[Code of Federal Regulations]
[Title 16, Volume 2]
[Revised as of January 1, 2007]
From the U.S. Government Printing Office via GPO Access
[CITE: 16CFR1500.43a]

[Page 442-449]
 
                     TITLE 16--COMMERCIAL PRACTICES
 
             CHAPTER II--CONSUMER PRODUCT SAFETY COMMISSION
 
PART 1500_HAZARDOUS SUBSTANCES AND ARTICLES; ADMINISTRATION AND 
ENFORCEMENT REGULATIONS--Table of Contents
 
Sec.  1500.43a  Method of test for flashpoint of volatile flammable materials.

    (a) Scope. (1) This method describes the test procedure which the 
Commission will use for the determination of the flashpoint of volatile 
flammable materials, using a Setaflash \1\ low-range closed tester, or 
an apparatus producing equivalent results. The method described in this 
section is essentially a Setaflash equilibrium procedure which closely 
parallels the test method designated ASTM D 3828-81, ``Standard Test 
Methods for Flash Point by

[[Page 443]]

Setaflash Closed Tester,'' published by the American Society for Testing 
and Materials (ASTM), 1916 Race Street, Philadelphia, Pennsylvania 
19103. Manufacturers and labelers of products subject to labeling and 
other requirements under the Federal Hazardous Substances Act may use 
other apparatus and/or test methods which produce equivalent results.
---------------------------------------------------------------------------

    \1\ Setaflash is a registered trademark of Stanhope-Seta Limited, 
Surrey, England.
---------------------------------------------------------------------------

    (2) At the option of the user, the procedures described in this 
section may be used to determine the actual flashpoint temperature of a 
sample or to determine whether a product will or will not flash at a 
specified temperature (flash/no flash).
    (3) If the substance to be tested has a viscosity greater than 150 
Stokes at 77 [deg]F (25 [deg]C), see paragraph (n) of this section for 
modifications to the testing procedure.
    (4) If the Commission has reason to believe on the basis of reliable 
experience or other relevant information or data that the flammability 
hazard of a substance is greater or less than its flammability 
classification based on flashpoint temperature determined in accordance 
with this Sec.  1500.43a and that the substance should be reclassified, 
the Commission will initiate a rulemaking proceeding for 
reclassification of the substance. Product manufacturers and labelers 
may use reliable experience or other relevant information or data in 
addition to the flashpoint temperature of a substance as a basis for 
compliance with any applicable requirements of the Federal Hazardous 
Substances Act in the absence of a rule issued by the Commission to 
reclassify the substance.
    (b) Summary of test methods. (1) Method A--Flash/No Flash Test. A 
specified volume of sample is introduced by a syringe into the cup of 
the apparatus that is set and maintained at the specified temperature. 
After a specific time a test flame is applied and an observation made as 
to whether or not a flash occurred. Test procedures are set forth in 
detail in Sec.  1500.43a(i).
    (2) Method B--Finite (or Actual) Flashpoint. (i) A specified voume 
of sample is introduced into the cup of the apparatus that is maintained 
at the expected flashpoint. After a specified time a test flame is 
applied and the observation made whether or not a flash occurred.
    (ii) The specimen is removed from the cup, the cup cleaned, and the 
cup temperature adjusted 5 [deg]C (9 [deg]F), lower or higher depending 
on whether or not a flash occurred previously. A fresh specimen is 
introduced and tested. This procedure is repeated until the flashpoint 
is established within 5 [deg]C (9 [deg]F).
    (iii) The procedure is then repeated at 1 [deg]C (2 [deg]F) 
intervals until the flashpoint is determined to the nearest 1 [deg]C (2 
[deg]F).
    (iv) If improved accuracy is desired the procedure is repeated at 
0.5 [deg]C (1 [deg]F). Test procedures are set forth in detail at Sec.  
1500.43a(j).
    (3) The test procedures will be modified, where necessary, to ensure 
that the results obtained reflect the hazard of the substance under 
reasonably foreseeable conditions of use. Thus, for example, the 
material, if a mixture, will normally be tested as it comes from the 
container, and/or after a period of evaporation. The period of 
evaporation for a material which is a mixture will normally be the time 
required for the mixture to evaporate in an open beaker under ambient 
conditions to 90 percent of its original volume, or a period of four 
hours, whichever occurs first. However, this period of evaporation will 
be changed if the results obtained do not represent the hazard of the 
substance under reasonably foreseeable conditions of use.
    (c) Definition of flashpoint. The lowest temperature of the sample, 
corrected to a barometric pressure of 101.3 kPa (760 mm Hg), at which 
application of a test flame causes the vapor of the sample to ignite 
under specified conditions of test. The sample is deemed to have flashed 
when a large flame appears and instantaneously propagates itself over 
the surface of the sample. Occasionally, particularly near actual 
flashpoint, the application of the test flame will cause a halo or an 
enlarged flame; this is not a flash and should be ignored.
    (d) Test apparatus. The test apparatus is an equilibrium closed-cup 
tester with a range up to 100 [deg]C (212 [deg]F). The essential 
dimensions and requirements are shown in figure 1 and table 3, and

