[Code of Federal Regulations]
[Title 40, Volume 19]
[Revised as of July 1, 2002]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR141.24]

[Page 363-372]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 141--NATIONAL PRIMARY DRINKING WATER REGULATIONS--Table of Contents
 
            Subpart C--Monitoring and Analytical Requirements
 
Sec. 141.24  Organic chemicals, sampling and analytical requirements.

    (a)-(d) [Reserved]
    (e) Analyses for the contaminants in this section shall be conducted 
using the following EPA methods or their equivalent as approved by EPA.
    (1) The following documents are incorporated by reference. This 
incorporation by reference was approved by the Director of the Federal 
Register in accordance with 5 U.S.C. 552(a) and 1 CFR Part 51. Copies 
may be inspected at EPA's Drinking Water Docket, 401 M St., SW., 
Washington, DC 20460; or at the Office of the Federal Register, 800 
North Capitol Street, NW., Suite 700, Washington, DC. Method 508A and 
515.1 are in Methods for the Determination of Organic Compounds in 
Drinking Water, EPA/600/4-88-039, December 1988, Revised, July 1991. 
Methods 547, 550 and 550.1 are in Methods for the Determination of 
Organic Compounds in Drinking Water--Supplement I, EPA/600-4-90-020, 
July 1990. Methods 548.1, 549.1, 552.1 and 555 are in Methods for the 
Determination of Organic Compounds in Drinking Water--Supplement II, 
EPA/600/R-92-129, August 1992. Methods 502.2, 504.1, 505, 506, 507, 508, 
508.1, 515.2, 524.2 525.2, 531.1, 551.1 and 552.2 are in Methods for the 
Determination of Organic Compounds in Drinking Water--Supplement III, 
EPA/600/R-95-131, August 1995. Method 1613 is titled ``Tetra-through 
Octa-Chlorinated Dioxins and Furans by Isotope-Dilution HRGC/HRMS'', 
EPA/821-B-94-005, October 1994. These documents are available from the 
National Technical Information Service, NTIS PB91-231480, PB91-146027, 
PB92-207703, PB95-261616 and PB95-104774, U.S. Department of Commerce, 
5285 Port Royal Road, Springfield, Virginia 22161. The toll-free number 
is 800-553-6847. Method 6651 shall be followed in accordance with 
Standard Methods for the Examination of Water and Wastewater, 18th 
edition, 1992 and 19th edition, 1995, American Public Health Association 
(APHA); either edition may be used. Method 6610 shall be followed in 
accordance with the Supplement to the 18th edition of Standard Methods 
for the Examination of Water and Wastewater, 1994 or with the 19th 
edition of Standard Methods for the Examination of Water and Wastewater, 
1995, APHA; either publication may be used. The APHA documents are 
available from APHA, 1015 Fifteenth Street NW., Washington, D.C. 20005. 
Other required analytical test procedures germane to the conduct of 
these analyses are contained in Technical Notes on Drinking Water 
Methods, EPA/600/R-94-173, October 1994, NTIS PB95-104766. EPA Methods 
515.3 and 549.2 are available from U.S. Environmental Protection Agency, 
National Exposure Research Laboratory (NERL)-Cincinnati, 26 West Martin 
Luther King Drive, Cincinnati, OH 45268. ASTM Method D 5317-93 is 
available in the Annual Book of ASTM Standards, 1996, Vol. 11.02, 
American Society for Testing and Materials, 100 Barr Harbor Drive, West 
Conshohocken, PA 19428, or in any edition published after 1993.


------------------------------------------------------------------------
                Contaminant                          Method \1\
------------------------------------------------------------------------
Benzene...................................  502.2, 524.2

Carbon tetrachloride                        502.2, 524.2, 551.1

Chlorobenzene                               502.2, 524.2
1,2-Dichlorobenzene.......................  502.2, 524.2
1,4-Dichlorobenzene.......................  502.2, 524.2
1,2-Dichloroethane........................  502.2, 524.2
cis-Dichloroethylene......................  502.2, 524.2
trans-Dichloroethylene....................  502.2, 524.2
Dichloromethane...........................  502.2, 524.2
1,2-Dichloropropane.......................  502.2, 524.2
Ethylbenzene..............................  502.2, 524.2
Styrene...................................  502.2, 524.2
Tetrachloroethylene.......................  502.2, 524.2, 551.1
1,1,1-Trichloroethane.....................  502.2, 524.2, 551.1
Trichloroethylene.........................  502.2, 524.2, 551.1
Toluene...................................  502.2, 524.2
1,2,4-Trichlorobenzene....................  502.2, 524.2
1,1-Dichloroethylene......................  502.2, 524.2
1,1,2-Trichloroethane.....................  502.2, 524.2, 551.1
Vinyl chloride............................  502.2, 524.2
Xylenes (total)...........................  502.2, 524.2
2,3,7,8-TCDD (dioxin).....................  1613
2,4-D\4\ (as acid, salts and esters)......  515.2, 555, 515.1, 515.3,
                                             D5317-93

