[Code of Federal Regulations]

[Title 40, Volume 22]

[Revised as of July 1, 2005]

From the U.S. Government Printing Office via GPO Access

[CITE: 40CFR141.23]



[Page 362-372]

 

                   TITLE 40--PROTECTION OF ENVIRONMENT

 

         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)

 

PART 141_NATIONAL PRIMARY DRINKING WATER REGULATIONS--Table of Contents

 

            Subpart C_Monitoring and Analytical Requirements

 

Sec. 141.23  Inorganic chemical sampling and analytical requirements.



    Community water systems shall conduct monitoring to determine 

compliance with the maximum contaminant levels specified in Sec. 141.62 

in accordance with this section. Non-transient, non-community water 

systems shall conduct monitoring to determine compliance with the 

maximum contaminant levels specified in Sec. 141.62 in accordance with 

this section. Transient, non-community water systems shall conduct 

monitoring to determine compliance with the nitrate and nitrite maximum 

contaminant levels in Sec. Sec. 141.11 and 141.62 (as appropriate) in 

accordance with this section.

    (a) Monitoring shall be conducted as follows:

    (1) Groundwater systems shall take a minimum of one sample at every 

entry point to the distribution system which is representative of each 

well after treatment (hereafter called a sampling point) beginning in 

the initial compliance period. The system shall take each sample at the 

same sampling point unless conditions make another sampling point more 

representative of each source or treatment plant.

    (2) Surface water systems shall take a minimum of one sample at 

every entry point to the distribution system after any application of 

treatment or in the distribution system at a point which is 

representative of each source after treatment (hereafter called a 

sampling point) beginning in the initial compliance period. The system 

shall take each sample at the same sampling point unless conditions make 

another sampling point more representative of each source or treatment 

plant.

    Note: For purposes of this paragraph, surface water systems include 

systems with a combination of surface and ground sources.

    (3) If a system draws water from more than one source and the 

sources are combined before distribution, the system must sample at an 

entry point to the distribution system during periods of normal 

operating conditions (i.e., when water is representative of all sources 

being used).

    (4) The State may reduce the total number of samples which must be 

analyzed by allowing the use of compositing. Composite samples from a 

maximum of five samples are allowed, provided that the detection limit 

of the method used for analysis is less than one-fifth of the MCL. 

Compositing of samples must be done in the laboratory.

    (i) If the concentration in the composite sample is greater than or 

equal to one-fifth of the MCL of any inorganic chemical, then a follow-

up sample must be taken within 14 days at each sampling point included 

in the composite. These samples must be analyzed for the contaminants 

which exceeded one-fifth of the MCL in the composite sample. Detection 

limits for each analytical method and MCLs for each inorganic 

contaminant are the following:



               Detection Limits for Inorganic Contaminants

------------------------------------------------------------------------

                                                             Detection

        Contaminant          MCL (mg/l)     Methodology     limit (mg/l)

------------------------------------------------------------------------

Antimony..................  0.006......  Atomic            0.003

                                          Absorption;

                                          Furnace.

                                         Atomic            0.0008 \5\

                                          Absorption;

                                          Platform.

                                         ICP-Mass          0.0004

                                          Spectrometry.

                                         Hydride-Atomic    0.001

                                          Absorption.

Arsenic...................  0.010 \6\..  Atomic            0.001

                                          Absorption;

                                          Furnace.

                                         Atomic            0.0005 \7\

                                          Absorption;

                                          Platform--Stabi

                                          lized

                                          Temperature.



[[Page 363]]





                                         Atomic            0.001

                                          Absorption;

                                          Gaseous Hydride.

                                         ICP-Mass          0.0014 \8\

                                          Spectrometry.

Asbestos..................  7 MFL \1\..  Transmission      0.01 MFL

                                          Electron

                                          Microscopy.

Barium....................  2..........  Atomic            0.002

                                          Absorption;

                                          furnace

                                          technique.

                                         Atomic            0.1

                                          Absorption;

                                          direct

                                          aspiration.

                                         Inductively       0.002 (0.001)

                                          Coupled Plasma.

Beryllium.................  0.004......  Atomic            0.0002

                                          Absorption;

                                          Furnace.

                                         Atomic            0.00002 \5\

                                          Absorption;

                                          Platform.

                                         Inductively       0.0003

                                          Coupled Plasma

                                          \2\.

                                         ICP-Mass          0.0003

                                          Spectrometry.

Cadmium...................  0.005......  Atomic            0.0001

                                          Absorption;

                                          furnace

                                          technique.

                                         Inductively       0.001

                                          Coupled Plasma.

Chromium..................  0.1........  Atomic            0.001

                                          Absorption;

                                          furnace

                                          technique.

                                         Inductively       0.007 (0.001)

                                          Coupled Plasma.

Cyanide...................  0.2........  Distillation,     0.02

                                          Spectrophotomet

                                          ric \3\.

                            ...........  Distillation,     0.005

                                          Automated,

                                          Spectrophotomet

                                          ric \3\.

                            ...........  Distillation,     0.02

                                          Amenable,

                                          Spectrophotomet

                                          ric \4\.

                            ...........  Distillation,     0.05

                                          Selective

                                          Electrode \3\.

                            ...........  UV,               0.0005

                                          Distillation,

                                          Spectrophotomet

                                          ric.

                            ...........  Distillation,     0.0006

                                          Spectrophotomet

                                          ric.

Mercury...................  0.002......  Manual Cold       0.0002

                                          Vapor Technique.

                                         Automated Cold    0.0002

                                          Vapor Technique.

