[Code of Federal Regulations]

[Title 40, Volume 12]

[Revised as of July 1, 2006]

From the U.S. Government Printing Office via GPO Access

[CITE: 40CFR63.1450]



[Page 34-38]

 

                   TITLE 40--PROTECTION OF ENVIRONMENT

 

         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)

 

PART 63_NATIONAL EMISSION STANDARDS FOR HAZARDOUS AIR POLLUTANTS FOR 

SOURCE CATEGORIES--Table of Contents

 

Subpart QQQ_National Emission Standards for Hazardous Air Pollutants for 

                         Primary Copper Smelting

 

Sec.  63.1450  What test methods and other procedures must I use to 

demonstrate initial compliance with the emission limitations?



    (a) Total particulate matter emission limits. You must conduct each 

performance test to determine compliance with the total particulate 

matter emission limits in Sec.  63.1444 or Sec.  63.1446 that apply to 

you according to the requirements for representative test conditions 

specified in Sec.  63.7(e)(1) and using the test methods and procedures 

in paragraphs (a)(1) through (5) of this section.

    (1) Determine the concentration of total particulate matter 

according to the test methods in appendix A to part 60 of this chapter 

as specified in paragraphs (a)(1)(i) through (iii) of this section.

    (i) Method 1 to select sampling port locations and the number of 

traverse points. Sampling ports must be located at the outlet of the 

control device and prior to any releases to the atmosphere.

    (ii) Method 2, 2F, or 2G to determine the volumetric flow rate of 

the stack gas.

    (iii) Method 3, 3A, or 3B to determine the dry molecular weight of 

the stack gas.

    (iv) Method 4 to determine the moisture content of the stack gas.

    (v) Method 5, 5D, or 17, as applicable, to determine the 

concentration of total particulate matter. You can also use ASTM D4536-

96 incorporated by reference in Sec.  63.14 as an alternative to the 

sampling equipment and operating procedures in Method 5 or 17 when 

testing a positive pressure baghouse, but you must use the sample 

traverse location and number of sampling points described in Method 5D.

    (2) As an alternative to using the applicable method specified in 

paragraph (a)(1)(v) of this section, you may determine total particulate 

matter emissions from the control device using Method 29 in appendix A 

of part 60 of this chapter provided that you follow the procedures and 

precautions prescribed in Method 29. If the control device is a positive 

pressure baghouse, you must also follow the measurement procedure 

specified in sections 4.1 through 4.3 of Method 5D.

    (3) You must conduct three separate test runs for each performance 

test. Each test run must have a minimum sampling time of 60 minutes and 

a minimum sampling volume of 0.85 dscm. For the purpose of determining 

compliance with the applicable total particulate matter emission limit, 

the arithmetic mean of the results for the three separate test runs is 

used.

    (4) For a venturi wet scrubber applied to emissions from an affected 

source and subject to operating limits in Sec.  63.1444(g) or Sec.  

63.1446(d) for pressure drop and scrubber water flow rate, you must 

establish site-specific operating limits according to the procedures in 

paragraph (a)(4)(i) and (ii) of this section.

    (i) Using the continuous parameter monitoring system (CPMS) required 

in Sec.  63.1452, measure and record the pressure drop and scrubber 

water flow rate during each run of the particulate matter performance 

test.

    (ii) Compute and record the hourly average pressure drop and 

scrubber water flow rate for each individual test run. Your operating 

limits are the lowest average pressure drop and scrubber water flow rate 

value in any of the three runs that meet the applicable emission limit.

    (5) For a control device other than a baghouse or venturi wet 

scrubber applied to emissions from an affected source and subject to 

site-specific operating limit(s) in Sec.  63.1444(h) or Sec.  63.1446(e) 

for appropriate, site-specific operating parameters that are 

representative and reliable indicators of the control device 

performance, you must establish a site-specific operating limit(s) 

according to the procedures in paragraph (a)(5)(i) through (iv) of this 

section.

    (i) Select one or more operating parameters, as appropriate for the 

control device design, that can be used as representative and reliable 

indicators of the control device operation.



[[Page 35]]



    (ii) Using the CPMS required in Sec.  63.1452, measure and record 

the selected operating parameters for the control device during each run 

of the total particulate matter performance test.

