 B          ..................
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
    Atomic Absorption; Furnace.......  ...........  D-2972-93C           3113 B               ..................
    Hydride Atomic Absorption........  ...........  D-2972-93B           3114 B               ..................
Asbestos:
    Transmission Electron Microscopy.    \9\ 100.1  ...................  ...................  ..................
    Transmission Electron Microscopy.   \10\ 100.2  ...................  ...................  ..................
Barium:
    Inductively Coupled Plasma.......    \2\ 200.7  ...................  3120 B               ..................
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Direct........  ...........  ...................  3111 D               ..................
    Atomic Absorption; Furnace.......  ...........  ...................  3113 B               ..................
Beryllium:
    Inductively Coupled Plasma.......    \2\ 200.7  ...................  3120 B               ..................
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
    Atomic Absorption; Furnace.......  ...........  D3645-93B            3113 B               ..................
Cadmium:
    Inductively Coupled Plasma.......    \2\ 200.7  ...................  ...................  ..................
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
    Atomic Absorption; Furnace.......  ...........  ...................  3113 B               ..................
Calcium:
    EDTA titrimetric.................  ...........  D511-93A             3500-Ca D
    Atomic absorption; direct          ...........  D511-93B             3111 B               ..................
     aspiration.
    Inductively-coupled plasma.......    \2\ 200.7  ...................  3120 B               ..................
Chromium:
    Inductively Coupled Plasma.......    \2\ 200.7  ...................  3120 B               ..................
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
    Atomic Absorption; Furnace.......  ...........  ...................  3113 B               ..................
Copper:

[[Page 359]]


    Atomic absorption; furnace.......  ...........  D1688-95C            3113 B               ..................
    Atomic absorption; direct          ...........  D1688-95A            3111 B               ..................
     aspiration.
    ICP..............................    \2\ 200.7  ...................  3120 B               ..................
    ICP--Mass spectrometry...........    \2\ 200.8  ...................  ...................  ..................
    Atomic absorption; platform......    \2\ 200.9  ...................  ...................  ..................
Conductivity Conductance.............  ...........  D1125-95A            2510 B               ..................
Cyanide:
    Manual Distillation followed by..  ...........  D2036-91A            4500-CN- C
        Spectrophotometric, Amenable.  ...........  D2036-91B            4500-CN- G
        Spectrophotometric...........
            Manual...................  ...........  D2036-91A            4500-CN- E           I-3300-85 \5\
            Semi-automated...........    \6\ 335.4
    Selective Electrode..............  ...........  ...................  4500-CN- F
Fluoride:
    Ion Chromatography...............    \6\ 300.0  D4327-91             4110 B
    Manual Distill.; Color. SPADNS...  ...........  ...................  4500-F- B, D
    Manual Electrode.................  ...........  D1179-93B            4500-F- C
    Automated Electrode..............  ...........  ...................  ...................  380-75WE \11\
    Automated Alizarin...............  ...........  ...................  4500-F- E            129-71W \11\
Lead:
    Atomic absorption; furnace.......  ...........  D3559-95D            3113 B
    ICP-Mass spectrometry............    \2\ 200.8
    Atomic absorption; platform......    \2\ 200.9
    Differential Pulse Anodic          ...........  ...................  ...................  Method 1001 \15\
     Stripping Voltammetry.
Magnesium:
    Atomic Absorption................  ...........  D 511-93 B           3111 B
    ICP..............................    \2\ 200.7  ...................  3120 B
    Complexation Titrimetric Methods.  ...........  D 511-93 A           3500-Mg E
Mercury:
    Manual, Cold Vapor...............    \2\ 245.1  D3223-91             3112 B
    Automated, Cold Vapor............    \1\ 245.2
    ICP-Mass Spectrometry............    \2\ 200.8
Nickel:
    Inductively Coupled Plasma.......    \2\ 200.7  ...................  3120 B
    ICP-Mass Spectrometry............    \2\ 200.8
    Atomic Absorption; Platform......    \2\ 200.9
    Atomic Absorption; Direct........  ...........  ...................  3111 B
    Atomic Absorption; Furnace.......  ...........  ...................  3113 B
Nitrate:
    Ion Chromatography...............    \6\ 300.0  D4327-91             4110 B               B-1011 \8\
    Automated Cadmium Reduction......    \6\ 353.2  D3867-90A            4500-NO3- F
    Ion Selective Electrode..........  ...........  ...................  4500-NO3- D          601 \7\
    Manual Cadmium Reduction.........  ...........  D3867-90B            4500-NO3- E
Nitrite:
    Ion Chromatography...............    \6\ 300.0  D4327-91             4110 B               B-1011 \8\
    Automated Cadmium Reduction......    \6\ 353.2  D3867-90A            4500-NO3- F
    Manual Cadmium Reduction.........  ...........  D3867-90B            4500-NO3- E
    Spectrophotometric...............  ...........  ...................  4500-NO2- B
Orthophosphate: \12\
    Colorimetric, automated, ascorbic    \6\ 365.1  ...................  4500-P F
     acid.
    Colorimetric, ascorbic acid,       ...........  D515-88A             4500-P E
     single reagent.

