[Code of Federal Regulations]

[Title 40, Volume 22]

[Revised as of July 1, 2005]

From the U.S. Government Printing Office via GPO Access

[CITE: 40CFR141.24]



[Page 372-381]

 

                   TITLE 40--PROTECTION OF ENVIRONMENT

 

         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)

 

PART 141_NATIONAL PRIMARY DRINKING WATER REGULATIONS--Table of Contents

 

            Subpart C_Monitoring and Analytical Requirements

 

Sec. 141.24  Organic chemicals, sampling and analytical requirements.



    (a)-(d) [Reserved]

    (e) Analyses for the contaminants in this section shall be conducted 

using the following EPA methods or their equivalent as approved by EPA.

    (1) The following documents are incorporated by reference. This 

incorporation by reference was approved by the Director of the Federal 

Register in accordance with 5 U.S.C. 552(a) and 1 CFR part 51. Copies 

may be inspected at EPA's Drinking Water Docket, 1301 Constitution 

Avenue, NW., EPA West, Room B102, Washington DC 20460 (Telephone: 202-

566-2426); or at the National Archives and Records Administration 

(NARA). For information on the availability of this material at NARA, 

call 202-741-6030, or go to: http://www.archives.gov/federal--register/

code--of--federal--regulations/ibr--locations.html. Method 508A and 

515.1 are in Methods for the Determination of Organic Compounds in 

Drinking Water, EPA/600/4-88-039, December 1988, Revised, July 1991. 

Methods 547, 550 and 550.1 are in Methods for the Determination of 

Organic Compounds in Drinking Water--Supplement I, EPA/600-4-90-020, 

July 1990. Methods 548.1, 549.1, 552.1 and 555 are in Methods for the 

Determination of Organic Compounds in Drinking Water--Supplement II, 

EPA/600/R-92-129, August 1992. Methods 502.2, 504.1, 505, 506, 507, 508, 

508.1, 515.2, 524.2, 525.2, 531.1, 551.1 and 552.2 are in Methods for 

the Determination of Organic Compounds in Drinking Water--Supplement 

III, EPA/600/R-95-131, August 1995. Method 1613 is titled ``Tetra-

through Octa-Chlorinated Dioxins and Furans by Isotope-Dilution HRGC/

HRMS'', EPA/821-B-94-005, October 1994. These documents are available 

from the National Technical Information Service, NTIS PB91-231480, PB91-

146027, PB92-207703, PB95-261616 and PB95-104774, U.S. Department of 

Commerce, 5285 Port Royal Road, Springfield, Virginia 22161. The toll-

free number is 800-553-6847. Method 6651 shall be followed in accordance 

with Standard Methods for the Examination of Water and Wastewater, 18th 

edition (1992), 19th edition (1995), or 20th edition (1998), American 

Public Health Association (APHA); any of these three editions may be 

used. Method 6610 shall be followed in accordance with Standard Methods 

for the Examination of Water and Wastewater, (18th Edition Supplement) 

(1994), or with the 19th edition (1995) or 20th edition (1998) of 

Standard Methods for the Examination of Water and Wastewater; any of 

these three editions may be used. The APHA documents are available from 

APHA, 1015 Fifteenth Street NW., Washington, D.C. 20005. Other required 

analytical test procedures germane to the conduct of these analyses are 

contained in Technical Notes on Drinking Water Methods, EPA/600/R-94-

173, October 1994, NTIS PB95-104766. EPA Methods 515.3 and 549.2 are 

available from U.S. Environmental Protection Agency, National Exposure 

Research Laboratory (NERL)-Cincinnati, 26 West Martin Luther King Drive, 

Cincinnati, OH 45268. ASTM Method D 5317-93 is available in the Annual 

Book of ASTM Standards, (1999), Vol. 11.02, ASTM International, 100 Barr 

Harbor Drive, West Conshohocken, PA 19428, or in any edition published 

after 1993. EPA Method 515.4, ``Determination of Chlorinated Acids in 

Drinking Water by Liquid-Liquid Microextraction, Derivatization and Fast 

Gas Chromatography with Electron Capture Detection,'' Revision 1.0, 

April 2000, EPA /815/B-00/001 can be accessed and downloaded directly 

on-line at www.epa.gov/safewater/methods/sourcalt.html. The Syngenta AG-

625, ``Atrazine in Drinking Water by Immunoassay'', February 2001 is 

available from Syngenta Crop Protection, Inc., 410 Swing Road, Post 

Office Box 18300, Greensboro, NC 27419, Phone number (336) 632-6000. 

Method 531.2 ``Measurement of N-methylcarbamoyloximes and N-

methylcarbamates in Water by Direct Aqueous Injection HPLC with



[[Page 373]]



Postcolumn Derivatization,'' Revision 1.0, September 2001, EPA 815/B/01/

002 can be accessed and downloaded directly on-line at www.epa.gov/

safewater/methods/sourcalt.html.



