[Code of Federal Regulations]
[Title 40, Volume 5]
[Revised as of July 1, 2007]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR53.55]

[Page 86-89]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 53_AMBIENT AIR MONITORING REFERENCE AND EQUIVALENT METHODS--Table of 
 
   Subpart E_Procedures for Testing Physical (Design) and Performance 
 
Sec. 53.55  Test for effect of variations in power line voltage and ambient 

temperature.

    (a) Overview. (1) This test procedure is a combined procedure to 
test various performance parameters under variations in power line 
voltage and ambient temperature. Tests shall be conducted in a 
temperature-controlled environment over four 6-hour time periods during 
which reference temperature and flow rate measurements shall be made at 
intervals not to exceed 5 minutes. Specific parameters to be evaluated 
at line voltages of 105 and 125 volts and temperatures of -20 [deg]C and 
+40 [deg]C are as follows:
    (i) Sample flow rate.
    (ii) Flow rate regulation.
    (iii) Flow rate measurement accuracy.
    (iv) Coefficient of variability measurement accuracy.
    (v) Ambient air temperature measurement accuracy.
    (vi) Proper operation of the sampler when exposed to power line 
voltage and ambient temperature extremes.
    (2) The performance parameters tested under this procedure, the 
corresponding minimum performance specifications, and the applicable 
test conditions are summarized in table E-1 of this subpart. Each 
performance parameter tested, as described or determined in the test 
procedure, must meet or exceed the associated performance specification 
given. The candidate sampler must meet all specifications for the 
associated PM2.5 or PM10-2.5 method (as 
applicable) to pass this test procedure.
    (b) Technical definition. Sample flow rate means the quantitative 
volumetric flow rate of the air stream caused by the sampler to enter 
the sampler inlet and pass through the sample filter, measured in actual 
volume units at the temperature and pressure of the air as it enters the 
inlet.
    (c) Required test equipment. (1) Environmental chamber or other 
temperature-controlled environment or environments, capable of obtaining 
and maintaining temperatures at -20 [deg]C and =40 [deg]C as required 
for the test with an accuracy of 2 [deg]C. The 
test environment(s) must be capable of maintaining these temperatures 
within the specified limits continuously with the additional heat load 
of the operating test sampler in the environment. Henceforth, where the 
test procedures specify a test or environmental ``chamber,'' an 
alternative temperature-controlled environmental area or areas may be 
substituted, provided the required test temperatures and all other test 
requirements are met.
    (2) Variable voltage AC power transformer, range 100 Vac to 130 Vac, 
with sufficient current capacity to operate the test sampler 
continuously under the test conditions.
    (3) Flow rate meter, suitable for measuring and recording the actual 
volumetric sample flow rate at the sampler downtube, with a minimum 
range of 10 to 25 actual L/min, 2 percent certified, NIST-traceable 
accuracy. Optional capability for continuous (analog) recording 
capability or digital recording at intervals not to exceed 5 minutes is 
recommended. While a flow meter which provides a direct indication of 
volumetric flow rate is preferred for this test, an alternative 
certified flow measurement device may be used as long as appropriate 
volumetric flow rate corrections are made based on measurements of 
actual ambient temperature and pressure conditions.
    (4) Ambient air temperature recorder, range -30 [deg]C to =50 
[deg]C, with a resolution of 0.1 [deg]C and certified accurate to within 
0.5 [deg]C. Ambient air temperature measurements must be made using 
continuous (analog) recording capability or digital recording at 
intervals not to exceed 5 minutes.