[[Page 360]]


    Colorimetric, phosphomolybdate;..  ...........  ...................  ...................  I-1602-85 \5\
        automated-segmented flow;....  ...........  ...................  ...................  I-2601-90 \5\
        automated discrete...........  ...........  ...................  ...................  I-2598-85 \5\
    Ion Chromatography...............    \6\ 300.0  D4327-91             4110 B
pH: Electrometric....................    \1\ 150.1  D1293-95             4500-H+ B
                                         \1\ 150.2  ...................  ...................  ..................
Selenium:
    Hydride-Atomic Absorption........  ...........  D3859-93A            3114 B               ..................
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
    Atomic Absorption; Furance.......  ...........  D3859-93B            3113 B               ..................
Silica:
    Colorimetric, molybdate blue;....  ...........  ...................  ...................  I-1700-85 \5\
        automated-segmented flow.....  ...........  ...................  ...................  I-2700-85 \5\
    Colorimetric.....................  ...........  D859-95              ...................  ..................
    Molybdosilicate..................  ...........  ...................  4500-Si D            ..................
Heteropoly blue......................  ...........  ...................  4500-Si E            ..................
    Automated method for molybdate-    ...........  ...................  4500-Si F            ..................
     reactive silica.
    Inductively-coupled plasma.......    \3\ 200.7  ...................  3120 B               ..................
Sodium:
    Inductively-coupled plasma.......    \2\ 200.7  ...................  ...................  ..................
    Atomic Absorption; direct          ...........  ...................  3111 B               ..................
     aspiration.
Temperature: Thermometric              ...........  ...................  2550                 ..................
Thallium:
    ICP-Mass Spectrometry............    \2\ 200.8  ...................  ...................  ..................
    Atomic Absorption; Platform......    \2\ 200.9  ...................  ...................  ..................
----------------------------------------------------------------------------------------------------------------
The procedures shall be done in accordance with the documents listed below. The incorporation by reference of
  the following documents listed in footnotes 1-11 and 15 was approved by the Director of the Federal Register
  in accordance with 5 U.S.C. 552(a) and 1 CFR part 51. Copies of the documents may be obtained from the sources
  listed below. Information regarding obtaining these documents can be obtained from the Safe Drinking Water
  Hotline at 800-426-4791. Documents may be inspected at EPA's Drinking Water Docket, 1200 Pennsylvania Ave.,
  NW., Washington, DC 20460 (Telephone: 202-260-3027); or at the Office of Federal Register, 800 North Capitol
  Street, NW., Suite 700, Washington, DC.
\1\ ``Methods for Chemical Analysis of Water and Wastes'', EPA/600/4-79/020, March 1983. Available at NTIS, PB84-
  128677.
\2\ ``Methods for the Determination of Metals in Environmental Samples--Supplement I'', EPA/600/R-94/111, May
  1994. Available at NTIS, PB95-125472.
