[Code of Federal Regulations]
[Title 40, Volume 7]
[Revised as of July 1, 2005]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR60.A6]

[Page 404]

		Appendix A-6 to Part 60--Test Methods 16 through 18

Method 16--Semicontinuous determination of sulfur emissions from 
stationary sources
Method 16A--Determination of total reduced sulfur emissions from 
stationary sources (impinger technique)
Method 16B--Determination of total reduced sulfur emissions from 
stationary sources



Method 17--Determination of particulate emissions from stationary 
sources (in-stack filtration method)
Method 18--Measurement of gaseous organic compound emissions by gas 
chromatography
    The test methods in this appendix are referred to in Sec. 60.8 
(Performance Tests) and Sec. 60.11 (Compliance With Standards and 
Maintenance Requirements) of 40 CFR part 60, subpart A (General 
Provisions). Specific uses of these test methods are described in the 
standards of performance contained in the subparts, beginning with 
Subpart D.
    Within each standard of performance, a section title ``Test Methods 
and Procedures'' is provided to: (1) Identify the test methods to be 
used as reference methods to the facility subject to the respective 
standard and (2) identify any special instructions or conditions to be 
followed when applying a method to the respective facility. Such 
instructions (for example, establish sampling rates, volumes, or 
temperatures) are to be used either in addition to, or as a substitute 
for procedures in a test method. Similarly, for sources subject to 
emission monitoring requirements, specific instructions pertaining to 
any use of a test method as a reference method are provided in the 
subpart or in Appendix B.
    Inclusion of methods in this appendix is not intended as an 
endorsement or denial of their applicability to sources that are not 
subject to standards of performance. The methods are potentially 
applicable to other sources; however, applicability should be confirmed 
by careful and appropriate evaluation of the conditions prevalent at 
such sources.
    The approach followed in the formulation of the test methods 
involves specifications for equipment, procedures, and performance. In 
concept, a performance specification approach would be preferable in all 
methods because this allows the greatest flexibility to the user. In 
practice, however, this approach is impractical in most cases because 
performance specifications cannot be established. Most of the methods 
described herein, therefore, involve specific equipment specifications 
and procedures, and only a few methods in this appendix rely on 
performance criteria.
    Minor changes in the test methods should not necessarily affect the 
validity of the results and it is recognized that alternative and 
equivalent methods exist. Section 60.8 provides authority for the 
Administrator to specify or approve (1) equivalent methods, (2) 
alternative methods, and (3) minor changes in the methodology of the 
test methods. It should be clearly understood that unless otherwise 
identified all such methods and changes must have prior approval of the 
Administrator. An owner employing such methods or deviations from the 
test methods without obtaining prior approval does so at the risk of 
subsequent disapproval and retesting with approved methods.
    Within the test methods, certain specific equipment or procedures 
are recognized as being acceptable or potentially acceptable and are 
specifically identified in the methods. The items identified as 
acceptable options may be used without approval but must be identified 
in the test report. The potentially approvable options are cited as 
``subject to the approval of the Administrator'' or as ``or 
equivalent.'' Such potentially approvable techniques or alternatives may 
be used at the discretion of the owner without prior approval. However, 
detailed descriptions for applying these potentially approvable 
techniques or alternatives are not provided in the test methods. Also, 
the potentially approvable options are not necessarily acceptable in all 
applications. Therefore, an owner electing to use such potentially 
approvable techniques or alternatives is responsible for: (1) assuring 
that the techniques or alternatives are in fact applicable and are 
properly executed; (2) including a written description of the 
alternative method in the test report (the written method must be clear 
and must be capable of being performed without additional instruction, 
and the degree of detail should be similar to the detail contained in 
the test methods); and (3) providing any rationale or supporting data 
necessary to show the validity of the alternative in the particular 
application. Failure to meet these requirements can result in the 
Administrator's disapproval of the alternative.

    Method 16--Semicontinuous Determination of Sulfur Emissions From 
                           Stationary Sources

    Note: This method does not include all of the specifications (e.g., 
equipment and supplies) and procedures (e.g., sampling and analytical) 
essential to its performance. Some material is incorporated by reference 
from other methods in this part. Therefore, to obtain reliable results, 
persons using this method should have a thorough knowledge of at least 
the following additional test methods: Method 1, Method 4, Method 15, 
and Method 16A.

                        1.0 Scope and Application

    1.1 Analytes.

------------------------------------------------------------------------
            Analyte                  CAS No.           Sensitivity
------------------------------------------------------------------------
Dimethyl disulfide [(CH3)2S2]..        62-49-20  50 ppb.
Dimethyl sulfide [(CH3)2S].....         75-18-3  50 ppb.

[[Page 405]]


Hydrogen sulfide [H2S].........       7783-06-4  50 ppb.
Methyl mercaptan [CH4S]........         74-93-1  50 ppb.
------------------------------------------------------------------------

    1.2 Applicability. This method is applicable for the determination 
of total reduced sulfur (TRS) compounds from recovery furnaces, lime 
kilns, and smelt dissolving tanks at kraft pulp mills and fuel gas 
combustion devices at petroleum refineries.

    Note: The method described below uses the principle of gas 
chromatographic (GC) separation and flame photometric detection (FPD). 
Since there are many systems or sets of operating conditions that 
represent useable methods of determining sulfur emissions, all systems 
which employ this principle, but differ only in details of equipment and 
operation, may be used as alternative methods, provided that the 
calibration precision and sample line loss criteria are met.

    1.3 Data Quality Objectives. Adherence to the requirements of this 
method will enhance the quality of the data obtained from air pollutant 
sampling methods.

                          2.0 Summary of Method

    2.1 A gas sample is extracted from the emission source and an 
aliquot is analyzed for hydrogen sulfide (H2S), methyl 
mercaptan (MeSH), dimethyl sulfide (DMS), and dimethyl disulfide (DMDS) 
by GC/FPD. These four compounds are known collectively as TRS.

                       3.0 Definitions [Reserved]

                            4.0 Interferences

    4.1 Moisture. Moisture condensation in the sample delivery system, 
the analytical column, or the FPD burner block can cause losses or 
interferences. This is prevented by maintaining the probe, filter box, 
and connections at a temperature of at least 120 [deg]C (248 [deg]F). 
Moisture is removed in the SO2 scrubber and heating the 
sample beyond this point is not necessary when the ambient temperature 
is above 0 [deg]C (32 [deg]F). Alternatively, moisture may be eliminated 
by heating the sample line, and by conditioning the sample with dry 
dilution air to lower its dew point below the operating temperature of 
the GC/FPD analytical system prior to analysis.
    4.2 Carbon Monoxide (CO) and Carbon Dioxide (CO2). CO and 
CO2 have a substantial desensitizing effect on the flame 
photometric detector even after dilution. Acceptable systems must 
demonstrate that they have eliminated this interference by some 
procedure such as eluting these compounds before any of the compounds to 
be measured. Compliance with this requirement can be demonstrated by 
submitting chromatograms of calibration gases with and without 
CO2 in the diluent gas. The CO2 level should be 
approximately 10 percent for the case with CO2 present. The 
two chromatograms should show agreement within the precision limits of 
Section 10.2.
    4.3 Particulate Matter. Particulate matter in gas samples can cause 
interference by eventual clogging of the analytical system. This 
interference is eliminated by using the Teflon filter after the probe.
    4.4 Sulfur Dioxide (SO2). Sulfur dioxide is not a 
specific interferant but may be present in such large amounts that it 
cannot effectively be separated from the other compounds of interest. 
The SO2 scrubber described in Section 6.1.3 will effectively 
remove SO2 from the sample.

                               5.0 Safety

    5.1 Disclaimer. This method may involve hazardous materials, 
operations, and equipment. This test method may not address all of the 
safety problems associated with its use. It is the responsibility of the 
user of this test method to establish appropriate safety and health 
practices and determine the applicability of regulatory limitations 
prior to performing this test method.
    5.2 Hydrogen Sulfide. A flammable, poisonous gas with the odor of 
rotten eggs. H2S is extremely hazardous and can cause 
collapse, coma, and death within a few seconds of one or two inhalations 
at sufficient concentrations. Low concentrations irritate the mucous 
membranes and may cause nausea, dizziness, and headache after exposure.

                       6.0 Equipment and Supplies

    6.1. Sample Collection. The following items are needed for sample 
collection.
    6.1.1 Probe. Teflon or Teflon-lined stainless steel. The probe must 
be heated to prevent moisture condensation. It must be designed to allow 
calibration gas to enter the probe at or near the sample point entry. 
Any portion of the probe that contacts the stack gas must be heated to 
prevent moisture condensation. Figure 16-1 illustrates the probe used in 
lime kilns and other sources where significant amounts of particulate 
matter are present. The probe is designed with the deflector shield 
placed between the sample and the gas inlet holes to reduce clogging of 
the filter and possible adsorption of sample gas. As an alternative, the 
probe described in Section 6.1.1 of Method 16A having a nozzle directed 
away from the gas stream may be used at sources having significant 
amounts of particulate matter.

[[Page 406]]

    6.1.2 Particulate Filter. 50-mm Teflon filter holder and a 1- to 2-
micron porosity Teflon filter (available through Savillex Corporation, 
5325 Highway 101, Minnetonka, Minnesota 55343). The filter holder must 
be maintained in a hot box at a temperature of at least 120 [deg]C (248 
[deg]F).
    6.1.3 SO2 Scrubber. Three 300-ml Teflon segmented 
impingers connected in series with flexible, thick-walled, Teflon 
tubing. (Impinger parts and tubing available through Savillex.) The 
first two impingers contain 100 ml of citrate buffer and the third 
impinger is initially dry. The tip of the tube inserted into the 
solution should be constricted to less than 3 mm (\1/8\ in.) ID and 
should be immersed to a depth of at least 5 cm (2 in.). Immerse the 
impingers in an ice water bath and maintain near 0 [deg]C (32 [deg]F). 
The scrubber solution will normally last for a 3-hour run before needing 
replacement. This will depend upon the effects of moisture and 
particulate matter on the solution strength and pH. Connections between 
the probe, particulate filter, and SO2 scrubber must be made 
of Teflon and as short in length as possible. All portions of the probe, 
particulate filter, and connections prior to the SO2 scrubber 
(or alternative point of moisture removal) must be maintained at a 
temperature of at least 120 [deg]C (248 [deg]F).
    6.1.4 Sample Line. Teflon, no greater than 1.3 cm (\1/2\ in.) ID. 
Alternative materials, such as virgin Nylon, may be used provided the 
line loss test is acceptable.
    6.1.5 Sample Pump. The sample pump must be a leakless Teflon-coated 
diaphragm type or equivalent.
    6.2 Analysis. The following items are needed for sample analysis:
    6.2.1 Dilution System. Needed only for high sample concentrations. 
The dilution system must be constructed such that all sample contacts 
are made of Teflon, glass, or stainless steel.
    6.2.2 Gas Chromatograph. The gas chromatograph must have at least 
the following components:
    6.2.2.1 Oven. Capable of maintaining the separation column at the 
proper operating temperature 1 [deg]C (2 [deg]F).
    6.2.2.2 Temperature Gauge. To monitor column oven, detector, and 
exhaust temperature 1 [deg]C (2 [deg]F).
    6.2.2.3 Flow System. Gas metering system to measure sample, fuel, 
combustion gas, and carrier gas flows.
    6.2.2.4 Flame Photometric Detector.
    6.2.2.4.1 Electrometer. Capable of full scale amplification of 
linear ranges of 10-9 to 10-4 amperes full scale.
    6.2.2.4.2 Power Supply. Capable of delivering up to 750 volts.
    6.2.2.4.3 Recorder. Compatible with the output voltage range of the 
electrometer.
    6.2.2.4.4 Rotary Gas Valves. Multiport Teflon-lined valves equipped 
with sample loop. Sample loop volumes must be chosen to provide the 
needed analytical range. Teflon tubing and fittings must be used 
throughout to present an inert surface for sample gas. The gas 
chromatograph must be calibrated with the sample loop used for sample 
analysis.
    6.2.3 Gas Chromatogram Columns. The column system must be 
demonstrated to be capable of resolving the four major reduced sulfur 
compounds: H2S, MeSH, DMS, and DMDS. It must also demonstrate 
freedom from known interferences. To demonstrate that adequate 
resolution has been achieved, submit a chromatogram of a calibration gas 
containing all four of the TRS compounds in the concentration range of 
the applicable standard. Adequate resolution will be defined as base 
line separation of adjacent peaks when the amplifier attenuation is set 
so that the smaller peak is at least 50 percent of full scale. Baseline 
separation is defined as a return to zero 5 
percent in the interval between peaks. Systems not meeting this criteria 
may be considered alternate methods subject to the approval of the 
Administrator.
    6.3 Calibration. A calibration system, containing the following 
components, is required (see Figure 16-2).
    6.3.1 Tube Chamber. Chamber of glass or Teflon of sufficient 
dimensions to house permeation tubes.
    6.3.2 Flow System. To measure air flow over permeation tubes at 
2 percent. Flow over the permeation device may 
also be determined using a soap bubble flowmeter.
    6.3.3 Constant Temperature Bath. Device capable of maintaining the 
permeation tubes at the calibration temperature within 0.1 [deg]C (0.2 
[deg]F).
    6.3.4 Temperature Gauge. Thermometer or equivalent to monitor bath 
temperature within 1 [deg]C (2 [deg]F).

                       7.0 Reagents and Standards

    7.1 Fuel. Hydrogen (H2), prepurified grade or better.
    7.2 Combustion Gas. Oxygen (O2) or air, research purity 
or better.
    7.3 Carrier Gas. Prepurified grade or better.
    7.4 Diluent (if required). Air containing less than 50 ppb total 
sulfur compounds and less than 10 ppmv each of moisture and total 
hydrocarbons.
    7.5 Calibration Gases
    7.5.1 Permeation tubes, one each of H2S, MeSH, DMS, and 
DMDS, gravimetrically calibrated and certified at some convenient 
operating temperature. These tubes consist of hermetically sealed FEP 
Teflon tubing in which a liquified gaseous substance is enclosed. The 
enclosed gas permeates through the tubing wall at a constant rate. When 
the temperature is constant, calibration gases

[[Page 407]]

covering a wide range of known concentrations can be generated by 
varying and accurately measuring the flow rate of diluent gas passing 
over the tubes. These calibration gases are used to calibrate the GC/FPD 
system and the dilution system.
    7.5.2 Cylinder Gases. Cylinder gases may be used as alternatives to 
permeation devices. The gases must be traceable to a primary standard 
(such as permeation tubes) and not used beyond the certification 
expiration date.
    7.6 Citrate Buffer and Sample Line Loss Gas. Same as Method 15, 
Sections 7.6 and 7.7.

       8.0 Sample Collection, Preservation, Storage, and Transport

    Same as Method 15, Section 8.0, except that the references to the 
dilution system may not be applicable.

                           9.0 Quality Control

------------------------------------------------------------------------
                                 Quality control
            Section                  measure               Effect
------------------------------------------------------------------------
8.0...........................  Sample line loss   Ensures that
                                 check.             uncorrected negative
                                                    bias introduced by
                                                    sample loss is no
                                                    greater than 20
                                                    percent, and
                                                    provides for
                                                    correction of bias
                                                    of 20 percent or
                                                    less.
8.0...........................  Calibration drift  Ensures that bias
                                 test.              introduced by drift
                                                    in the measurement
                                                    system output during
                                                    the run is no
                                                    greater than 5
                                                    percent.
10.0..........................  Analytical         Ensures precision of
                                 calibration.       analytical results
                                                    within 5 percent.
------------------------------------------------------------------------

                  10.0 Calibration and Standardization

    Same as Method 15, Section 10.0, with the following addition and 
exceptions:
    10.1 Use the four compounds that comprise TRS instead of the three 
reduced sulfur compounds measured by Method 15.
    10.2 Flow Meter. Calibration before each test run is recommended, 
but not required; calibration following each test series is mandatory. 
Calibrate each flow meter after each complete test series with a wet-
test meter. If the flow measuring device differs from the wet-test meter 
by 5 percent or more, the completed test runs must be voided. 
Alternatively, the flow data that yield the lower flow measurement may 
be used. Flow over the permeation device may also be determined using a 
soap bubble flowmeter.

                        11.0 Analytical Procedure

    Sample collection and analysis are concurrent for this method (see 
Section 8.0).

                   12.0 Data Analysis and Calculations

    12.1 Concentration of Reduced Sulfur Compounds. Calculate the 
average concentration of each of the four analytes (i.e., DMDS, DMS, 
H2S, and MeSH) over the sample run (specified in Section 8.2 
of Method 15 as 16 injections).
[GRAPHIC] [TIFF OMITTED] TR17OC00.278

Where:

Si=Concentration of any reduced sulfur compound from the 
ith sample injection, ppm.
C=Average concentration of any one of the reduced sulfur compounds for 
the entire run, ppm.
N=Number of injections in any run period.

    12.2 TRS Concentration. Using Equation 16-2, calculate the TRS 
concentration for each sample run.
[GRAPHIC] [TIFF OMITTED] TR17OC00.279

Where:

CTRS=TRS concentration, ppmv.
CH2S=Hydrogen sulfide concentration, ppmv.
CMeSH=Methyl mercaptan concentration, ppmv.
CDMS=Dimethyl sulfide concentration, ppmv.
CDMDS=Dimethyl disulfide concentration, ppmv.
d=Dilution factor, dimensionless.

    12.3 Average TRS Concentration. Calculate the average TRS 
concentration for all sample runs performed.
[GRAPHIC] [TIFF OMITTED] TR17OC00.280

Where:


[[Page 408]]


Average TRS=Average total reduced sulfur in ppm.
TRSi=Total reduced sulfur in ppm as determined by Equation 
16-2.
N=Number of samples.
Bwo=Fraction of volume of water vapor in the gas stream as 
determined by Method 4--Determination of Moisture in Stack Gases.

                         13.0 Method Performance

    13.1 Analytical Range. The analytical range will vary with the 
sample loop size. Typically, the analytical range may extend from 0.1 to 
100 ppmv using 10- to 0.1-ml sample loop sizes. This eliminates the need 
for sample dilution in most cases.
    13.2 Sensitivity. Using the 10-ml sample size, the minimum 
detectable concentration is approximately 50 ppb.