----------------------------------------------------------------------------------------------------------------

            Contaminant                 EPA method 1        Standard methods          ASTM            Other

----------------------------------------------------------------------------------------------------------------

1. Benzene........................  502.2, 524.2........

2. Carbon tetrachloride...........  502.2, 524.2, 551.1.

3. Chlorobenzene..................  502.2, 524.2........

4. 1,2-Dichlorobenzene............  502.2, 524.2........

5. 1,4-Dichlorobenzene............  502.2, 524.2........

6. 1,2-Dichloroethane.............  502.2, 524.2........

7. cis-Dichloroethylene...........  502.2, 524.2........

8. trans-Dichloroethylene.........  502.2, 524.2........

9. Dichloromethane................  502.2, 524.2........

10. 1,2-Dichloropropane...........  502.2, 524.2........

11. Ethylbenzene..................  502.2, 524.2........

12. Styrene.......................  502.2, 524.2........

13. Tetrachloroethylene...........  502.2, 524.2, 551.1.

14. 1,1,1-Trichloroethane.........  502.2, 524.2, 551.1.

15. Trichloroethylene.............  502.2, 524.2, 551.1.

16. Toluene.......................  502.2, 524.2........

17. 1,2,4-Trichlorobenzene........  502.2, 524.2........

18. 1,1-Dichloroethylene..........  502.2, 524.2........

19. 1,1,2-Trichloroethane.........  502.2, 524.2, 551.1.

20. Vinyl chloride................  502.2, 524.2........

21. Xylenes (total)...............  502.2, 524.2........

22. 2,3,7,8-TCDD (dioxin).........  1613................

23. 2,4-D 4 (as acid, salts and     515.2, 555, 515.1,    ....................  D5317-93.......

 esters).                            515.3, 515.4.

24. 2,4,5-TP 4 (Silvex)...........  515.2, 555, 515.1,    ....................  D5317-93.......

                                     515.3, 515.4.

25. Alachlor 2....................  507, 525.2, 508.1,

                                     505, 551.1.

26. Atrazine 2....................  507, 525.2, 508.1,    ....................  ...............  Syngenta AG-

                                     505, 551.1.                                                  625.

27. Benzo(a)pyrene................  525.2, 550, 550.1...

28. Carbofuran....................  531.1, 531.2........  6610................

29. Chlordane.....................  508, 525.2, 508.1,

                                     505.

30. Dalapon.......................  552.1, 515.1, 552.2,

                                     515.3, 515.4.

31. Di(2-ethylhexyl)adipate.......  506, 525.2..........

32. Di(2-ethylhexyl)phthalate.....  506, 525.2..........

33. Dibromochloropropane (DBCP)...  504.1, 551.1........

34. Dinoseb 4.....................  515.2, 555, 515.1,

                                     515.3, 515.4.

35. Diquat........................  549.2...............

36. Endothall.....................  548.1...............

37. Endrin........................  508, 525.2, 508.1,

                                     505, 551.1.

38. Ethylene dibromide (EDB)......  504.1, 551.1........

39. Glyphosate....................  547.................  6651................

40. Heptachlor....................  508, 525.2, 508.1,

                                     505, 551.1.

41. Heptachlor Epoxide............  508, 525.2, 508.1,

                                     505, 551.1.

42. Hexachlorobenzene.............  508, 525.2, 508.1,

                                     505, 551.1.

43. Hexachlorocyclopentadiene.....  508, 525.2, 508.1,

                                     505, 551.1.

44. Lindane.......................  508, 525.2, 508.1,

                                     505, 551.1.

45. Methoxychlor..................  508, 525.2, 508.1,

                                     505, 551.1.

46. Oxamyl........................  531.1, 531.2........  6610................

47. PCBs 3 (as decachlorobiphenyl)  508A................

48. PCBs 3 (as Aroclors)..........  508.1, 508, 525.2,

                                     505.

49. Pentachlorophenol.............  515.2, 525.2, 555,    ....................  D5317-93.......

                                     515.1, 515.3, 515.4.

50. Picloram 4....................  515.2, 555, 515.1,    ....................  D5317-93.......

                                     515.3, 515.4.

51. Simazine 2....................  507, 525.2, 508.1,

                                     505, 551.1.

52. Toxaphene.....................  508, 508.1, 525.2,

                                     505.

53. Total Trihalomethanes.........  502.2, 524.2, 551.1.

----------------------------------------------------------------------------------------------------------------

\1\ For previously approved EPA methods which remain available for compliance monitoring until June 1, 2001, see

  paragraph (e)(2) of this section.

\2\ Substitution of the detector specified in Method 505, 507, 508 or 508.1 for the purpose of achieving lower

  detection limits is allowed as follows. Either an electron capture or nitrogen phosphorous detector may be

  used provided all regulatory requirements and quality control criteria are met.

\3\ PCBs are qualitatively identified as Aroclors and measured for compliance purposes as decachlorobiphenyl.

