[Code of Federal Regulations]
[Title 40, Volume 18]
[Revised as of July 1, 2003]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR92.126]

[Page 458-459]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 92--CONTROL OF AIR POLLUTION FROM LOCOMOTIVES AND LOCOMOTIVE ENGINES--
Table of Contents
 
                       Subpart B--Test Procedures
 
Sec. 92.126  Test run.

    (a) The following steps shall be taken for each test:
    (1) Prepare the locomotive, engine, dynamometer, (as applicable) and 
sampling system for the test. Change filters, etc. and leak check as 
necessary.
    (2) Connect sampling equipment as appropriate for the sampling 
procedure employed; i.e. raw or dilute (evacuated sample collection 
bags, particulate, and raw exhaust sampling equipment, particulate 
sample filters, fuel flow measurement equipment, etc.).
    (3) Start the particulate dilution tunnel, the sample pumps, the 
engine cooling fan(s) (engine dynamometer testing) and the data 
collection and sampling systems (except particulate sample collection). 
The heated components of any continuous sampling systems(s) (if 
applicable) shall be preheated to their designated operating 
temperatures before the test begins.
    (4) Adjust the sample flow rates to the desired flow rates and set 
gas flow measuring devices to zero (particulate dilution tunnel).
    (5) Read and record all required general and pre-test data (i.e., 
all required data other than data that can only be collected during or 
after the emission test).
    (6) Warm-up the locomotive or locomotive engines according to normal 
warm-up procedures.
    (7) Begin the EPA Test Sequence for Locomotives and Locomotive 
Engines (see Sec. 92.124). Record all required general and test data 
throughout the duration of the test sequence.
    (i) Mark the start of the EPA Test Sequence for Locomotives and 
Locomotive Engines on all data records.
    (ii) Begin emission measurement after completing the warmup phase of 
the EPA Test Sequence for Locomotives and Locomotive Engines, as 
specified in paragraph (b) of this section. Mark the start and end of 
each mode on all data records.
    (iii) A mode shall be voided where the requirements of this subpart 
that apply to that test mode are not met. This includes the following:
    (A) The data acquisition is terminated prematurely; or
    (B) For engine testing, the engine speed or power output exceeds the 
tolerance bands established for that mode; or
    (C) Measured concentrations exceed the range of the instrument; or
    (D) The test equipment malfunctions.
    (iv) Modes within the test sequence shall be repeated if it is 
voided during the performance of the test sequence. A mode can be 
repeated by:
    (A) Repeating the two preceding modes and then continuing with the 
test sequence, provided that the locomotive or engine is not shut down 
after the voided test mode; or

[[Page 459]]

    (B) Repeating the preceding mode and then continuing with the test 
sequence from that point, provided that the locomotive or engine is not 
operated in any mode with lower power than the preceding mode after the 
voided test mode. For example, if the Notch 2 mode is voided, then the 
locomotive or engine would be returned to Notch 1 while any repairs are 
made.
    (b) Sampling and measurement timing. (1) Gaseous emissions shall be 
sampled and measured continuously.
    (2)(i) Sampling of particulate emissions from the raw exhaust (for 
dilution) shall be conducted continuously.
    (ii) Sampling of particulates from the diluted exhaust shall begin 
within ten seconds after the beginning of each test mode, and shall end 
six minutes after the beginning of each test mode.
    (iii) Sampling of CO2 in the dilution air and diluted 
exhaust does not need to be continuous, but the measurements used for 
the calculations must be made after the first two minutes of each mode.
    (3) Fuel flow rate shall be measured continuously. The value 
reported for the fuel flow rate shall be a one-minute average of the 
instantaneous fuel flow measurements taken during the last minute of the 
minimum sampling period listed in Table B124-1 in Sec. 92.124; except 
for testing during idle modes, where it shall be a three-minute average 
of the instantaneous fuel flow measurements taken during the last three 
minutes of the minimum sampling period listed in Table B124-1 in 
Sec. 92.124. Sampling periods greater than one minute, but no greater 
than three minutes are allowed for modes 2, 3, and 4, where required by 
good engineering practice.
    (4) Engine power shall be measured continuously. The value reported 
for the engine power shall be a one-minute average of the instantaneous 
power measurements taken during the last minute of the minimum sampling 
period listed in Table B124-1 in Sec. 92.124.
    (c) Exhaust gas measurements. (1) Should the analyzer response 
exceed 100 percent of full scale or respond less than 15 percent of full 
scale, the next higher or lower analyzer range shall be used.
    (2) Each analyzer range that may be used during a test sequence must 
have the zero and span responses recorded prior to the execution of the 
test sequence. Only the range(s) used to measure the emissions during a 
test sequence are required to have their zero and span recorded after 
the completion of the test sequence.
    (3) It is permitted to change filter elements between test modes, 
provided such changes do not cause a mode to be voided.
    (4) A leak check is permitted between test modes, provided such 
changes do not cause a mode to be voided.
    (5) A hang-up check is permitted between test modes, provided such 
changes do not cause a mode to be voided.
    (6) If, during the emission measurement portions of a test, the 
value of the gauges downstream of the NDIR analyzer(s) differs by more 
than 2 inches of water from the pretest value, the test is 
void.
    (7)(i) For bag samples, as soon as possible transfer the exhaust and 
dilution air bag samples to the analytical system and process the 
samples.
    (ii) A stabilized reading of the exhaust sample bag on all 
applicable analyzers shall be made within 20 minutes of the end of the 
sample collection phase of the mode.