[[Page 444]]

are described in Sec.  1500.43a(m). Closed-cup flashpoint testers and 
accessories meeting these requirements are available from commercial 
suppliers and distributors of laboratory equipment.
    (e) Safety precautions. The operator must exercise and take 
appropriate safety precautions during the initial application of the 
test flame to the sample. Samples containing low-flash material may give 
an abnormally strong flash when the test flame is first applied.
    (f) Preparation of samples. (1) Erroneously high flashpoints may be 
obtained if precautions are not taken to avoid the loss of volatile 
material. In preliminary tests of materials taken directly from the 
container, do not open containers unnecessarily and make a transfer 
unless the sample temperature is at least 10 [deg]C (18 [deg]F) below 
the expected flashpoint. Do not use samples in leaky containers for this 
test.
    (2) Do not store samples in plastic (polyethylene, polypropylene, 
etc.) bottles since volatile material may diffuse through the walls of 
the bottle.
    (3) A 2-ml specimen is required for each test. If possible, obtain 
at least a 50-ml sample from the bulk test site and store in a clean, 
tightly closed container.
    (g) Preparation of apparatus. (1) Place the tester on a level, 
stable surface. Unless tests are made in a draft-free area, surround the 
tester on three sides with a shield for protection. Do not rely on tests 
made in a laboratory draft hood or near ventilators.
    (2) Read the manufacturer's instructions on the care and servicing 
of the instrument and for correct operation of its controls.
    (h) Calibration and standardization. (1) Before initial use 
determine and plot the relationship between the temperature control dial 
and the thermometer readings at each major (numbered) dial division as 
follows:
    Turn the temperature control knob \2\ fully counterclockwise (``O'' 
reading). Advance the temperature control knob clockwise until the 
indicator light is illuminated.\3\ Advance the knob clockwise to the 
next numbered line. After the thermometer mercury column ceases to 
advance, record the dial reading and the temperature. Advance the knob 
clockwise to the next numbered line. After the thermometer mercury 
column ceases to advance, read the dial reading and the temperature. 
Repeat this procedure through the full range of the instrument. Plot the 
dial readings versus the respective temperatures.
---------------------------------------------------------------------------

    \2\ If the instrument has two temperature control knobs, set the 
fine control (center, small knob) at its mid-position and allow it to 
remain there throughout the calibration. The calibration is determined 
by adjusting the coarse control (large, outer knob) only.
    \3\ When using the tester, it will be found that the indicator light 
may not illuminate and the temperature may not rise until a temperature 
control dial setting between one and two is reached.
---------------------------------------------------------------------------