  Users of Method 505 may have more difficulty in achieving the required detection limits than users of Methods

  508.1, 525.2 or 508.

\4\ Accurate determination of the chlorinated esters requires hydrolysis of the sample as described in EPA

  Methods 515.1, 515.2, 515.3, 515.4 and 555 and ASTM Method D5317-93.



    (2) The following EPA methods will remain available for compliance 

monitoring until June 1, 2001. The following



[[Page 374]]



documents are incorporated by reference. This incorporation by reference 

was approved by the Director of the Federal Register in accordance with 

5 U.S.C. 552(a) and 1 CFR Part 51. Copies may be inspected at EPA's 

Drinking Water Docket, 401 M St., SW., Washington, DC 20460; or at the 

National Archives and Records Administration (NARA). For information on 

the availability of this material at NARA, call 202-741-6030, or go to: 

http://www.archives.gov/federal--register/code--of--federal--

regulations/ibr--locations.html. EPA methods 502.2 Rev. 2.0, 505 Rev. 

2.0, 507 Rev. 2.0, 508 Rev. 3.0, 531.1 Rev. 3.0 are in ``Methods for the 

Determination of Organic Compounds in Drinking Water'', December 1988, 

revised July 1991; methods 506 and 551 are in ``Methods for the 

Determination of Organic Compounds in Drinking Water--Supplement I'', 

July 1990; methods 515.2 Rev. 1.0 and 524.2 Rev. 4.0 are in ``Methods 

for the Determination of Organic Compounds in Drinking Water--Supplement 

II,'' August 1992; and methods 504.1 Rev. 1.0, 508.1 Rev. 1.0, 525.2 

Rev.1.0 are available from US EPA NERL, Cincinnati, OH 45268

    (f) Beginning with the initial compliance period, analysis of the 

contaminants listed in Sec. 141.61(a) (1) through (21) for the purpose 

of determining compliance with the maximum contaminant level shall be 

conducted as follows:

    (1) Groundwater systems shall take a minimum of one sample at every 

entry point to the distribution system which is representative of each 

well after treatment (hereafter called a sampling point). Each sample 

must be taken at the same sampling point unless conditions make another 

sampling point more representative of each source, treatment plant, or 

within the distribution system.

    (2) Surface water systems (or combined surface/ground) shall take a 

minimum of one sample at points in the distribution system that are 

representative of each source or at each entry point to the distribution 

system after treatment (hereafter called a sampling point). Each sample 

must be taken at the same sampling point unless conditions make another 

sampling point more representative of each source, treatment plant, or 

within the distribution system.

    (3) If the system draws water from more than one source and the 

sources are combined before distribution, the system must sample at an 

entry point to the distribution system during periods of normal 

operating conditions (i.e., when water representative of all sources is 

being used).

    (4) Each community and non-transient non-community water system 

shall take four consecutive quarterly samples for each contaminant 

listed in Sec. 141.61(a) (2) through (21) during each compliance 

period, beginning in the initial compliance period.

    (5) If the initial monitoring for contaminants listed in Sec. 

141.61(a) (1) through (8) and the monitoring for the contaminants listed 

in Sec. 141.61(a) (9) through (21) as allowed in paragraph (f)(18) has 

been completed by December 31, 1992, and the system did not detect any 

contaminant listed in Sec. 141.61(a) (1) through (21), then each ground 

and surface water system shall take one sample annually beginning with 

the initial compliance period.

    (6) After a minimum of three years of annual sampling, the State may 

allow groundwater systems with no previous detection of any contaiminant 

listed in Sec. 141.61(a) to take one sample during each compliance 

period.

    (7) Each community and non-transient non-community ground water 

system which does not detect a contaminant listed in Sec. 141.61(a) (1) 

through (21) may apply to the State for a waiver from the requirements 

of paragraphs (f)(5) and (f)(6) of this section after completing the 

initial monitoring. (For purposes of this section, detection is defined 

as =0.0005 mg/l.) A waiver shall be effective for no more 

than six years (two compliance periods). States may also issue waivers 

to small systems for the initial round of monitoring for 1,2,4-

trichlorobenzene.

    (8) A State may grant a waiver after evaluating the following 

factor(s):

    (i) Knowledge of previous use (including transport, storage, or 

disposal) of the contaminant within the watershed or zone of influence 

of the system. If a determination by the State reveals no previous use 

of the contaminant within



[[Page 375]]



the watershed or zone of influence, a waiver may be granted.

    (ii) If previous use of the contaminant is unknown or it has been 

used previously, then the following factors shall be used to determine 

whether a waiver is granted.

    (A) Previous analytical results.

    (B) The proximity of the system to a potential point or non-point 

source of contamination. Point sources include spills and leaks of 

chemicals at or near a water treatment facility or at manufacturing, 

distribution, or storage facilities, or from hazardous and municipal 

waste landfills and other waste handling or treatment facilities.