[[Page 364]]


2,4,5-TP \4\ (Silvex).....................  515.2, 555, 515.1, 515.3,
                                             D5317-93
Alachlor \2\..............................  507, 525.2, 508.1, 505,
                                             551.1
Atrazine \2\..............................  507, 525.2, 508.1, 505,
                                             551.1
Benzo(a)pyrene............................  525.2, 550, 550.1
Carbofuran................................  531.1, 6610
Chlordane.................................  508, 525.2, 508.1, 505
Dalapon...................................  552.1, 515.1, 552.2, 515.3
Di(2-ethylhexyl)adipate...................  506, 525.2
Di(2-ethylhexyl)phthalate.................  506, 525.2
Dibromochloropropane (DBCP)...............  504.1, 551.1
Dinoseb \4\...............................  515.2, 555, 515.1, 515.3
Diquat....................................  549.2
Endothall.................................  548.1
Endrin....................................  508, 525.2, 508.1, 505,
                                             551.1
Ethylene dibromide (EDB)..................  504.1, 551.1
Glyphosate................................  547, 6651
Heptachlor................................  508, 525.2, 508.1, 505,
                                             551.1
Heptachlor Epoxide........................  508, 525.2, 508.1, 505,
                                             551.1
Hexachlorobenzene.........................  508, 525.2, 508.1, 505,
                                             551.1
Hexachlorocyclopentadiene.................  508, 525.2, 508.1, 505,
                                             551.1
Lindane...................................  508, 525.2, 508.1, 505,
                                             551.1
Methoxychlor..............................  508, 525.2, 508.1, 505,
                                             551.1
Oxamyl....................................  531.1, 6610
PCBs \3\ (as decachlorobiphenyl)..........  508A
    (as Aroclors).........................  508.1, 508, 525.2, 505
Pentachlorophenol.........................  515.2, 525.2, 555, 515.1,
                                             515.3, D5317-93
Picloram \4\..............................  515.2, 555, 515.1, 515.3,
                                             D5317-93
Simazine \2\..............................  507, 525.2, 508.1, 505,
                                             551.1
Toxaphene.................................  508, 508.1, 525.2, 505
Total Trihalomethanes.....................  502.2, 524.2, 551.1
------------------------------------------------------------------------
\1\ For previously approved EPA methods which remain available for
  compliance monitoring until June 1, 2001, see paragraph (e)(2) of this
  section.
\2\ Substitution of the detector specified in Method 505, 507, 508 or
  508.1 for the purpose of achieving lower detection limits is allowed
  as follows. Either an electron capture or nitrogen phosphorous
  detector may be used provided all regulatory requirements and quality
  control criteria are met.
\3\ PCBs are qualitatively identified as Aroclors and measured for
  compliance purposes as decachlorobiphenyl. Users of Method 505 may
  have more difficulty in achieving the required detection limits than
  users of Methods 508.1, 525.2 or 508.
\4\ Accurate determination of the chlorinated esters requires hydrolysis
  of the sample as described in EPA Methods 515.1, 515.2, 515.3 and 555,
  and ASTM Method D 5317-93.

    (2) The following EPA methods will remain available for compliance 
monitoring until June 1, 2001. The following documents are incorporated 
by reference. This incorporation by reference was approved by the 
Director of the Federal Register in accordance with 5 U.S.C. 552(a) and 
1 CFR Part 51. Copies may be inspected at EPA's Drinking Water Docket, 
401 M St., SW., Washington, DC 20460; or at the Office of the Federal 
Register, 800 North Capitol Street, NW., Suite 700, Washington, DC. EPA 
methods 502.2 Rev. 2.0, 505 Rev. 2.0, 507 Rev. 2.0, 508 Rev. 3.0, 531.1 
Rev. 3.0 are in ``Methods for the Determination of Organic Compounds in 
Drinking Water'', December 1988, revised July 1991; methods 506 and 551 
are in ``Methods for the Determination of Organic Compounds in Drinking 
Water--Supplement I'', July 1990; methods 515.2 Rev. 1.0 and 524.2 Rev. 
4.0 are in ``Methods for the Determination of Organic Compounds in 
Drinking Water--Supplement II,'' August 1992; and methods 504.1 Rev. 
1.0, 508.1 Rev. 1.0, 525.2 Rev.1.0 are available from US EPA NERL, 
Cincinnati, OH 45268
    (f) Beginning with the initial compliance period, analysis of the 
contaminants listed in Sec. 141.61(a) (1) through (21) for the purpose 
of determining compliance with the maximum contaminant level shall be 
conducted as follows:
    (1) Groundwater systems shall take a minimum of one sample at every 
entry point to the distribution system which is representative of each 
well after treatment (hereafter called a sampling point). Each sample 
must be taken at the same sampling point unless conditions make another 
sampling point more representative of each source, treatment plant, or 
within the distribution system.
    (2) Surface water systems (or combined surface/ground) shall take a 
minimum of one sample at points in the distribution system that are 
representative of each source or at each entry point to the distribution 
system after treatment (hereafter called a sampling point). Each sample 
must be taken at the same sampling point unless conditions make another 
sampling point more representative of each source, treatment plant, or 
within the distribution system.
    (3) If the system draws water from more than one source and the 
sources are combined before distribution, the system must sample at an 
entry point to the distribution system during periods of normal 
operating conditions (i.e., when water representative of all sources is 
being used).
    (4) Each community and non-transient non-community water system 
shall take four consecutive quarterly samples for each contaminant 
listed in Sec. 141.61(a) (2) through (21) during each compliance period, 
beginning in the initial compliance period.
    (5) If the initial monitoring for contaminants listed in 
Sec. 141.61(a) (1)

