[Code of Federal Regulations]
[Title 21, Volume 7]
[Revised as of April 1, 2004]
From the U.S. Government Printing Office via GPO Access
[CITE: 21CFR606.3]

[Page 45-46]
 
                        TITLE 21--FOOD AND DRUGS
 
CHAPTER I--FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN 
                          SERVICES (CONTINUED)
 
PART 606_CURRENT GOOD MANUFACTURING PRACTICE FOR BLOOD AND BLOOD 
COMPONENTS--Table of Contents
 
                      Subpart A_General Provisions
 
Sec. 606.3  Definitions.




                      Subpart A_General Provisions

Sec.
606.3 Definitions.

                  Subpart B_Organization and Personnel

606.20 Personnel.

                     Subpart C_Plant and Facilities

606.40 Facilities.

                           Subpart D_Equipment

606.60 Equipment.
606.65 Supplies and reagents.

Subpart E [Reserved]



                Subpart F_Production and Process Controls

606.100 Standard operating procedures.
606.110 Plateletpheresis, leukapheresis, and plasmapheresis.

                   Subpart G_Finished Product Control

606.120 Labeling, general requirements.
606.121 Container label.
606.122 Instruction circular.

                      Subpart H_Laboratory Controls

606.140 Laboratory controls.
606.151 Compatibility testing.

                      Subpart I_Records and Reports

606.160 Records.
606.165 Distribution and receipt; procedures and records.
606.170 Adverse reaction file.
606.171 Reporting of product deviations by licensed manufacturers, 
          unlicenses registered blood establishments, and transfusion 
          services.

    Authority: 21 U.S.C. 321, 331, 351, 352, 355, 360, 360j, 371, 374; 
42 U.S.C. 216, 262, 263a, 264.

    Source: 40 FR 53532, Nov. 18, 1975, unless otherwise noted.



    As used in this part:

[[Page 46]]

    (a) Blood means whole blood collected from a single donor and 
processed either for transfusion or further manufacturing.
    (b) Unit means the volume of blood or one of its components in a 
suitable volume of anticoagulant obtained from a single collection of 
blood from one donor.
    (c) Component means that part of a single-donor's blood separated by 
physical or mechanical means.
    (d) Plasma for further manufacturing means that liquid portion of 
blood separated and used as material to prepare another product.
    (e) Plasmapheresis means the procedure in which blood is removed 
from the donor, the plasma is separated from the formed elements and at 
least the red blood cells are returned to the donor.
    (f) Plateletpheresis means the procedure in which blood is removed 
from a donor, a platelet concentrate is separated, and the remaining 
formed elements are returned to the donor along with a portion of the 
residual plasma.
    (g) Leukapheresis means the procedure in which blood is removed from 
the donor, a leukocyte concentrate is separated, and the remaining 
formed elements and residual plasma are returned to the donor.
    (h) Facilities means any area used for the collection, processing, 
compatibility testing, storage or distribution of blood and blood 
components.
    (i) Processing means any procedure employed after collection and 
before compatibility testing of blood and includes the identification of 
a unit of donor blood, the preparation of components from such unit of 
donor blood, serological testing, labeling and associated recordkeeping.
    (j) Compatibility testing means the procedures performed to 
establish the matching of a donor's blood or blood components with that 
of a potential recipient.
    (k) Distributed means:
    (1) The blood or blood components have left the control of the 
licensed manufacturer, unlicensed registered blood establishment, or 
transfusion service; or
    (2) The licensed manufacturer has provided Source Plasma or any 
other blood component for use in the manufacture of a licensed 
biological product.
    (l) Control means having responsibility for maintaining the 
continued safety, purity, and potency of the product and for compliance 
with applicable product and establishment standards, and for compliance 
with current good manufacturing practices.

[40 FR 53532, Nov. 18, 1975, as amended at 64 FR 45370, Aug. 19, 1999; 
65 FR 66635, Nov. 7, 2000; 66 FR 1835, Jan. 10, 2001; 66 FR 40889, Aug. 
6, 2001]