Nickel....................  xl.........  Atomic            0.001

                                          Absorption;

                                          Furnace.

                                         Atomic            0.0006 \5\

                                          Absorption;

                                          Platform.

                                         Inductively       0.005

                                          Coupled Plasma

                                          \2\.

                                         ICP-Mass          0.0005

                                          Spectrometry.

Nitrate...................  10 (as N)..  Manual Cadmium    0.01

                                          Reduction.

                                         Automated         0.01

                                          Hydrazine

                                          Reduction.

                                         Automated         0.05

                                          Cadmium

                                          Reduction.

                                         Ion Selective     1

                                          Electrode.

                                         Ion               0.01

                                          Chromatography.

Nitrite...................  1 (as N)...  Spectrophotometr  0.01

                                          ic.

                                         Automated         0.05

                                          Cadmium

                                          Reduction.

                                         Manual Cadmium    0.01

                                          Reduction.

                                         Ion               0.004

                                          Chromatography.

Selenium..................  0.05.......  Atomic            0.002

                                          Absorption;

                                          furnace.

                                         Atomic            0.002

                                          Absorption;

                                          gaseous hydride.

Thallium..................  0.002......  Atomic            0.001

                                          Absorption;

                                          Furnace.

                                         Atomic            0.0007 \5\

                                          Absorption;

                                          Platform.

                                         ICP-Mass          0.0003

                                          Spectrometry.

------------------------------------------------------------------------

\1\ MFL = million fibers per liter 10 [micro]m.

\2\ Using a 2X preconcentration step as noted in Method 200.7. Lower

  MDLs may be achieved when using a 4X preconcentration.

\3\ Screening method for total cyanides.

\4\ Measures ``free'' cyanides.

\5\ Lower MDLs are reported using stabilized temperature graphite

  furnace atomic absorption.

\6\ The value for arsenic is effective January 23, 2006. Unit then, the

  MCL is 0.05 mg/L.

\7\ The MDL reported for EPA method 200.9 (Atomic Absorption; Platform--

  Stablized Temperature) was determined using a 2x concentration step

  during sample digestion. The MDL determined for samples analyzed using

  direct analyses (i.e., no sample digestion) will be higher. Using

  multiple depositions, EPA 200.9 is capable of obtaining MDL of 0.0001

  mg/L.

\8\ Using selective ion monitoring, EPA Method 200.8 (ICP-MS) is capable

  of obtaining a MDL of 0.0001 mg/L.



    (ii) If the population served by the system is 3,300 

persons, then composit ing may only be permitted by the State at 

sampling points within a single system. In systems serving <=3,300 

persons, the State may permit compositing among different systems 

provided the 5-sample limit is maintained.

    (iii) If duplicates of the original sample taken from each sampling 

point used in the composite sample are available, the system may use 

these instead of resampling. The duplicates must be analyzed and the 

results reported to the State within 14 days after completing analysis 

of the composite sample, provided the holding time of the sample is not 

exceeded.

    (5) The frequency of monitoring for asbestos shall be in accordance 

with paragraph (b) of this section: the frequency of monitoring for 

antimony, arsenic, barium, beryllium, cadmium, chromium, cyanide, 

fluoride, mercury, nickel, selenium and thallium shall be in accordance 

with paragraph (c) of



[[Page 364]]



this section; the frequency of monitoring for nitrate shall be in 

accordance with paragraph (d) of this section; and the frequency of 

monitoring for nitrite shall be in accordance with paragraph (e) of this 

section.

    (b) The frequency of monitoring conducted to determine compliance 

with the maximum contaminant level for asbestos specified in Sec. 

141.62(b) shall be conducted as follows:

    (1) Each community and non-transient, non-community water system is 

required to monitor for asbestos during the first three-year compliance 

period of each nine-year compliance cycle beginning in the compliance 

period starting January 1, 1993.

    (2) If the system believes it is not vulnerable to either asbestos 

contamination in its source water or due to corrosion of asbestos-cement 

pipe, or both, it may apply to the State for a waiver of the monitoring 

requirement in paragraph (b)(1) of this section. If the State grants the 

waiver, the system is not required to monitor.

    (3) The State may grant a waiver based on a consideration of the 

following factors:

    (i) Potential asbestos contamination of the water source, and

    (ii) The use of asbestos-cement pipe for finished water distribution 

and the corrosive nature of the water.

    (4) A waiver remains in effect until the completion of the three-

year compliance period. Systems not receiving a waiver must monitor in 

accordance with the provisions of paragraph (b)(1) of this section.

    (5) A system vulnerable to asbestos contamination due solely to 

corrosion of asbestos-cement pipe shall take one sample at a tap served 

by asbestos-cement pipe and under conditions where asbestos 

contamination is most likely to occur.

    (6) A system vulnerable to asbestos contamination due solely to 

source water shall monitor in accordance with the provision of paragraph 

(a) of this section.

    (7) A system vulnerable to asbestos contamination due both to its 

source water supply and corrosion of asbestos-cement pipe shall take one 

sample at a tap served by asbestos-cement pipe and under conditions 

where asbestos contamination is most likely to occur.

    (8) A system which exceeds the maximum contaminant levels as 

determined in Sec. 141.23(i) of this section shall monitor quarterly 

beginning in the next quarter after the violation occurred.

    (9) The State may decrease the quarterly monitoring requirement to 

the frequency specified in paragraph (b)(1) of this section pro vided 

the State has determined that the sys tem is reliably and consistently 

below the maximum contaminant level. In no case can a State make this de 

term i na tion unless a groundwater system takes a minimum of two 

quarterly samples and a surface (or combined surface/ground) water 

system takes a minimum of four quarterly samples.