    (C) The environmental persistence and transport of the contaminants.

    (D) The number of persons served by the public water system and the 

proximity of a smaller system to a larger system.

    (E) How well the water source is protected against contamination, 

such as whether it is a surface or groundwater system. Groundwater 

systems must consider factors such as depth of the well, the type of 

soil, and wellhead protection. Surface water systems must consider 

watershed protection.

    (9) As a condition of the waiver a groundwater system must take one 

sample at each sampling point during the time the waiver is effective 

(i.e., one sample during two compliance periods or six years) and update 

its vulnerability assessment considering the factors listed in paragraph 

(f)(8) of this section. Based on this vulnerability assessment the State 

must reconfirm that the system is non-vulnerable. If the State does not 

make this reconfirmation within three years of the initial 

determination, then the waiver is invalidated and the system is required 

to sample annually as specified in paragraph (5) of this section.

    (10) Each community and non-transient non-community surface water 

system which does not detect a contaminant listed in Sec. 141.61(a) (1) 

through (21) may apply to the State for a waiver from the requirements 

of (f)(5) of this section after completing the initial monitoring. 

Composite samples from a maximum of five sampling points are allowed, 

provided that the detection limit of the method used for analysis is 

less than one-fifth of the MCL. Systems meeting this criterion must be 

determined by the State to be non-vulnerable based on a vulnerability 

assessment during each compliance period. Each system receiving a waiver 

shall sample at the frequency specified by the State (if any).

    (11) If a contaminant listed in Sec. 141.61(a) (2) through (21) is 

detected at a level exceeding 0.0005 mg/l in any sample, then:

    (i) The system must monitor quarterly at each sampling point which 

resulted in a detection.

    (ii) The State may decrease the quarterly monitoring requirement spe 

ci fied in paragraph (f)(11)(i) of this section provided it has 

determined that the system is reliably and consistently below the 

maximum contaminant level. In no case shall the State make this 

determination unless a groundwater system takes a minimum of two 

quarterly samples and a surface water system takes a minimum of four 

quarterly samples.

    (iii) If the State determines that the system is reliably and 

consistently below the MCL, the State may allow the system to monitor 

annually. Systems which monitor annually must monitor during the 

quarter(s) which previously yielded the highest analytical result.

    (iv) Systems which have three consecutive annual samples with no 

detection of a contaminant may apply to the State for a waiver as 

specified in paragraph (f)(7) of this section.

    (v) Groundwater systems which have detected one or more of the 

following two-carbon organic compounds: trichloroethylene, 

tetrachloroethylene, 1,2-dichloroethane, 1,1,1-trichloroethane, cis-1,2-

dichloroethylene, trans-1,2-dichloroethylene, or 1,1-dichloroethylene 

shall monitor quarterly for vinyl chloride. A vinyl chloride sample 

shall be taken at each sampling point at which one or more of the two-

carbon organic compounds was detected. If the results of the first 

analysis do not detect vinyl chloride, the State may reduce the 

quarterly monitoring frequency of vinyl chloride monitoring to one 

sample during each compliance period. Surface water systems



[[Page 376]]



are required to monitor for vinyl chloride as specified by the State.

    (12) Systems which violate the requirements of Sec. 141.61(a) (1) 

through (21), as determined by paragraph (f)(15) of this section, must 

monitor quarterly. After a minimum of four consecutive quarterly samples 

which show the system is in compliance as specified in paragraph (f)(15) 

of this section the system and the State determines that the system is 

reliably and consistently below the maximum contaminant level, the 

system may monitor at the frequency and times specified in paragraph 

(f)(11)(iii) of this section.

    (13) The State may require a confirmation sample for positive or 

negative results. If a confirmation sample is required by the State, the 

result must be averaged with the first sampling result and the average 

is used for the compliance determination as specified by paragraph 

(f)(15). States have discretion to delete results of obvious sampling 

errors from this calculation.

    (14) The State may reduce the total number of samples a system must 

analyze by allowing the use of compositing. Composite samples from a 

maximum of five sampling points are allowed, provided that the detection 

limit of the method used for analysis is less than one-fifth of the MCL. 

Compositing of samples must be done in the laboratory and analyzed 

within 14 days of sample collection.

    (i) If the concentration in the composite sample is greater than or 

equal to 0.0005 mg/l for any contaminant listed in Sec. 141.61(a), then 

a follow-up sample must be taken within 14 days at each sampling point 

included in the composite, and be analyzed for that contaminant.

    (ii) If duplicates of the original sample taken from each sampling 

point used in the composite sample are available, the system may use 

these instead of resampling. The duplicates must be analyzed and the 

results reported to the State within 14 days after completing analysis 

of the composite sample, provided the holding time of the sample is not 

exceeded.

    (iii) If the population served by the system is  3,300 

persons, then compositing may only be permitted by the State at sampling 

points within a single system. In systems serving <= 3,300 persons, the 

State may permit compositing among different systems provided the 5-

sample limit is maintained.