[[Page 87]]

    (5) Barometer, range 600 mm Hg to 800 mm Hg, certified accurate to 2 
mm Hg. If the certified flow rate meter does not provide direct 
volumetric flow rate readings, ambient pressure measurements must be 
made using continuous (analog) recording capability or digital recording 
at intervals not to exceed 5 minutes.
    (6) Flow measurement adaptor (40 CFR part 50, appendix L, figure L-
30) or equivalent adaptor to facilitate measurement of sampler flow rate 
at the sampler downtube.
    (7) Means for creating an additional pressure drop of 55 mm Hg in 
the sampler to simulate a heavily loaded filter, such as an orifice or 
flow restrictive plate installed in the filter holder or a valve or 
other flow restrictor temporarily installed in the flow path near the 
filter.
    (8) AC RMS voltmeter, accurate to 1.0 volt.
    (9) Teflon sample filter, as specified in section 6 of 40 CFR part 
50, appendix L (if required).
    (d) Calibration of test measurement instruments. Submit 
documentation showing evidence of appropriately recent calibration, 
certification of calibration accuracy, and NIST-traceability (if 
required) of all measurement instruments used in the tests. The accuracy 
of flow rate meters shall be verified at the highest and lowest 
pressures and temperatures used in the tests and shall be checked at 
zero and at least one flow rate within 3 percent 
of 16.7 L/min within 7 days prior to use for this test. Where an 
instrument's measurements are to be recorded with an analog recording 
device, the accuracy of the entire instrument-recorder system shall be 
calibrated or verified.
    (e) Test setup. (1) Setup of the sampler shall be performed as 
required in this paragraph (e) and otherwise as described in the 
sampler's operation or instruction manual referred to in Sec. 
53.4(b)(3). The sampler shall be installed upright and set up in the 
temperature-controlled chamber in its normal configuration for 
collecting PM samples. A sample filter and (or) the device for creating 
an additional 55 mm Hg pressure drop shall be installed for the duration 
of these tests. The sampler's ambient temperature, ambient pressure, and 
flow measurement systems shall all be calibrated per the sampler's 
operating manual within 7 days prior to this test.
    (2) The inlet of the candidate sampler shall be removed and the flow 
measurement adaptor installed on the sampler's downtube. A leak check as 
described in the sampler's operation or instruction manual shall be 
conducted and must be properly passed before other tests are carried 
out.
    (3) The inlet of the flow measurement adaptor shall be connected to 
the outlet of the flow rate meter.
    (4) The ambient air temperature recorder shall be installed in the 
test chamber such that it will accurately measure the temperature of the 
air in the vicinity of the candidate sampler without being unduly 
affected by the chamber's air temperature control system.
    (f) Procedure. (1) Set up the sampler as specified in paragraph (e) 
of this section and otherwise prepare the sampler for normal sample 
collection operation as directed in the sampler's operation or 
instruction manual.
    (2) The test shall consist of four test runs, one at each of the 
following conditions of chamber temperature and electrical power line 
voltage (respectively):
    (i) -20 [deg]C 2 [deg]C and 105 1 Vac.
    (ii) -20 [deg]C 2 [deg]C and 125 1 Vac.
    (iii) =40 [deg]C 2 [deg]C and 105 1 Vac.
    (iv) =40 [deg]C 2 [deg]C and 125 1 Vac.
    (3) For each of the four test runs, set the selected chamber 
temperature and power line voltage for the test run. Upon achieving each 
temperature setpoint in the chamber, the candidate sampler and flow 
meter shall be thermally equilibrated for a period of at least 2 hours 
prior to the test run. Following the thermal conditioning time, set the 
sampler to automatically start a 6-hour sample collection period at a 
convenient time.
    (4) During each 6-hour test period:
    (i) Measure and record the sample flow rate with the flow rate meter 
at intervals not to exceed 5 minutes. If ambient temperature and 
pressure corrections are necessary to calculate volumetric flow rate, 
ambient temperature and pressure shall be measured at