    (2) Standardize the instrument using a sample of material meeting 
the specifications in table 1. If the average of two determinations 
falls within the acceptable limits the instrument is assumed to be 
operating properly. If the average of the two determinations does not 
fall within this range, check the manufacturer's operating and 
maintenance instructions and determine that they are being followed. In 
particular, be sure that the cup lid assembly makes a vapor-tight seal 
with the cup, the shutter provides a light-tight seal, and that adequate 
heat transfer paste surrounds the thermometer bulb and the immersed 
portion of the barrel.
    (i) Test Method A--for determining Flash/No Flash. (1) Determine the 
target flashpoint as follows:
    (i) Target flashpoint, [deg]C=Sc--0.25 (101.3--A)
    (ii) Target flashpoint, [deg]C=Sc--0.03 (760-B)
    (iii) Target flashpoint, [deg]F=Sf--0.06 (760-B)

where:

Sc=specification, or uncorrected target, flashpoint, [deg]C,
Sf=specification, or uncorrected target, flashpoint, [deg]F,
B=ambient barometric pressure, mm Hg,\4\ and
---------------------------------------------------------------------------

    \4\ The barometric pressure used in this calculation must be the 
ambient pressure for the laboratory at the time of test. Many aneroid 
barometers, such as those used at weather stations and airports, are 
precorrected to give sea-level readings; these must not be used.

---------------------------------------------------------------------------

[[Page 445]]

---------------------------------------------------------------------------
A=ambient barometer pressure, kPa.\4\

    (2) Inspect the inside of the sample cup, lid, and shutter mechanism 
for cleaniness and freedom from contamination. Use an absorbent paper 
tissue to wipe clean, if necessary. Put cover in place and lock 
securely. The filing orifice may be convenienty cleaned with a pipe 
cleaner.
    (3) Set the instrument at the target temperature.
    (i) For target temperature below ambient. The instrument power 
switch is to be in the off position. Fill the refigerant-charged cooling 
block with a suitable material.\5\ Raise the lid and shutter assembly, 
and position the base of the block in the sample cup, being careful not 
to injure or mar the cup. When the thermometer reads approximately 6 to 
10 [deg]C (10 to 20 [deg]F) below the target temperature, remove the 
cooling block and quickly dry the cup with a paper tissue to remove any 
moisture. Immediately close the lid and shutter assembly and secure. 
Prepare to introduce the sample using the syringe, both of which have 
been precooled to a temperature 5 to 10 [deg]C (10 to 20 [deg]F) below 
the target temperature.
---------------------------------------------------------------------------

    \5\ If the target or specification temperature is not less than 5 
[deg]C (40 [deg]F) crushed ice and water may be used as charging 
(cooling) fluid. If below 5 [deg]C (40 [deg]F), a suitable charging 
(cooling) fluid is solid carbon dioxide (dry ice) and acetone. If the 
refrigerant charged cooling module is unavailable, refer to the 
manufacturer's instruction manual for alternative methods of cooling.
---------------------------------------------------------------------------

    (A) Caution: Do not cool the sample block below -38 [deg]C, the 
freezing point of mercury.
    (B) Caution: Acetone is extremely flammable. Keep away from heat, 
sparks, and flames and keep container closed when not actually pouring 
acetone. Use only in a well-ventilated area. Avoid inhalation and 
contact with the eyes or skin. Use cloth or leather gloves, goggles or 
safety shield, and keep dry ice in a canvas bag, especially when 
cracking.
    (ii) For target temperature above ambient. Switch the instrument on 
and turn the coarse temperature control knob fully clockwise (full on) 
causing the indicator light to illuminate.\6\ When the thermometer 
indicates a temperature about 3 [deg]C (5 [deg]F) below the target (or 
specification) temperature, reduce the heat input to the sample cup by 
turning the coarse temperature control knob counter-clockwise to the 
desired control point (see Sec.  1500.43a(i)(1)). When the indicator 
light slowly cycles on and off read the temperature on the thermometer. 
If necessary, adjust the fine (center) temperature control knob to 
obtain the desired test (target) temperature. When the test temperature 
is reached and the indicator lamp slowly cycles on and off, prepare to 
introduce the sample.
---------------------------------------------------------------------------