    (iv) Compositing samples prior to GC analysis.

    (A) Add 5 ml or equal larger amounts of each sample (up to 5 samples 

are allowed) to a 25 ml glass syringe. Special precautions must be made 

to maintain zero headspace in the syringe.

    (B) The samples must be cooled at 4 [deg]C during this step to 

minimize volatilization losses.

    (C) Mix well and draw out a 5-ml aliquot for analysis.

    (D) Follow sample introduction, purging, and desorption steps 

described in the method.

    (E) If less than five samples are used for compositing, a 

proportionately small syringe may be used.

    (v) Compositing samples prior to GC/MS analysis.

    (A) Inject 5-ml or equal larger amounts of each aqueous sample (up 

to 5 samples are allowed) into a 25-ml purging device using the sample 

introduction technique described in the method.

    (B) The total volume of the sample in the purging device must be 25 

ml.

    (C) Purge and desorb as described in the method.

    (15) Compliance with Sec. 141.61(a) (1) through (21) shall be 

determined based on the analytical results obtained at each sampling 

point. If one sampling point is in violation of an MCL, the system is in 

violation of the MCL.

    (i) For systems monitoring more than once per year, compliance with 

the MCL is determined by a running annual average at each sampling 

point.

    (ii) Systems monitoring annually or less frequently whose sample 

result exceeds the MCL must begin quarterly sampling. The system will 

not be considered in violation of the MCL until it has completed one 

year of quarterly sampling.

    (iii) If any sample result will cause the running annual average to 

exceed the MCL at any sampling point, the system is out of compliance 

with the MCL immediately.

    (iv) If a system fails to collect the required number of samples, 

compliance



[[Page 377]]



will be based on the total number of samples collected.

    (v) If a sample result is less than the detection limit, zero will 

be used to calculate the annual average.

    (16) [Reserved]

    (17) Analysis under this section shall only be conducted by 

laboratories that are certified by EPA or the State according to the 

following conditions (laboratories may conduct sample analysis under 

provisional certification until January 1, 1996):

    (i) To receive certification to conduct analyses for the 

contaminants in Sec. 141.61(a) (2) through (21) the laboratory must:

    (A) Analyze Performance Evaluation (PE) samples provided by EPA, the 

State, or by a third party (with the approval of the State or EPA) at 

least once a year by each method for which the laboratory desires 

certification.

    (B) Achieve the quantitative acceptance limits under paragraphs 

(f)(17)(i)(C) and (D) of this section for at least 80 percent of the 

regulated organic contaminants included in the PE sample.

    (C) Achieve quantitative results on the analyses performed under 

paragraph (f)(17)(i)(A) of this section that are within 20% of the actual amount of the substances in the 

Performance Evaluation sample when the actual amount is greater than or 

equal to 0.010 mg/l.

    (D) Achieve quantitative results on the analyses performed under 

paragraph (f)(17)(i)(A) of this section that are within 40 percent of the actual amount of the substances in the 

Performance Evaluation sample when the actual amount is less than 0.010 

mg/l.

    (E) Achieve a method detection limit of 0.0005 mg/l, according to 

the procedures in appendix B of part 136.

    (ii) To receive certification to conduct analyses for vinyl 

chloride, the laboratory must:

    (A) Analyze Performance Evaluation (PE) samples provided by EPA, the 

State, or by a third party (with the approval of the State or EPA) at 

least once a year by each method for which the laboratory desires 

certification.

    (B) Achieve quantitative results on the analyses performed under 

paragraph (f)(17)(ii)(A) of this section that are within 40 percent of the actual amount of vinyl chloride in the 

Performance Evaluation sample.

    (C) Achieve a method detection limit of 0.0005 mg/l, according to 

the procedures in appendix B of part 136.

    (D) Obtain certification for the contaminants listed in Sec. 

141.61(a)(2) through (21).

    (18) States may allow the use of monitoring data collected after 

January 1, 1988, required under section 1445 of the Act for purposes of 

initial monitoring compliance. If the data are generally consistent with 

the other requirements of this section, the State may use these data 

(i.e., a single sample rather than four quarterly samples) to satisfy 

the initial monitoring requirement of paragraph (f)(4) of this section. 

Systems which use grandfathered samples and did not detect any 

contaminant listed Sec. 141.61(a)(2) through (21) shall begin 

monitoring annually in accordance with paragraph (f)(5) of this section 

beginning with the initial compliance period.

    (19) States may increase required monitoring where necessary to 

detect variations within the system.

    (20) Each certified laboratory must determine the method detection 

limit (MDL), as defined in appendix B to part 136, at which it is 

capable of detecting VOCs. The acceptable MDL is 0.0005 mg/l. This 

concentration is the detection concentration for purposes of this 

section.

    (21) Each public water system shall monitor at the time designated 

by the State within each compliance period.