\3\ Annual Book of ASTM Standards, 1994 and 1996, Vols. 11.01 and 11.02, American Society for Testing and
  Materials. The previous versions of D1688-95A, D1688-95C (copper), D3559-95D (lead), D1293-95 (pH), D1125-91A
  (conductivity) and D859-94 (silica) are also approved. These previous versions D1688-90A, C; D3559-90D, D1293-
  84, D1125-91A and D859-88, respectively are located in the Annual Book of ASTM Standards, 1994, Vols. 11.01.
  Copies may be obtained from the American Society for Testing and Materials, 100 Barr Harbor Drive, West
  Conshohocken, PA 19428.
\4\ 18th and 19th editions of Standard Methods for the Examination of Water and Wastewater, 1992 and 1995,
  respectively, American Public Health Association; either edition may be used. Copies may be obtained from the
  American Public Health Association, 1015 Fifteenth Street NW, Washington, DC 20005.
\5\ Method I-2601-90, Methods for Analysis by the U.S. Geological Survey National Water Quality Laboratory--
  Determination of Inorganic and Organic Constituents in Water and Fluvial Sediments, Open File Report 93-125,
  1993; For Methods I-1030-85; I-1601-85; I-1700-85; I-2598-85; I-2700-85; and I-3300-85 See Techniques of Water
  Resources Investigation of the U.S. Geological Survey, Book 5, Chapter A-1, 3rd ed., 1989; Available from
  Information Services, U.S. Geological Survey, Federal Center, Box 25286, Denver, CO 80225-0425.
\6\ ``Methods for the Determination of Inorganic Substances in Environmental Samples'', EPA/600/R-93/100, August
  1993. Available at NTIS, PB94-120821.
\7\ The procedure shall be done in accordance with the Technical Bulletin 601 ``Standard Method of Test for
  Nitrate in Drinking Water'', July 1994, PN 221890-001, Analytical Technology, Inc. Copies may be obtained from
  ATI Orion, 529 Main Street, Boston, MA 02129.
\8\ Method B-1011, ``Waters Test Method for Determination of Nitrite/Nitrate in Water Using Single Column Ion
  Chromatography,'' August 1987. Copies may be obtained from Waters Corporation, Technical Services Division, 34
  Maple Street, Milford, MA 01757.
\9\ Method 100.1, ``Analytical Method For Determination of Asbestos Fibers in Water'', EPA/600/4-83/043, EPA,
  September 1983. Available at NTIS, PB83-260471.
\10\ 10 Method 100.2, ``Determination of Asbestos Structure Over 10-[mu]m In Length In Drinking Water'', EPA/600/
  R-94/134, June 1994. Available at NTIS, PB94-201902.
\11\ Industrial Method No. 129-71W, ``Fluoride in Water and Wastewater'', December 1972, and Method No. 380-
  75WE, ``Fluoride in Water and Wastewater'', February 1976, Technicon Industrial Systems. Copies may be
  obtained from Bran & Luebbe, 1025 Busch Parkway, Buffalo Grove, IL 60089.
\12\ Unfiltered, no digestion or hydrolysis.
\13\ Because MDLs reported in EPA Methods 200.7 and 200.9 were determined using a 2X preconcentration step
  during sample digestion, MDLs determined when samples are analyzed by direct analysis (i.e., no sample
  digestion) will be higher. For direct analysis of cadmium and arsenic by Method 200.7, and arsenic by Method
  3120 B sample preconcentration using pneumatic nebulization may be required to achieve lower detection limits.
  Preconcentration may also be required for direct analysis of antimony, lead, and thallium by Method 200.9;
  antimony and lead by Method 3113 B; and lead by Method D3559-90D unless multiple in-furnace depositions are
  made.