    (iii) Compute and record the hourly average value for each of the 

selected operating parameters for each individual test run. Your 

operating limits are the lowest value or the highest value, as 

appropriate for the selected operating parameter, measured in any of the 

three runs that meet the applicable emission limit.

    (iv) You must prepare written documentation to support your 

selection of the operating parameters used for the control device. This 

documentation must include a description of each selected parameter, a 

rationale for why you chose the parameter, a description of the method 

used to monitor the parameter, and the data recorded during the 

performance test and used to set the operating limit(s).

    (b) Nonsulfuric acid particulate matter emission limits. You must 

conduct each performance test to determine compliance with the 

nonsulfuric acid particulate matter emission limits in Sec.  63.1444 

that apply to you according to the requirements for representative test 

conditions specified in Sec.  63.7(e)(1) and using the test methods and 

procedures in paragraphs (b)(1) and (2) of this section.

    (1) Determine the concentration of nonsulfuric acid particulate 

matter according to the test methods in appendix A to part 60 of this 

chapter as specified in paragraphs (b)(1)(i) through (v) of this 

section.

    (i) Method 1 to select sampling port locations and the number of 

traverse points. Sampling ports must be located at the outlet of the 

control device and prior to any releases to the atmosphere.

    (ii) Method 2, 2F, or 2G to determine the volumetric flow rate of 

the stack gas.

    (iii) Method 3, 3A, or 3B to determine the dry molecular weight of 

the stack gas.

    (iv) Method 4 to determine the moisture content of the stack gas.

    (v) Method 5B to determine the nonsulfuric acid particulate matter 

emissions.

    (2) You must conduct three separate test runs for each performance 

test. Each test run must have a minimum sampling time of 240 minutes and 

a minimum sampling volume of 3.4 dscm. For the purpose of determining 

compliance with the nonsulfuric acid particulate matter emission limit, 

the arithmetic mean of the results for the three separate test runs is 

used.

    (c) Copper converter department capture system opacity limit. You 

must conduct each performance test to determine compliance with the 

opacity limit in Sec.  63.1444 using the test methods and procedures in 

paragraphs (c)(1) through (9) of this section.

    (1) You must conduct the performance test during the period when the 

primary copper smelter is operating under conditions representative of 

the smelter's normal blister copper production rate. You may not conduct 

a performance test during periods of startup, shutdown, or malfunction. 

Before conducting the performance test, you must prepare a written test 

plan specifying the copper production conditions to be maintained 

throughout the opacity observation period and including a copy of the 

written documentation you have prepared according to paragraph (a)(3) of 

this section to support the established operating limits for the copper 

converter department capture system. You must submit a copy of the test 

plan for review and approval by the Administrator or delegated 

authority. During the observation period, you must collect appropriate 

process information and copper converter department capture system 

operating information to prepare documentation sufficient to verify that 

all opacity observations were made during the copper production and 

capture system operating conditions specified in the approved test plan.

    (2) You must notify the Administrator or delegated authority before 

conducting the opacity observations to allow the Administrator or 

delegated authority the opportunity to have authorized representatives 

attend the test. Written notification of the location and scheduled date 

for conducting



[[Page 36]]



the opacity observations must be received by the Administrator on or 

before 30 calendar days before this scheduled date.

    (3) You must gather the data needed for determining compliance with 

the opacity limit using qualified visible emission observers and process 

monitors as described in paragraphs (c)(3)(i) and (ii) of this section.

    (i) Opacity observations must be performed by a sufficient number of 

qualified visible emission observers to obtain two complete concurrent 

sets of opacity readings for the required observation period. Each 

visible emission observer must be certified as a qualified observer by 

the procedure specified in section 3 of Method 9 in appendix A of part 

60 of this chapter. The entire set of readings during the required 

observation period does not need to be made by the same two observers. 

More than two observers may be used to allow for substitutions and 

provide for observer rest breaks. The owner or operator must obtain 

proof of current visible emission reading certification for each 

observer.

    (ii) A person (or persons) familiar with the copper production 

operations conducted at the smelter must serve as the indoor process 

monitor. The indoor process monitor is stationed at a location inside 

the building housing the batch copper converters such that he or she can 

visually observe and record operations that occur in the batch copper 

converter aisle during the times that the visible emission observers are 

making opacity readings. More than one indoor process monitor may be 

used to allow for substitutions and provide for rest breaks.