                  14.0 Pollution Prevention [Reserved]

                    15.0 Waste Management [Reserved]

                             16.0 References

    1. O'Keeffe, A.E., and G.C. Ortman. ``Primary Standards for Trace 
Gas Analysis.'' Analytical Chemical Journal, 38,76. 1966.
    2. Stevens, R.K., A.E. O'Keeffe, and G.C. Ortman. ``Absolute 
Calibration of a Flame Photometric Detector to Volatile Sulfur Compounds 
at Sub-Part-Per-Million Levels.'' Environmental Science and Technology, 
3:7. July 1969.
    3. Mulik, J.D., R.K. Stevens, and R. Baumgardner. ``An Analytical 
System Designed to Measure Multiple Malodorous Compounds Related to 
Kraft Mill Activities.'' Presented at the 12th Conference on Methods in 
Air Pollution and Industrial Hygiene Studies, University of Southern 
California, Los Angeles, CA. April 6-8, 1971.
    4. Devonald, R.H., R.S. Serenius, and A.D. McIntyre. ``Evaluation of 
the Flame Photometric Detector for Analysis of Sulfur Compounds.'' Pulp 
and Paper Magazine of Canada, 73,3. March 1972.
    5. Grimley, K.W., W.S. Smith, and R.M. Martin. ``The Use of a 
Dynamic Dilution System in the Conditioning of Stack Gases for Automated 
Analysis by a Mobile Sampling Van.'' Presented at the 63rd Annual APCA 
Meeting, St. Louis, MO. June 14-19, 1970.
    6. General Reference. Standard Methods of Chemical Analysis, Volumes 
III-A and III-B Instrumental Methods. Sixth Edition. Van Nostrand 
Reinhold Co.

         17.0 Tables, Diagrams, Flowcharts, and Validation Data

[[Page 409]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.281


[[Page 410]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.282

    Method 16A--Determination of Total Reduced Sulfur Emissions From 
                 Stationary Sources (Impinger Technique)

    Note: This method does not include all of the specifications (e.g., 
equipment and supplies) and procedures (e.g., sampling and analytical) 
essential to its performance. Some material is incorporated by reference 
from other methods in this part. Therefore, to obtain reliable results, 
persons using this method should have a thorough knowledge of at least 
the following additional test methods: Method 1, Method 6, and Method 
16.

                        1.0 Scope and Application

    1.1 Analytes.

[[Page 411]]



------------------------------------------------------------------------
              Analyte                   CAS No.          Sensitivity
------------------------------------------------------------------------
Total reduced sulfur (TRS)                     N/A  See Section 13.1.
 including:
    Dimethyl disulfide [(CH3)2S2].        62-49-20
    Dimethyl sulfide [(CH3)2S]....         75-18-3
    Hydrogen sulfide [H2S]........       7783-06-4
    Methyl mercaptan [CH4S].......         74-93-1
Reduced sulfur (RS) including:                 N/A
    H2S...........................       7783-06-4
    Carbonyl sulfide [COS]........        463-58-1
    Carbon disulfide [CS2]........         75-15-0
Reported as: Sulfur dioxide (SO2).       7449-09-5
------------------------------------------------------------------------

    1.2 Applicability. This method is applicable for the determination 
of TRS emissions from recovery boilers, lime kilns, and smelt dissolving 
tanks at kraft pulp mills, reduced sulfur compounds (H2S, 
carbonyl sulfide, and carbon disulfide from sulfur recovery units at 
onshore natural gas processing facilities, and from other sources when 
specified in an applicable subpart of the regulations. The flue gas must 
contain at least 1 percent oxygen for complete oxidation of all TRS to 
SO2.
    1.3 Data Quality Objectives. Adherence to the requirements of this 
method will enhance the quality of the data obtained from air pollutant 
sampling methods.

                          2.0 Summary of Method

    2.1 An integrated gas sample is extracted from the stack. 
SO2 is removed selectively from the sample using a citrate 
buffer solution. TRS compounds are then thermally oxidized to 
SO2, collected in hydrogen peroxide as sulfate, and analyzed 
by the Method 6 barium-thorin titration procedure.

                       3.0 Definitions [Reserved]

                            4.0 Interferences

    4.1 Reduced sulfur compounds other than those regulated by the 
emission standards, if present, may be measured by this method. 
Therefore, carbonyl sulfide, which is partially oxidized to 
SO2 and may be present in a lime kiln exit stack, would be a 
positive interferant.
    4.2 Particulate matter from the lime kiln stack gas (primarily 
calcium carbonate) can cause a negative bias if it is allowed to enter 
the citrate scrubber; the particulate matter will cause the pH to rise 
and H2S to be absorbed prior to oxidation. Furthermore, if 
the calcium carbonate enters the hydrogen peroxide impingers, the 
calcium will precipitate sulfate ion. Proper use of the particulate 
filter described in Section 6.1.3 will eliminate this interference.

                               5.0 Safety

    5.1 Disclaimer. This method may involve hazardous materials, 
operations, and equipment. This test method may not address all of the 
safety problems associated with its use. It is the responsibility of the 
user of this test method to establish appropriate safety and health 
practices and determine the applicability of regulatory limitations 
prior to performing this test method.
    5.2 Corrosive reagents. The following reagents are hazardous. 
Personal protective equipment and safe procedures are useful in 
preventing chemical splashes. If contact occurs, immediately flush with 
copious amounts of water for at least 15 minutes. Remove clothing under 
shower and decontaminate. Treat residual chemical burns as thermal 
burns.
    5.2.1 Hydrogen Peroxide (H2O2). Irritating to 
eyes, skin, nose, and lungs.
    5.2.2 Sodium Hydroxide (NaOH). Causes severe damage to eyes and 
skin. Inhalation causes irritation to nose, throat, and lungs. Reacts 
exothermically with limited amounts of water.
    5.2.3 Sulfuric Acid (H2SO4). Rapidly 
destructive to body tissue. Will cause third degree burns. Eye damage 
may result in blindness. Inhalation may be fatal from spasm of the 
larynx, usually within 30 minutes. May cause lung tissue damage with 
edema. 3 mg/m\3\ will cause lung damage in uninitiated. 1 mg/m\3\ for 8 
hours will cause lung damage or, in higher concentrations, death. 
Provide ventilation to limit inhalation. Reacts violently with metals 
and organics.
    5.3 Hydrogen Sulfide (H2S). A flammable, poisonous gas 
with the odor of rotten eggs. H2S is extremely hazardous and 
can cause collapse, coma, and death within a few seconds of one or two 
inhalations at sufficient concentrations. Low concentrations irritate 
the mucous membranes and may cause nausea, dizziness, and headache after 
exposure.

                       6.0 Equipment and Supplies

    6.1 Sample Collection. The sampling train is shown in Figure 16A-1 
and component parts are discussed below. Modifications to this sampling 
train are acceptable provided the system performance check is met (see 
Section 8.5).
    6.1.1 Probe. Teflon tubing, 6.4-mm (\1/4\-in.) diameter, 
sequentially wrapped with heat-resistant fiber strips, a rubberized heat 
tape (plug at one end), and heat-resistant adhesive tape. A flexible 
thermocouple or other suitable temperature measuring device

[[Page 412]]

should be placed between the Teflon tubing and the fiber strips so that 
the temperature can be monitored to prevent softening of the probe. The 
probe should be sheathed in stainless steel to provide in-stack 
rigidity. A series of bored-out stainless steel fittings placed at the 
front of the sheath will prevent moisture and particulate from entering 
between the probe and sheath. A 6.4-mm (\1/4\-in.) Teflon elbow (bored 
out) should be attached to the inlet of the probe, and a 2.54 cm (1 in.) 
piece of Teflon tubing should be attached at the open end of the elbow 
to permit the opening of the probe to be turned away from the 
particulate stream; this will reduce the amount of particulate drawn 
into the sampling train. The probe is depicted in Figure 16A-2.
    6.1.2 Probe Brush. Nylon bristle brush with handle inserted into a 
3.2-mm (\1/8\-in.) Teflon tubing. The Teflon tubing should be long 
enough to pass the brush through the length of the probe.
    6.1.3 Particulate Filter. 50-mm Teflon filter holder and a 1- to 2-
[micro]m porosity, Teflon filter (available through Savillex 
Corporation, 5325 Highway 101, Minnetonka, Minnesota 55343). The filter 
holder must be maintained in a hot box at a temperature sufficient to 
prevent moisture condensation. A temperature of 121 [deg]C (250 [deg]F) 
was found to be sufficient when testing a lime kiln under sub-freezing 
ambient conditions.
    6.1.4 SO2 Scrubber. Three 300-ml Teflon segmented 
impingers connected in series with flexible, thick-walled, Teflon 
tubing. (Impinger parts and tubing available through Savillex.) The 
first two impingers contain 100 ml of citrate buffer and the third 
impinger is initially dry. The tip of the tube inserted into the 
solution should be constricted to less than 3 mm (\1/8\-in.) ID and 
should be immersed to a depth of at least 5 cm (2 in.).
    6.1.5 Combustion Tube. Quartz glass tubing with an expanded 
combustion chamber 2.54 cm (1 in.) in diameter and at least 30.5 cm (12 
in.) long. The tube ends should have an outside diameter of 0.6 cm (\1/
4\ in.) and be at least 15.3 cm (6 in.) long. This length is necessary 
to maintain the quartz-glass connector near ambient temperature and 
thereby avoid leaks. Alternatively, the outlet may be constructed with a 
90-degree glass elbow and socket that would fit directly onto the inlet 
of the first peroxide impinger.
    6.1.6 Furnace. A furnace of sufficient size to enclose the 
combustion chamber of the combustion tube with a temperature regulator 
capable of maintaining the temperature at 800 100 
[deg]C (1472 180 [deg]F). The furnace operating 
temperature should be checked with a thermocouple to ensure accuracy.
    6.1.7 Peroxide Impingers, Stopcock Grease, Temperature Sensor, 
Drying Tube, Valve, Pump, and Barometer. Same as Method 6, Sections 
6.1.1.2, 6.1.1.4, 6.1.1.5, 6.1.1.6, 6.1.1.7, 6.1.1.8, and 6.1.2, 
respectively, except that the midget bubbler of Method 6, Section 
6.1.1.2 is not required.
    6.1.8 Vacuum Gauge. At least 760 mm Hg (30 in. Hg) gauge.
    6.1.9 Rate Meter. Rotameter, or equivalent, accurate to within 5 
percent at the selected flow rate of approximately 2 liters/min (4.2 
ft\3\/hr).
    6.1.10 Volume Meter. Dry gas meter capable of measuring the sample 
volume under the sampling conditions of 2 liters/min (4.2 ft\3\/hr) with 
an accuracy of 2 percent.
    6.2 Sample Recovery. Polyethylene Bottles, 250-ml (one per sample).
    6.3 Sample Preparation and Analysis. Same as Method 6, Section 6.3, 
except a 10-ml buret with 0.05-ml graduations is required, and the 
spectrophotometer is not needed.

                       7.0 Reagents and Standards

    Note: Unless otherwise indicated, all reagents must conform to the 
specifications established by the Committee on Analytical Reagents of 
the American Chemical Society. When such specifications are not 
available, the best available grade must be used.

    7.1 Sample Collection. The following reagents are required for 
sample analysis:
    7.1.1 Water. Same as in Method 6, Section 7.1.1.
    7.1.2 Citrate Buffer. Dissolve 300 g of potassium citrate (or 284 g 
of sodium citrate) and 41 g of anhydrous citric acid in 1 liter of water 
(200 ml is needed per test). Adjust the pH to between 5.4 and 5.6 with 
potassium citrate or citric acid, as required.
    7.1.3 Hydrogen Peroxide, 3 percent. Same as in Method 6, Section 
7.1.3 (40 ml is needed per sample).
    7.1.4 Recovery Check Gas. Hydrogen sulfide (100 ppmv or less) in 
nitrogen, stored in aluminum cylinders. Verify the concentration by 
Method 11 or by gas chromatography where the instrument is calibrated 
with an H2S permeation tube as described below. For Method 
11, the relative standard deviation should not exceed 5 percent on at 
least three 20-minute runs.

    Note: Alternatively, hydrogen sulfide recovery gas generated from a 
permeation device gravimetrically calibrated and certified at some 
convenient operating temperature may be used. The permeation rate of the 
device must be such that at a dilution gas flow rate of 3 liters/min 
(6.4 ft\3\/hr), an H2S concentration in the range of the 
stack gas or within 20 percent of the standard can be generated.

    7.1.5 Combustion Gas. Gas containing less than 50 ppb reduced sulfur 
compounds and less than 10 ppmv total hydrocarbons. The gas may be 
generated from a clean-air system that purifies ambient air and consists 
of the following components: Diaphragm pump, silica gel drying tube, 
activated charcoal

[[Page 413]]

tube, and flow rate measuring device. Flow from a compressed air 
cylinder is also acceptable.
    7.2 Sample Recovery and Analysis. Same as Method 6, Sections 7.2.1 
and 7.3, respectively.

       8.0 Sample Collection, Preservation, Storage, and Transport

    8.1 Preparation of Sampling Train.
    8.1.1 For the SO2 scrubber, measure 100 ml of citrate 
buffer into the first and second impingers; leave the third impinger 
empty. Immerse the impingers in an ice bath, and locate them as close as 
possible to the filter heat box. The connecting tubing should be free of 
loops. Maintain the probe and filter temperatures sufficiently high to 
prevent moisture condensation, and monitor with a suitable temperature 
sensor.
    8.1.2 For the Method 6 part of the train, measure 20 ml of 3 percent 
hydrogen peroxide into the first and second midget impingers. Leave the 
third midget impinger empty, and place silica gel in the fourth midget 
impinger. Alternatively, a silica gel drying tube may be used in place 
of the fourth impinger. Maintain the oxidation furnace at 800 100 [deg]C (1472 180 [deg]F). 
Place crushed ice and water around all impingers.
    8.2 Citrate Scrubber Conditioning Procedure. Condition the citrate 
buffer scrubbing solution by pulling stack gas through the Teflon 
impingers and bypassing all other sampling train components. A purge 
rate of 2 liters/min for 10 minutes has been found to be sufficient to 
obtain equilibrium. After the citrate scrubber has been conditioned, 
assemble the sampling train, and conduct (optional) a leak-check as 
described in Method 6, Section 8.2.
    8.3 Sample Collection. Same as in Method 6, Section 8.3, except the 
sampling rate is 2 liters/min (10 percent) for 1 
or 3 hours. After the sample is collected, remove the probe from the 
stack, and conduct (mandatory) a post-test leak-check as described in 
Method 6, Section 8.2. The 15-minute purge of the train following 
collection should not be performed. After each 3-hour test run (or after 
three 1-hour samples), conduct one system performance check (see Section 
8.5) to determine the reduced sulfur recovery efficiency through the 
sampling train. After this system performance check and before the next 
test run, rinse and brush the probe with water, replace the filter, and 
change the citrate scrubber (optional but recommended).

    Note: In Method 16, a test run is composed of 16 individual analyses 
(injects) performed over a period of not less than 3 hours or more than 
6 hours. For Method 16A to be consistent with Method 16, the following 
may be used to obtain a test run: (1) collect three 60-minute samples or 
(2) collect one 3-hour sample. (Three test runs constitute a test.)

    8.4 Sample Recovery. Disconnect the impingers. Quantitatively 
transfer the contents of the midget impingers of the Method 6 part of 
the train into a leak-free polyethylene bottle for shipment. Rinse the 
three midget impingers and the connecting tubes with water and add the 
washings to the same storage container. Mark the fluid level. Seal and 
identify the sample container.
    8.5 System Performance Check.
    8.5.1 A system performance check is done (1) to validate the 
sampling train components and procedure (prior to testing; optional) and 
(2) to validate a test run (after a run). Perform a check in the field 
prior to testing consisting of at least two samples (optional), and 
perform an additional check after each 3 hour run or after three 1-hour 
samples (mandatory).
    8.5.2 The checks involve sampling a known concentration of 
H2S and comparing the analyzed concentration with the known 
concentration. Mix the H2S recovery check gas (Section 7.1.4) 
and combustion gas in a dilution system such as that shown in Figure 
16A-3. Adjust the flow rates to generate an H2S concentration 
in the range of the stack gas or within 20 percent of the applicable 
standard and an oxygen concentration greater than 1 percent at a total 
flow rate of at least 2.5 liters/min (5.3 ft\3\/hr). Use Equation 16A-3 
to calculate the concentration of recovery gas generated. Calibrate the 
flow rate from both sources with a soap bubble flow meter so that the 
diluted concentration of H2S can be accurately calculated.
    8.5.3 Collect 30-minute samples, and analyze in the same manner as 
the emission samples. Collect the sample through the probe of the 
sampling train using a manifold or some other suitable device that will 
ensure extraction of a representative sample.
    8.5.4 The recovery check must be performed in the field prior to 
replacing the SO2 scrubber and particulate filter and before 
the probe is cleaned. Use Equation 16A-4 (see Section 12.5) to calculate 
the recovery efficiency. Report the recovery efficiency with the 
emission data; do not correct the emission data for the recovery 
efficiency. A sample recovery of 100 20 percent 
must be obtained for the emission data to be valid. However, if the 
recovery efficiency is not in the 100 20 percent 
range but the results do not affect the compliance or noncompliance 
status of the affected facility, the Administrator may decide to accept 
the results of the compliance test.

                           9.0 Quality Control

[[Page 414]]



------------------------------------------------------------------------
                                 Quality control
            Section                  measure               Effect
------------------------------------------------------------------------
8.5...........................  System             Ensure validity of
                                 performance        sampling train
                                 check.             components and
                                                    analytical
                                                    procedure.
8.2, 10.0.....................  Sampling           Ensure accurate
                                 equipment leak-    measurement of stack
                                 check and          gas flow rate,
                                 calibration.       sample volume.
10.0..........................  Barium standard    Ensure precision of
                                 solution           normality
                                 standardization.   determination.
11.1..........................  Replicate          Ensure precision of
                                 titrations.        titration
                                                    determinations.
11.2..........................  Audit sample       Evaluate analyst's
                                 analysis.          technique and
                                                    standards
                                                    preparation.
------------------------------------------------------------------------

                            10.0 Calibration

    Same as Method 6, Section 10.0.

                        11.0 Analytical Procedure

    11.1 Sample Loss Check and Sample Analysis. Same as Method 6, 
Sections 11.1 and 11.2, respectively, with the following exception: for 
1-hour sampling, take a 40-ml aliquot, add 160 ml of 100 percent 
isopropanol and four drops of thorin.
    11.2 Audit Sample Analysis. Same as Method 6, Section 11.3.

                   12.0 Data Analysis and Calculations

    In the calculations, at least one extra decimal figure should be 
retained beyond that of the acquired data. Figures should be rounded off 
after final calculations.
    12.1 Nomenclature.

CTRS=Concentration of TRS as SO2, dry basis 
corrected to standard conditions, ppmv.
CRG(act)=Actual concentration of recovery check gas (after 
dilution), ppm.
CRG(m)=Measured concentration of recovery check gas 
generated, ppm.
CH2S=Verified concentration of H2S recovery gas.
N=Normality of barium perchlorate titrant, milliequivalents/ml.
Pbar=Barometric pressure at exit orifice of the dry gas 
meter, mm Hg (in. Hg).
Pstd=Standard absolute pressure, 760 mm Hg (29.92 in. Hg).
QH2S=Calibrated flow rate of H2S recovery gas, 
liters/min.
QCG=Calibrated flow rate of combustion gas, liters/min.
R=Recovery efficiency for the system performance check, percent.
Tm=Average dry gas meter absolute temperature, [deg]K 
([deg]R).
Tstd=Standard absolute temperature, 293 [deg]K (528 [deg]R).
Va=Volume of sample aliquot titrated, ml.
Vm=Dry gas volume as measured by the dry gas meter, liters 
(dcf).
Vm(std)=Dry gas volume measured by the dry gas meter, 
corrected to standard conditions, liters (dscf).
Vsoln=Total volume of solution in which the sulfur dioxide 
sample is contained, 100 ml.
Vt=Volume of barium perchlorate titrant used for the sample, 
ml (average of replicate titrations).
Vtb=Volume of barium perchlorate titrant used for the blank, 
ml.
Y=Dry gas meter calibration factor.
32.03=Equivalent weight of sulfur dioxide, mg/meq.