    (10) If monitoring data collected after January 1, 1990 are 

generally consistent with the requirements of Sec. 141.23(b), then the 

State may allow systems to use that data to satisfy the monitoring 

requirement for the initial compliance period beginning January 1, 1993.

    (c) The frequency of monitoring conducted to determine compliance 

with the maximum contaminant levels in Sec. 141.62 for antimony, 

arsenic, barium, beryllium, cadmium, chromium, cyanide, fluoride, 

mercury, nickel, selenium and thallium shall be as follows:

    (1) Groundwater systems shall take one sample at each sampling point 

during each compliance period. Surface water systems (or combined 

surface/ground) shall take one sample annually at each sampling point.

    (2) The system may apply to the State for a waiver from the 

monitoring frequencies specified in paragraph (c)(1) of this section. 

States may grant a public water system a waiver for monitoring of 

cyanide, provided that the State determines that the system is not 

vulnerable due to lack of any industrial source of cyanide.

    (3) A condition of the waiver shall require that a system shall take 

a minimum of one sample while the waiver is effective. The term during 

which the waiver is effective shall not exceed one compliance cycle 

(i.e., nine years).



[[Page 365]]



    (4) The State may grant a waiver provided surface water systems have 

monitored annually for at least three years and groundwater systems have 

conducted a minimum of three rounds of monitoring. (At least one sample 

shall have been taken since January 1, 1990). Both surface and 

groundwater systems shall demonstrate that all previous analytical 

results were less than the maximum contaminant level. Systems that use a 

new water source are not eligible for a waiver until three rounds of 

monitoring from the new source have been completed.

    (5) In determining the appropriate reduced monitoring frequency, the 

State shall consider:

    (i) Reported concentrations from all previous monitoring;

    (ii) The degree of variation in reported concentrations; and

    (iii) Other factors which may affect contaminant concentrations such 

as changes in groundwater pumping rates, changes in the system's 

configuration, changes in the system's operating procedures, or changes 

in stream flows or characteristics.

    (6) A decision by the State to grant a waiver shall be made in 

writing and shall set forth the basis for the determination. The 

determination may be initiated by the State or upon an application by 

the public water system. The public water system shall specify the basis 

for its request. The State shall review and, where appropriate, revise 

its determination of the appropriate monitoring frequency when the 

system submits new monitoring data or when other data relevant to the 

system's appropriate monitoring frequency become available.

    (7) Systems which exceed the maximum contaminant levels as 

calculated in Sec. 141.23(i) of this section shall monitor quarterly 

beginning in the next quarter after the violation occurred.

    (8) The State may decrease the quarterly monitoring requirement to 

the frequencies specified in paragraphs (c)(1) and (c)(2) of this 

section provided it has determined that the system is reliably and 

consistently below the maximum contaminant level. In no case can a State 

make this determination unless a groundwater system takes a minimum of 

two quarterly samples and a surface water system takes a minimum of four 

quarterly samples.

    (9) All new systems or systems that use a new source of water that 

begin operation after January 22, 2004 must demonstrate compliance with 

the MCL within a period of time specified by the State. The system must 

also comply with the initial sampling frequencies specified by the State 

to ensure a system can demonstrate compliance with the MCL. Routine and 

increased monitoring frequencies shall be conducted in accordance with 

the requirements in this section.

    (d) All public water systems (community; non-transient, non-

community; and transient, non-community systems) shall monitor to 

determine compliance with the maximum contaminant level for nitrate in 

Sec. 141.62.

    (1) Community and non-transient, non-community water systems served 

by groundwater systems shall monitor annually beginning January 1, 1993; 

systems served by surface water shall monitor quarterly beginning 

January 1, 1993.

    (2) For community and non-transient, non-community water systems, 

the repeat monitoring frequency for groundwater systems shall be 

quarterly for at least one year following any one sample in which the 

concentration is =50 percent of the MCL. The State may allow 

a groundwater system to reduce the sampling frequency to annually after 

four consecutive quarterly samples are reliably and consistently less 

than the MCL.

    (3) For community and non-transient, non-community water systems, 

the State may allow a surface water system to reduce the sampling 

frequency to annually if all analytical results from four consecutive 

quarters are <50 percent of the MCL. A surface water system shall return 

to quarterly monitoring if any one sample is =50 percent of 

the MCL.

    (4) Each transient non-community water system shall monitor annually 

beginning January 1, 1993.



[[Page 366]]



    (5) After the initial round of quarterly sampling is completed, each 

community and non-transient non-community system which is monitoring 

annually shall take subsequent samples during the quarter(s) which 

previously resulted in the highest analytical result.

    (e) All public water systems (community; non-transient, non-

community; and transient, non-community systems) shall monitor to 

determine compliance with the maximum contaminant level for nitrite in 

Sec. 141.62(b).

    (1) All public water systems shall take one sample at each sampling 

point in the compliance period beginning January 1, 1993 and ending 

December 31, 1995.

    (2) After the initial sample, systems where an analytical result for 

nitrite is <50 percent of the MCL shall monitor at the frequency 

specified by the State.

    (3) For community, non-transient, non-community, and transient non-

community water systems, the repeat monitoring frequency for any water 

system shall be quarterly for at least one year following any one sample 

in which the concentration is [gteqt]50 percent of the MCL. The State 

may allow a system to reduce the sampling frequency to annually after 

determining the system is reliably and consistently less than the MCL.