    (4) You must make all opacity observations using Method 9 in 

appendix A to part 60 of this chapter and following the procedures 

described in paragraphs (c)(4)(i) and (ii) of this section.

    (i) Each visible emission observer must make his or her readings at 

a position from the outside of the building that houses the copper 

converter department such that the observer's line-of-sight is 

approximately perpendicular to the longer axis of the converter 

building, and the observer has an unobstructed view of the building roof 

monitor sections or roof exhaust fan outlets that are positioned over 

each of the batch copper converters inside the building. Opacity 

readings can only be made during those times when the observer's 

position meets the sun orientation and other conditions specified in 

section 2.1 of Method 9.

    (ii) At 15-second intervals, each visible emission observer views 

the building roof monitor sections or roof exhaust fan outlets that are 

positioned over each of the batch copper converters inside the building 

and reads the opacity of the visible plumes. If no plume is visible, the 

observer records zero as the opacity value for the 15-second interval. 

In situations when it is possible for an observer to distinguish two or 

more visible emission plumes from the building roof monitor sections or 

roof exhaust fan outlets, the observer must identify, to the extent 

feasible, the plume having the highest opacity and record his or her 

opacity reading for that plume as the opacity value for the 15-second 

interval.

    (5) You must make opacity observations for a period of sufficient 

duration to obtain a minimum of 120 1-minute intervals during which at 

least one copper converter is blowing and no interferences have occurred 

from other copper production events, as specified in paragraph (c)(7) of 

this section, which generate visible emissions inside the building that 

potentially can interfere with the visible emissions from the converter 

capture systems as seen by the outside observers. To obtain the required 

number of 1-minute intervals, the observation period may be divided into 

two or more segments performed on the same day or on different days if 

conditions prevent the required number of opacity readings from being 

obtained during one continuous time period. Examples of these conditions 

include, but are not limited to, changes in the sun's orientation 

relative to visible emission observers' positions such that the Method 9 

conditions are no longer met or an unexpected thunder storm. If the 

total observation period is divided into two or more segments, all 

opacity observations must be made during the same set of copper 

production conditions described in your approved test plan as required 

by paragraph (c)(1) of this section.



[[Page 37]]



    (6) You must gather indoor process information during all times that 

the visible emission observers are making opacity readings outside the 

building housing the copper converter department. The indoor process 

monitor must continually observe the operations occurring in the copper 

converter department and prepare a written record of his or her 

observations using the procedure specified in paragraphs (c)(6)(i) 

through (iv) of this section.

    (i) At the beginning of each observation period or segment, the 

clock time setting on the watch or clock to be used by the indoor 

process monitor must be synchronized with the clock time settings for 

the timepieces to be used by the outdoor opacity observers.

    (ii) During each period or segment when opacity readings are being 

made by the visible emission observers, the indoor process monitor must 

continuously observe the operations occurring in the copper converter 

department and record his or her observations in a log book, on data 

sheets, or other type of permanent written format.

    (iii) When a batch copper converter is blowing, a record must be 

prepared for the converter that includes, but is not limited to, the 

clock times for when blowing begins and when blowing ends and the 

converter blowing rate. This information may be recorded by the indoor 

process monitor or by a separate, automated computer data system.

    (iv) The process monitor must record each event other than converter 

blowing that occurs in or nearby the converter aisle that he or she 

observes to generate visible emissions inside the building. The recorded 

entry for each event must include, but is not limited to, a description 

of the event and the clock times when the event begins and when the 

event ends.

    (7) You must prepare a summary of the data for the entire 

observation period using the information recorded during the observation 

period by the outdoor visible emission observers and the indoor process 

monitor and the procedure specified in paragraphs (c)(7)(i) through (iv) 

of this section.

    (i) Using the field data sheets, identify the 1-minute clock times 

for which a total of eight opacity readings were made and recorded by 

both observers at 15-second intervals according to the test procedures 

(i.e., a total of four opacity values have been recorded for the 1-

minute interval by each of the two observers). Calculate the average of 

the eight 15-second interval readings recorded on the field data sheets 

by the two observers during the clock time minute interval (add the four 

consecutive 15-second interval opacity readings made by Observer A 

during the specified clock time minute, plus the four consecutive 15-

second interval opacity readings made by Observer B during the same 

clock time minute, and divide the resulting total by eight). Record the 

clock time and the opacity average for the 1-minute interval on a data 

summary sheet. Figure 1 of this subpart shows an example of the format 

for the data summary sheet you may use, but are not required to use.