    12.2 Dry Sample Gas Volume, Corrected to Standard Conditions.
    [GRAPHIC] [TIFF OMITTED] TR17OC00.283
    
Where:

K1=0.3855 [deg]K/mm Hg for metric units,
    =17.65 [deg]R/in. Hg for English units.

    12.3 Concentration of TRS as ppm SO2.
    [GRAPHIC] [TIFF OMITTED] TR17OC00.284
    
Where:

[[Page 415]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.285

    12.4 Concentration of Recovery Gas Generated in the System 
Performance Check.
[GRAPHIC] [TIFF OMITTED] TR17OC00.286

    12.5 Recovery Efficiency for the System Performance Check.
    [GRAPHIC] [TIFF OMITTED] TR17OC00.287
    
                         13.0 Method Performance

    13.1 Analytical Range. The lower detectable limit is 0.1 ppmv 
SO2 when sampling at 2 liters/min (4.2 ft\3\/hr) for 3 hours 
or 0.3 ppmv when sampling at 2 liters/min (4.2 ft\3\/hr) for 1 hour. The 
upper concentration limit of the method exceeds the TRS levels generally 
encountered at kraft pulp mills.
    13.2 Precision. Relative standard deviations of 2.0 and 2.6 percent 
were obtained when sampling a recovery boiler for 1 and 3 hours, 
respectively.
    13.3 Bias.
    13.3.1 No bias was found in Method 16A relative to Method 16 in a 
separate study at a recovery boiler.
    13.3.2 Comparison of Method 16A with Method 16 at a lime kiln 
indicated that there was no bias in Method 16A. However, instability of 
the source emissions adversely affected the comparison. The precision of 
Method 16A at the lime kiln was similar to that obtained at the recovery 
boiler (Section 13.2.1).
    13.3.3 Relative standard deviations of 2.7 and 7.7 percent have been 
obtained for system performance checks.

                  14.0 Pollution Prevention [Reserved]

                    15.0 Waste Management [Reserved]

                       16.0 Alternative Procedures

    As an alternative to the procedures specified in Section 7.1.4, the 
following procedure may be used to verify the H2S 
concentration of the recovery check gas.
    16.1 Summary. The H2S is collected from the calibration 
gas cylinder and is absorbed in zinc acetate solution to form zinc 
sulfide. The latter compound is then measured iodometrically.
    16.2 Range. The procedure has been examined in the range of 5 to 
1500 ppmv.
    16.3 Interferences. There are no known interferences to this 
procedure when used to analyze cylinder gases containing H2S 
in nitrogen.
    16.4 Precision and Bias. Laboratory tests have shown a relative 
standard deviation of less than 3 percent. The procedure showed no bias 
when compared to a gas chromatographic method that used gravimetrically 
certified permeation tubes for calibration.
    16.5 Equipment and Supplies.
    16.5.1 Sampling Apparatus. The sampling train is shown in Figure 
16A-4. Its component parts are discussed in Sections 16.5.1.1 through 
16.5.2.
    16.5.1.1 Sampling Line. Teflon tubing (\1/4\-in.) to connect the 
cylinder regulator to the sampling valve.
    16.5.1.2 Needle Valve. Stainless steel or Teflon needle valve to 
control the flow rate of gases to the impingers.
    16.5.1.3 Impingers. Three impingers of approximately 100-ml 
capacity, constructed to permit the addition of reagents through the gas 
inlet stem. The impingers shall be connected in series with leak-free 
glass or Teflon connectors. The impinger bottoms have a standard 24/25 
ground-glass fitting. The stems are from standard 6.4-mm (\1/4\-in.) 
ball joint midget impingers, custom lengthened by about 1 in. When 
fitted together, the stem end should be approximately 1.27 cm (\1/2\ 
in.) from the bottom (Southern Scientific, Inc., Micanopy, Florida: Set 
Number S6962-048). The third in-line impinger acts as a drop-out bottle.
    16.5.1.4 Drying Tube, Rate Meter, and Barometer. Same as Method 11, 
Sections 6.1.5, 6.1.8, and 6.1.10, respectively.
    16.5.1.5 Cylinder Gas Regulator. Stainless steel, to reduce the 
pressure of the gas stream entering the Teflon sampling line to a safe 
level.
    16.5.1.6 Soap Bubble Meter. Calibrated for 100 and 500 ml, or two 
separate bubble meters.
    16.5.1.7 Critical Orifice. For volume and rate measurements. The 
critical orifice may be fabricated according to Section 16.7.3 and must 
be calibrated as specified in Section 16.12.4.
    16.5.1.8 Graduated Cylinder. 50-ml size.
    16.5.1.9 Volumetric Flask. 1-liter size.
    16.5.1.10 Volumetric Pipette. 15-ml size.

[[Page 416]]

    16.5.1.11 Vacuum Gauge. Minimum 20 in. Hg capacity.
    16.5.1.12 Stopwatch.
    16.5.2 Sample Recovery and Analysis.
    16.5.2.1 Erlenmeyer Flasks. 125- and 250-ml sizes.
    16.5.2.2 Pipettes. 2-, 10-, 20-, and 100-ml volumetric.
    16.5.2.3 Burette. 50-ml size.
    16.5.2.4 Volumetric Flask. 1-liter size.
    16.5.2.5 Graduated Cylinder. 50-ml size.
    16.5.2.6 Wash Bottle.
    16.5.2.7 Stirring Plate and Bars.
    16.6 Reagents and Standards. Unless otherwise indicated, all 
reagents must conform to the specifications established by the Committee 
on Analytical Reagents of the American Chemical Society, where such 
specifications are available. Otherwise, use the best available grade.
    16.6.1 Water. Same as Method 11, Section 7.1.3.
    16.6.2 Zinc Acetate Absorbing Solution. Dissolve 20 g zinc acetate 
in water, and dilute to 1 liter.
    16.6.3 Potassium Bi-iodate [KH(IO3)2] 
Solution, Standard 0.100 N. Dissolve 3.249 g anhydrous 
KH(IO3)2 in water, and dilute to 1 liter.
    16.6.4 Sodium Thiosulfate (Na2S2O3) 
Solution, Standard 0.1 N. Same as Method 11, Section 7.3.2. Standardize 
according to Section 16.12.2.
    16.6.5 Na2S2O3 Solution, Standard 
0.01 N. Pipette 100.0 ml of 0.1 N 
Na2S2O3 solution into a 1-liter 
volumetric flask, and dilute to the mark with water.
    16.6.6 Iodine Solution, 0.1 N. Same as Method 11, Section 7.2.3.
    16.6.7 Standard Iodine Solution, 0.01 N. Same as in Method 11, 
Section 7.2.4. Standardize according to Section 16.12.3.
    16.6.8 Hydrochloric Acid (HCl) Solution, 10 Percent by Weight. Add 
230 ml concentrated HCl (specific gravity 1.19) to 770 ml water.
    16.6.9 Starch Indicator Solution. To 5 g starch (potato, arrowroot, 
or soluble), add a little cold water, and grind in a mortar to a thin 
paste. Pour into 1 liter of boiling water, stir, and let settle 
overnight. Use the clear supernatant. Preserve with 1.25 g salicylic 
acid, 4 g zinc chloride, or a combination of 4 g sodium propionate and 2 
g sodium azide per liter of starch solution. Some commercial starch 
substitutes are satisfactory.
    16.7 Pre-test Procedures.
    16.7.1 Selection of Gas Sample Volumes. This procedure has been 
validated for estimating the volume of cylinder gas sample needed when 
the H2S concentration is in the range of 5 to 1500 ppmv. The 
sample volume ranges were selected in order to ensure a 35 to 60 percent 
consumption of the 20 ml of 0.01 N iodine (thus ensuring a 0.01 N 
Na2S2O3 titer of approximately 7 to 12 
ml). The sample volumes for various H2S concentrations can be 
estimated by dividing the approximate ppm-liters desired for a given 
concentration range by the H2S concentration stated by the 
manufacturer. For example, for analyzing a cylinder gas containing 
approximately 10 ppmv H2S, the optimum sample volume is 65 
liters (650 ppm-liters/10 ppmv). For analyzing a cylinder gas containing 
approximately 1000 ppmv H2S, the optimum sample volume is 1 
liter (1000 ppm-liters/1000 ppmv).

------------------------------------------------------------------------
                                                            Approximate
    Approximate cylinder gas H2S concentration (ppmv)       ppm-liters
                                                              desired
------------------------------------------------------------------------
5 to <30................................................             650
30 to <500..............................................             800
500 to <1500............................................            1000
------------------------------------------------------------------------

    16.7.2 Critical Orifice Flow Rate Selection. The following table 
shows the ranges of sample flow rates that are desirable in order to 
ensure capture of H2S in the impinger solution. Slight 
deviations from these ranges will not have an impact on measured 
concentrations.

------------------------------------------------------------------------
                                            Critical  orifice  flow rate
  Cylinder gas H2S concentration (ppmv)               (ml/min)
------------------------------------------------------------------------
5 to 50 ppmv.............................  1500 500
50 to 250 ppmv...........................  500 250
250 to <1000 ppmv........................  200 50
1000 ppmv.....................  75 25
------------------------------------------------------------------------

    16.7.3 Critical Orifice Fabrication. Critical orifice of desired 
flow rates may be fabricated by selecting an orifice tube of desired 
length and connecting \1/16\-in.x\1/4\-in. (0.16 cmx0.64 cm) reducing 
fittings to both ends. The inside diameters and lengths of orifice tubes 
needed to obtain specific flow rates are shown below.

----------------------------------------------------------------------------------------------------------------
                                                                                  Flowrate  (ml/      Altech
                 Tube  (in. OD)                   Tube  (in. ID)   Length  (in.)       min)         Catalog No.
----------------------------------------------------------------------------------------------------------------
\1/16\..........................................           0.007             1.2              85          301430
\1/16\..........................................           0.01              3.2             215          300530
\1/16\..........................................           0.01              1.2             350          300530
\1/16\..........................................           0.02              1.2            1400          300230
----------------------------------------------------------------------------------------------------------------

    16.7.4 Determination of Critical Orifice Approximate Flow Rate. 
Connect the critical orifice to the sampling system as shown in Figure 
16A-4 but without the H2S cylinder.

[[Page 417]]

Connect a rotameter in the line to the first impinger. Turn on the pump, 
and adjust the valve to give a reading of about half atmospheric 
pressure. Observe the rotameter reading. Slowly increase the vacuum 
until a stable flow rate is reached, and record this as the critical 
vacuum. The measured flow rate indicates the expected critical flow rate 
of the orifice. If this flow rate is in the range shown in Section 
16.7.2, proceed with the critical orifice calibration according to 
Section 16.12.4.
    16.7.5 Determination of Approximate Sampling Time. Determine the 
approximate sampling time for a cylinder of known concentration. Use the 
optimum sample volume obtained in Section 16.7.1.
[GRAPHIC] [TIFF OMITTED] TR17OC00.288

    16.8 Sample Collection.
    16.8.1 Connect the Teflon tubing, Teflon tee, and rotameter to the 
flow control needle valve as shown in Figure 16A-4. Vent the rotameter 
to an exhaust hood. Plug the open end of the tee. Five to 10 minutes 
prior to sampling, open the cylinder valve while keeping the flow 
control needle valve closed. Adjust the delivery pressure to 20 psi. 
Open the needle valve slowly until the rotameter shows a flow rate 
approximately 50 to 100 ml above the flow rate of the critical orifice 
being used in the system.
    16.8.2 Place 50 ml of zinc acetate solution in two of the impingers, 
connect them and the empty third impinger (dropout bottle) and the rest 
of the equipment as shown in Figure 16A-4. Make sure the ground-glass 
fittings are tight. The impingers can be easily stabilized by using a 
small cardboard box in which three holes have been cut, to act as a 
holder. Connect the Teflon sample line to the first impinger. Cover the 
impingers with a dark cloth or piece of plastic to protect the absorbing 
solution from light during sampling.
    16.8.3 Record the temperature and barometric pressure. Note the gas 
flow rate through the rotameter. Open the closed end of the tee. Connect 
the sampling tube to the tee, ensuring a tight connection. Start the 
sampling pump and stopwatch simultaneously. Note the decrease in flow 
rate through the excess flow rotameter. This decrease should equal the 
known flow rate of the critical orifice being used. Continue sampling 
for the period determined in Section 16.7.5.
    16.8.4 When sampling is complete, turn off the pump and stopwatch. 
Disconnect the sampling line from the tee and plug it. Close the needle 
valve followed by the cylinder valve. Record the sampling time.
    16.9 Blank Analysis. While the sample is being collected, run a 
blank as follows: To a 250-ml Erlenmeyer flask, add 100 ml of zinc 
acetate solution, 20.0 ml of 0.01 N iodine solution, and 2 ml HCl 
solution. Titrate, while stirring, with 0.01 N 
Na2S2O3 until the solution is light 
yellow. Add starch, and continue titrating until the blue color 
disappears. Analyze a blank with each sample, as the blank titer has 
been observed to change over the course of a day.

    Note: Iodine titration of zinc acetate solutions is difficult to 
perform because the solution turns slightly white in color near the end 
point, and the disappearance of the blue color is hard to recognize. In 
addition, a blue color may reappear in the solution about 30 to 45 
seconds after the titration endpoint is reached. This should not be 
taken to mean the original endpoint was in error. It is recommended that 
persons conducting this test perform several titrations to be able to 
correctly identify the endpoint. The importance of this should be 
recognized because the results of this analytical procedure are 
extremely sensitive to errors in titration.

    16.10 Sample Analysis. Sample treatment is similar to the blank 
treatment. Before detaching the stems from the bottoms of the impingers, 
add 20.0 ml of 0.01 N iodine solution through the stems of the impingers 
holding the zinc acetate solution, dividing it between the two (add 
about 15 ml to the first impinger and the rest to the second). Add 2 ml 
HCl solution through the stems, dividing it as with the iodine. 
Disconnect the sampling line, and store the impingers for 30 minutes. At 
the end of 30 minutes, rinse the impinger stems into the impinger 
bottoms. Titrate the impinger contents with 0.01 N 
Na2S2O3. Do not transfer the contents 
of the impinger to a flask because this may result in a loss of iodine 
and cause a positive bias.
    16.11 Post-test Orifice Calibration. Conduct a post-test critical 
orifice calibration run using the calibration procedures outlined in 
Section 16.12.4. If the Qstd obtained before and after the 
test differs by more than 5 percent, void the sample; if not, proceed to 
perform the calculations.
    16.12 Calibrations and Standardizations.
    16.12.1 Rotameter and Barometer. Same as Method 11, Sections 10.1.3 
and 10.1.4.
    16.12.2 Na2S2O3 Solution, 0.1 N. 
Standardize the 0.1 N Na2S2O3 solution 
as follows:

[[Page 418]]

To 80 ml water, stirring constantly, add 1 ml concentrated 
H2SO4, 10.0 ml of 0.100 N 
KH(IO3)2 and 1 g potassium iodide. Titrate 
immediately with 0.1 N Na2S2O3 until 
the solution is light yellow. Add 3 ml starch solution, and titrate 
until the blue color just disappears. Repeat the titration until 
replicate analyses agree within 0.05 ml. Take the average volume of 
Na2S2O3 consumed to calculate the 
normality to three decimal figures using Equation 16A-5.
    16.12.3 Iodine Solution, 0.01 N. Standardize the 0.01 N iodine 
solution as follows: Pipet 20.0 ml of 0.01 N iodine solution into a 125-
ml Erlenmeyer flask. Titrate with standard 0.01 N 
Na2S2O3 solution until the solution is 
light yellow. Add 3 ml starch solution, and continue titrating until the 
blue color just disappears. If the normality of the iodine tested is not 
0.010, add a few ml of 0.1 N iodine solution if it is low, or a few ml 
of water if it is high, and standardize again. Repeat the titration 
until replicate values agree within 0.05 ml. Take the average volume to 
calculate the normality to three decimal figures using Equation 16A-6.
    16.12.4 Critical Orifice. Calibrate the critical orifice using the 
sampling train shown in Figure 16A-4 but without the H2S 
cylinder and vent rotameter. Connect the soap bubble meter to the Teflon 
line that is connected to the first impinger. Turn on the pump, and 
adjust the needle valve until the vacuum is higher than the critical 
vacuum determined in Section 16.7.4. Record the time required for gas 
flow to equal the soap bubble meter volume (use the 100-ml soap bubble 
meter for gas flow rates below 100 ml/min, otherwise use the 500-ml soap 
bubble meter). Make three runs, and record the data listed in Table 16A-
1. Use these data to calculate the volumetric flow rate of the orifice.
    16.13 Calculations.
    16.13.1 Nomenclature.

Bwa=Fraction of water vapor in ambient air during orifice 
calibration.
CH2S=H2S concentration in cylinder gas, 
ppmv.
[GRAPHIC] [TIFF OMITTED] TR17OC00.289

Ma=Molecular weight of ambient air saturated at impinger 
temperature, g/g-mole.
Ms=Molecular weight of sample gas (nitrogen) saturated at 
impinger temperature, g/g-mole.

    Note: (For tests carried out in a laboratory where the impinger 
temperature is 25 [deg]C, Ma=28.5 g/g-mole and 
Ms=27.7 g/g-mole.)

NI=Normality of standard iodine solution (0.01 N), g-eq/
liter.
NT=Normality of standard 
Na2S2O3 solution (0.01 N), g-eq/liter.
Pbar=Barometric pressure, mm Hg.
Pstd=Standard absolute pressure, 760 mm Hg.
Qstd=Average volumetric flow rate through critical orifice, 
liters/min.
Tamb=Absolute ambient temperature, [deg]K.
Tstd=Standard absolute temperature, 293 [deg]K.
[thetas]s=Sampling time, min.
[thetas]sb=Time for soap bubble meter flow rate measurement, 
min.
Vm(std)=Sample gas volume measured by the critical orifice, 
corrected to standard conditions, liters.
Vsb=Volume of gas as measured by the soap bubble meter, ml.
Vsb(std)=Volume of gas as measured by the soap bubble meter, 
corrected to standard conditions, liters.
VI=Volume of standard iodine solution (0.01 N) used, ml.
VT=Volume of standard 
Na2S2O3 solution (0.01 N) used, ml.
VTB=Volume of standard 
Na2S2O3 solution (0.01 N) used for the 
blank, ml.