[[Page 365]]

through (8) and the monitoring for the contaminants listed in 
Sec. 141.61(a) (9) through (21) as allowed in paragraph (f)(18) has been 
completed by December 31, 1992, and the system did not detect any 
contaminant listed in Sec. 141.61(a) (1) through (21), then each ground 
and surface water system shall take one sample annually beginning with 
the initial compliance period.
    (6) After a minimum of three years of annual sampling, the State may 
allow groundwater systems with no previous detection of any contaiminant 
listed in Sec. 141.61(a) to take one sample during each compliance 
period.
    (7) Each community and non-transient non-community ground water 
system which does not detect a contaminant listed in Sec. 141.61(a) (1) 
through (21) may apply to the State for a waiver from the requirements 
of paragraphs (f)(5) and (f)(6) of this section after completing the 
initial monitoring. (For purposes of this section, detection is defined 
as <0.0005 mg/l.) A waiver shall be effective for no more than six years 
(two compliance periods). States may also issue waivers to small systems 
for the initial round of monitoring for 1,2,4-trichlorobenzene.
    (8) A State may grant a waiver after evaluating the following 
factor(s):
    (i) Knowledge of previous use (including transport, storage, or 
disposal) of the contaminant within the watershed or zone of influence 
of the system. If a determination by the State reveals no previous use 
of the contaminant within the watershed or zone of influence, a waiver 
may be granted.
    (ii) If previous use of the contaminant is unknown or it has been 
used previously, then the following factors shall be used to determine 
whether a waiver is granted.
    (A) Previous analytical results.
    (B) The proximity of the system to a potential point or non-point 
source of contamination. Point sources include spills and leaks of 
chemicals at or near a water treatment facility or at manufacturing, 
distribution, or storage facilities, or from hazardous and municipal 
waste landfills and other waste handling or treatment facilities.
    (C) The environmental persistence and transport of the contaminants.
    (D) The number of persons served by the public water system and the 
proximity of a smaller system to a larger system.
    (E) How well the water source is protected against contamination, 
such as whether it is a surface or groundwater system. Groundwater 
systems must consider factors such as depth of the well, the type of 
soil, and wellhead protection. Surface water systems must consider 
watershed protection.
    (9) As a condition of the waiver a groundwater system must take one 
sample at each sampling point during the time the waiver is effective 
(i.e., one sample during two compliance periods or six years) and update 
its vulnerability assessment considering the factors listed in paragraph 
(f)(8) of this section. Based on this vulnerability assessment the State 
must reconfirm that the system is non-vulnerable. If the State does not 
make this reconfirmation within three years of the initial 
determination, then the waiver is invalidated and the system is required 
to sample annually as specified in paragraph (5) of this section.
    (10) Each community and non-transient non-community surface water 
system which does not detect a contaminant listed in Sec. 141.61(a) (1) 
through (21) may apply to the State for a waiver from the requirements 
of (f)(5) of this section after completing the initial monitoring. 
Composite samples from a maximum of five sampling points are allowed, 
provided that the detection limit of the method used for analysis is 
less than one-fifth of the MCL. Systems meeting this criterion must be 
determined by the State to be non-vulnerable based on a vulnerability 
assessment during each compliance period. Each system receiving a waiver 
shall sample at the frequency specified by the State (if any).
    (11) If a contaminant listed in Sec. 141.61(a) (2) through (21) is 
detected at a level exceeding 0.0005 mg/l in any sample, then:
    (i) The system must monitor quarterly at each sampling point which 
resulted in a detection.
    (ii) The State may decrease the quarterly monitoring requirement 
speci fied in paragraph (f)(11)(i) of this section provided it has 
determined that