    (4) Systems which are monitoring annually shall take each subsequent 

sample during the quarter(s) which previously resulted in the highest 

analytical result.

    (f) Confirmation samples:

    (1) Where the results of sampling for antimony, arsenic, asbestos, 

barium, beryllium, cadmium, chromium, cyanide, fluoride, mercury, 

nickel, selenium or thallium indicate an exceedance of the maximum 

contaminant level, the State may require that one additional sample be 

collected as soon as possible after the initial sample was taken (but 

not to exceed two weeks) at the same sampling point.

    (2) Where nitrate or nitrite sampling results indicate an exceedance 

of the maximum contaminant level, the system shall take a confirmation 

sample within 24 hours of the system's receipt of notification of the 

analytical results of the first sample. Systems unable to comply with 

the 24-hour sampling requirement must immediately notify persons served 

by the public water system in accordance with Sec. 141.202 and meet 

other Tier 1 public notification requirements under Subpart Q of this 

part. Systems exercising this option must take and analyze a 

confirmation sample within two weeks of notification of the analytical 

results of the first sample.

    (3) If a State-required confirmation sample is taken for any 

contaminant, then the results of the initial and confirmation sample 

shall be aver aged. The resulting average shall be used to determine the 

system's compliance in accordance with paragraph (i) of this section. 

States have the discretion to delete results of obvious sampling errors.

    (g) The State may require more frequent monitoring than specified in 

paragraphs (b), (c), (d) and (e) of this section or may require 

confirmation samples for positive and negative results at its 

discretion.

    (h) Systems may apply to the State to conduct more frequent 

monitoring than the minimum monitoring frequencies specified in this 

section.

    (i) Compliance with Sec. Sec. 141.11 or 141.62(b) (as appropriate) 

shall be determined based on the analytical result(s) obtained at each 

sampling point.

    (1) For systems which are conducting monitoring at a frequency 

greater than annual, compliance with the maximum contaminant levels for 

antimony, arsenic, asbestos, barium, beryllium, cadmium, chromium, 

cyanide, fluoride, mercury, nickel, selenium or thallium is determined 

by a running annual average at any sampling point. If the average at any 

sampling point is greater than the MCL, then the system is out of 

compliance. If any one sample would cause the annual average to be 

exceeded, then the system is out of compliance immediately. Any sample 

below the method detection limit shall be calculated at zero for the 

purpose of determining the annual average. If a system fails to collect 

the required number of samples, compliance (average concentration) will 

be based on the total number of samples collected.



[[Page 367]]



    (2) For systems which are monitoring annually, or less frequently, 

the system is out of compliance with the maximum contaminant levels for 

antimony, arsenic, asbestos, barium, beryllium, cadmium, chromium, 

cyanide, fluoride, mercury, nickel, selenium or thallium if the level of 

a contaminant is greater than the MCL. If confirmation samples are 

required by the State, the determination of compliance will be based on 

the annual average of the initial MCL exceedance and any State-required 

confirmation samples. If a system fails to collect the required number 

of samples, compliance (average concentration) will be based on the 

total number of samples collected.

    (3) Compliance with the maximum contaminant levels for nitrate and 

nitrate is determined based on one sample if the levels of these 

contaminants are below the MCLs. If the levels of nitrate and/or nitrite 

exceed the MCLs in the initial sample, a confirmation sample is required 

in accordance with paragraph (f)(2) of this section, and compliance 

shall be determined based on the average of the initial and confirmation 

samples.

    (4) Arsenic sampling results will be reported to the nearest 0.001 

mg/L.

    (j) Each public water system shall monitor at the time designated by 

the State during each compliance period.

    (k) Inorganic analysis:

    (1) Analysis for the following contaminants shall be conducted in 

accordance with the methods in the following table, or their equivalent 

as determined by EPA. Criteria for analyzing arsenic, barium, beryllium, 

cadmium, calcium, chromium, copper, lead, nickel, selenium, sodium, and 

thallium with digestion or directly without digestion, and other 

analytical test procedures are contained in Technical Notes on Drinking 

Water Methods, EPA-600/R-94-173, October 1994. This document also 

contains approved analytical test methods which remain available for 

compliance monitoring until July 1, 1996. These methods will not be 

available for use after July 1, 1996. This document is available from 

the National Technical Information Service, NTIS PB95-104766, U.S. 

Department of Commerce, 5285 Port Royal Road, Springfield, Virginia 

22161. The toll-free number is 800-553-6847.



----------------------------------------------------------------------------------------------------------------

  Contaminant and  methodology                                    SM \4\  (18th,   SM \4\  (20th

              \13\                      EPA          ASTM \3\        19th ed.)         ed.)            Other

----------------------------------------------------------------------------------------------------------------

1. Alkalinity:

    Titrimetric.................  ..............  D1067--92B....  2320 B........  2320 B

    Electrometric titration.....  ..............  ..............  ..............  ..............  I-1030-85 \5\

2. Antimony:

    Inductively Coupled Plasma    200.8 \2\.....

     (ICP)--Mass Spectrometry.

    Hydride-Atomic Absorption...  ..............  D3697-92

    Atomic Absorption; Platform.  200.9 \2\

    Atomic Absorption; Furnace..  ..............  ..............  3113 B

3. Arsenic: \14\

    Inductively Coupled Plasma    200.7 \2\.....  ..............  3120 B........  3120 B

     \15\.

    ICP-Mass Spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

    Atomic Absorption; Furnace..  ..............  D2972-97C.....  3113 B

    Hydride Atomic Absorption...  ..............  D2972-97B.....  3114 B

4. Asbestos:

    Transmission Electron         100.1 \9\.....

