[Code of Federal Regulations]
[Title 40, Volume 30]
[Revised as of July 1, 2004]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR1065.250]

[Page 662-663]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 1065_TEST PROCEDURES AND EQUIPMENT--Table of Contents
 
                Subpart C_Test Fuels and Analytical Gases
 
Sec. 1065.250  Analytical gases.

    Analytical gases that you use to comply with this part must meet the 
accuracy and purity specifications of this section. You must record the 
expiration date specified by the gas supplier and may not use any gas 
after the expiration date.
    (a) Pure gases. Use the ``pure gases'' shown in the following table:

                                             Table 1 of Sec. 1065.250--Concentration Limits for Pure Gases
--------------------------------------------------------------------------------------------------------------------------------------------------------
                                                                Maximum contaminant concentrations
             Gas type              -------------------------------------------------------------------------------------------       Oxygen content
                                        Organic carbon        Carbon monoxide         Carbon dioxide       Nitric oxide (NO)
--------------------------------------------------------------------------------------------------------------------------------------------------------
Purified Nitrogen.................  1 ppmC...............  1 ppm................  400 ppm..............  0.1 ppm.............  NA.
-----------------------------------
Purified Oxygen...................  NA...................  NA...................  NA...................  NA..................  99.5-100.0%.
-----------------------------------
Purified Synthetic Air, or Zero-    1 ppmC...............  1 ppm................  400 ppm..............  0.1 ppm.............  18-21%.
 Grade Air.
--------------------------------------------------------------------------------------------------------------------------------------------------------

    (b) Fuel for flame ionization detectors. Use a hydrogen-helium 
mixture as the fuel. Make sure the mixture contains 40  2 percent hydrogen and no more than 1 ppmC of organic 
carbon or 400 ppm of CO2.
    (c) Calibration and span gases. Apply the following provisions to 
calibration and span gases:
    (1) Use the following gas mixtures, as applicable, for calibrating 
and spanning your analytical instruments:
    (i) Propane in purified synthetic air. You may ask us to allow you 
to use propane in purified nitrogen for high concentrations of propane.
    (ii) CO in purified nitrogen.

[[Page 663]]

    (iii) NO and NO2 in purified nitrogen (the amount of 
NO2 in this calibration gas must not exceed 5 percent of the 
NO content).
    (iv) Oxygen in purified nitrogen.
    (v) CO2 in purified nitrogen.
    (vi) Methane in purified synthetic air.
    (2) The calibration gases in paragraph (c)(1) of this section must 
be traceable to within one percent of NIST gas standards or other gas 
standards we have approved. Span gases in paragraph (c)(1) of this 
section must be accurate to within two percent of true concentration, 
where true concentration refers to NIST gas standards, or other gas 
standards we have approved. Record concentrations of calibration gas as 
volume percent or volume ppm.
    (3) You may use gases for species other than those in paragraph 
(c)(1) of this section (such as methanol in air gases used to determine 
response factors), as long as they meet the following criteria:
    (i) They are traceable to within 2 percent of 
NIST gas standards or other standards we have approved.
    (ii) They remain within 2 percent of the 
labeled concentration. Show this by measuring quarterly with a precision 
of 2 percent (two standard deviations) or by using 
another method we approve. You may take multiple measurements. If the 
true concentration of the gas changes by more than two percent, but less 
than ten percent, you may relabel the gas with the new concentration.
    (4) You may generate calibration and span gases using precision 
blending devices (gas dividers) to dilute gases with purified nitrogen 
or with purified synthetic air. Make sure the mixing device produces a 
concentration of blended calibration gases that is accurate to within 
 1.5 percent. To do so, you must know the 
concentration of primary gases used for blending to an accuracy of at 
least  1 percent, traceable to NIST gas standards 
or other gas standards we have approved. For each calibration 
incorporating a blending device, verify the blending accuracy between 15 
and 50 percent of full scale. You may optionally check the blending 
device with an instrument that is linear by nature (for example, using 
NO gas with a CLD). Adjust the instrument's span value with the span gas 
connected directly to it. Check the blending device at the used settings 
to ensure that the difference between nominal values and measured 
concentrations at each point stays within  0.5 
percent of the nominal value.
    (d) Oxygen interference gases. Gases to check oxygen interference 
are mixtures of oxygen, nitrogen, and propane. The oxygen concentration 
must be 20-22 percent and the propane concentration must be 50-90 
percent of the maximum value in the most typically used FID range. 
Independently measure the concentration of total hydrocarbons plus 
impurities by chromatographic analysis or by dynamic blending.