    (22) All new systems or systems that use a new source of water that 

begin operation after January 22, 2004 must demonstrate compliance with 

the MCL within a period of time specified by the State. The system must 

also comply with the initial sampling frequencies specified by the State 

to ensure a system can demonstrate compliance with the MCL. Routine and 

increased monitoring frequencies shall be conducted in accordance with 

the requirements in this section.

    (g) [Reserved]



[[Page 378]]



    (h) Analysis of the contaminants listed in Sec. 141.61(c) for the 

purposes of determining compliance with the maximum contaminant level 

shall be conducted as follows: \7\

---------------------------------------------------------------------------



    \7\ Monitoring for the contaminants aldicarb, aldicarb sulfoxide, 

and aldicarb sulfone shall be conducted in accordance with Sec. 141.40.

---------------------------------------------------------------------------



    (1) Groundwater systems shall take a minimum of one sample at every 

entry point to the distribution system which is representative of each 

well after treatment (hereafter called a sampling point). Each sample 

must be taken at the same sampling point unless conditions make another 

sampling point more representative of each source or treatment plant.

    (2) Surface water systems shall take a minimum of one sample at 

points in the distribution system that are representative of each source 

or at each entry point to the distribution system after treatment 

(hereafter called a sampling point). Each sample must be taken at the 

same sampling point unless conditions make another sampling point more 

representative of each source or treatment plant.

    Note: For purposes of this paragraph, surface water systems include 

systems with a combination of surface and ground sources.

    (3) If the system draws water from more than one source and the 

sources are combined before distribution, the system must sample at an 

entry point to the distribution system during periods of normal 

operating conditions (i.e., when water representative of all sources is 

being used).

    (4) Monitoring frequency:

    (i) Each community and non-transient non-community water system 

shall take four consecutive quarterly samples for each contaminant 

listed in Sec. 141.61(c) during each compliance period beginning with 

the initial compliance period.

    (ii) Systems serving more than 3,300 persons which do not detect a 

contaminant in the initial compliance period may reduce the sampling 

frequency to a minimum of two quarterly samples in one year during each 

repeat compliance period.

    (iii) Systems serving less than or equal to 3,300 persons which do 

not detect a contaminant in the initial compliance period may reduce the 

sampling frequency to a minimum of one sample during each repeat 

compliance period.

    (5) Each community and non-transient water system may apply to the 

State for a waiver from the requirement of paragraph (h)(4) of this 

section. A system must reapply for a waiver for each compliance period.

    (6) A State may grant a waiver after evaluating the following 

factor(s): Knowledge of previous use (including transport, storage, or 

disposal) of the contaminant within the watershed or zone of influence 

of the system. If a determination by the State reveals no previous use 

of the contaminant within the watershed or zone of influence, a waiver 

may be granted. If previous use of the contaminant is unknown or it has 

been used previously, then the following factors shall be used to 

determine whether a waiver is granted.

    (i) Previous analytical results.

    (ii) The proximity of the system to a potential point or non-point 

source of contamination. Point sources include spills and leaks of 

chemicals at or near a water treatment facility or at manufacturing, 

distribution, or storage facilities, or from hazardous and municipal 

waste landfills and other waste handling or treatment facilities. Non-

point sources include the use of pesticides to control insect and weed 

pests on agricultural areas, forest lands, home and gardens, and other 

land application uses.

    (iii) The environmental persistence and transport of the pesticide 

or PCBs.

    (iv) How well the water source is protected against contamination 

due to such factors as depth of the well and the type of soil and the 

integrity of the well casing.

    (v) Elevated nitrate levels at the water supply source.

    (vi) Use of PCBs in equipment used in the production, storage, or 

distribution of water (i.e., PCBs used in pumps, transformers, etc.).

    (7) If an organic contaminant listed in Sec. 141.61(c) is detected 

(as defined by paragraph (h)(18) of this section) in any sample, then:

    (i) Each system must monitor quarterly at each sampling point which 

resulted in a detection.



[[Page 379]]



    (ii) The State may decrease the quarterly monitoring requirement 

specified in paragraph (h)(7)(i) of this section provided it has 

determined that the system is reliably and consistently below the 

maximum contaminant level. In no case shall the State make this 

determination unless a groundwater system takes a minimum of two 

quarterly samples and a surface water system takes a minimum of four 

quarterly samples.

    (iii) After the State determines the system is reliably and 

consistently below the maximum contaminant level the State may allow the 

system to monitor annually. Systems which monitor annually must monitor 

during the quarter that previously yielded the highest analytical 

result.

    (iv) Systems which have 3 consecutive annual samples with no 

detection of a contaminant may apply to the State for a waiver as 

specified in paragraph (h)(6) of this section.

    (v) If monitoring results in detection of one or more of certain 

related contaminants (aldicarb, aldicarb sulfone, aldicarb sulfoxide and 

heptachlor, heptachlor epoxide), then subsequent monitoring shall 

analyze for all related contaminants.