    16.13.2 Normality of Standard 
Na2S2O3 Solution (0.1 N).
[GRAPHIC] [TIFF OMITTED] TR17OC00.290

    16.13.3 Normality of Standard Iodine Solution (0.01 N).
      
      

[[Page 419]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.291

    16.13.4 Sample Gas Volume.
    [GRAPHIC] [TIFF OMITTED] TR17OC00.292
    
    16.13.5 Concentration of H2S in the Gas Cylinder.

                             17.0 References
[GRAPHIC] [TIFF OMITTED] TR17OC00.293

    1. American Public Health Association, American Water Works 
Association, and Water Pollution Control Federation. Standard Methods 
for the Examination of Water and Wastewater. Washington, DC. American 
Public Health Association. 1975. pp. 316-317.
    2. American Society for Testing and Materials. Annual Book of ASTM 
Standards. Part 31: Water, Atmospheric Analysis. Philadelphia, PA. 1974. 
pp. 40-42.
    3. Blosser, R.O. A Study of TRS Measurement Methods. National 
Council of the Paper Industry for Air and Stream Improvement, Inc., New 
York, NY. Technical Bulletin No. 434. May 1984. 14 pp.
    4. Blosser, R.O., H.S. Oglesby, and A.K. Jain. A Study of Alternate 
SO2 Scrubber Designs Used for TRS Monitoring. A Special 
Report by the National Council of the Paper Industry for Air and Stream 
Improvement, Inc., New York, NY. July 1977.
    5. Curtis, F., and G.D. McAlister. Development and Evaluation of an 
Oxidation/Method 6 TRS Emission Sampling Procedure. Emission Measurement 
Branch, Emission Standards and Engineering Division, U.S. Environmental 
Protection Agency, Research Triangle Park, NC 27711. February 1980.
    6. Gellman, I. A Laboratory and Field Study of Reduced Sulfur 
Sampling and Monitoring Systems. National Council of the Paper Industry 
for Air and Stream Improvement, Inc., New York, NY. Atmospheric Quality 
Improvement Technical Bulletin No. 81. October 1975.
    7. Margeson, J.H., J.E. Knoll, and M.R. Midgett. A Manual Method for 
TRS Determination. Source Branch, Quality Assurance Division, U.S. 
Environmental Protection Agency, Research Triangle Park, NC 27711.
    8. National Council of the Paper Industry for Air and Stream 
Improvement. An Investigation of H2S and SO2. 
Calibration Cylinder Gas Stability and Their Standardization Using Wet 
Chemical Techniques. Special Report 76-06. New York, NY. August 1976.
    9. National Council of the Paper Industry for Air and Stream 
Improvement. Wet Chemical Method for Determining the H2S 
Concentration of Calibration Cylinder Gases. Technical Bulletin Number 
450. New York, NY. January 1985. 23 pp.
    10. National Council of the Paper Industry for Air and Stream 
Improvement. Modified Wet Chemical Method for Determining the 
H2S Concentration of Calibration Cylinder Gases. Draft 
Report. New York, NY. March 1987. 29 pp.

         18.0 Tables, Diagrams, Flowcharts, and Validation Data

[[Page 420]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.294


[[Page 421]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.295


[[Page 422]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.296


[[Page 423]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.297

Date____________________________________________________________________
Critical orifice ID_____________________________________________________
Soap bubble meter volume, Vsb---- liters
Time, [thetas]sb
Run no. 1 ---- min ---- sec
Run no. 2 ---- min ---- sec
Run no. 3 ---- min ---- sec
Average ---- min ---- sec
Covert the seconds to fraction of minute:
Time=---- min + ---- Sec/60=---- min
Barometric pressure, Pbar=---- mm Hg
Ambient temperature, t amb=273 + ---- [deg]C=---- [deg]K=---- 
mm Hg. (This should be approximately 0.4 times barometric pressure.)
Pump vacuum,

[[Page 424]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.298

             Table 16A-1. Critical Orifice Calibration Data

    Method 16B--Determination of Total Reduced Sulfur Emissions From 
                           Stationary Sources

    Note: This method does not include all of the specifications (e.g., 
equipment and supplies) and procedures (e.g., sampling and analytical) 
essential to its performance. Some material is incorporated by reference 
from other methods in this part. Therefore, to obtain reliable results, 
persons using this method should have a knowledge of at least the 
following additional test methods: Method 6C, Method 16, and Method 16A.

                        1.0 Scope and Application

    1.1 Analytes.

------------------------------------------------------------------------
                         Analyte                              CAS No.
------------------------------------------------------------------------
Total reduced sulfur (TRS) including:                                N/A
    Dimethyl disulfide (DMDS), [(CH3)2S2]...............        62-49-20
    Dimethyl sulfide (DMS), [(CH3)2S]...................         75-18-3
    Hydrogen sulfide (H2S)..............................       7783-06-4
    Methyl mercaptan (MeSH), [CH4S].....................         74-93-1
Reported as: Sulfur dioxide (SO2).......................       7449-09-5
------------------------------------------------------------------------

    1.2 Applicability. This method is applicable for determining TRS 
emissions from recovery furnaces (boilers), lime kilns, and smelt 
dissolving tanks at kraft pulp mills, and from other sources when 
specified in an applicable subpart of the regulations. The flue gas must 
contain at least 1 percent oxygen for complete oxidation of all TRS to 
SO2.
    1.3 Data Quality Objectives. Adherence to the requirements of this 
method will enhance the quality of the data obtained from air pollutant 
sampling methods.

                          2.0 Summary of Method

    2.1 An integrated gas sample is extracted from the stack. The 
SO2 is removed selectively from the sample using a citrate 
buffer solution. The TRS compounds are then thermally oxidized to 
SO2 and analyzed as SO2 by gas chromatography (GC) 
using flame photometric detection (FPD).

                       3.0 Definitions [Reserved]

                            4.0 Interferences

    4.1 Reduced sulfur compounds other than those regulated by the 
emission standards, if present, may be measured by this method. 
Therefore, carbonyl sulfide, which is partially oxidized to 
SO2 and may be present in a lime kiln exit stack, would be a 
positive interferant.
    4.2 Particulate matter from the lime kiln stack gas (primarily 
calcium carbonate) can cause a negative bias if it is allowed to enter 
the citrate scrubber; the particulate matter will cause the pH to rise 
and H2S to be absorbed before oxidation. Proper use of the 
particulate filter, described in Section 6.1.3 of Method 16A, will 
eliminate this interference.
    4.3 Carbon monoxide (CO) and carbon dioxide (CO2) have 
substantial desensitizing effects on the FPD even after dilution. 
Acceptable systems must demonstrate that they have eliminated this 
interference by some procedure such as eluting these compounds before 
the SO2. Compliance with this requirement can be demonstrated 
by submitting chromatograms of calibration gases with and without 
CO2 in diluent gas. The CO2 level should be 
approximately 10 percent for the case with CO2 present. The 
two chromatograms should show agreement within the precision limits of 
Section 13.0.

                               5.0 Safety

    5.1 Disclaimer. This method may involve hazardous materials, 
operations, and equipment. This test method may not address all of the 
safety problems associated with its use. It is the responsibility of the 
user of this test method to establish appropriate safety and health 
practices and determine the applicability of regulatory limitations 
prior to performing this test method.
    5.2 Hydrogen Sulfide (H2S). A flammable, poisonous gas 
with the odor of rotten eggs. H2S is extremely hazardous and 
can cause collapse, coma, and death within a few seconds of one or two 
inhalations at sufficient concentrations. Low concentrations irritate

[[Page 425]]

the mucous membranes and may cause nausea, dizziness, and headache after 
exposure.

                       6.0 Equipment and Supplies

    6.1 Sample Collection. The sampling train is shown in Figure 16B-1. 
Modifications to the apparatus are accepted provided the system 
performance check in Section 8.4.1 is met.
    6.1.1 Probe, Probe Brush, Particulate Filter, SO2 
Scrubber, Combustion Tube, and Furnace. Same as in Method 16A, Sections 
6.1.1 to 6.1.6.
    6.1.2 Sampling Pump. Leakless Teflon-coated diaphragm type or 
equivalent.
    6.2 Analysis.
    6.2.1 Dilution System (optional), Gas Chromatograph, Oven, 
Temperature Gauges, Flow System, Flame Photometric Detector, 
Electrometer, Power Supply, Recorder, Calibration System, Tube Chamber, 
Flow System, and Constant Temperature Bath. Same as in Method 16, 
Sections 6.2.1, 6.2.2, and 6.3.
    6.2.2 Gas Chromatograph Columns. Same as in Method 16, Section 
6.2.3. Other columns with demonstrated ability to resolve SO2 
and be free from known interferences are acceptable alternatives. Single 
column systems such as a 7-ft Carbsorb B HT 100 column have been found 
satisfactory in resolving SO2 from CO2.

                       7.0 Reagents and Standards

    Same as in Method 16, Section 7.0, except for the following:
    7.1 Calibration Gas. SO2 permeation tube gravimetrically 
calibrated and certified at some convenient operating temperature. These 
tubes consist of hermetically sealed FEP Teflon tubing in which a 
liquefied gaseous substance is enclosed. The enclosed gas permeates 
through the tubing wall at a constant rate. When the temperature is 
constant, calibration gases covering a wide range of known 
concentrations can be generated by varying and accurately measuring the 
flow rate of diluent gas passing over the tubes. In place of 
SO2 permeation tubes, cylinder gases containing 
SO2 in nitrogen may be used for calibration. The cylinder gas 
concentration must be verified according to Section 8.2.1 of Method 6C. 
The calibration gas is used to calibrate the GC/FPD system and the 
dilution system.
    7.2 Recovery Check Gas.
    7.2.1 Hydrogen sulfide [100 parts per million by volume (ppmv) or 
less] in nitrogen, stored in aluminum cylinders. Verify the 
concentration by Method 11, the procedure discussed in Section 16.0 of 
Method 16A, or gas chromatography where the instrument is calibrated 
with an H2S permeation tube as described below. For the wet-
chemical methods, the standard deviation should not exceed 5 percent on 
at least three 20-minute runs.
    7.2.2 Hydrogen sulfide recovery gas generated from a permeation 
device gravimetrically calibrated and certified at some convenient 
operation temperature may be used. The permeation rate of the device 
must be such that at a dilution gas flow rate of 3 liters/min (64 ft\3\/
hr), an H2S concentration in the range of the stack gas or 
within 20 percent of the emission standard can be generated.
    7.3 Combustion Gas. Gas containing less than 50 ppbv reduced sulfur 
compounds and less than 10 ppmv total hydrocarbons. The gas may be 
generated from a clean-air system that purifies ambient air and consists 
of the following components: diaphragm pump, silica gel drying tube, 
activated charcoal tube, and flow rate measuring device. Gas from a 
compressed air cylinder is also acceptable.

       8.0 Sample Collection, Preservation, Storage, and Transport

    8.1 Pretest Procedures. Same as in Method 15, Section 8.1.
    8.2 Sample Collection. Before any source sampling is performed, 
conduct a system performance check as detailed in Section 8.4.1 to 
validate the sampling train components and procedures. Although this 
test is optional, it would significantly reduce the possibility of 
rejecting tests as a result of failing the post-test performance check. 
At the completion of the pretest system performance check, insert the 
sampling probe into the test port making certain that no dilution air 
enters the stack though the port. Condition the entire system with 
sample for a minimum of 15 minutes before beginning analysis. If the 
sample is diluted, determine the dilution factor as in Section 10.4 of 
Method 15.
    8.3 Analysis. Inject aliquots of the sample into the GC/FPD analyzer 
for analysis. Determine the concentration of SO2 directly 
from the calibration curves or from the equation for the least-squares 
line.
    8.4. Post-Test Procedures
    8.4.1 System Performance Check. Same as in Method 16A, Section 8.5. 
A sufficient number of sample injections should be made so that the 
precision requirements of Section 13.2 are satisfied.
    8.4.2 Determination of Calibration Drift. Same as in Method 15, 
Section 8.3.2.

                           9.0 Quality Control

------------------------------------------------------------------------
                                 Quality control
            Section                  measure               Effect
------------------------------------------------------------------------
8.2, 8.3......................  System             Ensure validity of
                                 performance        sampling train
                                 check.             components and
                                                    analytical
                                                    procedure.

[[Page 426]]


8.1...........................  Sampling           Ensure accurate
                                 equipment leak-    measurement of stack
                                 check and          gas flow rate,
                                 calibration.       sample volume.
10.0..........................  Analytical         Ensure precision of
                                 calibration.       analytical results
                                                    within 5 percent.
------------------------------------------------------------------------

                            10.0 Calibration

    Same as in Method 16, Section 10, except SO2 is used 
instead of H2S.

                        11.0 Analytical Procedure

    11.1 Sample collection and analysis are concurrent for this method 
(see section 8.3).
    12.0 Data Analysis and Calculations
12.1 Nomenclature.

CSO2=Sulfur dioxide concentration, 
ppmv.
CTRS=Total reduced sulfur concentration 
as determined by Equation 16B-1, ppmv.
d=Dilution factor, dimensionless.
N=Number of samples.

    12.2 SO2 Concentration. Determine the concentration of 
SO2, CSO2, directly from the 
calibration curves. Alternatively, the concentration may be calculated 
using the equation for the least-squares line.
    12.3 TRS Concentration.
    [GRAPHIC] [TIFF OMITTED] TR17OC00.299
    
    12.4 Average TRS Concentration
    [GRAPHIC] [TIFF OMITTED] TR17OC00.300
    
                        13.0 Method Performance.

    13.1 Range and Sensitivity. Coupled with a GC using a 1-ml sample 
size, the maximum limit of the FPD for SO2 is approximately 
10 ppmv. This limit is extended by diluting the sample gas before 
analysis or by reducing the sample aliquot size. For sources with 
emission levels between 10 and 100 ppm, the measuring range can be best 
extended by reducing the sample size.
    13.2 GC/FPD Calibration and Precision. A series of three consecutive 
injections of the sample calibration gas, at any dilution, must produce 
results which do not vary by more than 5 percent from the mean of the 
three injections.
    13.3 Calibration Drift. The calibration drift determined from the 
mean of the three injections made at the beginning and end of any run or 
series of runs within a 24-hour period must not exceed 5 percent.
    13.4 System Calibration Accuracy. Losses through the sample 
transport system must be measured and a correction factor developed to 
adjust the calibration accuracy to 100 percent.
    13.5 Field tests between this method and Method 16A showed an 
average difference of less than 4.0 percent. This difference was not 
determined to be significant.

                  14.0 Pollution Prevention [Reserved]

                    15.0 Waste Management [Reserved]

                             16.0 References

    1. Same as in Method 16, Section 16.0.
    2. National Council of the Paper Industry for Air and Stream 
Improvement, Inc, A Study of TRS Measurement Methods. Technical Bulletin 
No. 434. New York, NY. May 1984. 12p.
    3. Margeson, J.H., J.E. Knoll, and M.R. Midgett. A Manual Method for 
TRS Determination. Draft available from the authors. Source Branch, 
Quality Assurance Division, U.S. Environmental Protection Agency, 
Research Triangle Park, NC 27711.

         17.0 Tables, Diagrams, Flowcharts, and Validation Data

[[Page 427]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.301

Method 17--Determination of Particulate Matter Emissions From Stationary 
                                 Sources

    Note: This method does not include all of the specifications (e.g., 
equipment and supplies) and procedures (e.g., sampling and analytical) 
essential to its performance. Some material is incorporated by reference 
from other methods in this part. Therefore, to obtain reliable results, 
persons using this method should have a thorough knowledge of at least 
the following additional test methods: Method 1, Method 2, Method 3, 
Method 5.

                        1.0 Scope and Application

    1.1 Analyte. Particulate matter (PM). No CAS number assigned.


[[Page 428]]


    Note: Particulate matter is not an absolute quantity. It is a 
function of temperature and pressure. Therefore, to prevent variability 
in PM emission regulations and/or associated test methods, the 
temperature and pressure at which PM is to be measured must be carefully 
defined. Of the two variables (i.e., temperature and pressure), 
temperature has the greater effect upon the amount of PM in an effluent 
gas stream; in most stationary source categories, the effect of pressure 
appears to be negligible. In Method 5, 120 [deg]C (248 [deg]F) is 
established as a nominal reference temperature. Thus, where Method 5 is 
specified in an applicable subpart of the standard, PM is defined with 
respect to temperature. In order to maintain a collection temperature of 
120 [deg]C (248 [deg]F), Method 5 employs a heated glass sample probe 
and a heated filter holder. This equipment is somewhat cumbersome and 
requires care in its operation. Therefore, where PM concentrations (over 
the normal range of temperature associated with a specified source 
category) are known to be independent of temperature, it is desirable to 
eliminate the glass probe and the heating systems, and to sample at 
stack temperature.

    1.2 Applicability. This method is applicable for the determination 
of PM emissions, where PM concentrations are known to be independent of 
temperature over the normal range of temperatures characteristic of 
emissions from a specified source category. It is intended to be used 
only when specified by an applicable subpart of the standards, and only 
within the applicable temperature limits (if specified), or when 
otherwise approved by the Administrator. This method is not applicable 
to stacks that contain liquid droplets or are saturated with water 
vapor. In addition, this method shall not be used as written if the 
projected cross-sectional area of the probe extension-filter holder 
assembly covers more than 5 percent of the stack cross-sectional area 
(see Section 8.1.2).
    1.3 Data Quality Objectives. Adherence to the requirements of this 
method will enhance the quality of the data obtained from air pollutant 
sampling methods.

                          2.0 Summary of Method

    2.1 Particulate matter is withdrawn isokinetically from the source 
and collected on a glass fiber filter maintained at stack temperature. 
The PM mass is determined gravimetrically after the removal of 
uncombined water.

                             3.0 Definitions

    Same as Method 5, Section 3.0.

                      4.0 Interferences [Reserved]

                               5.0 Safety

    5.1 Disclaimer. This method may involve hazardous materials, 
operations, and equipment. This test method may not address all of the 
safety problems associated with its use. It is the responsibility of the 
user of this test method to establish appropriate safety and health 
practices and determine the applicability of regulatory limitations 
prior to performing this test method.