[[Page 88]]

the same frequency as that of the certified flow rate measurements. Note 
and record the actual start and stop times for the 6-hour flow rate test 
period.
    (ii) Determine and record the ambient (chamber) temperature 
indicated by the sampler and the corresponding ambient (chamber) 
temperature measured by the ambient temperature recorder specified in 
paragraph (c)(4) of this section at intervals not to exceed 5 minutes.
    (iii) Measure the power line voltage to the sampler at intervals not 
greater than 1 hour.
    (5) At the end of each test run, terminate the sample period (if not 
automatically terminated by the sampler) and download all archived 
instrument data from the test sampler.
    (g) Test results. For each of the four test runs, examine the 
chamber temperature measurements and the power line voltage 
measurements. Verify that the temperature and line voltage met the 
requirements specified in paragraph (f) of this section at all times 
during the test run. If not, the test run is not valid and must be 
repeated. Determine the test results as follows:
    (1) Mean sample flow rate. (i) From the certified measurements 
(Qref) of the test sampler flow rate, tabulate each flow rate 
measurement in units of L/min. If ambient temperature and pressure 
corrections are necessary to calculate volumetric flow rate, each 
measured flow rate shall be corrected using its corresponding 
temperature and pressure measurement values. Calculate the mean flow 
rate for each sample period (Qref,ave) as follows:
                               Equation 11
[GRAPHIC] [TIFF OMITTED] TR18JY97.073

where:

n equals the number of discrete certified flow rate measurements over 
each 6-hour test period.

    (ii)(A) Calculate the percent difference between this mean flow rate 
value and the design value of 16.67 L/min, as follows:
                               Equation 12
[GRAPHIC] [TIFF OMITTED] TR18JY97.074

    (B) To successfully pass this test, the percent difference 
calculated in Equation 12 of this paragraph (g)(1)(ii) must be within 
5 percent for each test run.
    (2) Sample flow rate regulation. (i) From the certified measurements 
of the test sampler flow rate, calculate the sample coefficient of 
variation of the discrete measurements as follows:
                               Equation 13
[GRAPHIC] [TIFF OMITTED] TR18JY97.075

    (ii) To successfully pass this test, the calculated coefficient of 
variation for the certified flow rates must not exceed 2 percent.
    (3) Flow rate measurement accuracy. (i) Using the mean volumetric 
flow rate reported by the candidate test sampler at the completion of 
each 6-hour test (Qind,ave), determine the accuracy of the 
reported mean flow rate as:
                               Equation 14
[GRAPHIC] [TIFF OMITTED] TR18JY97.076

    (ii) To successfully pass this test, the percent difference 
calculated in Equation 14 of this paragraph (g)(3) shall not exceed 2 
percent for each test run.
    (4) Flow rate coefficient of variation measurement accuracy. (i) 
Using the flow rate coefficient of variation indicated by the candidate 
test sampler (%CVind), determine the accuracy of the reported 
coefficient of variation as:
                               Equation 15
[GRAPHIC] [TIFF OMITTED] TR18JY97.077

    (ii) To successfully pass this test, the absolute difference 
calculated in Equation 15 of this paragraph (g)(4) must not exceed 0.3 
(CV%) for each test run.
    (5) Ambient temperature measurement accuracy. (i) Calculate the 
absolute

[[Page 89]]

value of the difference between the mean ambient air temperature 
indicated by the test sampler and the mean ambient (chamber) air 
temperature measured with the ambient air temperature recorder as:
[GRAPHIC] [TIFF OMITTED] TR17OC06.037

Where:
Tind,ave = The mean ambient air temperature indicated by the 
          test sampler, [deg]C; and
Tref,ave = The mean ambient air temperature measured by the 
          reference temperature instrument, [deg]C.

    (ii) The calculated temperature difference must be less than 2 
[deg]C for each test run.
    (6) Sampler functionality. To pass the sampler functionality test, 
the following two conditions must both be met for each test run:
    (i) The sampler must not shutdown during any portion of the 6-hour 
test.
    (ii) An inspection of the downloaded data from the test sampler 
verifies that all the data are consistent with normal operation of the 
sampler.

[62 FR 38799, July 18, 1997, as amended at 71 FR 61291, Oct. 17, 2006]