     Microscopy.

    Transmission Electron         100.2 \10\....

     Microscopy.

5. Barium:

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B

    ICP-Mass Spectrometry.......  200.8 \2\.....

    Atomic Absorption; Direct...  ..............  ..............  3111 D

    Atomic Absorption; Furnace..  ..............  ..............  3113 B

6. Beryllium:

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B

    ICP-Mass Spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

    Atomic Absorption; Furnace..  ..............  D3645--97B....  3113 B



[[Page 368]]





7. Cadmium:

    Inductively Coupled Plasma..  200.7 \2\

    ICP-Mass Spectrometry.......  200.8 \2\

    Atomic Absorption; Platform.  200.9 \2\

    Atomic Absorption; Furnace..  ..............  ..............  3113 B........

8. Calcium:

    EDTA titrimetric............  ..............  D511--93A.....  3500-Ca D.....  3500-Ca B.....

    Atomic Absorption; Direct     ..............  D511--93B.....  3111 B........

     Aspiration.

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B........

9. Chromium:

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B........

    ICP-Mass Spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

    Atomic Absorption; Furnace..  ..............  ..............  3113 B........

10. Copper:

    Atomic Absorption; Furnace..  ..............  D1688-95C.....  3113 B........

    Atomic Absorption; Direct     ..............  D1688-95A.....  3111 B........

     Aspiration.

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B........

    ICP-Mass spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

11. Conductivity:

    Conductance.................  ..............  D1125-95A.....  2510 B........  2510 B........

12. Cyanide:

    Manual Distillation followed  ..............  D2036-98A.....  4500-CN- C....  4500-CN- C....

     by.

        Spectrophotometric        ..............  D2036-98A.....  4500-CN- E....  4500-CN- E....  I-3300-85 \5\

         Manual.

        Spectrophotometric Semi-  335.4 \6\.....

         automated.

        Spectrophotometric,       ..............  D2036-98B.....  4500-CN- G....  4500-CN- G....

         Amenable.

    Selective Electrode.........  ..............  ..............  4500-CN- F....  4500-CN- F....

    UV/Distillation/              ..............  ..............  ..............  ..............  Kelada 01 \17\

     Spectrophotometric.

    Distillation/                 ..............  ..............  ..............  ..............  QuikChem 10-

     Spectrophotometric.                                                                           204-00-1-X

                                                                                                   \18\

13. Fluoride:

    Ion Chromatography..........  300.0 \6\.....  D4327-97......  4110 B........  4110 B........

    Manual Distill.; Color.       ..............  ..............  4500-F- B,D...  4500-F- B,D...  ..............

     SPADNS.

    Manual Electrode............  ..............  D1179-93B.....  4500-F- C.....  4500-F- C.....  ..............

    Automated Electrode.........  ..............  ..............  ..............  ..............  380-75WE \11\

    Automated Alizarin..........  ..............  ..............  4500-F- E.....  4500-F- E.....  29-71W \11\

14. Lead:

    Atomic Absorption; Furnace..  ..............  D3559-96D.....  3113 B........

    ICP-Mass spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

    Differential Pulse Anodic     ..............  ..............  ..............  ..............  Method 1001

     Stripping Voltammetry.                                                                        \16\

15. Magnesium:

    Atomic Absorption...........  ..............  D511-93 B.....  3111 B........

    ICP.........................  200.7 \2\.....  ..............  3120 B........  3120 B........

    Complexation Titrimetric      ..............  D511-93 A.....  3500-Mg E.....  3500-Mg B.....

     Methods.

16. Mercury:

    Manual, Cold Vapor..........  245.1 \2\.....  D3223-97......  3112 B........

    Automated, Cold Vapor.......  245.2 \1\.....

    ICP-Mass Spectrometry.......  200.8 \2\.....

17. Nickel:

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B........

    ICP-Mass Spectrometry.......  200.8 \2\.....



[[Page 369]]





    Atomic Absorption; Platform.  200.9 \2\.....

    Atomic Absorption; Direct...  ..............  ..............  3111 B........

    Atomic Absorption; Furnace..  ..............  ..............  3113 B........

18. Nitrate:

    Ion Chromatography..........  300.0 \6\.....  D4327-97......  4110 B........  4110 B........  B-1011 \8\

    Automated Cadmium Reduction.  353.2 \6\.....  D3867-90A.....  4500-NO3- F...  4500-NO3- F...

    Ion Selective Electrode.....  ..............  ..............  4500-NO3- D...  4500-NO3- D...  601 \7\

    Manual Cadmium Reduction....  ..............  D3867-90B.....  4500-NO3- E...  4500-NO3- E...

19. Nitrite:

    Ion Chromatography..........  300.0 \6\.....  D4327-97......  4110 B........  4110 B........  B-1011 \8\

    Automated Cadmium Reduction.  353.2 \6\.....  D3867-90A.....  4500-NO3-.....  4500-NO3- F...

    Manual Cadmium Reduction....  ..............  D3867-90B.....  4500-NO3- E...  4500-NO3- E...  ..............

    Spectrophotometric..........  ..............  ..............  4500-NO2- B...  4500- NO2- B..

20. Ortho-phosphate: \12\

    Colorimetric, Automated,      365.1 \6\.....  ..............  4500-P F......  4500-P F......

     Ascorbic Acid.

    Colorimetric, ascorbic acid,  ..............  D515-88A......  4500-P E......  4500-P E......

     single reagent.