[[Page 366]]

the system is reliably and consistently below the maximum contaminant 
level. In no case shall the State make this determination unless a 
groundwater system takes a minimum of two quarterly samples and a 
surface water system takes a minimum of four quarterly samples.
    (iii) If the State determines that the system is reliably and 
consistently below the MCL, the State may allow the system to monitor 
annually. Systems which monitor annually must monitor during the 
quarter(s) which previously yielded the highest analytical result.
    (iv) Systems which have three consecutive annual samples with no 
detection of a contaminant may apply to the State for a waiver as 
specified in paragraph (f)(7) of this section.
    (v) Groundwater systems which have detected one or more of the 
following two-carbon organic compounds: trichloroethylene, 
tetrachloroethylene, 1,2-dichloroethane, 1,1,1-trichloroethane, cis-1,2-
dichloroethylene, trans-1,2-dichloroethylene, or 1,1-dichloroethylene 
shall monitor quarterly for vinyl chloride. A vinyl chloride sample 
shall be taken at each sampling point at which one or more of the two-
carbon organic compounds was detected. If the results of the first 
analysis do not detect vinyl chloride, the State may reduce the 
quarterly monitoring frequency of vinyl chloride monitoring to one 
sample during each compliance period. Surface water systems are required 
to monitor for vinyl chloride as specified by the State.
    (12) Systems which violate the requirements of Sec. 141.61(a) (1) 
through (21), as determined by paragraph (f)(15) of this section, must 
monitor quarterly. After a minimum of four consecutive quarterly samples 
which show the system is in compliance as specified in paragraph (f)(15) 
of this section the system and the State determines that the system is 
reliably and consistently below the maximum contaminant level, the 
system may monitor at the frequency and times specified in paragraph 
(f)(11)(iii) of this section.
    (13) The State may require a confirmation sample for positive or 
negative results. If a confirmation sample is required by the State, the 
result must be averaged with the first sampling result and the average 
is used for the compliance determination as specified by paragraph 
(f)(15). States have discretion to delete results of obvious sampling 
errors from this calculation.
    (14) The State may reduce the total number of samples a system must 
analyze by allowing the use of compositing. Composite samples from a 
maximum of five sampling points are allowed, provided that the detection 
limit of the method used for analysis is less than one-fifth of the MCL. 
Compositing of samples must be done in the laboratory and analyzed 
within 14 days of sample collection.
    (i) If the concentration in the composite sample is greater than or 
equal to 0.0005 mg/l for any contaminant listed in Sec. 141.61(a), then 
a follow-up sample must be taken within 14 days at each sampling point 
included in the composite, and be analyzed for that contaminant.
    (ii) If duplicates of the original sample taken from each sampling 
point used in the composite sample are available, the system may use 
these instead of resampling. The duplicates must be analyzed and the 
results reported to the State within 14 days after completing analysis 
of the composite sample, provided the holding time of the sample is not 
exceeded.
    (iii) If the population served by the system is  3,300 
persons, then compositing may only be permitted by the State at sampling 
points within a single system. In systems serving [le] 3,300 persons, 
the State may permit compositing among different systems provided the 5-
sample limit is maintained.
    (iv) Compositing samples prior to GC analysis.
    (A) Add 5 ml or equal larger amounts of each sample (up to 5 samples 
are allowed) to a 25 ml glass syringe. Special precautions must be made 
to maintain zero headspace in the syringe.
    (B) The samples must be cooled at 4  deg.C during this step to 
minimize volatilization losses.
    (C) Mix well and draw out a 5-ml aliquot for analysis.
    (D) Follow sample introduction, purging, and desorption steps 
described in the method.

[[Page 367]]