    (ii) Using the data summary sheets prepared according to paragraph 

(c)(7)(i) of this section and the process information recorded according 

to paragraph (c)(6)(iii) of this section, identify those 1-minute 

intervals for which at least one of the batch copper converters was 

blowing.

    (iii) Using the data summary sheets prepared according to paragraph 

(c)(7)(ii) of this section and the process information recorded 

according to paragraph (c)(6)(iv) of this section, identify the 1-minute 

intervals during which at least one copper converter was blowing but 

none of the interference events listed in paragraphs (c)(7)(iii)(A) 

through (F) of this section occurred. Other ancillary activities not 

listed but conducted in or adjacent to the converter aisle during the 

opacity observations are not considered to be interference events (e.g., 

converter aisle cleaning, placement of smoking ladles or skulls on the 

converter aisle floor).

    (A) Charging of copper matte, reverts, or other materials to a batch 

copper converter;

    (B) Skimming slag or other molten materials from a batch copper 

converter;

    (C) Pouring of blister copper or other molten materials from a batch 

copper converter;



[[Page 38]]



    (D) Return of slag or other molten materials to the flash smelting 

furnace or slag cleaning vessel;

    (E) Roll-out or roll-in of the batch copper converter; or

    (F) Smoke and fumes generated inside the converter building by 

operation of the smelting furnace, the slag cleaning vessel (if used), 

anode refining and casting processes that drift into the copper 

converter department.

    (iv) Using the data summary sheets prepared according to paragraph 

(c)(7)(iii) of this section, up to five 1-minute intervals following an 

interference event may be eliminated from data used for the compliance 

determination calculation specified in paragraph (c)(8) of this section 

by applying a time delay factor. The time delay factor must be a 

constant number of minutes not to exceed 5 minutes that is added to the 

clock time recorded when cessation of the interference event occurs. The 

same time delay factor must be used for all interference events (i.e., a 

constant time delay factor for the smelter of 1 minute, 2 minutes, 3 

minutes, 4 minutes, or 5 minutes). The number of minutes to be used for 

the time delay factor is determined based on the site-specific equipment 

and converter building configuration. An explanation of the rationale 

for selecting the value used for the time delay factor must be prepared 

and included in the test report.

    (8) You must use the data summary prepared in paragraph (c)(7) of 

this section to calculate the average opacity value for a minimum of 120 

1-minute intervals during which at least one copper converter was 

blowing with no interference events as determined according to 

paragraphs (c)(7)(iii) and (iv) of this section. Average opacity is 

calculated using Equation 1 of this section:

[GRAPHIC] [TIFF OMITTED] TR12JN02.026



Where



VEave = Average opacity to be used for compliance 

determination (percent);

n = Total number of 1-minute intervals during which at least one copper 

converter was blowing with no interference events as determined 

according to paragraphs (c)(7)(iii) and (iv) of this section (at least 

120 1-minute intervals);

i = 1-minute interval ``i'' during which at least one copper converter 

was blowing with no interference events as determined according to 

paragraphs (c)(7)(iii) and (iv) of this section; and

VEi = Average opacity value calculated for the eight opacity 

readings recorded during 1-minute interval ``i'' (percent).



    (9) You must certify that the copper converter department capture 

system operated during the performance test at the operating limits 

established in your capture system operation and maintenance plan using 

the procedure specified in paragraphs (c)(9)(i) through (iv) of this 

section.

    (i) Concurrent with all opacity observations, measure and record 

values for each of the operating limit parameters in your capture system 

operation and maintenance plan according to the monitoring requirements 

specified in Sec.  63.1452(a).

    (ii) For any dampers that are manually set and remain in the same 

position at all times the capture system is operating, the damper 

position must be visually checked and recorded at the beginning and end 

of each opacity observation period segment.

    (iii) Review the recorded monitoring data. Identify and explain any 

times during batch copper converter blowing when the capture system 

operated outside the applicable operating limits.

    (iv) Certify in your performance test report that during all 

observation period segments, the copper converter department capture 

system was operating at the values or settings established in your 

capture system operation and maintenance plan.