[[Page 361]]


\14\ If ultrasonic nebulization is used in the determination of arsenic by Methods 200.7, 200.8, or SM 3120 B,
  the arsenic must be in the pentavalent state to provide uniform signal response. For methods 200.7 and 3120 B,
  both samples and standards must be diluted in the same mixed acid matrix concentration of nitric and
  hydrochloric acid with the addition of 100 [mu]L of 30% hydrogen peroxide per 100ml of solution. For direct
  analysis of arsenic with method 200.8 using ultrasonic nebulization, samples and standards must contain one mg/
  L of sodium hypochlorite.
\15\ The description for Method Number 1001 for lead is available from Palintest, LTD, 21 Kenton Lands Road,
  P.O. Box 18395, Erlanger, KY 41018. Or from the Hach Company, P.O. Box 389, Loveland, CO 8053.

    (2) Sample collection for antimony, arsenic, asbestos, barium, 
beryllium, cadmium, chromium, cyanide, fluoride, mercury, nickel, 
nitrate, nitrite, selenium, and thallium under this section shall be 
conducted using the sample preservation, container, and maximum holding 
time procedures specified in the table below:


----------------------------------------------------------------------------------------------------------------
              Contaminant                    Preservative \1\            Container \2\            Time \3\
----------------------------------------------------------------------------------------------------------------
Antimony..............................  HNO\3\....................  P or G................  6 months
Arsenic...............................  Conc HNO3 to pH <2........  P or G................  6 months
Asbestos..............................  4  deg.C..................  P or G................  48 hours \4\
Barium................................  HNO\3\....................  P or G................  6 months
Beryllium.............................  HNO\3\....................  P or G................  6 months
Cadmium...............................  HNO\3\....................  P or G................  6 months
Chromium..............................  HNO\3\....................  P or G................  6 months
Cyanide...............................  4  deg.C, NaOH............  P or G................  14 days
Fluoride..............................  None......................  P or G................  1 month
Mercury...............................  HNO\3\....................  P or G................  28 days
Nickel................................  HNO\3\....................  P or G................  6 months
Nitrate...............................  4  deg.C..................  P or G................  48 hours \5\
Nitrate-Nitrite \6\...................  H\2\SO\4\.................  P or G................  28 days
Nitrite...............................  4 deg.C...................  P or G................  48 hours
Selenium..............................  HNO\3\....................  P or G................  6 months
Thallium..............................  HNO\3\....................  P or G................  6 months
----------------------------------------------------------------------------------------------------------------
\1\ For cyanide determinations samples must be adjusted with sodium hydroxide to pH 12 at the time off
  collection. When chilling is indicated the sample must be shipped and stored at 4  deg.C or less.
  Acidification of nitrate or metals samples may be with a concentrated acid or a dilute (50% by volume)
  solution of the applicable concentrated acid. Acidification of samples for metals analysis is encouraged and
  allowed at the laboratory rather than at the time of sampling provided the shipping time and other
  instructions in Section 8.3 of EPA Methods 200.7 or 200.8 or 200.9 are followed.
\2\ P=plastic, hard or soft; G=glass, hard or soft.
\3\ In all cases samples should be analyzed as soon after collection as possible. Follow additional (if any)
  information on preservation, containers or holding times that is specified in method.
\4\ Instructions for containers, preservation procedures and holding times as specified in Method 100.2 must be
  adhered to for all compliance analyses including those conducted with Method 100.1.
\5\ If the sample is chlorinated, the holding time for an unacidified sample kept at 4  deg.C is extended to 14
  days.
\6\ Nitrate-Nitrite refers to a measurement of total nitrate.

    (3) Analysis under this section shall only be conducted by 
laboratories that have been certified by EPA or the State. Laboratories 
may conduct sample analysis under provisional certification until 
January 1, 1996. To receive certification to conduct analyses for 
antimony, arsenic, asbestos, barium, beryllium, cadmium, chromium, 
cyanide, fluoride, mercury, nickel, nitrate, nitrite and selenium and 
thallium, the laboratory must:
    (i) Analyze Performance Evaluation (PE) samples provided by EPA, the 
State or by a third party (with the approval of the State or EPA) at 
least once a year.
    (ii) For each contaminant that has been included in the PE sample 
and for each method for which the laboratory desires certification 
achieve quantitative results on the analyses that are within the 
following acceptance limits:


------------------------------------------------------------------------
             Contaminant                       Acceptance limit
------------------------------------------------------------------------
Antimony............................  30 at [gE]0.006 mg/1
Arsenic.............................  30 at [ge]0.003 mg/L
Asbestos............................  2 standard deviations based on
                                       study statistics.
Barium..............................  15% at [gE]0.15 mg/1
Beryllium...........................  15% at [gE]0.001 mg/1
Cadmium.............................  20% at [gE]0.002 mg/1
Chromium............................  15% at [gE]0.01 mg/1
Cyanide.............................  25% at [gE]0.1 mg/1
Fluoride............................  10% at [gE]1 to 10 mg/
                                       1
Mercury.............................  30% at [gE]0.0005 mg/1
Nickel..............................  15% at [gE]0.01 mg/1
Nitrate.............................  10% at [gE]0.4 mg/1
Nitrite.............................  15% at [gE]0.4 mg/1
Selenium............................  20% at [gE]0.01 mg/1
Thallium............................  30% at [gE]0.002 mg/1
------------------------------------------------------------------------


    (l) Analyses for the purpose of determining compliance with 
Sec. 141.11 shall be conducted using the requirements specified in 
paragraphs (l) through (q) of this section.
    (1) Analyses for all community water systems utilizing surface water 
sources shall be completed by June 24, 1978. These analyses shall be 
repeated at yearly intervals.
    (2) Analyses for all community water systems utilizing only ground 
water sources shall be completed by June 24, 1979. These analyses shall 
be repeated at three-year intervals.
    (3) For non-community water systems, whether supplied by surface or 
ground sources, analyses for nitrate shall be completed by December 24, 
1980. These analyses shall be repeated at intervals determined by the 
State.

