[Code of Federal Regulations]
[Title 40, Volume 28]
[Revised as of July 1, 2002]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR796.1050]

[Page 80-82]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 796--CHEMICAL FATE TESTING GUIDELINES--Table of Contents
 
               Subpart B--Physical and Chemical Properties
 
Sec. 796.1050  Absorption in aqueous solution: Ultraviolet/visible spectra.


    (a) Introductory information--(1) Guidance information. (i) 
Molecular formula.
    (ii) Structural formula.
    (2) Standard documents. The spectrophotometric method is based on 
national standards and consensus methods which are applied to measure 
the absorption spectra.
    (b) Method--(1)(i) Introduction, purpose, scope, relevance, 
application and limits of test. (A) The primary environmental purpose in 
determining the ultraviolet-visible (UV-VIS) absorption spectrum of a 
chemical compound is to have some indication of the wavelengths at which 
the compounds may be susceptible to photochemical degradation. Since 
photochemical degradation is likely to occur in both the atmosphere and 
the aquatic environment, spectra appropriate to these media will be 
informative concerning the need for further persistence testing.
    (B) Degradation will depend upon the total energy absorbed in 
specific wavelength regions. Such energy absorption is characterized by 
both molar absorption coefficient (molar extinction coefficient) and 
band width. However, the absence of measurable absorption does not 
preclude the possibility of photodegradation.
    (ii) Definitions and units. The UV-VIS absorption spectrum of a 
solution is a function of the concentration, c1, expressed in 
mol/L, of all absorbing species present; the path length, d, of the 
spectrophotometer cell, expressed in cm; and the molar absorption 
(extinction) coefficient, 
[]i, of each 
species. The absorbance (optical density) A of the solution is then 
given by:
[GRAPHIC] [TIFF OMITTED] TC15NO91.045


For a resolvable absorbance peak, the band width 
[] is the 
wavelength range, expressed in nm=10-9 m, of the peak at half 
the absorbance maximum.
    (iii) Reference substances. (A) The reference substances need not be 
employed in all cases when investigating a new substance. They are 
provided primarily so that calibration of the method may be performed 
from time to time and to offer the chance to compare the results when 
another method is applied.
    (B) Reference compounds appropriate for the calibration of the 
system are:
    (1) Potassium dichromate (in 0.005 mol/L, 
H2SO4 solution) from J.A.A. Ketelaar, paragraph 
(d)(2) of this section:

log []..............................
[] in nm..........................


    (2) Fluoranthene (in methanol) from C.R.C. Atlas of Spectral Data, 
paragraph (d)(3) of this section:

log []..............
[    237    236    288    339    357
 <] in nm...................


    (3) 4-nitrophenol (in methanol) from C.R.C. Atlas of Spectral Data, 
paragraph (d)(3) of this section:

log []....    3.88    4.04
[]     288     311
 in nm..................................................



See also paragraph (d)(1) of this section.
    (iv) Principle of the test method. This method utilizes a double-
beam spectrophotometer which records only the absorption differences 
between the blank and test solutions to give the spectrum of the 
chemical being tested.

[[Page 81]]