    (8) Systems which violate the requirements of Sec. 141.61(c) as 

determined by paragraph (h)(11) of this section must monitor quarterly. 

After a minimum of four quarterly samples show the system is in 

compliance and the State determines the system is reliably and 

consistently below the MCL, as specified in paragraph (h)(11) of this 

section, the system shall monitor at the frequency specified in 

paragraph (h)(7)(iii) of this section.

    (9) The State may require a confirmation sample for positive or 

negative results. If a confirmation sample is required by the State, the 

result must be averaged with the first sampling result and the average 

used for the compliance determination as specified by paragraph (h)(11) 

of this section. States have discretion to delete results of obvious 

sampling errors from this calculation.

    (10) The State may reduce the total number of samples a system must 

analyze by allowing the use of compositing. Composite samples from a 

maximum of five sampling points are allowed, provided that the detection 

limit of the method used for analysis is less than one-fifth of the MCL. 

Compositing of samples must be done in the laboratory and analyzed 

within 14 days of sample collection.

    (i) If the concentration in the composite sample detects one or more 

contaminants listed in Sec. 141.61(c), then a follow-up sample must be 

taken within 14 days at each sampling point included in the composite, 

and be analyzed for that contaminant.

    (ii) If duplicates of the original sample taken from each sampling 

point used in the composite sample are available, the system may use 

these instead of resampling. The duplicates must be analyzed and the 

results reported to the State within 14 days after completion of the 

composite analysis or before the holding time for the initial sample is 

exceeded whichever is sooner.

    (iii) If the population served by the system is 3,300 

persons, then compositing may only be permitted by the State at sampling 

points within a single system. In systems serving <= 3,300 persons, the 

State may permit compositing among different systems provided the 5-

sample limit is maintained.

    (11) Compliance with Sec. 141.61(c) shall be determined based on 

the analytical results obtained at each sampling point. If one sampling 

point is in violation of an MCL, the system is in violation of the MCL.

    (i) For systems monitoring more than once per year, compliance with 

the MCL is determined by a running annual average at each sampling 

point.

    (ii) Systems monitoring annually or less frequently whose sample 

result exceeds the regulatory detection level as defined by paragraph 

(h)(18) of this section must begin quarterly sampling. The system will 

not be considered in violation of the MCL until it has completed one 

year of quarterly sampling.

    (iii) If any sample result will cause the running annual average to 

exceed the MCL at any sampling point, the system is out of compliance 

with the MCL immediately.

    (iv) If a system fails to collect the required number of samples, 

compliance



[[Page 380]]



will be based on the total number of samples collected.

    (v) If a sample result is less than the detection limit, zero will 

be used to calculate the annual average.

    (12) [Reserved]

    (13) Analysis for PCBs shall be conducted as follows using the 

methods in paragraph (e) of this section:

    (i) Each system which monitors for PCBs shall analyze each sample 

using either Method 508.1, 525.2, 508 or 505. Users of Method 505 may 

have more difficulty in achieving the required Aroclor detection limits 

than users of Methods 508.1, 525.2 or 508.

    (ii) If PCBs (as one of seven Aroclors) are detected (as designated 

in this paragraph) in any sample analyzed using Method 505 or 508, the 

system shall reanalyze the sample using Method 508A to quantitate PCBs 

(as decachlorobiphenyl).



------------------------------------------------------------------------

                                                        Detection limit

                       Aroclor                               (mg/l)

------------------------------------------------------------------------

1016.................................................        0.00008

1221.................................................        0.02

1232.................................................        0.0005

1242.................................................        0.0003

1248.................................................        0.0001

1254.................................................        0.0001

1260.................................................        0.0002

------------------------------------------------------------------------



    (iii) Compliance with the PCB MCL shall be determined based upon the 

quantitative results of analyses using Method 508A.

    (14) If monitoring data collected after January 1, 1990, are 

generally consistent with the requirements of Sec. 141.24(h), then the 

State may allow systems to use that data to satisfy the monitoring 

requirement for the initial compliance period beginning January 1, 1993.

    (15) The State may increase the required monitoring frequency, where 

necessary, to detect variations within the system (e.g., fluctuations in 

concentration due to seasonal use, changes in water source).

    (16) The State has the authority to determine compliance or initiate 

enforcement action based upon analytical results and other information 

compiled by their sanctioned representatives and agencies.

    (17) Each public water system shall monitor at the time designated 

by the State within each compliance period.

    (18) Detection as used in this paragraph shall be defined as greater 

than or equal to the following concentrations for each contaminant.