    \6\ The target temperature may be attained by originally turning the 
coarse temperature control knob to the proper setting (see Sec.  
1500.43a(h)(1) for the temperature desired rather than the maximum 
setting (full on). The elapsed time to reach the temperature will be 
greater, except for maximum temperature. However, less attention will be 
required during the intervening period.
---------------------------------------------------------------------------

    (4) Charge the syringe with a 2-ml specimen of the sample \7\ to be 
tested; transfer the syringe to the filling orifice, taking care not to 
lose any sample; discharge the test specimen into the cup by fully 
depressing the syringe plunger, remove the syringe.
---------------------------------------------------------------------------

    \7\ For target or expected temperatures below ambient, both syringe 
and sample must be precooled to cup temperature (see Sec.  
1500.43a(i)(3)(i)) before the specimen is taken.
---------------------------------------------------------------------------

    (5)(i) Set the timer \8\ by rotating its knob clockwise to its stop. 
Open the gas control valve and light the pilot and test flames. Adjust 
the test flame with the pinch valve to conform to the size of the 4-mm 
(5/32-in.) gage.
---------------------------------------------------------------------------

    \8\ For target temperatures below ambient, do not set the timer. 
Adjust the test flame and allow the temperature to rise under ambient 
conditions until the target temperature is reached. Immediately apply 
the test flame as detailed.
---------------------------------------------------------------------------

    (ii) After the time signal indicates the specimen is at test 
temperature \8\, apply the test flame by slowly and uniformly opening 
the shutter and closing it completely over a period of approximately 
2\1/2\ s.\9\ Watch closely for a flash at the cup openings.
---------------------------------------------------------------------------

    \9\ Never apply the test flame to the specimen more than once. Fresh 
portions of the sample must be used for each test.

---------------------------------------------------------------------------

[[Page 446]]

    (iii) The sample is deemed to have flashed when a large flame 
appears and instantaneously propagates itself over the surface of the 
sample (see Sec.  1500.43a(c)).
    (6) Record the test results as ``flash'' or ``no flash'' and the 
test temperature.
    (7) Turn off the pilot and test flames using the gas control valve. 
Remove the sample and clean the instrument. It may be necessary to allow 
the cup temperature to decline to a safe level before cleaning.
    (j) Test Method B--for determining Finite or Actual Flashpoint. (1) 
Inspect the inside of the sample cup, lid, and shutter mechanism for 
cleanliness and freedom from contamination. Use an absorbent paper 
tissue to wipe clean, if necessary. Put cover in place and lock 
securely. The filling orifice may be conveniently cleaned with a pipe 
cleaner.
    (2) For expected flashpoints below ambient. (i) The instrument power 
switch is to be in off position. Fill the refrigerant-charged cooling 
block with a suitable material.\5\ Raise the lid and shutter assembly, 
and position the base of the block in the sample cup, being careful not 
to injure or mar the cup. When the thermometer reaches a temperature 5 
to 10 [deg]C (10 to 20 [deg]F) below the expected flashpoint, remove the 
cooling block and quickly dry the cup with a paper tissue to remove any 
moisture. Immediately close the lid and shutter assembly and secure. 
Prepare to introduce the sample using the syringe, both of which have 
been precooled to a temperature 5 to 10 [deg]C (10 to 20 [deg]F) below 
the expected temperature (See Sec.  1500.43a(j)(5)).
    (ii) Caution: Do not cool the sample block below -38 [deg]C, the 
freezing point of mercury.
    (3) For tests where the expected flashpoint is above ambient. Turn 
the coarse temperature control knob fully clockwise (full on) causing 
the indicator light to illuminate. When the thermometer reaches a 
temperature 3 [deg]C (5 [deg]F) below the estimated flashpoint, turn the 
coarse temperature knob counter-clockwise to the dial reading 
representing the estimated flashpoint temperature as shown on the 
calibration curve (See Sec.  1500.43a(h)(1)). When the indicator light 
slowly cycles on and off, read the temperature on the thermometer. If 
necessary, adjust the fine temperature control knob to obtain the exact 
desired temperature.
    (4)(i) Charge the syringe \7\ with a 2 ml specimen of the sample \7\ 
to be tested; transfer the syringe to the filling orifice, taking care 
not to lose any sample; discharge the test specimen into the cup by 
fully depressing the syringe plunger; remove the syringe.
    (ii) Set the timer \10\ by rotating its knob clockwise to its stop. 
Open the gas control valve and ignite the pilot and test flames. Adjust 
the test flame with the pinch valve to conform to the size of the 4-mm 
(\5/32\-in.) gage.
    (iii) After the audible time signal indicates the specimen is at 
test temperature,\10\ apply the test flame by slowly and uniformly 
opening the shutter and then closing it completely over a period of 
approximately 2\1/2\ s. Watch closely for a flash at the cup opening.
---------------------------------------------------------------------------