                       6.0 Equipment and Supplies

    6.1 Sampling Train. A schematic of the sampling train used in this 
method is shown in Figure 17-1. The sampling train components and 
operation and maintenance are very similar to Method 5, which should be 
consulted for details.
    6.1.1 Probe Nozzle, Differential Pressure Gauge, Metering System, 
Barometer, Gas Density Determination Equipment. Same as in Method 5, 
Sections 6.1.1, 6.1.4, 6.1.8, 6.1.9, and 6.1.10, respectively.
    6.1.2 Filter Holder. The in-stack filter holder shall be constructed 
of borosilicate or quartz glass, or stainless steel. If a gasket is 
used, it shall be made of silicone rubber, Teflon, or stainless steel. 
Other holder and gasket materials may be used, subject to the approval 
of the Administrator. The filter holder shall be designed to provide a 
positive seal against leakage from the outside or around the filter.
    6.1.3 Probe Extension. Any suitable rigid probe extension may be 
used after the filter holder.
    6.1.4 Pitot Tube. Same as in Method 5, Section 6.1.3.
    6.1.4.1 It is recommended (1) that the pitot tube have a known 
baseline coefficient, determined as outlined in Section 10 of Method 2; 
and (2) that this known coefficient be preserved by placing the pitot 
tube in an interference-free arrangement with respect to the sampling 
nozzle, filter holder, and temperature sensor (see Figure 17-1). Note 
that the 1.9 cm (\3/4\-in.) free-space between the nozzle and pitot tube 
shown in Figure 17-1, is based on a 1.3 cm (\1/2\-in.) ID nozzle. If the 
sampling train is designed for sampling at higher flow rates than that 
described in APTD-0581, thus necessitating the use of larger sized 
nozzles, the free-space shall be 1.9 cm (\3/4\-in.) with the largest 
sized nozzle in place.
    6.1.4.2 Source-sampling assemblies that do not meet the minimum 
spacing requirements of Figure 17-1 (or the equivalent of these 
requirements, e.g., Figure 2-4 of Method 2) may be used; however, the 
pitot tube coefficients of such assemblies shall be determined by 
calibration, using methods subject to the approval of the Administrator.

[[Page 429]]

    6.1.5 Condenser. It is recommended that the impinger system or 
alternatives described in Method 5 be used to determine the moisture 
content of the stack gas. Flexible tubing may be used between the probe 
extension and condenser. Long tubing lengths may affect the moisture 
determination.
    6.2 Sample Recovery. Probe-liner and probe-nozzle brushes, wash 
bottles, glass sample storage containers, petri dishes, graduated 
cylinder and/or balance, plastic storage containers, funnel and rubber 
policeman, funnel. Same as in Method 5, Sections 6.2.1 through 6.2.8, 
respectively.
    6.3 Sample Analysis. Glass weighing dishes, desiccator, analytical 
balance, balance, beakers, hygrometer, temperature sensor. Same as in 
Method 5, Sections 6.3.1 through 6.3.7, respectively.

                       7.0 Reagents and Standards

    7.1 Sampling. Filters, silica gel, water, crushed ice, stopcock 
grease. Same as in Method 5, Sections 7.1.1, 7.1.2, 7.1.3, 7.1.4, and 
7.1.5, respectively. Thimble glass fiber filters may also be used.
    7.2 Sample Recovery. Acetone (reagent grade). Same as in Method 5, 
Section 7.2.
    7.3 Sample Analysis. Acetone and Desiccant. Same as in Method 5, 
Sections 7.3.1 and 7.3.2, respectively.

       8.0 Sample Collection, Preservation, Storage, and Transport

    8.1 Sampling.
    8.1.1 Pretest Preparation. Same as in Method 5, Section 8.1.1.
    8.1.2 Preliminary Determinations. Same as in Method 5, Section 
8.1.2, except as follows: Make a projected-area model of the probe 
extension-filter holder assembly, with the pitot tube face openings 
positioned along the centerline of the stack, as shown in Figure 17-2. 
Calculate the estimated cross-section blockage, as shown in Figure 17-2. 
If the blockage exceeds 5 percent of the duct cross sectional area, the 
tester has the following options exist: (1) a suitable out-of-stack 
filtration method may be used instead of in-stack filtration; or (2) a 
special in-stack arrangement, in which the sampling and velocity 
measurement sites are separate, may be used; for details concerning this 
approach, consult with the Administrator (see also Reference 1 in 
Section 17.0). Select a probe extension length such that all traverse 
points can be sampled. For large stacks, consider sampling from opposite 
sides of the stack to reduce the length of probes.
    8.1.3 Preparation of Sampling Train. Same as in Method 5, Section 
8.1.3, except the following: Using a tweezer or clean disposable 
surgical gloves, place a labeled (identified) and weighed filter in the 
filter holder. Be sure that the filter is properly centered and the 
gasket properly placed so as not to allow the sample gas stream to 
circumvent the filter. Check filter for tears after assembly is 
completed. Mark the probe extension with heat resistant tape or by some 
other method to denote the proper distance into the stack or duct for 
each sampling point. Assemble the train as in Figure 17-1, using a very 
light coat of silicone grease on all ground glass joints and greasing 
only the outer portion (see APTD-0576) to avoid possibility of 
contamination by the silicone grease. Place crushed ice around the 
impingers.
    8.1.4 Leak-Check Procedures. Same as in Method 5, Section 8.1.4, 
except that the filter holder is inserted into the stack during the 
sampling train leak-check. To do this, plug the inlet to the probe 
nozzle with a material that will be able to withstand the stack 
temperature. Insert the filter holder into the stack and wait 
approximately 5 minutes (or longer, if necessary) to allow the system to 
come to equilibrium with the temperature of the stack gas stream.
    8.1.5 Sampling Train Operation. The operation is the same as in 
Method 5. Use a data sheet such as the one shown in Figure 5-3 of Method 
5, except that the filter holder temperature is not recorded.
    8.1.6 Calculation of Percent Isokinetic. Same as in Method 5, 
Section 12.11.
    8.2 Sample Recovery.
    8.2.1 Proper cleanup procedure begins as soon as the probe extension 
assembly is removed from the stack at the end of the sampling period. 
Allow the assembly to cool.
    8.2.2 When the assembly can be safely handled, wipe off all external 
particulate matter near the tip of the probe nozzle and place a cap over 
it to prevent losing or gaining particulate matter. Do not cap off the 
probe tip tightly while the sampling train is cooling down as this would 
create a vacuum in the filter holder, forcing condenser water backward.
    8.2.3 Before moving the sample train to the cleanup site, disconnect 
the filter holder-probe nozzle assembly from the probe extension; cap 
the open inlet of the probe extension. Be careful not to lose any 
condensate, if present. Remove the umbilical cord from the condenser 
outlet and cap the outlet. If a flexible line is used between the first 
impinger (or condenser) and the probe extension, disconnect the line at 
the probe extension and let any condensed water or liquid drain into the 
impingers or condenser. Disconnect the probe extension from the 
condenser; cap the probe extension outlet. After wiping off the silicone 
grease, cap off the condenser inlet. Ground glass stoppers, plastic 
caps, or serum caps (whichever are appropriate) may be used to close 
these openings.
    8.2.4 Transfer both the filter holder-probe nozzle assembly and the 
condenser to the cleanup area. This area should be clean and protected 
from the wind so that the chances

[[Page 430]]

of contaminating or losing the sample will be minimized.
    8.2.5 Save a portion of the acetone used for cleanup as a blank. 
Take 200 ml of this acetone from the wash bottle being used and place it 
in a glass sample container labeled ``acetone blank.'' Inspect the train 
prior to and during disassembly and not any abnormal conditions. Treat 
the sample as discussed in Method 5, Section 8.2.

                     9.0 Quality Control [Reserved]

                  10.0 Calibration and Standardization

    The calibrations of the probe nozzle, pitot tube, metering system, 
temperature sensors, and barometer are the same as in Method 5, Sections 
10.1 through 10.3, 10.5, and 10.6, respectively.

                        11.0 Analytical Procedure

    Same as in Method 5, Section 11.0. Analytical data should be 
recorded on a form similar to that shown in Figure 5-6 of Method 5.

                  12.0 Data Analysis and Calculations.

    Same as in Method 5, Section 12.0.

                   13.0 Method Performance [Reserved]

                  14.0 Pollution Prevention [Reserved]

                    15.0 Waste Management [Reserved]

                       16.0 Alternative Procedures

    Same as in Method 5, Section 16.0.

                             17.0 References

    Same as in Method 5, Section 17.0, with the addition of the 
following:

    1. Vollaro, R.F. Recommended Procedure for Sample Traverses in Ducts 
Smaller than 12 Inches in Diameter. U.S. Environmental Protection 
Agency, Emission Measurement Branch. Research Triangle Park, NC. 
November 1976.

         18.0 Tables, Diagrams, Flowcharts, and Validation Data

[[Page 431]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.302


[[Page 432]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.303

  Method 18--Measurement of Gaseous Organic Compound Emissions By Gas 
                             Chromatography

    Note: This method is not inclusive with respect to specifications 
(e.g., equipment and supplies) and procedures (e.g., sampling and 
analytical) essential to its performance. Some material is incorporated 
by reference from other methods in this part. Therefore, to obtain 
reliable results, persons using this method should have a thorough 
knowledge of at least the following additional test methods: Method 1, 
Method 2, Method 3.
    Note: This method should not be attempted by persons unfamiliar with 
the performance characteristics of gas chromatography, nor by those 
persons who are unfamiliar with source sampling. Particular care

[[Page 433]]

should be exercised in the area of safety concerning choice of equipment 
and operation in potentially explosive atmospheres.

                        1.0 Scope and Application

    1.1 Analyte. Total gaseous organic compounds.
    1.2 Applicability.
    1.2.1 This method is designed to measure gaseous organics emitted 
from an industrial source. While designed for ppm level sources, some 
detectors are quite capable of detecting compounds at ambient levels, 
e.g., ECD, ELCD, and helium ionization detectors. Some other types of 
detectors are evolving such that the sensitivity and applicability may 
well be in the ppb range in only a few years.
    1.2.2 This method will not determine compounds that (1) are 
polymeric (high molecular weight), (2) can polymerize before analysis, 
or (3) have very low vapor pressures at stack or instrument conditions.
    1.3 Range. The lower range of this method is determined by the 
sampling system; adsorbents may be used to concentrate the sample, thus 
lowering the limit of detection below the 1 part per million (ppm) 
typically achievable with direct interface or bag sampling. The upper 
limit is governed by GC detector saturation or column overloading; the 
upper range can be extended by dilution of sample with an inert gas or 
by using smaller volume gas sampling loops. The upper limit can also be 
governed by condensation of higher boiling compounds.
    1.4 Sensitivity. The sensitivity limit for a compound is defined as 
the minimum detectable concentration of that compound, or the 
concentration that produces a signal-to-noise ratio of three to one. The 
minimum detectable concentration is determined during the presurvey 
calibration for each compound.

                          2.0 Summary of Method

    The major organic components of a gas mixture are separated by gas 
chromatography (GC) and individually quantified by flame ionization, 
photoionization, electron capture, or other appropriate detection 
principles. The retention times of each separated component are compared 
with those of known compounds under identical conditions. Therefore, the 
analyst confirms the identity and approximate concentrations of the 
organic emission components beforehand. With this information, the 
analyst then prepares or purchases commercially available standard 
mixtures to calibrate the GC under conditions identical to those of the 
samples. The analyst also determines the need for sample dilution to 
avoid detector saturation, gas stream filtration to eliminate 
particulate matter, and prevention of moisture condensation.

                       3.0 Definitions [Reserved]

                            4.0 Interferences

    4.1 Resolution interferences that may occur can be eliminated by 
appropriate GC column and detector choice or by shifting the retention 
times through changes in the column flow rate and the use of temperature 
programming.
    4.2 The analytical system is demonstrated to be essentially free 
from contaminants by periodically analyzing blanks that consist of 
hydrocarbon-free air or nitrogen.
    4.3 Sample cross-contamination that occurs when high-level and low-
level samples or standards are analyzed alternately is best dealt with 
by thorough purging of the GC sample loop between samples.
    4.4 To assure consistent detector response, calibration gases are 
contained in dry air. To adjust gaseous organic concentrations when 
water vapor is present in the sample, water vapor concentrations are 
determined for those samples, and a correction factor is applied.
    4.5 The gas chromatograph run time must be sufficient to clear all 
eluting peaks from the column before proceeding to the next run (in 
order to prevent sample carryover).

                               5.0 Safety

    5.1 Disclaimer. This method may involve hazardous materials, 
operations, and equipment. This test method may not address all of the 
safety problems associated with its use. It is the responsibility of the 
user of this test method to establish appropriate safety and health 
practices and determine the applicability of regulatory limitations 
prior to performing this test method. The analyzer users manual should 
be consulted for specific precautions to be taken with regard to the 
analytical procedure.

                       6.0 Equipment and Supplies

    6.1 Equipment needed for the presurvey sampling procedure can be 
found in Section 16.1.1.
    6.2 Equipment needed for the integrated bag sampling and analysis 
procedure can be found in Section 8.2.1.1.1.
    6.3 Equipment needed for direct interface sampling and analysis can 
be found in Section 8.2.2.1.
    6.4 Equipment needed for the dilution interface sampling and 
analysis can be found in Section 8.2.3.1.
    6.5 Equipment needed for adsorbent tube sampling and analysis can be 
found in Section 8.2.4.1.

                       7.0 Reagents and Standards

    7.1 Reagents needed for the presurvey sampling procedure can be 
found in Section 16.1.2.

[[Page 434]]

    7.2 Quality Assurance Audit Samples. When making compliance 
determinations, and upon availability, an audit sample may be obtained 
from the appropriate EPA Regional Office or from the responsible 
enforcement authority.

    Note: The responsible enforcement autority should be notified at 
least 30 days prior to the test date to allow sufficient time for sample 
delivery.

       8.0 Sample Collection, Preservation, Storage, and Transport

    8.2 Final Sampling and Analysis Procedure. Considering safety (flame 
hazards) and the source conditions, select an appropriate sampling and 
analysis procedure (Section 8.2.1, 8.2.2, 8.2.3 or 8.2.4). In situations 
where a hydrogen flame is a hazard and no intrinsically safe GC is 
suitable, use the flexible bag collection technique or an adsorption 
technique.
    8.2.1 Integrated Bag Sampling and Analysis.
    8.2.1.1 Evacuated Container Sampling Procedure. In this procedure, 
the bags are filled by evacuating the rigid air-tight container holding 
the bags. Use a field sample data sheet as shown in Figure 18-10. 
Collect triplicate samples from each sample location.
    8.2.1.1.1 Apparatus.
    8.2.1.1.1.1 Probe. Stainless steel, Pyrex glass, or Teflon tubing 
probe, according to the duct temperature, with Teflon tubing of 
sufficient length to connect to the sample bag. Use stainless steel or 
Teflon unions to connect probe and sample line.
    8.2.1.1.1.2 Quick Connects. Male (2) and female (2) of stainless 
steel construction.
    8.2.1.1.1.3 Needle Valve. To control gas flow.
    8.2.1.1.1.4 Pump. Leakless Teflon-coated diaphragm-type pump or 
equivalent. To deliver at least 1 liter/min.
    8.2.1.1.1.5 Charcoal Adsorption Tube. Tube filled with activated 
charcoal, with glass wool plugs at each end, to adsorb organic vapors.
    8.2.1.1.1.6 Flowmeter. 0 to 500-ml flow range; with manufacturer's 
calibration curve.
    8.2.1.1.2 Sampling Procedure. To obtain a sample, assemble the 
sample train as shown in Figure 18-9. Leak-check both the bag and the 
container. Connect the vacuum line from the needle valve to the Teflon 
sample line from the probe. Place the end of the probe at the centroid 
of the stack or at a point no closer to the walls than 1 m, and start 
the pump. Set the flow rate so that the final volume of the sample is 
approximately 80 percent of the bag capacity. After allowing sufficient 
time to purge the line several times, connect the vacuum line to the 
bag, and evacuate until the rotameter indicates no flow. Then position 
the sample and vacuum lines for sampling, and begin the actual sampling, 
keeping the rate proportional to the stack velocity. As a precaution, 
direct the gas exiting the rotameter away from sampling personnel. At 
the end of the sample period, shut off the pump, disconnect the sample 
line from the bag, and disconnect the vacuum line from the bag 
container. Record the source temperature, barometric pressure, ambient 
temperature, sampling flow rate, and initial and final sampling time on 
the data sheet shown in Figure 18-10. Protect the Tedlar bag and its 
container from sunlight. Record the time lapsed between sample 
collection and analysis, and then conduct the recovery procedure in 
Section 8.4.2.
    8.2.1.2 Direct Pump Sampling Procedure. Follow 8.2.1.1, except place 
the pump and needle valve between the probe and the bag. Use a pump and 
needle valve constructed of inert material not affected by the stack 
gas. Leak-check the system, and then purge with stack gas before 
connecting to the previously evacuated bag.
    8.2.1.3 Explosion Risk Area Bag Sampling Procedure. Follow 8.2.1.1 
except replace the pump with another evacuated can (see Figure 18-9a). 
Use this method whenever there is a possibility of an explosion due to 
pumps, heated probes, or other flame producing equipment.
    8.2.1.4 Other Modified Bag Sampling Procedures. In the event that 
condensation is observed in the bag while collecting the sample and a 
direct interface system cannot be used, heat the bag during collection, 
and maintain it at a suitably elevated temperature during all subsequent 
operations. (Note: Take care to leak-check the system prior to the 
dilutions so as not to create a potentially explosive atmosphere.) As an 
alternative, collect the sample gas, and simultaneously dilute it in the 
Tedlar bag.
    8.2.1.4.1 First Alternative Procedure. Heat the box containing the 
sample bag to 120 [deg]C (5 [deg]C). Then 
transport the bag as rapidly as possible to the analytical area while 
maintaining the heating, or cover the box with an insulating blanket. In 
the analytical area, keep the box heated to 120 [deg]C (5 [deg]C) until analysis. Be sure that the method of 
heating the box and the control for the heating circuit are compatible 
with the safety restrictions required in each area.
    8.2.1.4.2 Second Alternative Procedure. Prefill the Tedlar bag with 
a known quantity of inert gas. Meter the inert gas into the bag 
according to the procedure for the preparation of gas concentration 
standards of volatile liquid materials (Section 10.1.2.2), but eliminate 
the midget impinger section. Take the partly filled bag to the source, 
and meter the source gas into the bag through heated sampling lines and 
a heated flowmeter, or Teflon positive displacement pump. Verify the 
dilution factors before sampling each bag