------------------------------------------------------------------------

                                                              Detection

                        Contaminant                           limit (mg/

                                                                  l)

------------------------------------------------------------------------

Alachlor...................................................   .0002

Aldicarb...................................................   .0005

Aldicarb sulfoxide.........................................   .0005

Aldicarb sulfone...........................................   .0008

Atrazine...................................................   .0001

Benzo[a]pyrene.............................................   .00002

Carbofuran.................................................   .0009

Chlordane..................................................   .0002

Dalapon....................................................   .001

1,2-Dibromo-3-chloropropane (DBCP).........................   .00002

Di (2-ethylhexyl) adipate..................................   .0006

Di (2-ethylhexyl) phthalate................................   .0006

Dinoseb....................................................   .0002

Diquat.....................................................   .0004

2,4-D......................................................   .0001

Endothall..................................................   .009

Endrin.....................................................   .00001

Ethylene dibromide (EDB)...................................   .00001

Glyphosate.................................................   .006

Heptachlor.................................................   .00004

Heptachlor epoxide.........................................   .00002

Hexachlorobenzene..........................................   .0001

Hexachlorocyclopentadiene..................................   .0001

Lindane....................................................   .00002

Methoxychlor...............................................   .0001

Oxamyl.....................................................   .002

Picloram...................................................   .0001

Polychlorinated biphenyls (PCBs) (as decachlorobiphenyl)...   .0001

Pentachlorophenol..........................................   .00004

Simazine...................................................   .00007

Toxaphene..................................................   .001

2,3,7,8-TCDD (Dioxin)......................................   .000000005

2,4,5-TP (Silvex)..........................................   .0002

------------------------------------------------------------------------





    (19) Anaylsis under this section shall only be conducted by 

laboratories that have received certification by EPA or the State and 

have met the following conditions:

    (i) To receive certification to conduct analyses for the 

contaminants in Sec. 141.61(c) the laboratory must:

    (A) Analyze Performance Evaluation (PE) samples provided by EPA, the 

State, or by a third party (with the approval of the State or EPA) at 

least once a year by each method for which the laboratory desires 

certification.

    (B) For each contaminant that has been included in the PE sample 

achieve quantitative results on the analyses that are within the 

following acceptance limits:



------------------------------------------------------------------------

                Contaminant                  Acceptance limits (percent)

------------------------------------------------------------------------

DBCP......................................  40

EDB.......................................  40.

Alachlor..................................  45.

Atrazine..................................  45.



[[Page 381]]





Benzo[a]pyrene............................  2 standard deviations.

Carbofuran................................  45.

Chlordane.................................  45.

Dalapon...................................  2 standard deviations.

Di(2-ethylhexyl)adipate...................  2 standard deviations.

Di(2-ethylhexyl)phthalate.................  2 standard deviations.

Dinoseb...................................  2 standard deviations.

Diquat....................................  2 standard deviations.

Endothall.................................  2 standard deviations.

Endrin....................................  30.

Glyphosate................................  2 standard deviations.

Heptachlor................................  45.

Heptachlor epoxide........................  45.

Hexachlorobenzene.........................  2 standard deviations.

Hexachloro- cyclopentadiene                 2 standard deviations.

Lindane...................................  45.

Methoxychlor..............................  45.

Oxamyl....................................  2 standard deviations.

PCBs (as Decachlorobiphenyl)                0-200.

Picloram..................................  2 standard deviations.

Simazine..................................  2 standard deviations.

Toxaphene.................................  45.

Aldicarb..................................  2 standard deviations.

Aldicarb sulfoxide........................  2 standard deviations.

Aldicarb sulfone..........................  2 standard deviations.

Pentachlorophenol.........................  50.

2,3,7,8-TCDD (Dioxin).....................  2 standard deviations.

2,4-D.....................................  50.

2,4,5-TP (Silvex).........................  50.

------------------------------------------------------------------------



    (ii) [Reserved]

    (20) All new systems or systems that use a new source of water that 

begin operation after January 22, 2004 must demonstrate compliance with 

the MCL within a period of time specified by the State. The system must 

also comply with the initial sampling frequencies specified by the State 

to ensure a system can demonstrate compliance with the MCL. Routine and 

increased monitoring frequencies shall be conducted in accordance with 

the requirements in this section.



(Approved by the Office of Management and Budget under control number 

2040-0090)



[40 FR 59570, Dec. 24, 1975, as amended at 44 FR 68641, Nov. 29, 1979; 

45 FR 57345, Aug. 27, 1980; 47 FR 10998, Mar. 12, 1982; 52 FR 25712, 

July 8, 1987; 53 FR 5147, Feb. 19, 1988; 53 FR 25110, July 1, 1988; 56 

FR 3583, Jan. 30, 1991; 56 FR 30277, July 1, 1991; 57 FR 22178, May 27, 

1992; 57 FR 31841, July 17, 1992; 59 FR 34323, July 1, 1994; 59 FR 

62468, Dec. 5, 1994; 60 FR 34085, June 29, 1995; 64 FR 67464, Dec. 1, 

1999; 65 FR 26022, May 4, 2000; 66 FR 7063, Jan. 22, 2001; 67 FR 65250, 

Oct. 23, 2002; 67 FR 65898, Oct. 29, 2002]