    \10\ For expected flashpoint below ambient, do not set the timing 
device. Adjust the test flame. Allow the temperature to rise under 
ambient conditions until the temperature reaches 5 [deg]C (9 [deg]F) 
below the expected flashpoint. Immediately apply the test flame.
---------------------------------------------------------------------------

    (iv) The sample is deemed to have flashed only if a large flame 
appears and instantaneously propagates itself over the surface of the 
sample. (See Sec.  1500.43a(c).)
    (v) Turn off the pilot and test flames using the gas control valve. 
When the cup temperature declines to a safe level, remove the sample and 
clean the instrument.
    (5)(i) If a flash was observed in Sec.  1500.43a(j)(4)(iii) repeat 
the procedure given in Sec.  1500.43a(j)(2) or (3), and in Sec.  
1500.43a(j)(4), testing a new specimen at a temperature 5 [deg]C (9 
[deg]F) below that at which the flash was observed.
    (ii) If necessary, repeat the procedure in Sec.  1500.43a(j)(5)(i), 
lowering the temperature 5 [deg]C (9 [deg]F) each time, until no flash 
is observed.\9\
    (iii) Proceed to Sec.  1500.43a(j)(7).
    (6)(i) If no flash was observed in Sec.  1500.43a(j)(4)(iii) repeat 
the procedure given in Sec.  1500.43a(j)(2) or (3), and in Sec.  
1500.43a(j)(4), testing a fresh specimen at a temperature 5 [deg]C (9 
[deg]F) above that

[[Page 447]]