    Colorimetric                  ..............  ..............  ..............  ..............  I-1601-85 \5\

     Phosphomolybdate;.

        Automated-segmented       ..............  ..............  ..............  ..............  I-2601-90 \5\

         Flow;.

        Automated Discrete......  ..............  ..............  ..............  ..............  I-2598-85 \5\

    Ion Chromatography..........  300.0 \6\.....  D4327-97......  4110 B........  4110 B........

21. pH:

    Electrometric...............  150.1 \1\.....  D1293-95......  4500-H+ B.....  4500-H+ B.....

                                  150.2 \1\.....

22. Selenium:

    Hydride-Atomic Absorption...  ..............  D3859-98A.....  3114 B........

    ICP-Mass Spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

    Atomic Absorption; Furnace..  ..............  D3859-98B.....  3113 B........

23. Silica:

    Colorimetric, Molybdate       ..............  ..............  ..............  ..............  I-1700-85 \5\

     Blue;.

        Automated-segmented Flow  ..............  ..............  ..............  ..............  I-2700-85 \5\

    Colorimetric................  ..............  D859-95.......

    Molybdosilicate.............  ..............  ..............  4500-Si D.....  4500-SiO2 C...

    Heteropoly Blue.............  ..............  ..............  4500-Si E.....  4500-SiO2 D...

    Automated for Molybdate-      ..............  ..............  4500-Si F.....  4500-SiO2 E...

     reactive Silica.

    Inductively Coupled Plasma..  200.7 \2\.....  ..............  3120 B........  3120 B........

24. Sodium:

    Inductively Coupled Plasma..  200.7 \2\.....

    Atomic Absorption; Direct     ..............  ..............  3111 B........

     Aspiration.

25. Temperature:

    Thermometric................  ..............  ..............  2550..........  2550..........

26. Thallium:

    ICP-Mass Spectrometry.......  200.8 \2\.....

    Atomic Absorption; Platform.  200.9 \2\.....

----------------------------------------------------------------------------------------------------------------

The procedures shall be done in accordance with the documents listed below. The incorporation by reference of

  the following documents listed in footnotes 1-11 and 16 was approved by the Director of the Federal Register

  in accordance with 5 U.S.C. 552(a) and 1 CFR part 51. Copies of the documents may be obtained from the sources

  listed below. Information regarding obtaining these documents can be obtained from the Safe Drinking Water

  Hotline at 800-426-4791. Documents may be inspected at EPA'sDrinking Water Docket, EPA West, 1301 Constitution

  Avenue, NW, Room B135, Washington, DC (Telephone: 202-566-2426); or at the National Archives and Records

  Administration (NARA). For information on the availability of this material at NARA, call 202-741-6030, or go

  to: http://www.archives.gov/federal--register/code--of--federal--regulations/ibr--locations.html.

\1\ ``Methods for Chemical Analysis of Water and Wastes'', EPA/600/4-79/020, March 1983. Available at NTIS, PB84-

  128677.

\2\ ``Methods for the Determination of Metals in Environmental Samples--Supplement I'', EPA/600/R-94/111, May

  1994. Available at NTIS, PB95-125472.



[[Page 370]]





\3\ Annual Book of ASTM Standards, 1994, 1996, or 1999, Vols. 11.01 and 11.02, ASTM International; any year

  containing the cited version of the method may be used. The previous versions of D1688-95A, D1688-95C

  (copper), D3559-95D (lead), D1293-95 (pH), D1125-91A (conductivity) and D859-94 (silica) are also approved.

  These previous versions D1688-90A, C; D3559-90D, D1293-84, D1125-91A and D859-88, respectively are located in

  the Annual Book of ASTM Standards, 1994, Vol. 11.01. Copies may be obtained from ASTM International, 100 Barr

  Harbor Drive, West Conshohocken, PA 19428.

\4\ Standard Methods for the Examination of Water and Wastewater, 18th edition (1992), 19th edition (1995), or

  20th edition (1998). American Public Health Association, 1015 Fifteenth Street, NW, Washington, DC 20005. The

  cited methods published in any of these three editions may be used, except that the versions of 3111 B, 3111

  D, 3113 B and 3114 B in the 20th edition may not be used.

\5\ Method I-2601-90, Methods for Analysis by the U.S. Geological Survey National Water Quality Laboratory--

  Determination of Inorganic and Organic Constituents in Water and Fluvial Sediment, Open File Report 93-125,

  1993; For Methods I-1030-85; I-1601-85; I-1700-85; I-2598-85; I-2700-85; and I-3300-85 See Techniques of Water

  Resources Investigation of the U.S. Geological Survey, Book 5, Chapter A-1, 3rd ed., 1989; Available from

  Information Services, U.S. Geological Survey, Federal Center, Box 25286, Denver, CO 80225-0425.

\6\ ``Methods for the Determination of Inorganic Substances in Environmental Samples'', EPA/600/R-93/100, August

  1993. Available at NTIS, PB94-120821.

\7\ The procedure shall be done in accordance with the Technical Bulletin 601 ``Standard Method of Test for

  Nitrate in Drinking Water'', July 1994, PN 221890-001, Analytical Technology, Inc. Copies may be obtained from

  ATI Orion, 529 Main Street, Boston, MA 02129.

\8\ Method B-1011, ``Waters Test Method for Determination of Nitrite/Nitrate in Water Using Single Column Ion

  Chromatography,'' August 1987. Copies may be obtained from Waters Corporation, Technical Services Division, 34

  Maple Street, Milford, MA 01757.