[[Page 362]]

    (4) The State has the authority to determine compliance or initiate 
enforcement action based upon analytical results and other information 
compiled by their sanctioned representatives and agencies.
    (m) If the result of an analysis made under paragraph (l) of this 
section indicates that the level of any contaminant listed in 
Sec. 141.11 exceeds the maximum contaminant level, the supplier of the 
water shall report to the State within 7 days and initiate three 
additional analyses at the same sampling point within one month.
    (n) When the average of four analyses made pursuant to paragraph (m) 
of this section, rounded to the same number of significant figures as 
the maximum contaminant level for the substance in question, exceeds the 
maximum contaminant level, the supplier of water shall notify the State 
pursuant to Sec. 141.31 and give notice to the public pursuant to 
subpart Q. Monitoring after public notification shall be at a frequency 
designated by the State and shall continue until the maximum contaminant 
level has not been exceeded in two successive samples or until a 
monitoring schedule as a condition to a variance, exemption or 
enforcement action shall become effective.
    (o) The provisions of paragraphs (m) and (n) of this section 
notwithstanding, compliance with the maximum contaminant level for 
nitrate shall be determined on the basis of the mean of two analyses. 
When a level exceeding the maximum contaminant level for nitrate is 
found, a second analysis shall be initiated within 24 hours, and if the 
mean of the two analyses exceeds the maximum contaminant level, the 
supplier of water shall report his findings to the State pursuant to 
Sec. 141.31 and shall notify the public pursuant to subpart Q.
    (p) For the initial analyses required by paragraph (l) (1), (2) or 
(3) of this section, data for surface waters acquired within one year 
prior to the effective date and data for ground waters acquired within 3 
years prior to the effective date of this part may be substituted at the 
discretion of the State.
    (q) [Reserved]

[56 FR 3579, Jan. 30, 1991, as amended at 56 FR 30274, July 1, 1991; 57 
FR 31838, July 17, 1992; 59 FR 34322, July 1, 1994; 59 FR 62466, Dec. 5, 
1994; 60 FR 33932, 34085, June 29, 1995; 64 FR 67461, Dec. 1, 1999; 65 
FR 26022, May 4, 2000; 66 FR 7061, Jan. 22, 2001]

    Effective Date Note: At 66 FR 7061, Jan. 22, 2001, Sec. 141.23 was 
amended by adding paragraph (c)(9) and revising paragraphs (i)(1) and 
(i)(2), effective Mar. 23, 2001, except for the amendments to paragraphs 
(i)(1) and (i)(2) which are effective Jan. 22, 2004. At 66 FR 16134, 
Mar. 23, 2001, the effective date was delayed until May 22, 2001. At 66 
FR 28350, May 22, 2001, the effective date for paragraph (c)(9) was 
delayed until Jan. 22, 2004. For the convenience of the user, the 
revised and added text is set forth as follows:

Sec. 141.23  Inorganic chemical sampling and analytical requirements.

                                * * * * *

    (c) * * *
    (9) All new systems or systems that use a new source of water that 
begin operation after January 22, 2004 must demonstrate compliance with 
the MCL within a period of time specified by the State. The system must 
also comply with the initial sampling frequencies specified by the State 
to ensure a system can demonstrate compliance with the MCL. Routine and 
increased monitoring frequencies shall be conducted in accordance with 
the requirements in this section.

                                * * * * *

    (i) * * *
    (1) For systems which are conducting monitoring at a frequency 
greater than annual, compliance with the maximum contaminant levels for 
antimony, arsenic, asbestos, barium, beryllium, cadmium, chromium, 
cyanide, fluoride, mercury, nickel, selenium or thallium is determined 
by a running annual average at any sampling point. If the average at any 
sampling point is greater than the MCL, then the system is out of 
compliance. If any one sample would cause the annual average to be 
exceeded, then the system is out of compliance immediately. Any sample 
below the method detection limit shall be calculated at zero for the 
purpose of determining the annual average. If a system fails to collect 
the required number of samples, compliance (average concentration) will 
be based on the total number of samples collected.
    (2) For systems which are monitoring annually, or less frequently, 
the system is out of compliance with the maximum contaminant levels for 
antimony, arsenic, asbestos,

[[Page 363]]

barium, beryllium, cadmium, chromium, cyanide, fluoride, mercury, 
nickel, selenium or thallium if the level of a contaminant is greater 
than the MCL. If confirmation samples are required by the State, the 
determination of compliance will be based on the annual average of the 
initial MCL exceedance and any State-required confirmation samples. If a 
system fails to collect the required number of samples, compliance 
(average concentration) will be based on the total number of samples 
collected.

                                * * * * *