    (E) If less than five samples are used for compositing, a 
proportionately small syringe may be used.
    (v) Compositing samples prior to GC/MS analysis.
    (A) Inject 5-ml or equal larger amounts of each aqueous sample (up 
to 5 samples are allowed) into a 25-ml purging device using the sample 
introduction technique described in the method.
    (B) The total volume of the sample in the purging device must be 25 
ml.
    (C) Purge and desorb as described in the method.
    (15) Compliance with Sec. 141.61(a) (1) through (21) shall be 
determined based on the analytical results obtained at each sampling 
point.
    (i) For systems which are conducting monitoring at a frequency 
greater than annual, compliance is determined by a running annual 
average of all samples taken at each sampling point. If the annual 
average of any sampling point is greater than the MCL, then the system 
is out of compliance. If the initial sample or a subsequent sample would 
cause the annual average to be exceeded, then the system is out of 
compliance immediately.
    (ii) If monitoring is conducted annually, or less frequently, the 
system is out of compliance if the level of a contaminant at any 
sampling point is greater than the MCL. If a confirmation sample is 
required by the State, the determination of compliance will be based on 
the average of two samples.
    (16) [Reserved]
    (17) Analysis under this section shall only be conducted by 
laboratories that are certified by EPA or the State according to the 
following conditions (laboratories may conduct sample analysis under 
provisional certification until January 1, 1996):
    (i) To receive certification to conduct analyses for the 
contaminants in Sec. 141.61(a) (2) through (21) the laboratory must:
    (A) Analyze Performance Evaluation (PE) samples provided by EPA, the 
State, or by a third party (with the approval of the State or EPA) at 
least once a year by each method for which the laboratory desires 
certification.
    (B) Achieve the quantitative acceptance limits under paragraphs 
(f)(17)(i)(C) and (D) of this section for at least 80 percent of the 
regulated organic contaminants included in the PE sample.
    (C) Achieve quantitative results on the analyses performed under 
paragraph (f)(17)(i)(A) of this section that are within 20% 
of the actual amount of the substances in the Performance Evaluation 
sample when the actual amount is greater than or equal to 0.010 mg/l.
    (D) Achieve quantitative results on the analyses performed under 
paragraph (f)(17)(i)(A) of this section that are within 40 
percent of the actual amount of the substances in the Performance 
Evaluation sample when the actual amount is less than 0.010 mg/l.
    (E) Achieve a method detection limit of 0.0005 mg/l, according to 
the procedures in appendix B of part 136.
    (ii) To receive certification to conduct analyses for vinyl 
chloride, the laboratory must:
    (A) Analyze Performance Evaluation (PE) samples provided by EPA, the 
State, or by a third party (with the approval of the State or EPA) at 
least once a year by each method for which the laboratory desires 
certification.
    (B) Achieve quantitative results on the analyses performed under 
paragraph (f)(17)(ii)(A) of this section that are within 40 
percent of the actual amount of vinyl chloride in the Performance 
Evaluation sample.
    (C) Achieve a method detection limit of 0.0005 mg/l, according to 
the procedures in appendix B of part 136.
    (D) Obtain certification for the contaminants listed in 
Sec. 141.61(a)(2) through (21).
    (18) States may allow the use of monitoring data collected after 
January 1, 1988, required under section 1445 of the Act for purposes of 
initial monitoring compliance. If the data are generally consistent with 
the other requirements of this section, the State may use these data 
(i.e., a single sample rather than four quarterly samples) to satisfy 
the initial monitoring requirement of paragraph (f)(4) of this section. 
Systems which use grandfathered samples and did not detect any 
contaminant listed Sec. 141.61(a)(2) through (21) shall begin monitoring 
annually in accordance

[[Page 368]]

with paragraph (f)(5) of this section beginning with the initial 
compliance period.
    (19) States may increase required monitoring where necessary to 
detect variations within the system.
    (20) Each certified laboratory must determine the method detection 
limit (MDL), as defined in appendix B to part 136, at which it is 
capable of detecting VOCs. The acceptable MDL is 0.0005 mg/l. This 
concentration is the detection concentration for purposes of this 
section.
    (21) Each public water system shall monitor at the time designated 
by the State within each compliance period.
    (g) [Reserved]
    (h) Analysis of the contaminants listed in Sec. 141.61(c) for the 
purposes of determining compliance with the maximum contaminant level 
shall be conducted as follows: \7\
---------------------------------------------------------------------------

    \7\ Monitoring for the contaminants aldicarb, aldicarb sulfoxide, 
and aldicarb sulfone shall be conducted in accordance with Sec. 141.40.
---------------------------------------------------------------------------

    (1) Groundwater systems shall take a minimum of one sample at every 
entry point to the distribution system which is representative of each 
well after treatment (hereafter called a sampling point). Each sample 
must be taken at the same sampling point unless conditions make another 
sampling point more representative of each source or treatment plant.
    (2) Surface water systems shall take a minimum of one sample at 
points in the distribution system that are representative of each source 
or at each entry point to the distribution system after treatment 
(hereafter called a sampling point). Each sample must be taken at the 
same sampling point unless conditions make another sampling point more 
representative of each source or treatment plant.
    Note: For purposes of this paragraph, surface water systems include 
systems with a combination of surface and ground sources.
    (3) If the system draws water from more than one source and the 
sources are combined before distribution, the system must sample at an 
entry point to the distribution system during periods of normal 
operating conditions (i.e., when water representative of all sources is 
being used).
    (4) Monitoring frequency:
    (i) Each community and non-transient non-community water system 
shall take four consecutive quarterly samples for each contaminant 
listed in Sec. 141.61(c) during each compliance period beginning with 
the initial compliance period.
    (ii) Systems serving more than 3,300 persons which do not detect a 
contaminant in the initial compliance period may reduce the sampling 
frequency to a minimum of two quarterly samples in one year during each 
repeat compliance period.
    (iii) Systems serving less than or equal to 3,300 persons which do 
not detect a contaminant in the initial compliance period may reduce the 
sampling frequency to a minimum of one sample during each repeat 
compliance period.
    (5) Each community and non-transient water system may apply to the 
State for a waiver from the requirement of paragraph (h)(4) of this 
section. A system must reapply for a waiver for each compliance period.
    (6) A State may grant a waiver after evaluating the following 
factor(s): Knowledge of previous use (including transport, storage, or 
disposal) of the contaminant within the watershed or zone of influence 
of the system. If a determination by the State reveals no previous use 
of the contaminant within the watershed or zone of influence, a waiver 
may be granted. If previous use of the contaminant is unknown or it has 
been used previously, then the following factors shall be used to 
determine whether a waiver is granted.
    (i) Previous analytical results.
    (ii) The proximity of the system to a potential point or non-point 
source of contamination. Point sources include spills and leaks of 
chemicals at or near a water treatment facility or at manufacturing, 
distribution, or storage facilities, or from hazardous and municipal 
waste landfills and other waste handling or treatment facilities. Non-
point sources include the use of pesticides to control insect and weed 
pests on agricultural areas, forest lands, home and gardens, and other 
land application uses.
    (iii) The environmental persistence and transport of the pesticide 
or PCBs.