at which the specimen was tested in Sec.  1500.43a(j)(4)(iii).
    (ii) If necessary repeat the procedure in Sec.  1500.43a(j)(6)(i), 
above, raising the temperature 5 [deg]C (9 [deg]F) each time until a 
flash is observed.\9\
    (7) Having established a flash within two temperatures 5 [deg]C (9 
[deg]F) apart, repeat the procedure at 1 [deg]C (2 [deg]F) intervals 
from the lower of the two temperatures until a flash is observed.\9\ 
Record the temperature of the test when this flash occurs as the 
flashpoint, allowing for any known thermometer correction. Record the 
barometric pressure.\4\
    (8) The flashpoint determined in Sec.  1500.43a(j)(7) will be to the 
nearest 1 [deg]C (2 [deg]F). If improved accuracy is desired (that is, 
to the nearest 0.5 [deg]C (1 [deg]F)), test a fresh specimen at a 
temperature 0.5 [deg]C (1 [deg]F) below that at which the flash was 
observed in Sec.  1500.43a(j)(7). If no flash is observed, the 
temperature recorded in Sec.  1500.43a(j)(7), is the flashpoint to the 
nearest 0.5 [deg]C (1 [deg]F). If a flash is observed at the lower 
temperature, record this latter temperature as the flashpoint.
    (9) Turn off the pilot and test flames using the gas control valve. 
When the cup temperature declines to a safe level, remove the sample and 
clean the instrument.
    (k) Calculations. If it is desired to correct the observed finite 
flashpoint for the effect of barometric pressure, proceed as follows: 
Observe and record the ambient barometric pressure\4\ at the time of the 
test. If the pressure differs from 101.3 kPa (760 mm Hg), correct the 
flashpoint as follows:
    (1) Corrected flashpoint ( [deg]C)=C+0.25 (101.3-A)
    (2) Corrected flashpoint ( [deg]F)=F+0.06 (760-B)
    (3) Corrected flashpoint ( [deg]C)=C+0.03 (760-B)
Where: F=Observed flashpoint, [deg]F,
C=observed flashpoint, [deg]C,
B=ambient barometric pressure, mm Hg; and
A=ambient barometric pressure, kPa.
    (l) Precision. The precision of the method as determined by 
statistical examination of interlaboratory results is as follows:
    (1) Repeatability. The difference between two test results obtained 
by the same operator with the same apparatus under constant operating 
conditions on identical test material, would, in the long run, in the 
normal and correct operation of the test method, exceed the values shown 
in table 2 only in 1 case in 20.
    (2) Reproducibility. The difference between two single and 
independent results obtained by different operators working in different 
laboratories on identical test material, would, in the long run, in the 
normal and correct operation of the test method, exceed the values shown 
in table 2 only in 1 case in 20.
    (m) Flash Test Apparatus. (1)(i) Unit consisting of an aluminum 
alloy or nonrusting metal block of suitable conductivity with a 
cylindrical depression, or sample cup, over which is fitted a cover. A 
thermometer is embedded in the block.
    (ii) The cover is fitted with an opening slide and a device capable 
of inserting an ignition flame (diameter 40.5 mm) 
into the well when the slide device shall intersect the plane of the 
underside of the cover. The cover is also provided with an orifice 
extending into the sample well for insertion of the test sample and also 
a suitable clamping device for securing the cover tightly to the metal 
block. The three openings in the cover shall be within the diameter of 
the sample well. When the slide is in the open position, the two 
openings in the slide shall coincide exactly with the two corresponding 
openings in the cover.
    (iii) Electrical heaters are attached to the bottom of the cup in a 
manner that provides efficient transfer of heat. An electronic heat 
control is required to hold the equilibrium temperature, in a draft-free 
area, within 0.1 [deg]C (0.2 [deg]F) for the low-temperature tester. A 
visual indicator lamp shows when energy is or is not being applied. 
Energy may be supplied from 120 or 240 V, 50 or 60 Hz main service.
    (2)(i) Test flame and pilot flame-regulatable test flame, for 
dipping into the sample cup to try for flash, and a pilot flame, to 
maintain the test flame, are required. These flames may be fueled by 
piped gas service. A gage ring 4mm (5/32 in.) in diameter, engraved on 
the lid near the test flame, is required to ensure uniformity in the 
size of the test flame.

[[Page 448]]

    (ii) Caution: Never recharge the self-contained gas tank at elevated 
temperature, or with the pilot or test flames lighted, nor in the 
vicinity of other flames.
    (iii) Audible Signal is required. The audiable signal is given after 
1 min in the case of the low-temperature tester.
    (iv) Syringe. 2ml capacity, equipped with a needle suitable for use 
with the apparatus, adjusted to deliver 2.000.05 
ml.
    (3) Essential dimensions of the test apparatus are set forth in 
table 3.
    (n) Testing high-viscosity liquids. (1) High-viscosity materials may 
be added to the cup by the following procedure:
    (i) Back load a 5 or 10-ml syringe with the sample to be tested and 
extrude 2 ml into the cup. Spread the specimen as evenly as possible 
over the bottom of the cup.
    (ii) If the sample cannot be loaded into a syringe and extruded, 
other means of adding the sample to the cup may be used such as a spoon. 
Add approximately 2 ml of material to the spoon and then push the 
material from the spoon into the cup.
    (iii) If the test specimen does not close the sampling port in the 
cup, seal the cup externally by suitable means.
    (2) Using the appropriate procedure, either Method A in Sec.  
1500.43a(i) or Method B in Sec.  1500.43a(j), determine the flashpoint 
of the specimen which has been added to the tester in accordance with 
Sec.  1500.43a(n)(i), except that the time specified is increased from 1 
to 5 minutes for samples at or above ambient temperature.

