[Code of Federal Regulations]
[Title 40, Volume 18]
[Revised as of July 1, 2003]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR92.116]

[Page 443-445]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 92--CONTROL OF AIR POLLUTION FROM LOCOMOTIVES AND LOCOMOTIVE ENGINES--
Table of Contents
 
                       Subpart B--Test Procedures
 
Sec. 92.116  Engine output measurement system calibrations.

    (a) General requirements for dynamometer calibration. (1) The engine 
flywheel torque and engine speed measurement transducers shall be 
calibrated with the calibration equipment described in this section.
    (2) The engine flywheel torque feedback signals to the cycle 
verification equipment shall be electronically checked before each test, 
and adjusted as necessary.
    (3) Other engine dynamometer system calibrations shall be performed 
as dictated by good engineering practice.
    (4) When calibrating the engine flywheel torque transducer, any 
lever arm used to convert a weight or a force through a distance into a 
torque shall be used in a horizontal position (5 degrees).
    (5) Calibrated resistors may not be used for engine flywheel torque 
transducer calibration, but may be used to span the transducer prior to 
engine testing.
    (b) Dynamometer calibration equipment--(1) Torque calibration 
equipment. Two techniques are allowed for torque calibration. Alternate 
techniques may be used if shown to yield equivalent accuracies. The NIST 
``true'' value torque is defined as the torque calculated by taking the 
product of an NIST traceable weight or force and a sufficiently accurate 
horizontal lever arm distance, corrected for the hanging torque of the 
lever arm.
    (i) The lever-arm dead-weight technique involves the placement of 
known weights at a known horizontal distance from the center of rotation 
of the

[[Page 444]]

torque measuring device. The equipment required is:
    (A) Calibration weights. A minimum of six calibration weights for 
each range of torque measuring device used are required. The weights 
must be approximately equally spaced and each must be traceable to NIST 
weights within 0.1 percent. Laboratories located in foreign countries 
may certify calibration weights to local government bureau standards. 
Certification of weight by state government Bureau of Weights and 
Measures is acceptable. Effects of changes in gravitational constant at 
the test site may be accounted for if desired.
    (B) Lever arm. A lever arm with a minimum length of 24 inches is 
required. The horizontal distance from the centerline of the engine 
torque measurement device to the point of weight application shall be 
accurate to within 0.10 inches. The arm must be balanced, or 
the hanging torque of the arm must be known to within 0.1 
ft-lbs.
    (ii) The transfer technique involves the calibration of a master 
load cell (i.e., dynamometer case load cell). This calibration can be 
done with known calibration weights at known horizontal distances, or by 
using a hydraulically actuated precalibrated master load cell. This 
calibration is then transferred to the flywheel torque measuring device. 
The technique involves the following steps:
    (A) A master load cell shall be either precalibrated or be 
calibrated per paragraph (b)(1)(i)(A) of this section with known weights 
traceable to NIST within 0.1 percent, and used with the lever arm(s) 
specified in this section. The dynamometer should be either running or 
vibrated during this calibration to minimize static hysteresis.
    (B) Transfer of calibration from the case or master load cell to the 
flywheel torque measuring device shall be performed with the dynamometer 
operating at a constant speed. The flywheel torque measurement device 
readout shall be calibrated to the master load cell torque readout at a 
minimum of six loads approximately equally spaced across the full useful 
ranges of both measurement devices. (Note that good engineering practice 
requires that both devices have approximately equal useful ranges of 
torque measurement.) The transfer calibration shall be performed in a 
manner such that the accuracy requirements of Sec. 92.106(b)(1)(ii) for 
the flywheel torque measurement device readout be met or exceeded.
    (iii) Other techniques may be used if shown to yield equivalent 
accuracy.
    (2) Speed calibration equipment. A 60 (or greater) tooth wheel in 
combination with a common mode rejection frequency counter is considered 
an absolute standard for engine or dynamometer speed.
    (c) Dynamometer calibration. (1) If necessary, follow the 
manufacturer's instructions for initial start-up and basic operating 
adjustments.
    (2) Check the dynamometer torque measurement for each range used by 
the following:
    (i) Warm up the dynamometer following the equipment manufacturer's 
specifications.
    (ii) Determine the dynamometer calibration moment arm. Equipment 
manufacturer's data, actual measurement, or the value recorded from the 
previous calibration used for this subpart may be used.
    (iii) Calculate the indicated torque (IT) for each calibration 
weight to be used by:

IT=calibration weight (lb)xcalibration moment arm (ft)

    (iv) Attach each calibration weight specified in paragraph 
(b)(1)(i)(A) of this section to the moment arm at the calibration 
distance determined in paragraph (b)(2)(ii)(B) of this section. Record 
the power measurement equipment response (ft-lb) to each weight.
    (v) For each calibration weight, compare the torque value measured 
in paragraph (b)(2)(iv) of this section to the calculated torque 
determined in paragraph (b)(2)(iii) of this section.
    (vi) The measured torque must be within 2 percent of the calculated 
torque.
    (vii) If the measured torque is not within 2 percent of the 
calculated torque, adjust or repair the system. Repeat the steps in 
paragraphs (b)(2)(i) through (b)(2)(vi) of this section with the 
adjusted or repaired system.

[[Page 445]]

    (3) Option. A master load-cell or transfer standard may be used to 
verify the in-use torque measurement system.
    (i) The master load-cell and read out system must be calibrated with 
weights at each test weight specified in paragraph (b)(1)(i)(A) of this 
section. The calibration weights must be traceable to within 0.1 percent 
of NIST weights.
    (ii) Warm up the dynamometer following the equipment manufacturer's 
specifications.
    (iii) Attach the master load-cell and loading system.
    (iv) Load the dynamometer to a minimum of 6 equally spaced torque 
values as indicated by the master load-cell for each in-use range used.
    (v) The in-use torque measurement must be within 2 percent of the 
torque measured by the master system for each load used.
    (vi) If the in-use torque is not within 2 percent of the master 
torque, adjust or repair the system. Repeat steps in paragraphs 
(b)(3)(ii) through (b)(3)(vi) of this section with the adjusted or 
repaired system.
    (4) The dynamometer calibration must be completed within 2 hours 
from the completion of the dynamometer warm-up.
    (d) Electrical load banks. Equipment used to measure the electrical 
power output dissipated by electrical load banks shall be calibrated as 
frequently as required by Sec. 92.115, using a calibration procedure 
that is consistent with good engineering practice and approved by the 
Administrator.