[[Page 369]]

    (iv) How well the water source is protected against contamination 
due to such factors as depth of the well and the type of soil and the 
integrity of the well casing.
    (v) Elevated nitrate levels at the water supply source.
    (vi) Use of PCBs in equipment used in the production, storage, or 
distribution of water (i.e., PCBs used in pumps, transformers, etc.).
    (7) If an organic contaminant listed in Sec. 141.61(c) is detected 
(as defined by paragraph (h)(18) of this section) in any sample, then:
    (i) Each system must monitor quarterly at each sampling point which 
resulted in a detection.
    (ii) The State may decrease the quarterly monitoring requirement 
specified in paragraph (h)(7)(i) of this section provided it has 
determined that the system is reliably and consistently below the 
maximum contaminant level. In no case shall the State make this 
determination unless a groundwater system takes a minimum of two 
quarterly samples and a surface water system takes a minimum of four 
quarterly samples.
    (iii) After the State determines the system is reliably and 
consistently below the maximum contaminant level the State may allow the 
system to monitor annually. Systems which monitor annually must monitor 
during the quarter that previously yielded the highest analytical 
result.
    (iv) Systems which have 3 consecutive annual samples with no 
detection of a contaminant may apply to the State for a waiver as 
specified in paragraph (h)(6) of this section.
    (v) If monitoring results in detection of one or more of certain 
related contaminants (aldicarb, aldicarb sulfone, aldicarb sulfoxide and 
heptachlor, heptachlor epoxide), then subsequent monitoring shall 
analyze for all related contaminants.
    (8) Systems which violate the requirements of Sec. 141.61(c) as 
determined by paragraph (h)(11) of this section must monitor quarterly. 
After a minimum of four quarterly samples show the system is in 
compliance and the State determines the system is reliably and 
consistently below the MCL, as specified in paragraph (h)(11) of this 
section, the system shall monitor at the frequency specified in 
paragraph (h)(7)(iii) of this section.
    (9) The State may require a confirmation sample for positive or 
negative results. If a confirmation sample is required by the State, the 
result must be averaged with the first sampling result and the average 
used for the compliance determination as specified by paragraph (h)(11) 
of this section. States have discretion to delete results of obvious 
sampling errors from this calculation.
    (10) The State may reduce the total number of samples a system must 
analyze by allowing the use of compositing. Composite samples from a 
maximum of five sampling points are allowed, provided that the detection 
limit of the method used for analysis is less than one-fifth of the MCL. 
Compositing of samples must be done in the laboratory and analyzed 
within 14 days of sample collection.
    (i) If the concentration in the composite sample detects one or more 
contaminants listed in Sec. 141.61(c), then a follow-up sample must be 
taken within 14 days at each sampling point included in the composite, 
and be analyzed for that contaminant.
    (ii) If duplicates of the original sample taken from each sampling 
point used in the composite sample are available, the system may use 
these instead of resampling. The duplicates must be analyzed and the 
results reported to the State within 14 days after completion of the 
composite analysis or before the holding time for the initial sample is 
exceeded whichever is sooner.
    (iii) If the population served by the system is 3,300 
persons, then compositing may only be permitted by the State at sampling 
points within a single system. In systems serving [le] 3,300 persons, 
the State may permit compositing among different systems provided the 5-
sample limit is maintained.
    (11) Compliance with Sec. 141.61(c) shall be determined based on the 
analytical results obtained at each sampling point.
    (i) For systems which are conducting monitoring at a frequency 
greater than annual, compliance is determined by a running annual 
average of all samples

[[Page 370]]

taken at each sampling point. If the annual average of any sampling 
point is greater than the MCL, then the system is out of compliance. If 
the initial sample or a subsequent sample would cause the annual average 
to be exceeded, then the system is out of compliance immediately. Any 
samples below the detection limit shall be calculated as zero for 
purposes of determining the annual average.
    (ii) If monitoring is conducted annually, or less frequently, the 
system is out of compliance if the level of a contaminant at any 
sampling point is greater than the MCL. If a confirmation sample is 
required by the State, the determination of compliance will be based on 
the average of two samples.
    (12) [Reserved]
    (13) Analysis for PCBs shall be conducted as follows using the 
methods in paragraph (e) of this section:
    (i) Each system which monitors for PCBs shall analyze each sample 
using either Method 508.1, 525.2, 508 or 505. Users of Method 505 may 
have more difficulty in achieving the required Aroclor detection limits 
than users of Methods 508.1, 525.2 or 508.
    (ii) If PCBs (as one of seven Aroclors) are detected (as designated 
in this paragraph) in any sample analyzed using Method 505 or 508, the 
system shall reanalyze the sample using Method 508A to quantitate PCBs 
(as decachlorobiphenyl).