                     Table 1--Calibration of Tester
Material..................................  p-xyleneA (Caution).B
Specific gravity. 15.6/15.6 [deg]C (60/60   0.850 to 0.866.
 [deg]F).
Boiling range.............................  2 [deg]C maximum including
                                             138.35 [deg]C (281.03
                                             [deg]F).
Freezing point............................  11.23 [deg]C (52.2 [deg]F)
                                             minimum.
Flashpoint [deg]C (acceptable range)......  25.60.5 (781 [Phi]).

A Available as Flash Point Check Fluid (p-xylene) from Special Products
  Div., Phillips Petroleum Co., Drawer `O,' Borger, Texas 79007.
B Caution: Handle xylene with care. Avoid inhalation; use only in a well-
  ventilated area. Avoid prolonged or repeated contact with skin. Keep
  away from flames and heat, except as necessary for the actual flash
  point determination.


               Table 2--Repeatability and Reproducibility
------------------------------------------------------------------------
                                        Repeatability,
     Temperature, [deg]C ( [deg]F)         [deg]C (     Reproducibility,
                                            [deg]F)     [deg]C ( [deg]F)
------------------------------------------------------------------------
20(68)................................       0.5(0.9)           1.4(2.6)
70(158)...............................       0.5(0.9)           2.9(5.3)
93(200)...............................       1.3(2.3)           4.9(8.8)
150(300)..............................       2.0(3.6)          7.5(13.5)
200(400)..............................       2.6(4.7)          9.9(17.9)
260(500)..............................       3.3(5.9)         12.4(22.3)
------------------------------------------------------------------------


        Table 3--Essential Dimensions of Flash Test Apparatus A,B
                       Sample Block
Block diameter............................................     61.5-62.5
Sample well diameter......................................   49.40-49.70
Sample well depth.........................................    9.70-10.00
Top of block to center of thermometer hole................   16.00-17.00
Diameter of thermometer hole (approx.)....................        7.0004
                           Cover
Large opening length......................................   12.42-12.47
Large opening width.......................................   10.13-10.18
Small opening length......................................     5.05-5.10
Small opening width.......................................     7.60-7.65
Distance between extreme edges of small openings..........   48.37-48.32
Filling orifice diameter..................................     4.00-4.50
Bore or filler tube.......................................     1.80-1.85
Maximum distance of filler tube from base of well with              0.75
 cover closed (max.)......................................
                           Slide
Large opening length......................................   12.42-12.47
Large opening width.......................................   10.13-10.18
Small opening length......................................     5.05-5.10
Small opening width.......................................     7.60-7.65
Near edge of large opening to end of slide................   12.80-12.85
Extremes of large and small openings......................   30.40-30.45
                            Jet
Length of jet.............................................   18.30-18.40
External diameter at end of jet...........................     2.20-2.60
Bore of jet...............................................     1.60-1.65
Height of jet center above top surface of cover...........   11.00-11.20
Jet pivot to center of block with cover closed............   12.68-12.72

A The O-seal or gasket which provides a seal when the cover is shut,
  should be made of a heat-resistant material capable of withstanding
  temperatures up to 150 [deg]C for the low-range apparatus.
B When in position, the thermometer bulb should be surrounded with heat-
  conducting thermoplastic compound, such as a paste comprised of zinc
  oxide and mineral oil.


[[Page 449]]

[GRAPHIC] [TIFF OMITTED] TC03OC91.054


[51 FR 28539, Aug. 8, 1986]