\9\ Method 100.1, ``Analytical Method For Determination of Asbestos Fibers in Water'', EPA/600/4-83/043, EPA,

  September 1983. Available at NTIS, PB83-260471.

\10\ Method 100.2, ``Determination of Asbestos Structure Over 10 [micro]m In Length In Drinking Water'', EPA/600/

  R-94/134, June 1994. Available at NTIS, PB94-201902.

\11\ Industrial Method No. 129-71W, ``Fluoride in Water and Wastewater'', December 1972, and Method No. 380-

  75WE, ``Fluoride in Water and Wastewater'', February 1976, Technicon Industrial Systems. Copies may be

  obtained from Bran & Luebbe, 1025 Busch Parkway, Buffalo Grove, IL 60089.

\12\ Unfiltered, no digestion or hydrolysis.

\13\ Because MDLs reported in EPA Methods 200.7 and 200.9 were determined using a 2X preconcentration step

  during sample digestion, MDLs determined when samples are analyzed by direct analysis (i.e., no sample

  digestion) will be higher. For direct analysis of cadmium and arsenic by Method 200.7, and arsenic by Method

  3120 B sample preconcentration using pneumatic nebulization may be required to achieve lower detection limits.

  Preconcentration may also be required for direct analysis of antimony, lead, and thallium by Method 200.9;

  antimony and lead by Method 3113 B; and lead by Method D3559-90D unless multiple in-furnace depositions are

  made.

\14\ If ultrasonic nebulization is used in the determination of arsenic by Methods 200.7, 200.8, or SM 3120 B,

  the arsenic must be in the pentavalent state to provide uniform signal response. For methods 200.7 and 3120 B,

  both samples and standards must be diluted in the same mixed acid matrix concentration of nitric and

  hydrochloric acid with the addition of 100 [micro]L of 30% hydrogen peroxide per 100ml of solution. For direct

  analysis of arsenic with method 200.8 using ultrasonic nebulization, samples and standards must contain one mg/

  L of sodium hypochlorite.

\15\ Starting January 23, 2006, analytical methods using the ICP-AES technology, may not be used because the

  detection limits for these methods are 0.008 mg/L or higher. This restriction means that the two ICP-AES

  methods (EPA Method 200.7 and SM 3120 B) approved for use for the MCL of 0.05 mg/L may not be used for

  compliance determinations for the revised MCL of 0.010 mg/L. However, prior to January 23, 2006, systems may

  have compliance samples analyzed with these less sensitive methods.

\16\ The description for Method Number 1001 for lead is available from Palintest, LTD, 21 Kenton Lands Road,

  P.O. Box 18395, Erlanger, KY 41018. Or from the Hach Company, P.O. Box 389, Loveland, CO 80539.

\17\ The description for the Kelada 01 Method, ``Kelada Automated Test Methods for Total Cyanide, Acid

  Dissociable Cyanide, And Thiocyanate'', Revision 1.2, August 2001, EPA  821-B-01-009 for cyanide is

  available from the National Technical Information Service (NTIS), PB 2001-108275, 5285 Port Royal Road,

  Springfield, VA 22161. The toll free telephone number is 800-553-6847.

\18\ The description for the QuikChem Method 10-204-00-1-X, ``Digestion and distillation of total cyanide in

  drinking and wastewaters using MICRO DIST and determination of cyanide by flow injection analysis'', Revision

  2.1, November 30, 2000 for cyanide is available from Lachat Instruments, 6645 W. Mill Rd., Milwaukee, WI

  53218, USA. Phone: 414-358-4200.



    (2) Sample collection for antimony, arsenic, asbestos, barium, 

beryllium, cadmium, chromium, cyanide, fluoride, mercury, nickel, 

nitrate, nitrite, selenium, and thallium under this section shall be 

conducted using the sample preservation, container, and maximum holding 

time procedures specified in the table below:



------------------------------------------------------------------------

                                                  Container

         Contaminant           Preservative \1\      \2\       Time \3\

------------------------------------------------------------------------

Antimony.....................  HNO\3\..........  P or G....  6 months

Arsenic......................  Conc HNO3 to pH   P or G....  6 months

                                <2.

Asbestos.....................  4 [deg]C........  P or G....  48 hours

                                                              \4\

Barium.......................  HNO\3\..........  P or G....  6 months

Beryllium....................  HNO\3\..........  P or G....  6 months

Cadmium......................  HNO\3\..........  P or G....  6 months

Chromium.....................  HNO\3\..........  P or G....  6 months

Cyanide......................  4 [deg]C, NaOH..  P or G....  14 days

Fluoride.....................  None............  P or G....  1 month

Mercury......................  HNO\3\..........  P or G....  28 days

Nickel.......................  HNO\3\..........  P or G....  6 months

Nitrate......................  4 [deg]C........  P or G....  48 hours

                                                              \5\

Nitrate-Nitrite \6\..........  H\2\SO\4\.......  P or G....  28 days

Nitrite......................  4[deg]C.........  P or G....  48 hours

Selenium.....................  HNO\3\..........  P or G....  6 months

Thallium.....................  HNO\3\..........  P or G....  6 months

------------------------------------------------------------------------

\1\ For cyanide determinations samples must be adjusted with sodium

  hydroxide to pH 12 at the time off collection. When chilling is

  indicated the sample must be shipped and stored at 4 [deg]C or less.