------------------------------------------------------------------------
                                                        Detection limit
                       Aroclor                               (mg/l)
------------------------------------------------------------------------
1016.................................................        0.00008
1221.................................................        0.02
1232.................................................        0.0005
1242.................................................        0.0003
1248.................................................        0.0001
1254.................................................        0.0001
1260.................................................        0.0002
------------------------------------------------------------------------

    (iii) Compliance with the PCB MCL shall be determined based upon the 
quantitative results of analyses using Method 508A.
    (14) If monitoring data collected after January 1, 1990, are 
generally consistent with the requirements of Sec. 141.24(h), then the 
State may allow systems to use that data to satisfy the monitoring 
requirement for the initial compliance period beginning January 1, 1993.
    (15) The State may increase the required monitoring frequency, where 
necessary, to detect variations within the system (e.g., fluctuations in 
concentration due to seasonal use, changes in water source).
    (16) The State has the authority to determine compliance or initiate 
enforcement action based upon analytical results and other information 
compiled by their sanctioned representatives and agencies.
    (17) Each public water system shall monitor at the time designated 
by the State within each compliance period.
    (18) Detection as used in this paragraph shall be defined as greater 
than or equal to the following concentrations for each contaminant.

------------------------------------------------------------------------
                                                              Detection
                        Contaminant                           limit (mg/
                                                                  l)
------------------------------------------------------------------------
Alachlor...................................................   .0002
Aldicarb...................................................   .0005
Aldicarb sulfoxide.........................................   .0005
Aldicarb sulfone...........................................   .0008
Atrazine...................................................   .0001
Benzo[a]pyrene.............................................   .00002
Carbofuran.................................................   .0009
Chlordane..................................................   .0002
Dalapon....................................................   .001
1,2-Dibromo-3-chloropropane (DBCP).........................   .00002
Di (2-ethylhexyl) adipate..................................   .0006
Di (2-ethylhexyl) phthalate................................   .0006
Dinoseb....................................................   .0002
Diquat.....................................................   .0004
2,4-D......................................................   .0001
Endothall..................................................   .009
Endrin.....................................................   .00001
Ethylene dibromide (EDB)...................................   .00001
Glyphosate.................................................   .006
Heptachlor.................................................   .00004
Heptachlor epoxide.........................................   .00002
Hexachlorobenzene..........................................   .0001
Hexachlorocyclopentadiene..................................   .0001
Lindane....................................................   .00002
Methoxychlor...............................................   .0001
Oxamyl.....................................................   .002
Picloram...................................................   .0001
Polychlorinated biphenyls (PCBs) (as decachlorobiphenyl)...   .0001
Pentachlorophenol..........................................   .00004
Simazine...................................................   .00007
Toxaphene..................................................   .001
2,3,7,8-TCDD (Dioxin)......................................   .000000005
2,4,5-TP (Silvex)..........................................   .0002
------------------------------------------------------------------------


    (19) Anaylsis under this section shall only be conducted by 
laboratories that have received certification by EPA or the State and 
have met the following conditions:
    (i) To receive certification to conduct analyses for the 
contaminants in Sec. 141.61(c) the laboratory must:
    (A) Analyze Performance Evaluation (PE) samples provided by EPA, the

[[Page 371]]

State, or by a third party (with the approval of the State or EPA) at 
least once a year by each method for which the laboratory desires 
certification.
    (B) For each contaminant that has been included in the PE sample 
achieve quantitative results on the analyses that are within the 
following acceptance limits:

------------------------------------------------------------------------
                Contaminant                  Acceptance limits (percent)
------------------------------------------------------------------------
DBCP......................................  40
EDB.......................................  40.
Alachlor..................................  45.
Atrazine..................................  45.
Benzo[a]pyrene............................  2 standard deviations.
Carbofuran................................  45.
Chlordane.................................  45.
Dalapon...................................  2 standard deviations.
Di(2-ethylhexyl)adipate...................  2 standard deviations.
Di(2-ethylhexyl)phthalate.................  2 standard deviations.
Dinoseb...................................  2 standard deviations.
Diquat....................................  2 standard deviations.
Endothall.................................  2 standard deviations.
Endrin....................................  30.
Glyphosate................................  2 standard deviations.
Heptachlor................................  45.
Heptachlor epoxide........................  45.
Hexachlorobenzene.........................  2 standard deviations.
Hexachloro- cyclopentadiene                 2 standard deviations.
Lindane...................................  45.
Methoxychlor..............................  45.
Oxamyl....................................  2 standard deviations.
PCBs (as Decachlorobiphenyl)                0-200.
Picloram..................................  2 standard deviations.
Simazine..................................  2 standard deviations.
Toxaphene.................................  45.
Aldicarb..................................  2 standard deviations.
Aldicarb sulfoxide........................  2 standard deviations.
Aldicarb sulfone..........................  2 standard deviations.
Pentachlorophenol.........................  50.
2,3,7,8-TCDD (Dioxin).....................  2 standard deviations.
2,4-D.....................................  50.
2,4,5-TP (Silvex).........................  50.
------------------------------------------------------------------------