[[Page 435]]

through dilution and analysis of gases of known concentration.
    8.2.1.5 Analysis of Bag Samples.
    8.2.1.5.1 Apparatus. Same as Section 8.1. A minimum of three gas 
standards are required.
    8.2.1.5.2 Procedure.
    8.2.1.5.2.1 Establish proper GC operating conditions as described in 
Section 10.2, and record all data listed in Figure 18-7. Prepare the GC 
so that gas can be drawn through the sample valve. Flush the sample loop 
with calibration gas mixture, and activate the valve (sample pressure at 
the inlet to the GC introduction valve should be similar during 
calibration as during actual sample analysis). Obtain at least three 
chromatograms for the mixture. The results are acceptable when the peak 
areas for the three injections agree to within 5 percent of their 
average. If they do not agree, run additional samples or correct the 
analytical techniques until this requirement is met. Then analyze the 
other two calibration mixtures in the same manner. Prepare a calibration 
curve as described in Section 10.2.
    8.2.1.5.2.2 Analyze the two field audit samples as described in 
Section 9.2 by connecting each Tedlar bag containing an audit gas 
mixture to the sampling valve. Calculate the results; record and report 
the data to the audit supervisor.
    8.2.1.5.2.3 Analyze the three source gas samples by connecting each 
bag to the sampling valve with a piece of Teflon tubing identified with 
that bag. Analyze each bag sample three times. Record the data in Figure 
18-11. If certain items do not apply, use the notation ``N.A.'' If the 
bag has been maintained at an elevated temperature as described in 
Section 8.2.1.4, determine the stack gas water content by Method 4. 
After all samples have been analyzed, repeat the analysis of the mid-
level calibration gas for each compound. Compare the average response 
factor of the pre- and post-test analysis for each compound. If they 
differ by 5percent, analyze the other calibration gas levels 
for that compound, and prepare a calibration curve using all the pre- 
and post-test calibration gas mixture values. If the two response factor 
averages (pre-and post-test) differ by less than 5 percent from their 
mean value, the tester has the option of using only the pre-test 
calibration curve to generate the concentration values.
    8.2.1.6 Determination of Bag Water Vapor Content. Measure the 
ambient temperature and barometric pressure near the bag. From a water 
saturation vapor pressure table, determine and record the water vapor 
content of the bag as a decimal figure. (Assume the relative humidity to 
be 100 percent unless a lesser value is known.) If the bag has been 
maintained at an elevated temperature as described in Section 8.2.1.4, 
determine the stack gas water content by Method 4.
    8.2.1.7 Audit Gas Analysis. Immediately prior to the analysis of the 
stack gas samples, perform audit analyses as described in Section 9.2.
    8.2.1.8 Emission Calculations. From the calibration curve described 
in Section 8.2.1.5, select the value of Cs that corresponds 
to the peak area. Calculate the concentration Cc in ppm, dry 
basis, of each organic in the sample using Equation 18-5 in Section 
12.6.
    8.2.2 Direct Interface Sampling and Analysis Procedure. The direct 
interface procedure can be used provided that the moisture content of 
the gas does not interfere with the analysis procedure, the physical 
requirements of the equipment can be met at the site, and the source gas 
concentration falls within the linear range of the detector. Adhere to 
all safety requirements with this method.
    8.2.2.1 Apparatus.
    8.2.2.1.1 Probe. Constructed of stainless steel, Pyrex glass, or 
Teflon tubing as dictated by duct temperature and reactivity of target 
compounds. A filter or glass wool plug may be needed if particulate is 
present in the stack gas. If necessary, heat the probe with heating tape 
or a special heating unit capable of maintaining a temperature greater 
than 110 [deg]C.
    8.2.2.1.2 Sample Lines. 6.4-mm OD (or other diameter as needed) 
Teflon lines, heat-traced to prevent condensation of material (greater 
than 110 [deg]C).
    8.2.2.1.3 Quick Connects. To connect sample line to gas sampling 
valve on GC instrument and to pump unit used to withdraw source gas. Use 
a quick connect or equivalent on the cylinder or bag containing 
calibration gas to allow connection of the calibration gas to the gas 
sampling valve.
    8.2.2.1.4 Thermocouple Readout Device. Potentiometer or digital 
thermometer, to measure source temperature and probe temperature.
    8.2.2.1.5 Heated Gas Sampling Valve. Of two-position, six-port 
design, to allow sample loop to be purged with source gas or to direct 
source gas into the GC instrument.
    8.2.2.1.6 Needle Valve. To control gas sampling rate from the 
source.
    8.2.2.1.7 Pump. Leakless Teflon-coated diaphragm-type pump or 
equivalent, capable of at least 1 liter/minute sampling rate.
    8.2.2.1.8 Flowmeter. Of suitable range to measure sampling rate.
    8.2.2.1.9 Charcoal Adsorber. To adsorb organic vapor vented from the 
source to prevent exposure of personnel to source gas.
    8.2.2.1.10 Gas Cylinders. Carrier gas, oxygen and fuel as needed to 
run GC and detector.
    8.2.2.1.11 Gas Chromatograph. Capable of being moved into the field, 
with detector, heated gas sampling valve, column required

[[Page 436]]

to complete separation of desired components, and option for temperature 
programming.
    8.2.2.1.12 Recorder/Integrator. To record results.
    8.2.2.2 Procedure. Calibrate the GC using the procedures in Section 
8.2.1.5.2.1. To obtain a stack gas sample, assemble the sampling system 
as shown in Figure 18-12. Make sure all connections are tight. Turn on 
the probe and sample line heaters. As the temperature of the probe and 
heated line approaches the target temperature as indicated on the 
thermocouple readout device, control the heating to maintain a 
temperature greater than 110 [deg]C. Conduct a 3-point calibration of 
the GC by analyzing each gas mixture in triplicate. Generate a 
calibration curve. Place the inlet of the probe at the centroid of the 
duct, or at a point no closer to the walls than 1 m, and draw source gas 
into the probe, heated line, and sample loop. After thorough flushing, 
analyze the stack gas sample using the same conditions as for the 
calibration gas mixture. For each run, sample, analyze, and record five 
consecutive samples. A test consists of three runs (five samples per run 
times three runs, for a total of fifteen samples). After all samples 
have been analyzed, repeat the analysis of the mid-level calibration gas 
for each compound. For each calibration standard, compare the pre- and 
post-test average response factors (RF) for each compound. If the two 
calibration RF values (pre- and post-analysis) differ by more than 5 
percent from their mean value, then analyze the other calibration gas 
levels for that compound and determine the stack gas sample 
concentrations by comparison to both calibration curves (this is done by 
preparing a calibration curve using all the pre and post-test 
calibration gas mixture values). If the two calibration RF values differ 
by less than 5 percent from their mean value, the tester has the option 
of using only the pre-test calibration curve to generate the 
concentration values. Record this calibration data and the other 
required data on the data sheet shown in Figure 18-11, deleting the 
dilution gas information.

    Note: Take care to draw all samples, calibration mixtures, and 
audits through the sample loop at the same pressure.

    8.2.2.3 Determination of Stack Gas Moisture Content. Use Method 4 to 
measure the stack gas moisture content.
    8.2.2.4 Quality Assurance. Same as Section 8.2.1.7. Introduce the 
audit gases in the sample line immediately following the probe.
    8.2.2.5 Emission Calculations. Same as Section 8.2.1.8.
    8.2.3 Dilution Interface Sampling and Analysis Procedure. Source 
samples that contain a high concentration of organic materials may 
require dilution prior to analysis to prevent saturating the GC 
detector. The apparatus required for this direct interface procedure is 
basically the same as that described in the Section 8.2.2, except a 
dilution system is added between the heated sample line and the gas 
sampling valve. The apparatus is arranged so that either a 10:1 or 100:1 
dilution of the source gas can be directed to the chromatograph. A pump 
of larger capacity is also required, and this pump must be heated and 
placed in the system between the sample line and the dilution apparatus.
    8.2.3.1 Apparatus. The equipment required in addition to that 
specified for the direct interface system is as follows:
    8.2.3.1.1 Sample Pump. Leakless Teflon-coated diaphragm-type that 
can withstand being heated to 120 [deg]C and deliver 1.5 liters/minute.
    8.2.3.1.2 Dilution Pumps. Two Model A-150 Komhyr Teflon positive 
displacement type delivering 150 cc/minute, or equivalent. As an option, 
calibrated flowmeters can be used in conjunction with Teflon-coated 
diaphragm pumps.
    8.2.3.1.3 Valves. Two Teflon three-way valves, suitable for 
connecting to Teflon tubing.
    8.2.3.1.4 Flowmeters. Two, for measurement of diluent gas.
    8.2.3.1.5 Diluent Gas with Cylinders and Regulators. Gas can be 
nitrogen or clean dry air, depending on the nature of the source gases.
    8.2.3.1.6 Heated Box. Suitable for being heated to 120 [deg]C, to 
contain the three pumps, three-way valves, and associated connections. 
The box should be equipped with quick connect fittings to facilitate 
connection of: (1) the heated sample line from the probe, (2) the gas 
sampling valve, (3) the calibration gas mixtures, and (4) diluent gas 
lines. A schematic diagram of the components and connections is shown in 
Figure 18-13. The heated box shown in Figure 18-13 is designed to 
receive a heated line from the probe. An optional design is to build a 
probe unit that attaches directly to the heated box. In this way, the 
heated box contains the controls for the probe heaters, or, if the box 
is placed against the duct being sampled, it may be possible to 
eliminate the probe heaters. In either case, a heated Teflon line is 
used to connect the heated box to the gas sampling valve on the 
chromatograph.

    Note: Care must be taken to leak-check the system prior to the 
dilutions so as not to create a potentially explosive atmosphere.

    8.2.3.2 Procedure.
    8.2.3.2.1 Assemble the apparatus by connecting the heated box, shown 
in Figure 18-13, between the heated sample line from the probe and the 
gas sampling valve on the chromatograph. Vent the source gas from the 
gas sampling valve directly to the charcoal filter, eliminating the pump 
and rotameter. Heat the sample probe, sample line, and

[[Page 437]]

heated box. Insert the probe and source thermocouple at the centroid of 
the duct, or to a point no closer to the walls than 1 m. Measure the 
source temperature, and adjust all heating units to a temperature 0 to 3 
[deg]C above this temperature. If this temperature is above the safe 
operating temperature of the Teflon components, adjust the heating to 
maintain a temperature high enough to prevent condensation of water and 
organic compounds (greater than 110 [deg]C). Calibrate the GC through 
the dilution system by following the procedures in Section 8.2.1.5.2.1. 
Determine the concentration of the diluted calibration gas using the 
dilution factor and the certified concentration of the calibration gas. 
Record the pertinent data on the data sheet shown in Figure 18-11.
    8.2.3.2.2 Once the dilution system and GC operations are 
satisfactory, proceed with the analysis of source gas, maintaining the 
same dilution settings as used for the standards.
    8.2.3.2.3 Analyze the audit samples using either the dilution 
system, or directly connect to the gas sampling valve as required. 
Record all data and report the results to the audit supervisor.
    8.2.3.3 Determination of Stack Gas Moisture Content. Same as Section 
8.2.2.3.
    8.2.3.4 Quality Assurance. Same as Section 8.2.2.4.
    8.2.3.5 Emission Calculations. Same as section 8.2.2.5, with the 
dilution factor applied.
    8.2.4 Adsorption Tube Procedure. Any commercially available 
adsorbent is allowed for the purposes of this method, as long as the 
recovery study criteria in Section 8.4.3 are met. Help in choosing the 
adsorbent may be found by calling the distributor, or the tester may 
refer to National Institute for Occupational Safety and Health (NIOSH) 
methods for the particular organics to be sampled. For some adsorbents, 
the principal interferent will be water vapor. If water vapor is thought 
to be a problem, the tester may place a midget impinger in an ice bath 
before the adsorbent tubes. If this option is chosen, the water catch in 
the midget impinger shall be analyzed for the target compounds. Also, 
the spike for the recovery study (in Section 8.4.3) shall be conducted 
in both the midget impinger and the adsorbent tubes. The combined 
recovery (add the recovered amount in the impinger and the adsorbent 
tubes to calculate R) shall then meet the criteria in Section 8.4.3.

    Note: Post-test leak-checks are not allowed for this technique since 
this can result in sample contamination.

    8.2.4.1 Additional Apparatus. The following items (or equivalent) 
are suggested.
    8.2.4.1.1 Probe. Borosilicate glass or stainless steel, 
approximately 6-mm ID, with a heating system if water condensation is a 
problem, and a filter (either in-stack or out-of-stack, heated to stack 
temperature) to remove particulate matter. In most instances, a plug of 
glass wool is a satisfactory filter.
    8.2.4.1.2 Flexible Tubing. To connect probe to adsorption tubes. Use 
a material that exhibits minimal sample adsorption.
    8.2.4.1.3 Leakless Sample Pump. Flow controlled, constant rate pump, 
with a set of limiting (sonic) orifices.
    8.2.4.1.4 Bubble-Tube Flowmeter. Volume accuracy within 1 percent, 
to calibrate pump.
    8.2.4.1.5 Stopwatch. To time sampling and pump rate calibration.
    8.2.4.1.6 Adsorption Tubes. Precleaned adsorbent, with mass of 
adsorbent to be determined by calculating breakthrough volume and 
expected concentration in the stack.
    8.2.4.1.7 Barometer. Accurate to 5 mm Hg, to measure atmospheric 
pressure during sampling and pump calibration.
    8.2.4.1.8 Rotameter. O to 100 cc/min, to detect changes in flow rate 
during sampling.
    8.2.4.2 Sampling and Analysis.
    8.2.4.2.1 Calibrate the pump and limiting orifice flow rate through 
adsorption tubes with the bubble tube flowmeter before sampling. The 
sample system can be operated as a ``recirculating loop'' for this 
operation. Record the ambient temperature and barometric pressure. Then, 
during sampling, use the rotameter to verify that the pump and orifice 
sampling rate remains constant.
    8.2.4.2.2 Use a sample probe, if required, to obtain the sample at 
the centroid of the duct, or at a point no closer to the walls than 1 m. 
Minimize the length of flexible tubing between the probe and adsorption 
tubes. Several adsorption tubes can be connected in series, if the extra 
adsorptive capacity is needed. Adsorption tubes should be maintained 
vertically during the test in order to prevent channeling. Provide the 
gas sample to the sample system at a pressure sufficient for the 
limiting orifice to function as a sonic orifice. Record the total time 
and sample flow rate (or the number of pump strokes), the barometric 
pressure, and ambient temperature. Obtain a total sample volume 
commensurate with the expected concentration(s) of the volatile 
organic(s) present, and recommended sample loading factors (weight 
sample per weight adsorption media). Laboratory tests prior to actual 
sampling may be necessary to predetermine this volume. If water vapor is 
present in the sample at concentrations above 2 to 3 percent, the 
adsorptive capacity may be severely reduced. Operate the gas 
chromatograph according to the manufacturer's instructions. After 
establishing optimum conditions, verify and document these conditions 
during all operations. Calibrate the instrument. Analyze the audit 
samples (see Section 16.1.4.3), then the emission samples.
    8.2.4.3 Standards and Calibration. If using thermal desorption, 
obtain calibration gases

[[Page 438]]

using the procedures in Section 10.1. If using solvent extraction, 
prepare liquid standards in the desorption solvent. Use a minimum of 
three different standards; select the concentrations to bracket the 
expected average sample concentration. Perform the calibration before 
and after each day's sample analyses using the procedures in Section 
8.2.1.5.2.1.
    8.2.4.4 Quality Assurance.
    8.2.4.4.1 Determine the recovery efficiency of the pollutants of 
interest according to Section 8.4.3.
    8.2.4.4.2 Determination of Sample Collection Efficiency (Optional). 
If sample breakthrough is thought to be a problem, a routine procedure 
for determining breakthrough is to analyze the primary and backup 
portions of the adsorption tubes separately. If the backup portion 
exceeds 10 percent of the total amount (primary and back-up), it is 
usually a sign of sample breakthrough. For the purposes of this method, 
only the recovery efficiency value (Section 8.4.3) is used to determine 
the appropriateness of the sampling and analytical procedure.
    8.2.4.4.3 Volume Flow Rate Checks. Perform this check immediately 
after sampling with all sampling train components in place. Use the 
bubble-tube flowmeter to measure the pump volume flow rate with the 
orifice used in the test sampling, and record the result. If it has 
changed by more than 5 but less than 20 percent, calculate an average 
flow rate for the test. If the flow rate has changed by more than 20 
percent, recalibrate the pump and repeat the sampling.
    8.2.4.4.4 Calculations. Correct all sample volumes to standard 
conditions. If a sample dilution system has been used, multiply the 
results by the appropriate dilution ratio. Correct all results according 
to the applicable procedure in Section 8.4.3. Report results as ppm by 
volume, dry basis.
    8.3 Reporting of Results. At the completion of the field analysis 
portion of the study, ensure that the data sheets shown in Figure 18-11 
have been completed. Summarize this data on the data sheets shown in 
Figure 18-15.
    8.4 Recovery Study. After conducting the presurvey and identifying 
all of the pollutants of interest, conduct the appropriate recovery 
study during the test based on the sampling system chosen for the 
compounds of interest.
    8.4.1 Recovery Study for Direct Interface or Dilution Interface 
Sampling. If the procedures in Section 8.2.2 or 8.2.3 are to be used to 
analyze the stack gas, conduct the calibration procedure as stated in 
Section 8.2.2.2 or 8.2.3.2, as appropriate. Upon successful completion 
of the appropriate calibration procedure, attach the mid-level 
calibration gas for at least one target compound to the inlet of the 
probe or as close as possible to the inlet of the probe, but before the 
filter. Repeat the calibration procedure by sampling and analyzing the 
mid-level calibration gas through the entire sampling and analytical 
system in triplicate. The mean of the calibration gas response sampled 
through the probe shall be within 10 percent of the analyzer response. 
If the difference in the two means is greater than 10 percent, check for 
leaks throughout the sampling system and repeat the analysis of the 
standard through the sampling system until this criterion is met.
    8.4.2 Recovery Study for Bag Sampling.
    8.4.2.1 Follow the procedures for the bag sampling and analysis in 
Section 8.2.1. After analyzing all three bag samples, choose one of the 
bag samples and tag this bag as the spiked bag. Spike the chosen bag 
sample with a known mixture (gaseous or liquid) of all of the target 
pollutants. The theoretical concentration, in ppm, of each spiked 
compound in the bag shall be 40 to 60 percent of the average 
concentration measured in the three bag samples. If a target compound 
was not detected in the bag samples, the concentration of that compound 
to be spiked shall be 5 times the limit of detection for that compound. 
Store the spiked bag for the same period of time as the bag samples 
collected in the field. After the appropriate storage time has passed, 
analyze the spiked bag three times. Calculate the average fraction 
recovered (R) of each spiked target compound with the equation in 
Section 12.7.
    8.4.2.2 For the bag sampling technique to be considered valid for a 
compound, 0.70 <= R <= 1.30. If the R value does not meet this criterion 
for a target compound, the sampling technique is not acceptable for that 
compound, and therefore another sampling technique shall be evaluated 
for acceptance (by repeating the recovery study with another sampling 
technique). Report the R value in the test report and correct all field 
measurements with the calculated R value for that compound by using the 
equation in Section 12.8.
    8.4.3 Recovery Study for Adsorption Tube Sampling. If following the 
adsorption tube procedure in Section 8.2.4, conduct a recovery study of 
the compounds of interest during the actual field test. Set up two 
identical sampling trains. Collocate the two sampling probes in the 
stack. The probes shall be placed in the same horizontal plane, where 
the first probe tip is 2.5 cm from the outside edge of the other. One of 
the sampling trains shall be designated the spiked train and the other 
the unspiked train. Spike all of the compounds of interest (in gaseous 
or liquid form) onto the adsorbent tube(s) in the spiked train before 
sampling. The mass of each spiked compound shall be 40 to 60 percent of 
the mass expected to be collected with the unspiked train. Sample the 
stack

[[Page 439]]

gas into the two trains simultaneously. Analyze the adsorbents from the 
two trains utilizing identical analytical procedures and 
instrumentation. Determine the fraction of spiked compound recovered (R) 
using the equations in Section 12.9.
    8.4.3.1 Repeat the procedure in Section 8.4.3 twice more, for a 
total of three runs. In order for the adsorbent tube sampling and 
analytical procedure to be acceptable for a compound, 0.70<=R<=1.30 (R 
in this case is the average of three runs). If the average R value does 
not meet this criterion for a target compound, the sampling technique is 
not acceptable for that compound, and therefore another sampling 
technique shall be evaluated for acceptance (by repeating the recovery 
study with another sampling technique). Report the R value in the test 
report and correct all field measurements with the calculated R value 
for that compound by using the equation in Section 12.8.