    (v) Quality criteria--Reproducibility and sensitivity. (A) 
Reproducibility and sensitivity, need not be measured directly. Instead, 
the accuracy of the system in measuring the spectra of reference 
compounds will be defined so as to assure appropriate reproducibility 
and sensitivity. It is preferable to use a recording double-beam 
spectrophotometer to obtain the UV-VIS spectrum of the test compound. 
Such an instrument should have a photometric accuracy of [plusmn]0.02 
units over the absorbance range of 0 to 2 units. It should be capable of 
recording absorbances at wavelengths of 200 to 750 nanometers nm with a 
wavelength accuracy of [plusmn]0.5 nm. The cells employed with the 
instrument must necessarily be transparent over this wavelength range 
and must have a path length determined to within 1 percent. To ensure 
that the instrument is performing satisfactorily, spectra for test 
solutions of K2Cr2O7 (for absorbance 
accuracy) and holmium glass (for wavelength accuracy) should be run 
periodically.
    (B) In the event that a recording double-beam instrument is not 
available, it will be necessary to determine the absorbance of the test 
solution in a single-beam instrument at 5-nm intervals over the entire 
wavelength range and at 1-nm intervals where there are indicated 
absorbance maxima. Wavelength and absorbance tests should be done as 
with the double-beam instrument.
    (2) Description of the test procedure--(i) Preparation--(A) 
Preparation of test solutions. (1) Solutions should be prepared by 
accurately weighing an appropriate amount of the purest form of the test 
substance available. This should be made up in a concentration which 
will result in at least one absorbance maximum in the range 0.5 to 1.5 
units.
    (2) The absorption of a compound is due to its particular chemical 
form. It is often the case that different forms are present, depending 
on whether the medium is acidic, basic, or neutral. Consequently, 
spectra under all three conditions are required where solubility and 
concentration allow. Where it is not possible to obtain sufficient 
concentrations in any of the aqueous media, a suitable organic solvent 
should be used (methanol preferred).
    (3) The acid medium should have a pH of less than 2, and the basic 
medium should be at least pH 10. The solvent for the neutral solution, 
and for preparing the acidic and basic ones, should be distilled water, 
transparent to ultraviolet radiation down to 200 nm. If methanol must be 
used, acidic and basic solutions can be prepared by adding 10 percent by 
volume of HCl or NaOH in aqueous solution ([HCl], [NaOH]=1 mol/L).
    (4) In theory, all chemical species other than that being tested are 
present in both beams and would therefore not appear in the recorded 
spectrum of a double-beam instrument. In practice, because the solvent 
is usually present in great excess, there is a threshold value of 
wavelength below which it is not possible to record the spectrum of the 
test chemical. Such a wavelength will be a property of the solvent or of 
the test medium. In general, distilled water is useful from 200 nm 
(dissolved ions will often increase this), methanol from 210 nm, hexane 
from 210 nm, acetonitrile from 215 nm and dichloromethane from 235 nm.
    (B) Blank solutions. A blank must be prepared which contains the 
solvent and all chemical species other than the test chemical. The 
absorption spectrum of this solution should be recorded in a manner 
identical to that of the test solution and preferably on the same chart. 
This ``baseline'' spectrum should never record an absorbance reading 
varying more than [plusmn]0.05 from the nominal zero value.
    (C) Cells. Cell pathlengths are usually between 0.1 cm and 10 cm. 
Cell lengths should be selected to permit recording of at least one 
maximum in the absorbance range of 0.5 to 1.5 units. Which set of cells 
should be used will be governed by the concentration and the absorbance 
of the test solution as indicated by the Beer-Lambert Law. The cells 
should be transparent over the range of the spectrum being recorded, and 
the path-lengths should be known to an accuracy of at least 1 per cent. 
Cells should be thoroughly cleaned in an appropriate manner (chromic 
acid is useful for quartz cells) and rinsed several times with the test 
or blank solutions.
    (ii) Performance of the test. Both cells to be employed should be 
rinsed with

[[Page 82]]

the blank solution and then filled with same. The instrument should be 
set to scan at a rate appropriate for the required wavelength resolution 
and the spectrum of the blank recorded. The sample cell should then be 
rinsed and filled with the test solution and the scanning repeated, 
preferably on the same spectrum chart, to display the baseline. The test 
should be carried out at 25 [deg]C.
    (c) Data and reporting--(1) Treatment of results. (i) The molar 
absorption coefficient [] 
should be calculated for all absorbance maxima of the test substance. 
The formula for this calculation is:
[GRAPHIC] [TIFF OMITTED] TC15NO91.046


where the quantities are as defined above (see Definitions and units).
    (ii) For each peak which is capable of being resolved, either as 
recorded or by extrapolated symmetrical peaks, the bandwidth should be 
recorded.
    (2) Test report. (i) The report should contain a copy of each of the 
three spectra (3 pH conditions). If neither water nor methanol solutions 
are feasible, there will be only one spectrum. Spectra should include a 
readable wave-length scale. Each spectrum should be clearly marked with 
the test conditions.
    (ii) For each maximum in each spectrum, the 
[] value and bandwidth 
(when applicable) should be calculated and reported, along with the 
wavelength of the maximum. This should be presented in tabular form.
    (iii) The various test conditions should be included, such as scan 
speed, the name and model of the spectrophotom-eter, the slit width 
(where available), cell type and path length, the concentrations of the 
test substance, and the nature and acidity of the solvent medium. A 
recent test spectrum on appropriate reference materials for photometric 
and wavelength accuracy should also be submitted (see Reproducibility 
and sensitivity).
    (d) Literature references. For additional background information on 
this test guideline, the following references should be consulted:
    (1) Milazzo, G., Caroli, S., Palumbo-Doretti, M., Violante, N., 
Analytical Chemistry, 49: 711 (1977).
    (2) Katelaar, J.A.A., Photoelectric Spectrometry Group Bulletin, 8, 
(Cambridge, 1955).
    (3) Chemical Rubber Company, Atlas of Spectral Data, (Cliffland, 
Ohio).

[50 FR 39472, Sept. 27, 1985]