[Code of Federal Regulations]
[Title 40, Volume 18]
[Revised as of July 1, 2003]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR92.115]
[Page 443]
TITLE 40--PROTECTION OF ENVIRONMENT
CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
PART 92--CONTROL OF AIR POLLUTION FROM LOCOMOTIVES AND LOCOMOTIVE ENGINES--
Table of Contents
Subpart B--Test Procedures
Sec. 92.115 Calibrations; frequency and overview.
(a) Calibrations shall be performed as specified in Secs. 92.116
through 92.122.
(b) At least monthly or after any maintenance which could alter
calibration, perform the periodic calibrations required by
Sec. 92.118(a)(2) (certain analyzers may require more frequent
calibration depending on the equipment and use). Exception: the water
rejection ratio and the CO2 rejection ratio on all NDIR
analyzers is only required to be performed quarterly.
(c) At least monthly or after any maintenance which could alter
calibration, calibrate the engine dynamometer flywheel torque and speed
measurement transducers, as specified in Sec. 92.116.
(d) At least monthly or after any maintenance which could alter
calibration, check the oxides of nitrogen converter efficiency, as
specified in Sec. 92.121.
(e) At least weekly or after any maintenance which could alter
calibration, check the dynamometer (if used) shaft torque feedback
signal at steady-state conditions by comparing:
(1) Shaft torque feedback to dynamometer beam load; or
(2) By comparing in-line torque to armature current; or
(3) By checking the in-line torque meter with a dead weight per
Sec. 92.116(b)(1).
(f) At least quarterly or after any maintenance which could alter
calibration, calibrate the fuel flow measurement system as specified in
Sec. 92.107.
(g) At least annually or after any maintenance which could alter
calibration, calibrate the electrical output measurement system for the
electrical load bank used for locomotive testing.
(h) Sample conditioning columns, if used in the CO analyzer train,
should be checked at a frequency consistent with observed column life or
when the indicator of the column packing begins to show deterioration.
(i) For equipment not addressed in Secs. 92.116 through 92.122
calibrations shall be performed at least as often as required by the
equipment manufacturer or as necessary according to good practices. The
calibrations shall be performed in accordance with procedures specified
by the equipment manufacturer.
(j) Where testing is conducted intermittently, calibrations are not
required during period in which no testing is conducted, provided that
times between the most recent calibrations and the date of any test does
not exceed the calibration period. For example, if it has been more than
one month since the analyzers have been calibrated (as specified in
paragraph (c) of this section) then they must be calibrated prior to the
start of testing.