[Code of Federal Regulations]
[Title 40, Volume 5]
[Revised as of July 1, 2007]
From the U.S. Government Printing Office via GPO Access
[CITE: 40CFR53.56]

[Page 89-92]
 
                   TITLE 40--PROTECTION OF ENVIRONMENT
 
         CHAPTER I--ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)
 
PART 53_AMBIENT AIR MONITORING REFERENCE AND EQUIVALENT METHODS--Table of 
 
   Subpart E_Procedures for Testing Physical (Design) and Performance 
 
Sec. 53.56  Test for effect of variations in ambient pressure.

    (a) Overview. (1) This test procedure is designed to test various 
sampler performance parameters under variations in ambient (barometric) 
pressure. Tests shall be conducted in a pressure-controlled environment 
over two 6-hour time periods during which reference pressure and flow 
rate measurements shall be made at intervals not to exceed 5 minutes. 
Specific parameters to be evaluated at operating pressures of 600 and 
800 mm Hg are as follows:
    (i) Sample flow rate.
    (ii) Flow rate regulation.
    (iii) Flow rate measurement accuracy.
    (iv) Coefficient of variability measurement accuracy.
    (v) Ambient pressure measurement accuracy.
    (vi) Proper operation of the sampler when exposed to ambient 
pressure extremes.
    (2) The performance parameters tested under this procedure, the 
corresponding minimum performance specifications, and the applicable 
test conditions are summarized in table E-1 of this subpart. Each 
performance parameter tested, as described or determined in the test 
procedure, must meet or exceed the associated performance specification 
given. The candidate sampler must meet all specifications for the 
associated PM2.5 or PM10-2.5 method (as 
applicable) to pass this test procedure.
    (b) Technical definition. Sample flow rate means the quantitative 
volumetric flow rate of the air stream caused by the sampler to enter 
the sampler inlet and pass through the sample filter, measured in actual 
volume units at the temperature and pressure of the air as it enters the 
inlet.
    (c) Required test equipment. (1) Hypobaric chamber or other 
pressure-controlled environment or environments, capable of obtaining 
and maintaining pressures at 600 mm Hg and 800 mm Hg required for the 
test with an accuracy of 5 mm Hg. Henceforth, where the test procedures 
specify a test or environmental chamber, an alternative pressure-
controlled environmental area or areas may be substituted, provided the 
test pressure requirements are met. Means for simulating ambient 
pressure using a closed-loop sample air system may also be approved for 
this test; such a proposed method for simulating the test pressure 
conditions may be described and submitted to EPA at the address given in 
Sec. 53.4(a) prior to conducting the test for a specific individual 
determination of acceptability.
    (2) Flow rate meter, suitable for measuring and recording the actual 
volumetric sampler flow rate at the sampler downtube, with a minimum 
range of 10 to 25 L/min, 2 percent certified, NIST-traceable accuracy. 
Optional capability for continuous (analog) recording capability or 
digital recording at intervals not to exceed 5 minutes is recommended. 
While a flow meter which provides a direct indication of volumetric flow 
rate is preferred for this test, an alternative certified flow 
measurement device may be used as long as appropriate volumetric

[[Page 90]]

flow rate corrections are made based on measurements of actual ambient 
temperature and pressure conditions.
    (3) Ambient air temperature recorder (if needed for volumetric 
corrections to flow rate measurements) with a range -30 [deg]C to =50 
[deg]C, certified accurate to within 0.5 [deg]C. If the certified flow 
meter does not provide direct volumetric flow rate readings, ambient 
temperature measurements must be made using continuous (analog) 
recording capability or digital recording at intervals not to exceed 5 
minutes.
    (4) Barometer, range 600 mm Hg to 800 mm Hg, certified accurate to 2 
mm Hg. Ambient air pressure measurements must be made using continuous 
(analog) recording capability or digital recording at intervals not to 
exceed 5 minutes.
    (5) Flow measurement adaptor (40 CFR part 50, appendix L, figure L-
30) or equivalent adaptor to facilitate measurement of sampler flow rate 
at the sampler downtube.
    (6) Means for creating an additional pressure drop of 55 mm Hg in 
the sampler to simulate a heavily loaded filter, such as an orifice or 
flow restrictive plate installed in the filter holder or a valve or 
other flow restrictor temporarily installed in the flow path near the 
filter.
    (7) Teflon sample filter, as specified in section 6 of 40 CFR part 
50, appendix L (if required).
    (d) Calibration of test measurement instruments. Submit 
documentation showing evidence of appropriately recent calibration, 
certification of calibration accuracy, and NIST-traceability (if 
required) of all measurement instruments used in the tests. The accuracy 
of flow rate meters shall be verified at the highest and lowest 
pressures and temperatures used in the tests and shall be checked at 
zero and at least one flow rate within 3 percent 
of 16.7 L/min within 7 days prior to use for this test. Where an 
instrument's measurements are to be recorded with an analog recording 
device, the accuracy of the entire instrument-recorder system shall be 
calibrated or verified.
    (e) Test setup. (1) Setup of the sampler shall be performed as 
required in this paragraph (e) and otherwise as described in the 
sampler's operation or instruction manual referred to in Sec. 
53.4(b)(3). The sampler shall be installed upright and set up in the 
pressure-controlled chamber in its normal configuration for collecting 
PM samples. A sample filter and (or) the device for creating an 
additional 55 mm Hg pressure drop shall be installed for the duration of 
these tests. The sampler's ambient temperature, ambient pressure, and 
flow measurement systems shall all be calibrated per the sampler's 
operating manual within 7 days prior to this test.
    (2) The inlet of the candidate sampler shall be removed and the flow 
measurement adaptor installed on the sampler's downtube. A leak check as 
described in the sampler's operation or instruction manual shall be 
conducted and must be properly passed before other tests are carried 
out.
    (3) The inlet of the flow measurement adaptor shall be connected to 
the outlet of the flow rate meter.
    (4) The barometer shall be installed in the test chamber such that 
it will accurately measure the air pressure to which the candidate 
sampler is subjected.
    (f) Procedure. (1) Set up the sampler as specified in paragraph (e) 
of this section and otherwise prepare the sampler for normal sample 
collection operation as directed in the sampler's operation or 
instruction manual.
    (2) The test shall consist of two test runs, one at each of the 
following conditions of chamber pressure:
    (i) 600 mm Hg.
    (ii) 800 mm Hg.
    (3) For each of the two test runs, set the selected chamber pressure 
for the test run. Upon achieving each pressure setpoint in the chamber, 
the candidate sampler shall be pressure-equilibrated for a period of at 
least 30 minutes prior to the test run. Following the conditioning time, 
set the sampler to automatically start a 6-hour sample collection period 
at a convenient time.
    (4) During each 6-hour test period:
    (i) Measure and record the sample flow rate with the flow rate meter 
at intervals not to exceed 5 minutes. If ambient temperature and 
pressure corrections are necessary to calculate volumetric flow rate, 
ambient temperature and pressure shall be measured at