                           9.0 Quality Control

    9.1 Miscellaneous Quality Control Measures

------------------------------------------------------------------------
                                 Quality control
            Section                  measure               Effect
------------------------------------------------------------------------
8.4.1.........................  Recovery study     Ensure that there are
                                 for direct         no significant leaks
                                 interface or       in the sampling
                                 dilution           system.
                                 interface
                                 sampling.
8.4.2.........................  Recovery study     Demonstrate that
                                 for bag sampling.  proper sampling/
                                                    analysis procedures
                                                    were selected.
8.4.3.........................  Recovery study     Demonstrate that
                                 for adsorption     proper sampling/
                                 tube sampling.     analysis procedures
                                                    were selected.
------------------------------------------------------------------------

    9.2 Quality Assurance for Laboratory Procedures. Immediately after 
the preparation of the calibration curves, the analysis audit described 
in 40 CFR Part 61, Appendix C, Procedure 2: ``Procedure for Field 
Auditing GC Analysis,'' should be performed if audit materials are 
available. The information required to document the analysis of the 
audit samples has been included on the example data sheets shown in 
Figures 18-3 and 18-7. The audit analyses should agree with the 
certified audit concentrations within 10 percent. Audit sample results 
shall be submitted according to directions provided with the audit 
samples.

                  10.0 Calibration and Standardization.

    10.1 Calibration Standards. Obtain calibration gas standards for 
each target compound to be analyzed. Commercial cylinder gases certified 
by the manufacturer to be accurate to within 1 percent of the certified 
label value are preferable, although cylinder gases certified by the 
manufacturer to 2 percent accuracy are allowed. Another option allowed 
by this method is for the tester to obtain high concentration certified 
cylinder gases and then use a dilution system meeting the requirements 
of Test Method 205, 40 CFR Part 51, Appendix M to make multi-level 
calibration gas standards. Prepare or obtain enough calibration 
standards so that there are three different concentrations of each 
organic compound expected to be measured in the source sample. For each 
organic compound, select those concentrations that bracket the 
concentrations expected in the source samples. A calibration standard 
may contain more than one organic compound. If samples are collected in 
adsorbent tubes and extracted using solvent extraction, prepare or 
obtain standards in the same solvent used for the sample extraction 
procedure. Verify the stability of all standards for the time periods 
they are used.
    10.2 Preparation of Calibration Curves.
    10.2.1 Establish proper GC conditions, then flush the sampling loop 
for 30 seconds. Allow the sample loop pressure to equilibrate to 
atmospheric pressure, and activate the injection valve. Record the 
standard concentration, attenuator factor, injection time, chart speed, 
retention time, peak area, sample loop temperature, column temperature, 
and carrier gas flow rate. Analyze each standard in triplicate.
    10.2.2 Repeat this procedure for each standard. Prepare a graphical 
plot of concentration (Cs) versus the calibration area 
values. Perform a regression analysis, and draw the least square line.

                       11.0 Analytical Procedures

    11.1 Analysis Development
    11.1.1 Selection of GC Parameters
    11.1.1.1 Column Choice. Based on the initial contact with plant 
personnel concerning the plant process and the anticipated emissions, 
choose a column that provides good resolution and rapid analysis time. 
The choice of an appropriate column can be aided by a literature search, 
contact with manufacturers of GC columns, and discussion with personnel 
at the emission source.

    Note: Most column manufacturers keep excellent records on their 
products. Their technical service departments may be able to recommend 
appropriate columns and detector type for separating the anticipated 
compounds, and they may be able to provide information on interferences, 
optimum operating conditions, and column limitations. Plants with 
analytical laboratories may be

[[Page 440]]

able to provide information on their analytical procedures.

    11.1.1.2 Preliminary GC Adjustment. Using the standards and column 
obtained in Section 11.1.1.1, perform initial tests to determine 
appropriate GC conditions that provide good resolution and minimum 
analysis time for the compounds of interest.
    11.1.1.3 Preparation of Presurvey Samples. If the samples were 
collected on an adsorbent, extract the sample as recommended by the 
manufacturer for removal of the compounds with a solvent suitable to the 
type of GC analysis. Prepare other samples in an appropriate manner.
    11.1.1.4 Presurvey Sample Analysis.
    11.1.1.4.1 Before analysis, heat the presurvey sample to the duct 
temperature to vaporize any condensed material. Analyze the samples by 
the GC procedure, and compare the retention times against those of the 
calibration samples that contain the components expected to be in the 
stream. If any compounds cannot be identified with certainty by this 
procedure, identify them by other means such as GC/mass spectroscopy 
(GC/MS) or GC/infrared techniques. A GC/MS system is recommended.
    11.1.1.4.2 Use the GC conditions determined by the procedure of 
Section 11.1.1.2 for the first injection. Vary the GC parameters during 
subsequent injections to determine the optimum settings. Once the 
optimum settings have been determined, perform repeat injections of the 
sample to determine the retention time of each compound. To inject a 
sample, draw sample through the loop at a constant rate (100 ml/min for 
30 seconds). Be careful not to pressurize the gas in the loop. Turn off 
the pump and allow the gas in the sample loop to come to ambient 
pressure. Activate the sample valve, and record injection time, loop 
temperature, column temperature, carrier flow rate, chart speed, and 
attenuator setting. Calculate the retention time of each peak using the 
distance from injection to the peak maximum divided by the chart speed. 
Retention times should be repeatable within 0.5 seconds.
    11.1.1.4.3 If the concentrations are too high for appropriate 
detector response, a smaller sample loop or dilutions may be used for 
gas samples, and, for liquid samples, dilution with solvent is 
appropriate. Use the standard curves (Section 10.2) to obtain an 
estimate of the concentrations.
    11.1.1.4.4 Identify all peaks by comparing the known retention times 
of compounds expected to be in the retention times of peaks in the 
sample. Identify any remaining unidentified peaks which have areas 
larger than 5 percent of the total using a GC/MS, or estimation of 
possible compounds by their retention times compared to known compounds, 
with confirmation by further GC analysis.

                   12.0 Data Analysis and Calculations

    12.1 Nomenclature.

Bws=Water vapor content of the bag sample or stack gas, 
proportion by volume.
Cs=Concentration of the organic from the calibration curve, 
ppm.
Gv=Gas volume or organic compound injected, ml.
Lv=Liquid volume of organic injected, [micro]l.
M=Molecular weight of organic, g/g-mole.
ms=Total mass of compound measured on adsorbent with spiked 
train ([micro]g).
mu=Total mass of compound measured on adsorbent with unspiked 
train ([micro]g).
mv=Mass per volume of spiked compound measured ([micro]g/L).
Pi=Barometric or absolute sample loop pressure at time of 
sample analysis, mm Hg.
Pm=Absolute pressure of dry gas meter, mm Hg.
Pr=Reference pressure, the barometric pressure or absolute 
sample loop pressure recorded during calibration, mm Hg.
Ps=Absolute pressure of syringe before injection, mm Hg.
qc=Flow rate of the calibration gas to be diluted.
qc1=Flow rate of the calibration gas to be diluted in stage 
1.
qc2=Flow rate of the calibration gas to be diluted in stage 
2.
qd=Diluent gas flow rate.
qd1=Flow rate of diluent gas in stage 1.
qd2=Flow rate of diluent gas in stage 2.
s=Theoretical concentration (ppm) of spiked target compound in the bag.
S=Theoretical mass of compound spiked onto adsorbent in spiked train 
([micro]g).
t=Measured average concentration (ppm) of target compound and source 
sample (analysis results subsequent to bag spiking)
Ti=Sample loop temperature at the time of sample analysis, 
[deg]K.
Tm=Absolute temperature of dry gas meter, [deg]K.
Ts=Absolute temperature of syringe before injection, [deg]K.
u=Source sample average concentration (ppm) of target compound in the 
bag (analysis results before bag spiking).
Vm=Gas volume indicated by dry gas meter, liters.
vs=volume of stack gas sampled with spiked train (L).
vu=volume of stack gas sampled with unspiked train (L).
X=Mole or volume fraction of the organic in the calibration gas to be 
diluted.
Y=Dry gas meter calibration factor, dimensionless.
[micro]l=Liquid organic density as determined, g/ml.

24.055=Ideal gas molar volume at 293 [deg]K and 760 mm Hg, liters/g-
mole.

1000=Conversion factor, ml/liter.
10\6\=Conversion to ppm.


[[Page 441]]


    12.2 Calculate the concentration, Cs, in ppm using the 
following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.304

    12.3 Calculate the concentration, Cs, in ppm of the 
organic in the final gas mixture using the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.305

    12.4 Calculate each organic standard concentration, Cs, 
in ppm using the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.306

    12.5 Calculate each organic standard concentration, Cs, 
in ppm using the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.307

    12.6 Calculate the concentration, Cc, in ppm, dry basis, 
of each organic is the sample using the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.308

    12.7 Calculate the average fraction recovered (R) of each spiked 
target compound using the following equation:

[[Page 442]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.309

    12.8 Correct all field measurements with the calculated R value for 
that compound using the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.310

    12.9 Determine the mass per volume of spiked compound measured using 
the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.311

    12.10 Calculate the fraction of spiked compound recovered, R, using 
the following equation:
[GRAPHIC] [TIFF OMITTED] TR17OC00.312

                         13.0 Method Performance

    13.1 Since a potential sample may contain a variety of compounds 
from various sources, a specific precision limit for the analysis of 
field samples is impractical. Precision in the range of 5 to 10 percent 
relative standard deviation (RSD) is typical for gas chromatographic 
techniques, but an experienced GC operator with a reliable instrument 
can readily achieve 5 percent RSD. For this method, the following 
combined GC/operator values are required.
    (a) Precision. Triplicate analyses of calibration standards fall 
within 5 percent of their mean value.
    (b) Accuracy. Analysis results of prepared audit samples are within 
10 percent of preparation values.
    (c) Recovery. After developing an appropriate sampling and 
analytical system for the pollutants of interest, conduct the procedure 
in Section 8.4. Conduct the appropriate recovery study in Section 8.4 at 
each sampling point where the method is being applied. Submit the data 
and results of the recovery procedure with the reporting of results 
under Section 8.3.

                  14.0 Pollution Prevention [Reserved]

                    15.0 Waste Management [Reserved]

                       16.0 Alternative Procedures

    16.1 Optional Presurvey and Presurvey Sampling.

    Note: Presurvey screening is optional. Presurvey sampling should be 
conducted for sources where the target pollutants are not known from 
previous tests and/or process knowledge.

    Perform a presurvey for each source to be tested. Refer to Figure 
18-1. Some of the information can be collected from literature surveys 
and source personnel. Collect gas samples that can be analyzed to 
confirm the identities and approximate concentrations of the organic 
emissions.
    16.1.1 Apparatus. This apparatus list also applies to Sections 8.2 
and 11.
    16.1.1.1 Teflon Tubing. (Mention of trade names or specific products 
does not constitute endorsement by the U.S. Environmental Protection 
Agency.) Diameter and length determined by connection requirements of 
cylinder regulators and the GC. Additional tubing is necessary to 
connect the GC sample loop to the sample.
    16.1.1.2 Gas Chromatograph. GC with suitable detector, columns, 
temperature-controlled sample loop and valve assembly, and temperature 
programmable oven, if necessary. The GC shall achieve sensitivity 
requirements for the compounds under study.
    16.1.1.3 Pump. Capable of pumping 100 ml/min. For flushing sample 
loop.
    16.1.1.4 Flow Meter. To measure flow rates.
    16.1.1.5 Regulators. Used on gas cylinders for GC and for cylinder 
standards.
    16.1.1.6 Recorder. Recorder with linear strip chart is minimum 
acceptable. Integrator (optional) is recommended.
    16.1.1.7 Syringes. 0.5-ml, 1.0- and 10-microliter size, calibrated, 
maximum accuracy (gas tight) for preparing calibration standards. Other 
appropriate sizes can be used.

[[Page 443]]

    16.1.1.8 Tubing Fittings. To plumb GC and gas cylinders.
    16.1.1.9 Septa. For syringe injections.
    16.1.1.10 Glass Jars. If necessary, clean, colored glass jars with 
Teflon-lined lids for condensate sample collection. Size depends on 
volume of condensate.
    16.1.1.11 Soap Film Flowmeter. To determine flow rates.
    16.1.1.12 Tedlar Bags. 10- and 50-liter capacity, for preparation of 
standards.
    16.1.1.13 Dry Gas Meter with Temperature and Pressure Gauges. 
Accurate to 2 percent, for preparation of gas 
standards.
    16.1.1.14 Midget Impinger/Hot Plate Assembly. For preparation of gas 
standards.
    16.1.1.15 Sample Flasks. For presurvey samples, must have gas-tight 
seals.
    16.1.1.16 Adsorption Tubes. If necessary, blank tubes filled with 
necessary adsorbent (charcoal, Tenax, XAD-2, etc.) for presurvey 
samples.
    16.1.1.17 Personnel Sampling Pump. Calibrated, for collecting 
adsorbent tube presurvey samples.
    16.1.1.18 Dilution System. Calibrated, the dilution system is to be 
constructed following the specifications of an acceptable method.
    16.1.1.19 Sample Probes. Pyrex or stainless steel, of sufficient 
length to reach centroid of stack, or a point no closer to the walls 
than 1 m.
    16.1.1.20 Barometer. To measure barometric pressure.
    16.1.2 Reagents.
    16.1.2.1 Water. Deionized distilled.
    16.1.2.2 Methylene chloride.
    16.1.2.3 Calibration Gases. A series of standards prepared for every 
compound of interest.
    16.1.2.4 Organic Compound Solutions. Pure (99.9 percent), or as pure 
as can reasonably be obtained, liquid samples of all the organic 
compounds needed to prepare calibration standards.
    16.1.2.5 Extraction Solvents. For extraction of adsorbent tube 
samples in preparation for analysis.
    16.1.2.6 Fuel. As recommended by the manufacturer for operation of 
the GC.
    16.1.2.7 Carrier Gas. Hydrocarbon free, as recommended by the 
manufacturer for operation of the detector and compatibility with the 
column.
    16.1.2.8 Zero Gas. Hydrocarbon free air or nitrogen, to be used for 
dilutions, blank preparation, and standard preparation.
    16.1.3 Sampling.
    16.1.3.1 Collection of Samples with Glass Sampling Flasks. Presurvey 
samples may be collected in precleaned 250-ml double-ended glass 
sampling flasks. Teflon stopcocks, without grease, are preferred. Flasks 
should be cleaned as follows: Remove the stopcocks from both ends of the 
flasks, and wipe the parts to remove any grease. Clean the stopcocks, 
barrels, and receivers with methylene chloride (or other non-target 
pollutant solvent, or heat and humidified air). Clean all glass ports 
with a soap solution, then rinse with tap and deionized distilled water. 
Place the flask in a cool glass annealing furnace, and apply heat up to 
500 [deg]C. Maintain at this temperature for 1 hour. After this time 
period, shut off and open the furnace to allow the flask to cool. Return 
the stopcocks to the flask receivers. Purge the assembly with high-
purity nitrogen for 2 to 5 minutes. Close off the stopcocks after 
purging to maintain a slight positive nitrogen pressure. Secure the 
stopcocks with tape. Presurvey samples can be obtained either by drawing 
the gases into the previously evacuated flask or by drawing the gases 
into and purging the flask with a rubber suction bulb.
    16.1.3.1.1 Evacuated Flask Procedure. Use a high-vacuum pump to 
evacuate the flask to the capacity of the pump; then close off the 
stopcock leading to the pump. Attach a 6-mm outside diameter (OD) glass 
tee to the flask inlet with a short piece of Teflon tubing. Select a 6-
mm OD borosilicate sampling probe, enlarged at one end to a 12-mm OD and 
of sufficient length to reach the centroid of the duct to be sampled. 
Insert a glass wool plug in the enlarged end of the probe to remove 
particulate matter. Attach the other end of the probe to the tee with a 
short piece of Teflon tubing. Connect a rubber suction bulb to the third 
leg of the tee. Place the filter end of the probe at the centroid of the 
duct, and purge the probe with the rubber suction bulb. After the probe 
is completely purged and filled with duct gases, open the stopcock to 
the grab flask until the pressure in the flask reaches duct pressure. 
Close off the stopcock, and remove the probe from the duct. Remove the 
tee from the flask and tape the stopcocks to prevent leaks during 
shipment. Measure and record the duct temperature and pressure.
    16.1.3.1.2 Purged Flask Procedure. Attach one end of the sampling 
flask to a rubber suction bulb. Attach the other end to a 6-mm OD glass 
probe as described in Section 8.3.3.1.1. Place the filter end of the 
probe at the centroid of the duct, or at a point no closer to the walls 
than 1 m, and apply suction with the bulb to completely purge the probe 
and flask. After the flask has been purged, close off the stopcock near 
the suction bulb, and then close off the stopcock near the probe. Remove 
the probe from the duct, and disconnect both the probe and suction bulb. 
Tape the stopcocks to prevent leakage during shipment. Measure and 
record the duct temperature and pressure.
    16.1.3.2 Flexible Bag Procedure. Tedlar or aluminized Mylar bags can 
also be used to obtain the presurvey sample. Use new bags, and leak-
check them before field use. In addition, check the bag before use for 
contamination by filling it with nitrogen or air, and

[[Page 444]]

analyzing the gas by GC at high sensitivity. Experience indicates that 
it is desirable to allow the inert gas to remain in the bag about 24 
hours or longer to check for desorption of organics from the bag. Follow 
the leak-check and sample collection procedures given in Section 8.2.1.
    16.1.3.3 Determination of Moisture Content. For combustion or water-
controlled processes, obtain the moisture content from plant personnel 
or by measurement during the presurvey. If the source is below 59 
[deg]C, measure the wet bulb and dry bulb temperatures, and calculate 
the moisture content using a psychrometric chart. At higher 
temperatures, use Method 4 to determine the moisture content.
    16.1.4 Determination of Static Pressure. Obtain the static pressure 
from the plant personnel or measurement. If a type S pitot tube and an 
inclined manometer are used, take care to align the pitot tube 90[deg] 
from the direction of the flow. Disconnect one of the tubes to the 
manometer, and read the static pressure; note whether the reading is 
positive or negative.
    16.1.5 Collection of Presurvey Samples with Adsorption Tube. Follow 
Section 8.2.4 for presurvey sampling.