    (ii) [Reserved]

(Approved by the Office of Management and Budget under control number 
2040-0090)

[40 FR 59570, Dec. 24, 1975, as amended at 44 FR 68641, Nov. 29, 1979; 
45 FR 57345, Aug. 27, 1980; 47 FR 10998, Mar. 12, 1982; 52 FR 25712, 
July 8, 1987; 53 FR 5147, Feb. 19, 1988; 53 FR 25110, July 1, 1988; 56 
FR 3583, Jan. 30, 1991; 56 FR 30277, July 1, 1991; 57 FR 22178, May 27, 
1992; 57 FR 31841, July 17, 1992; 59 FR 34323, July 1, 1994; 59 FR 
62468, Dec. 5, 1994; 60 FR 34085, June 29, 1995; 64 FR 67464, Dec. 1, 
1999; 65 FR 26022, May 4, 2000]

    Effective Date Note: At 66 FR 7063, Jan. 22, 2001, Sec. 141.24 was 
amended by adding a new sentence to the end of paragraph (f)(15) 
introductory text; revising paragraphs (f)(15)(i) and (f)(15)(ii) and 
adding new paragraphs (f)(15)(iii) through (f)(15)(v); adding paragraph 
(f)(22); adding a new sentence to the end of paragraph (h)(11) 
introductory text; revising paragraphs (h)(11)(i) and (h)(11)(ii) and 
adding new paragraphs (h)(11)(iii) through (h)(11)(v); and adding 
paragraph (h)(20), effective Mar. 23, 2001, except for the amendments to 
(f)(15), (h)(11), and (h)(20), which are effective Jan. 22, 2004. At 66 
FR 16134, Mar. 23, 2001, the effective date was delayed until May 22, 
2001. At 66 FR 28350, May 22, 2001, the effective date of (f)(22) was 
also delayed until Jan. 22, 2004. For the convenience of the user, the 
revised and added text is set forth as follows:

Sec. 141.24  Organic chemicals other than total trihalomethanes, 
          sampling and analytical methods.

    (f) * * *
    (15) * * * If one sampling point is in violation of an MCL, the 
system is in violation of the MCL.
    (i) For systems monitoring more than once per year, compliance with 
the MCL is determined by a running annual average at each sampling 
point.
    (ii) Systems monitoring annually or less frequently whose sample 
result exceeds the MCL must begin quarterly sampling. The system will 
not be considered in violation of the MCL until it has completed one 
year of quarterly sampling.
    (iii) If any sample result will cause the running annual average to 
exceed the MCL at any sampling point, the system is out of compliance 
with the MCL immediately.
    (iv) If a system fails to collect the required number of samples, 
compliance will be based on the total number of samples collected.
    (v) If a sample result is less than the detection limit, zero will 
be used to calculate the annual average.

                                * * * * *

    (22) All new systems or systems that use a new source of water that 
begin operation after January 22, 2004 must demonstrate compliance with 
the MCL within a period of time specified by the State. The system must 
also comply with the initial sampling frequencies specified by the State 
to ensure a system can demonstrate compliance with the MCL. Routine and 
increased monitoring frequencies shall be conducted in accordance with 
the requirements in this section.

                                * * * * *

    (h) * * *
    (11)* * * If one sampling point is in violation of an MCL, the 
system is in violation of the MCL.
    (i) For systems monitoring more than once per year, compliance with 
the MCL is determined by a running annual average at each sampling 
point.
    (ii) Systems monitoring annually or less frequently whose sample 
result exceeds the regulatory detection level as defined by

[[Page 372]]

paragraph (h)(18) of this section must begin quarterly sampling. The 
system will not be considered in violation of the MCL until it has 
completed one year of quarterly sampling.
    (iii) If any sample result will cause the running annual average to 
exceed the MCL at any sampling point, the system is out of compliance 
with the MCL immediately.
    (iv) If a system fails to collect the required number of samples, 
compliance will be based on the total number of samples collected.
    (v) If a sample result is less than the detection limit, zero will 
be used to calculate the annual average.

                                * * * * *

    (20) All new systems or systems that use a new source of water that 
begin operation after January 22, 2004 must demonstrate compliance with 
the MCL within a period of time specified by the State. The system must 
also comply with the initial sampling frequencies specified by the State 
to ensure a system can demonstrate compliance with the MCL. Routine and 
increased monitoring frequencies shall be conducted in accordance with 
the requirements in this section.