  Acidification of nitrate or metals samples may be with a concentrated

  acid or a dilute (50% by volume) solution of the applicable

  concentrated acid. Acidification of samples for metals analysis is

  encouraged and allowed at the laboratory rather than at the time of

  sampling provided the shipping time and other instructions in Section

  8.3 of EPA Methods 200.7 or 200.8 or 200.9 are followed.

\2\ P=plastic, hard or soft; G=glass, hard or soft.

\3\ In all cases samples should be analyzed as soon after collection as

  possible. Follow additional (if any) information on preservation,

  containers or holding times that is specified in method.

\4\ Instructions for containers, preservation procedures and holding

  times as specified in Method 100.2 must be adhered to for all

  compliance analyses including those conducted with Method 100.1.



[[Page 371]]





\5\ If the sample is chlorinated, the holding time for an unacidified

  sample kept at 4 [deg]C is extended to 14 days.

\6\ Nitrate-Nitrite refers to a measurement of total nitrate.



    (3) Analysis under this section shall only be conducted by 

laboratories that have been certified by EPA or the State. Laboratories 

may conduct sample analysis under provisional certification until 

January 1, 1996. To receive certification to conduct analyses for 

antimony, arsenic, asbestos, barium, beryllium, cadmium, chromium, 

cyanide, fluoride, mercury, nickel, nitrate, nitrite and selenium and 

thallium, the laboratory must:

    (i) Analyze Performance Evaluation (PE) samples provided by EPA, the 

State or by a third party (with the approval of the State or EPA) at 

least once a year.

    (ii) For each contaminant that has been included in the PE sample 

and for each method for which the laboratory desires certification 

achieve quantitative results on the analyses that are within the 

following acceptance limits:



------------------------------------------------------------------------

             Contaminant                       Acceptance limit

------------------------------------------------------------------------

Antimony............................  30 at =0.006 mg/1

Arsenic.............................  30 at =0.003 mg/L

Asbestos............................  2 standard deviations based on

                                       study statistics.

Barium..............................  15% at =0.15 mg/1

Beryllium...........................  15% at =0.001 mg/1

Cadmium.............................  20% at =0.002 mg/1

Chromium............................  15% at =0.01 mg/1

Cyanide.............................  25% at =0.1 mg/1

Fluoride............................  10% at =1 to 10 mg/1

Mercury.............................  30% at =0.0005 mg/1

Nickel..............................  15% at =0.01 mg/1

Nitrate.............................  10% at =0.4 mg/1

Nitrite.............................  15% at =0.4 mg/1

Selenium............................  20% at =0.01 mg/1

Thallium............................  30% at =0.002 mg/1

------------------------------------------------------------------------



    (l) Analyses for the purpose of determining compliance with Sec. 

141.11 shall be conducted using the requirements specified in paragraphs 

(l) through (q) of this section.

    (1) Analyses for all community water systems utilizing surface water 

sources shall be completed by June 24, 1978. These analyses shall be 

repeated at yearly intervals.

    (2) Analyses for all community water systems utilizing only ground 

water sources shall be completed by June 24, 1979. These analyses shall 

be repeated at three-year intervals.

    (3) For non-community water systems, whether supplied by surface or 

ground sources, analyses for nitrate shall be completed by December 24, 

1980. These analyses shall be repeated at intervals determined by the 

State.

    (4) The State has the authority to determine compliance or initiate 

enforcement action based upon analytical results and other information 

compiled by their sanctioned representatives and agencies.

    (m) If the result of an analysis made under paragraph (l) of this 

section indicates that the level of any contaminant listed in Sec. 

141.11 exceeds the maximum contaminant level, the supplier of the water 

shall report to the State within 7 days and initiate three additional 

analyses at the same sampling point within one month.

    (n) When the average of four analyses made pursuant to paragraph (m) 

of this section, rounded to the same number of significant figures as 

the maximum contaminant level for the substance in question, exceeds the 

maximum contaminant level, the supplier of water shall notify the State 

pursuant to Sec. 141.31 and give notice to the public pursuant to 

subpart Q. Monitoring after public notification shall be at a frequency 

designated by the State and shall continue until the maximum contaminant 

level has not been exceeded in two successive samples or until a 

monitoring schedule as a condition to a variance, exemption or 

enforcement action shall become effective.

    (o) The provisions of paragraphs (m) and (n) of this section 

notwithstanding, compliance with the maximum contaminant level for 

nitrate shall be determined on the basis of the mean of two analyses. 

When a level exceeding the maximum contaminant level for nitrate is 

found, a second analysis shall be initiated within 24 hours, and if the 

mean of the two analyses exceeds the maximum contaminant level, the 

supplier of water shall report his findings to the State pursuant to 

Sec. 141.31 and shall notify the public pursuant to subpart Q.

    (p) For the initial analyses required by paragraph (l) (1), (2) or 

(3) of this section, data for surface waters acquired within one year 

prior to the effective date and data for ground waters



[[Page 372]]



acquired within 3 years prior to the effective date of this part may be 

substituted at the discretion of the State.

    (q) [Reserved]



[56 FR 3579, Jan. 30, 1991, as amended at 56 FR 30274, July 1, 1991; 57 

FR 31838, July 17, 1992; 59 FR 34322, July 1, 1994; 59 FR 62466, Dec. 5, 

1994; 60 FR 33932, 34085, June 29, 1995; 64 FR 67461, Dec. 1, 1999; 65 

FR 26022, May 4, 2000; 66 FR 7061, Jan. 22, 2001; 67 FR 65246, Oct. 23, 

2002; 67 FR 65897, Oct. 29, 2002; 67 FR 68911, Nov. 13, 2002; 68 FR 

14506, Mar. 25, 2003]