                             17.0 References

    1. American Society for Testing and Materials. C1 Through C5 
Hydrocarbons in the Atmosphere by Gas Chromatography. ASTM D 2820-72, 
Part 23. Philadelphia, Pa. 23:950-958. 1973.
    2. Corazon, V.V. Methodology for Collecting and Analyzing Organic 
Air Pollutants. U.S. Environmental Protection Agency. Research Triangle 
Park, N.C. Publication No. EPA-600/2-79-042. February 1979.
    3. Dravnieks, A., B.K. Krotoszynski, J. Whitfield, A. O'Donnell, and 
T. Burgwald. Environmental Science and Technology. 5(12):1200-1222. 
1971.
    4. Eggertsen, F.T., and F.M. Nelsen. Gas Chromatographic Analysis of 
Engine Exhaust and Atmosphere. Analytical Chemistry. 30(6): 1040-1043. 
1958.
    5. Feairheller, W.R., P.J. Marn, D.H. Harris, and D.L. Harris. 
Technical Manual for Process Sampling Strategies for Organic Materials. 
U.S. Environmental Protection Agency. Research Triangle Park, N.C. 
Publication No. EPA 600/2-76-122. April 1976. 172 p.
    6. Federal Register, 39 FR 9319-9323. 1974.
    7. Federal Register, 39 FR 32857-32860. 1974.
    8. Federal Register, 23069-23072 and 23076-23090. 1976.
    9. Federal Register, 46569-46571. 1976.
    10. Federal Register, 41771-41776. 1977.
    11. Fishbein, L. Chromatography of Environmental Hazards, Volume II. 
Elesevier Scientific Publishing Company. New York, N.Y. 1973.
    12. Hamersma, J.W., S.L. Reynolds, and R.F. Maddalone. EPA/IERL-RTP 
Procedures Manual: Level 1 Environmental Assessment. U.S. Environmental 
Protection Agency. Research Triangle Park, N.C. Publication No. EPA 600/
276-160a. June 1976. 130 p.
    13. Harris, J.C., M.J. Hayes, P.L. Levins, and D.B. Lindsay. EPA/
IERL-RTP Procedures for Level 2 Sampling and Analysis of Organic 
Materials. U.S. Environmental Protection Agency. Research Triangle Park, 
N.C. Publication No. EPA 600/7-79-033. February 1979. 154 p.
    14. Harris, W.E., H.W. Habgood. Programmed Temperature Gas 
Chromatography. John Wiley and Sons, Inc. New York. 1966.
    15. Intersociety Committee. Methods of Air Sampling and Analysis. 
American Health Association. Washington, D.C. 1972.
    16. Jones, P.W., R.D. Grammer, P.E. Strup, and T.B. Stanford. 
Environmental Science and Technology. 10:806-810. 1976.
    17. McNair Han Bunelli, E.J. Basic Gas Chromatography. Consolidated 
Printers. Berkeley. 1969.
    18. Nelson, G.O. Controlled Test Atmospheres, Principles and 
Techniques. Ann Arbor. Ann Arbor Science Publishers. 1971. 247 p.
    19. NIOSH Manual of Analytical Methods, Volumes 1, 2, 3, 4, 5, 6, 7. 
U.S. Department of Health and Human Services, National Institute for 
Occupational Safety and Health. Center for Disease Control. 4676 
Columbia Parkway, Cincinnati, Ohio 45226. April 1977--August 1981. May 
be available from the Superintendent of Documents, Government Printing 
Office, Washington, D.C. 20402. Stock Number/Price:

Volume 1--O17-033-00267-3/$13
Volume 2--O17-033-00260-6/$11
Volume 3--O17-033-00261-4/$14
Volume 4--O17-033-00317-3/$7.25
Volume 5--O17-033-00349-1/$10
Volume 6--O17-033-00369-6/$9
Volume 7--O17-033-00396-5/$7

Prices subject to change. Foreign orders add 25 percent.
    20. Schuetzle, D., T.J. Prater, and S.R. Ruddell. Sampling and 
Analysis of Emissions from Stationary Sources; I. Odor and Total 
Hydrocarbons. Journal of the Air Pollution Control Association. 25(9): 
925-932. 1975.
    21. Snyder, A.D., F.N. Hodgson, M.A. Kemmer and J.R. McKendree. 
Utility of Solid Sorbents for Sampling Organic Emissions from Stationary 
Sources. U.S. Environmental Protection Agency. Research Triangle Park, 
N.C. Publication No. EPA 600/2-76-201. July 1976. 71 p.
    22. Tentative Method for Continuous Analysis of Total Hydrocarbons 
in the Atmosphere. Intersociety Committee, American Public Health 
Association. Washington, D.C. 1972. p. 184-186.

[[Page 445]]

    23. Zwerg, G. CRC Handbook of Chromatography, Volumes I and II. 
Sherma, Joseph (ed.). CRC Press. Cleveland. 1972.

         18.0 Tables, Diagrams, Flowcharts, and Validation Data

I. Name of company______________________________________________________
Date____________________________________________________________________
Address_________________________________________________________________
________________________________________________________________________
Contracts_______________________________________________________________
Phone___________________________________________________________________
Process to be sampled___________________________________________________
________________________________________________________________________
________________________________________________________________________
Duct or vent to be sampled______________________________________________
________________________________________________________________________
II. Process description_________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
Raw material____________________________________________________________
________________________________________________________________________
________________________________________________________________________
Products________________________________________________________________
________________________________________________________________________
________________________________________________________________________

Operating cycle
Check: Batch -------- Continuous -------- Cyclic --------
Timing of batch or cycle________________________________________________
Best time to test_______________________________________________________

III. Sampling site______________________________________________________
A. Description__________________________________________________________
Site decription_________________________________________________________
Duct shape and size_____________________________________________________
Material________________________________________________________________
Wall thickness -------- inches
Upstream distance -------- inches -------- diameter
Downstream distance -------- inches -------- diameter
Size of port____________________________________________________________
Size of access area_____________________________________________________
Hazards -------- Ambient temp. -------- [deg]F

B. Properties of gas stream
Temperature -------- [deg]C -------- [deg]F, Data source --------
Velocity --------, Data source --------
Static pressure -------- inches H2O, Data source --------
Moisture content --------%, Data source --------
Particulate content --------, Data source--------

Gaseous components
N2 -------- % Hydrocarbons -------- ppm
O2 --------% --------
CO -------- % -------- --------
CO2 -------- % -------- --------
SO2 -------- % -------- --------

Hydrocarbon components
-------- -------- ppm
-------- -------- ppm
-------- -------- ppm
-------- -------- ppm
-------- -------- ppm
-------- -------- ppm

C. Sampling considerations
Location to set up GC___________________________________________________
________________________________________________________________________
Special hazards to be considered________________________________________
________________________________________________________________________
Power available at duct_________________________________________________
Power available for GC__________________________________________________
Plant safety requirements_______________________________________________
________________________________________________________________________
Vehicle traffic rules___________________________________________________
________________________________________________________________________
Plant entry requirements________________________________________________
________________________________________________________________________
Security agreements_____________________________________________________
________________________________________________________________________
Potential problems______________________________________________________
________________________________________________________________________

D. Site diagrams. (Attach additional sheets if required).

               Figure 18-1. Preliminary Survey Data Sheet

Components to be analyzed and Expected concentration
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
Suggested chromatographic column________________________________________
________________________________________________________________________
Column flow rate -- ml/min
Head pressure -------- mm Hg

Column temperature: Isothermal -------- [deg]C, Programmed from -------- 
[deg]C to -------- [deg]C at -------- [deg]C/min
Injection port/sample loop temperature -------- [deg]C
Detector temperature -------- [deg]C
Detector flow rates: Hydrogen -------- ml/min., head pressure -------- 
mm Hg, Air/Oxygen -------- ml/min., head pressure -------- mm Hg.
Chart speed -------- inches/minute
Compound data:
Compound and Retention time and Attenuation
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________
________________________________________________________________________

           Figure 18-2. Chromatographic Conditions Data Sheet

[[Page 446]]



                   Figure 18-3. Preparation of Standards in Tedlar Bags and Calibration Curve
----------------------------------------------------------------------------------------------------------------
                                                                                  Standards
                                                           -----------------------------------------------------
                                                             Mixture 1               i>2               i>3
----------------------------------------------------------------------------------------------------------------
Standards Preparation Data:
    Organic:
        Bag number or identification......................
        Dry gas meter calibration factor..................
        Final meter reading (liters)......................
        Initial meter reading (liters)....................
        Metered volume (liters)...........................
        Average meter temperature ([deg]K)................
        Average meter pressure, gauge (mm Hg).............
        Average atmospheric perssure (mm Hg)..............
        Average meter pressure, absolute (mm Hg)..........
        Syringe temperature ([deg]K) (see Section
         10.1.2.1)........................................
        Syringe pressure, absolute (mm Hg) (see Section
         10.1.2.1)........................................
        Volume of gas in syringe (ml) (Section 10.1.2.1)..
        Density of liquid organic (g/ml) (Section
         10.1.2.1)........................................
        Volume of liquid in syringe (ml) (Section
         10.1.2.1)........................................
GC Operating Conditions:
    Sample loop volume (ml)...............................
    Sample loop temperature ( [deg]C).....................
    Carrier gas flow rate (ml/min)........................
Column temperature:
    Initial ( [deg]C).....................................
    Rate change ( [deg]C/min).............................
    Final ( [deg]C).......................................
Organic Peak Identification and Calculated Concentrations:
    Injection time (24 hour clock)........................
    Distance to peak (cm).................................
    Chart speed (cm/min)..................................
    Organic retention time (min)..........................
    Attenuation factor....................................
    Peak height (mm)......................................
    Peak area (mm2).......................................
    Peak area * attenuation factor (mm2)..................
    Calculated concentration (ppm) (Equation 18-3 or 18-4)
----------------------------------------------------------------------------------------------------------------
Plot peak area * attenuation factor against calculated concentration to obtain calibration curve.

Flowmeter number or identification______________________________________
Flowmeter Type__________________________________________________________
Method: Bubble meter---- Spirometer---- Wet test meter ----
Readings at laboratory conditions:
Laboratory temperature (Tlab) ---- [deg]K
Laboratory barometric pressure (Plab)---- mm Hg
Flow data:

                                                    Flowmeter
----------------------------------------------------------------------------------------------------------------
         Reading (as marked)                     Temp. ([deg]K)                      Pressure (absolute)
----------------------------------------------------------------------------------------------------------------

-------------------------------------

-------------------------------------

-------------------------------------

-------------------------------------

----------------------------------------------------------------------------------------------------------------


                                               Calibration Device
----------------------------------------------------------------------------------------------------------------
             Time (min)                           Gas volume a                           Flow rate b
----------------------------------------------------------------------------------------------------------------

-------------------------------------

-------------------------------------

-------------------------------------

[[Page 447]]



-------------------------------------

----------------------------------------------------------------------------------------------------------------
 a Vol. of gas may be measured in milliliters, liters or cubic feet.
b Convert to standard conditions (20 [deg]C and 760 mm Hg). Plot flowmeter reading against flow rate (standard
  conditions), and draw a smooth curve. If the flowmeter being calibrated is a rotameter or other flow device
  that is viscosity dependent, it may be necessary to generate a ``family'' of calibration curves that cover the
  operating pressure and temperature ranges of the flowmeter. While the following technique should be verified
  before application, it may be possible to calculate flow rate reading for rotameters at standard conditions
  Qstd as follows:

  [GRAPHIC] [TIFF OMITTED] TR17OC00.313
  

------------------------------------------------------------------------
 Flow rate (laboratory conditions)        Flow rate (STD conditions)
------------------------------------------------------------------------

------------------------------------

------------------------------------

------------------------------------

------------------------------------

------------------------------------------------------------------------

                   Figure 18-4. Flowmeter Calibration

[[Page 448]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.314


[[Page 449]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.315


                            Preparation of Standards by Dilution of Cylinder Standard
                   [Cylinder Standard: Organic -------- Certified Concentration -------- ppm]
----------------------------------------------------------------------------------------------------------------
                                                                         Date:
     Standards preparation data:      --------------------------------------------------------------------------
                                              Mixture 1                Mixture 2                Mixture 3
----------------------------------------------------------------------------------------------------------------
Stage 1:
    Standard gas flowmeter reading...
    Diluent gas flowmeter reading
    Laboratory temperature ([deg]K)
    Barometric pressure (mm Hg)
    Flowmeter gage pressure (mm Hg)
    Flow rate cylinder gas at
     standard conditions (ml/min)
    Flow rate diluent gas at standard
     conditions (ml/min)
    Calculated concentration (ppm)
Stage 2 (if used):
    Standard gas flowmeter reading
    Diluent gas flowmeter reading
    Flow rate Stage 1 gas at standard
     conditions (ml/min)
    Flow rate diluent gas at standard
     conditions
    Calculated concentration (ppm)
GC Operating Conditions:
    Sample loop volume (ml)
    Sample loop temperature ( [deg]C)
    Carrier gas flow rate (ml/min)
Column temperature:
    Initial ( [deg]C)

[[Page 450]]


    Program rate ( [deg]C/min)
    Final ( [deg]C)
Organic Peak Identification and
 Calculated Concentrations:
    Injection time (24-hour clock)
    Distance to peak (cm)
    Chart speed (cm/min)
    Retention time (min)
    Attenuation factor
    Peak area (mm \2\)
    Peak area *attenuation factor
----------------------------------------------------------------------------------------------------------------
Plot peak area *attenuation factor against calculated concentration to obtain calibration curve.

    Figure 18-7. Standards Prepared by Dilution of Cylinder Standard
[GRAPHIC] [TIFF OMITTED] TR17OC00.316


[[Page 451]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.317


[[Page 452]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.318


                                     Plant-------- Date-------- Site--------
----------------------------------------------------------------------------------------------------------------
                                                                     Sample 1        Sample 2        Sample 3
----------------------------------------------------------------------------------------------------------------
Source temperature ( [deg]C)....................................  ..............  ..............  ..............
Barometric pressure (mm Hg).....................................  ..............  ..............  ..............
Ambient temperature ( [deg]C)...................................  ..............  ..............  ..............
Sample flow rate (appr.)........................................  ..............  ..............  ..............
Bag number......................................................  ..............  ..............  ..............
Start time......................................................  ..............  ..............  ..............
Finish time.....................................................  ..............  ..............  ..............
----------------------------------------------------------------------------------------------------------------

   Figure 18-10. Field Sample Data Sheet--Tedlar Bag Collection Method

         Plant ------------------ Date ---------------- Location
                        ------------------------


1. General information:
    Source temperature ( [deg]C)......................  ................
    Probe temperature ( [deg]C).......................  ................
    Ambient temperature ( [deg]C).....................  ................
    Atmospheric pressure (mm).........................  ................
    Source pressure ('Hg).............................  ................
    Absolute source pressure (mm).....................  ................
    Sampling rate (liter/min).........................  ................
    Sample loop volume (ml)...........................  ................
    Sample loop temperature ( [deg]C).................  ................
    Columnar temperature:
        Initial ( [deg]C) time (min)..................  ................

[[Page 453]]


        Program rate ( [deg]C/min)....................  ................
        Final ( [deg]C)/time (min)....................  ................
    Carrier gas flow rate (ml/min)....................  ................
    Detector temperature ( [deg]C)....................  ................
    Injection time (24-hour basis)....................  ................
    Chart Speed (mm/min)..............................  ................
    Dilution gas flow rate (ml/min)...................  ................
    Dilution gas used (symbol)........................  ................
    Dilution ratio....................................  ................



                                     2. Field Analysis Data--Calibration Gas
                                     2. [Run No. -------- Time ------------]
----------------------------------------------------------------------------------------------------------------
       Components              Area             Attenuation            A x A Factor            Conc.-- (ppm)
----------------------------------------------------------------------------------------------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
------------------------
                         ...............  ......................  ......................  ......................
----------------------------------------------------------------------------------------------------------------

                Figure 18-11. Field Analysis Data Sheets

[[Page 454]]

[GRAPHIC] [TIFF OMITTED] TR17OC00.319


[[Page 455]]


[GRAPHIC] [TIFF OMITTED] TR17OC00.320


                                Gaseous Organic Sampling and Analysis Check List
                                [Respond with initials or number as appropriate]

                                                                                                Date

1. Presurvey data:
    A. Grab sample collected........................  [squ]                         ------
        B. Grab sample analyzed for composition.....  [squ]                         ------
        Method GC...................................  [squ]                         ------
            GC/MS...................................  [squ]                         ------
            Other...................................  [squ]                         ------
    C. GC-FID analysis performed....................  [squ]                         ------
2. Laboratory calibration data:
    A. Calibration curves prepared..................  [squ]                         ------
        Number of components........................  [squ]                         ------
        Number of concentrations/component (3         [squ]                         ------
         required).
    B. Audit samples (optional):
    Analysis completed..............................  [squ]                         ------
    Verified for concentration......................  [squ]                         ------
    OK obtained for field work......................  [squ]                         ------
3. Sampling procedures:
    A. Method:
        Bag sample..................................  [squ]                         ------
        Direct interface............................  [squ]                         ------
        Dilution interface..........................  [squ]                         ------
    B. Number of samples collected..................  [squ]                         ------
4. Field Analysis:
    A. Total hydrocarbon analysis performed.........  [squ]                         ------
    B. Calibration curve prepared...................  [squ]                         ------
        Number of components........................  [squ]                         ------
        Number of concentrations per component (3     [squ]                         ------
         required).



[[Page 456]]

               Gaseous Organic Sampling and Analysis Data

Plant___________________________________________________________________
Date____________________________________________________________________
Location________________________________________________________________

----------------------------------------------------------------------------------------------------------------
            Source sample 1                       Source sample 2                     Source sample 3
---------------------------------------------------------------------------------------------------------------
1. General information:
    Source temperature ( [deg]C)......  ..................................  ..................................
    Probe temperature ( [deg]C).......  ..................................  ..................................
    Ambient temperature ( [deg]C).....  ..................................  ..................................
    Atmospheric pressure (mm Hg)......  ..................................  ..................................
    Source pressure (mm Hg)...........  ..................................  ..................................
    Sampling rate (ml/min)............  ..................................  ..................................
    Sample loop volume (ml)...........  ..................................  ..................................
    Sample loop temperature ( [deg]C).  ..................................  ..................................
    Sample collection time (24-hr       ..................................  ..................................
     basis).
    Column temperature:
        Initial ( [deg]C).............  ..................................  ..................................
        Program rate ( [deg]C/min)....  ..................................  ..................................
        Final ( [deg]C)...............  ..................................  ..................................
    Carrier gas flow rate (ml/min)....  ..................................  ..................................
    Detector temperature ( [deg]C)....  ..................................  ..................................
    Chart speed (cm/min)..............  ..................................  ..................................
    Dilution gas flow rate (ml/min)...  ..................................  ..................................
    Diluent gas used (symbol).........  ..................................  ..................................
    Dilution ratio....................  ..................................  ..................................
Performed by: (signature):------------------------ Date:------------------------
----------------------------------------------------------------------------------------------------------------

                Figure 18-14. Sampling and Analysis Sheet

[36 FR 24877, Dec. 23, 1971]

    Editorial Note: For Federal Register citations affecting part 60, 
appendix A-6, see the List of CFR Sections Affected, which appears in 
the Finding Aids section of the printed volume and on GPO Access.