[[Page 91]]

the same frequency as that of the certified flow rate measurements. Note 
and record the actual start and stop times for the 6-hour flow rate test 
period.
    (ii) Determine and record the ambient (chamber) pressure indicated 
by the sampler and the corresponding ambient (chamber) pressure measured 
by the barometer specified in paragraph (c)(4) of this section at 
intervals not to exceed 5 minutes.
    (5) At the end of each test period, terminate the sample period (if 
not automatically terminated by the sampler) and download all archived 
instrument data for the test run from the test sampler.
    (g) Test results. For each of the two test runs, examine the chamber 
pressure measurements. Verify that the pressure met the requirements 
specified in paragraph (f) of this section at all times during the test. 
If not, the test run is not valid and must be repeated. Determine the 
test results as follows:
    (1) Mean sample flow rate. (i) From the certified measurements 
(Qref) of the test sampler flow rate, tabulate each flow rate 
measurement in units of L/min. If ambient temperature and pressure 
corrections are necessary to calculate volumetric flow rate, each 
measured flow rate shall be corrected using its corresponding 
temperature and pressure measurement values. Calculate the mean flow 
rate for the sample period (Qref,ave) as follows:
                               Equation 17
[GRAPHIC] [TIFF OMITTED] TR18JY97.079

where:

n equals the number of discrete certified flow measurements over the 6-
hour test period.

    (ii)(A) Calculate the percent difference between this mean flow rate 
value and the design value of 16.67 L/min, as follows:
                               Equation 18
[GRAPHIC] [TIFF OMITTED] TR18JY97.080

    (B) To successfully pass this test, the percent difference 
calculated in Equation 18 of this paragraph (g)(1) must be within 5 percent for each test run.
    (2) Sample flow rate regulation. (i) From the certified measurements 
of the test sampler flow rate, calculate the sample coefficient of 
variation of the discrete measurements as follows:
                               Equation 19
[GRAPHIC] [TIFF OMITTED] TR18JY97.081

    (ii) To successfully pass this test, the calculated coefficient of 
variation for the certified flow rates must not exceed 2 percent.
    (3) Flow rate measurement accuracy. (i) Using the mean volumetric 
flow rate reported by the candidate test sampler at the completion of 
each 6-hour test (Qind,ave), determine the accuracy of the 
reported mean flow rate as:
                               Equation 20
[GRAPHIC] [TIFF OMITTED] TR18JY97.082

    (ii) To successfully pass this test, the percent difference 
calculated in Equation 20 of this paragraph (g)(3) shall not exceed 2 
percent for each test run.
    (4) Flow rate CV measurement accuracy. (i) Using the flow rate 
coefficient of variation indicated by the candidate test sampler at the 
completion of the 6-hour test (%CVind), determine the 
accuracy of the reported coefficient of variation as:
                               Equation 21
[GRAPHIC] [TIFF OMITTED] TR18JY97.083

    (ii) To successfully pass this test, the absolute difference in 
values calculated in Equation 21 of this paragraph (g)(4) must not 
exceed 0.3 (CV%) for each test run.

[[Page 92]]

    (5) Ambient pressure measurement accuracy. (i) Calculate the 
absolute difference between the mean ambient air pressure indicated by 
the test sampler and the ambient (chamber) air pressure measured with 
the reference barometer as:
                               Equation 22
[GRAPHIC] [TIFF OMITTED] TR18JY97.084

where:

Pind,ave = mean ambient pressure indicated by the test 
sampler, mm Hg; and
Pref,ave = mean barometric pressure measured by the reference 
barometer, mm Hg.

    (ii) The calculated pressure difference must be less than 10 mm Hg 
for each test run to pass the test.
    (6) Sampler functionality. To pass the sampler functionality test, 
the following two conditions must both be met for each test run:
    (i) The sampler must not shut down during any part of the 6-hour 
tests; and
    (ii) An inspection of the downloaded data from the test sampler 
verifies that all the data are consistent with normal operation of the 
sampler.

[62 FR 38799, July 18, 1997; 63 FR 7714, Feb. 17, 1998, as amended at 71 
FR 61292, Oct. 17, 2006]